腕位置変化がPATおよびPTT計測に与える影響について

PAT and PTT are considered one of the effective methods for cuffless blood pressure estimation. On the other hand, the arterial properties are extremely sensitive to measurement position. For this reason, PAT and PTT may also be affected by the arm position. The aim of this study was to verify the e...

Full description

Saved in:
Bibliographic Details
Published inTransactions of Japanese Society for Medical and Biological Engineering Vol. 54Annual; no. 26AM-Abstract; p. S34
Main Authors 田村, 俊世, 水谷, 孝一, 関根, 正樹, 前田, 祐佳
Format Journal Article
LanguageJapanese
Published 公益社団法人 日本生体医工学会 2016
Japanese Society for Medical and Biological Engineering
Online AccessGet full text
ISSN1347-443X
1881-4379
DOI10.11239/jsmbe.54Annual.S34

Cover

More Information
Summary:PAT and PTT are considered one of the effective methods for cuffless blood pressure estimation. On the other hand, the arterial properties are extremely sensitive to measurement position. For this reason, PAT and PTT may also be affected by the arm position. The aim of this study was to verify the effect of the arm position changes on monitoring PAT and PTT. Eleven healthy volunteers participated in this experiment. Subjects were asked to put their right arm down and up. The results showed that the changes of PTT during the arm position change revealed the effect of blood pressure change. From a standpoint of estimating blood pressure, PPT was more useful than PAT.