Model‐based iterative reconstruction for single‐shot EPI at 7T

Purpose To describe a model‐based reconstruction strategy for single‐shot echo planar imaging (EPI) that intrinsically accounts for k‐space nonuniformity, Nyquist ghosting, and geometric distortions during rather than before or after image reconstruction. Methods Ramp sampling and inhomogeneous B0 f...

Full description

Saved in:
Bibliographic Details
Published inMagnetic resonance in medicine Vol. 78; no. 6; pp. 2250 - 2264
Main Authors Yarach, Uten, In, Myung‐Ho, Chatnuntawech, Itthi, Bilgic, Berkin, Godenschweger, Frank, Mattern, Hendrik, Sciarra, Alessandro, Speck, Oliver
Format Journal Article
LanguageEnglish
Published United States Wiley Subscription Services, Inc 01.12.2017
Subjects
Online AccessGet full text
ISSN0740-3194
1522-2594
1522-2594
DOI10.1002/mrm.26633

Cover

More Information
Summary:Purpose To describe a model‐based reconstruction strategy for single‐shot echo planar imaging (EPI) that intrinsically accounts for k‐space nonuniformity, Nyquist ghosting, and geometric distortions during rather than before or after image reconstruction. Methods Ramp sampling and inhomogeneous B0 field‐induced distortion cause the EPI samples to lie on a non‐Cartesian grid, thus requiring the nonuniform fast Fourier transform. Additionally, a 2D Nyquist ghost phase correction without the need for extra navigator acquisition is included in the proposed reconstruction. Coil compression is also incorporated to reduce the computational load. The proposed method is applied to phantom and human brain MRI data. Results The results demonstrate that Nyquist ghosting and geometric distortions are reduced by the proposed reconstruction. The proposed 2D phase correction is superior to a conventional 1D correction. The reductions of both artifacts lead to improved temporal signal‐to‐noise ratio (tSNR). The virtual coil results suggest that the processing time can be reduced by up to 75%, with a mean tSNR loss of only 3.2% when using 8‐virtual instead of 32‐physical coils for twofold undersampled data. Conclusion The proposed reconstruction improves the quality (ghosting, geometry, and tSNR) of EPI without requiring calibration data for Nyquist ghost correction. Magn Reson Med 78:2250–2264, 2017. © 2017 International Society for Magnetic Resonance in Medicine.
Bibliography:Grant sponsor: Supported by the German Research Foundation DFG (SP632‐4); the National Institutes of Health (R01‐DA021146); and the FP7 Marie Curie Actions of the European Commission (FP7‐PEOPLE‐2012‐ITN‐316716).