Clinical Validation of a Miniature Nuclear Probe System for Continuous On-line Monitoring of Cardiac Function and ST-Segment

A new, miniature cesium iodide/photodiode nuclear probe (the "Cardioscint") has been developed for continuous on-line measurement of left ventricular function and the ST-segment. Ejection fraction (EF) measurements in 77 patients were compared with gated equilibrium radionuclide ventriculo...

Full description

Saved in:
Bibliographic Details
Published inThe Journal of nuclear medicine (1978) Vol. 32; no. 1; pp. 37 - 43
Main Authors Broadhurst, Paul, Cashman, Peter, Crawley, John, Raftery, Edward, Lahiri, Avijit
Format Journal Article
LanguageEnglish
Published Reston, VA Soc Nuclear Med 01.01.1991
Society of Nuclear Medicine
Subjects
Online AccessGet full text
ISSN0161-5505
1535-5667

Cover

More Information
Summary:A new, miniature cesium iodide/photodiode nuclear probe (the "Cardioscint") has been developed for continuous on-line measurement of left ventricular function and the ST-segment. Ejection fraction (EF) measurements in 77 patients were compared with gated equilibrium radionuclide ventriculograms. The probe was positioned over the left ventricle by first using a blind positioning algorithm and then by using the gamma camera. Background was measured both manually and automatically. There was good correlation between probe (positioned blind) and gamma camera EF with both manual (r = 0.80, n = 65) and automatic (r = 0.78, n = 66) backgrounds. Use of the gamma camera did not significantly alter the results. Correlation between the probe stroke counts and thermodilution-derived stroke index during atrial pacing in six subjects was also satisfactory (r = 0.69, n = 102). Thus, the Cardioscint is able to provide a reliable estimate of EF and can track rapid changes in cardiac volumes.
Bibliography:ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Feature-1
content type line 23
ISSN:0161-5505
1535-5667