세척/흡입 밸브 유동해석
In order to develop a valve for a laparoscopic surgical irrigation/suction device, three shapes of valve were designed, and CDF analysis was performed on each design shape to get the relation between flow rate and pressure drop to determine the best shape. Steady-state incompressible flow analysis w...
Saved in:
| Published in | 한국전산유체공학회지 Vol. 29; no. 3; pp. 163 - 172 |
|---|---|
| Main Authors | , |
| Format | Journal Article |
| Language | Korean |
| Published |
한국전산유체공학회
01.09.2024
|
| Subjects | |
| Online Access | Get full text |
| ISSN | 1598-6071 1598-6071 |
| DOI | 10.6112/kscfe.2024.29.3.163 |
Cover
| Summary: | In order to develop a valve for a laparoscopic surgical irrigation/suction device, three shapes of valve were designed, and CDF analysis was performed on each design shape to get the relation between flow rate and pressure drop to determine the best shape. Steady-state incompressible flow analysis was performed using the Fluent s/w, and the Realizable turbulence model and standard wall function were used for turbulent flow analysis. We were able to develop a valve that is easy on the fingers and can replace the existing tube compression valve. KCI Citation Count: 0 |
|---|---|