Reconstructing Quantitative Cerebral Perfusion Images Directly From Measured Sinogram Data Acquired Using C-arm Cone-Beam CT
To shorten the door-to-puncture time for better treating patients with acute ischemic stroke, it is highly desired to obtain quantitative cerebral perfusion images using C-arm cone-beam computed tomography (CBCT) equipped in the interventional suite. However, limited by the slow gantry rotation spee...
Saved in:
Main Authors | , , , , , , , |
---|---|
Format | Journal Article |
Language | English |
Published |
06.12.2024
|
Subjects | |
Online Access | Get full text |
DOI | 10.48550/arxiv.2412.05084 |
Cover
Summary: | To shorten the door-to-puncture time for better treating patients with acute
ischemic stroke, it is highly desired to obtain quantitative cerebral perfusion
images using C-arm cone-beam computed tomography (CBCT) equipped in the
interventional suite. However, limited by the slow gantry rotation speed, the
temporal resolution and temporal sampling density of typical C-arm CBCT are
much poorer than those of multi-detector-row CT in the diagnostic imaging
suite. The current quantitative perfusion imaging includes two cascaded steps:
time-resolved image reconstruction and perfusion parametric estimation. For
time-resolved image reconstruction, the technical challenge imposed by poor
temporal resolution and poor sampling density causes inaccurate quantification
of the temporal variation of cerebral artery and tissue attenuation values. For
perfusion parametric estimation, it remains a technical challenge to
appropriately design the handcrafted regularization for better solving the
associated deconvolution problem. These two challenges together prevent
obtaining quantitatively accurate perfusion images using C-arm CBCT. The
purpose of this work is to simultaneously address these two challenges by
combining the two cascaded steps into a single joint optimization problem and
reconstructing quantitative perfusion images directly from the measured
sinogram data. In the developed direct cerebral perfusion parametric image
reconstruction technique, TRAINER in short, the quantitative perfusion images
have been represented as a subject-specific conditional generative model
trained under the constraint of the time-resolved CT forward model, perfusion
convolutional model, and the subject's own measured sinogram data. Results
shown in this paper demonstrated that using TRAINER, quantitative cerebral
perfusion images can be accurately obtained using C-arm CBCT in the
interventional suite. |
---|---|