Validation of a deep learning computer aided system for CT based lung nodule detection, classification, and growth rate estimation in a routine clinical population

In this study, we evaluated a commercially available computer assisted diagnosis system (CAD). The deep learning algorithm of the CAD was trained with a lung cancer screening cohort and developed for detection, classification, quantification, and growth of actionable pulmonary nodules on chest CT sc...

Full description

Saved in:
Bibliographic Details
Published inPloS one Vol. 17; no. 5; p. e0266799
Main Authors Murchison, John T., Ritchie, Gillian, Senyszak, David, Nijwening, Jeroen H., van Veenendaal, Gerben, Wakkie, Joris, van Beek, Edwin J. R.
Format Journal Article
LanguageEnglish
Published United States Public Library of Science 05.05.2022
Public Library of Science (PLoS)
Subjects
Online AccessGet full text
ISSN1932-6203
1932-6203
DOI10.1371/journal.pone.0266799

Cover

More Information
Summary:In this study, we evaluated a commercially available computer assisted diagnosis system (CAD). The deep learning algorithm of the CAD was trained with a lung cancer screening cohort and developed for detection, classification, quantification, and growth of actionable pulmonary nodules on chest CT scans. Here, we evaluated the CAD in a retrospective cohort of a routine clinical population. In total, a number of 337 scans of 314 different subjects with reported nodules of 3-30 mm in size were included into the evaluation. Two independent thoracic radiologists alternately reviewed scans with or without CAD assistance to detect, classify, segment, and register pulmonary nodules. A third, more experienced, radiologist served as an adjudicator. In addition, the cohort was analyzed by the CAD alone. The study cohort was divided into five different groups: 1) 178 CT studies without reported pulmonary nodules, 2) 95 studies with 1-10 pulmonary nodules, 23 studies from the same patients with 3) baseline and 4) follow-up studies, and 5) 18 CT studies with subsolid nodules. A reference standard for nodules was based on majority consensus with the third thoracic radiologist as required. Sensitivity, false positive (FP) rate and Dice inter-reader coefficient were calculated. After analysis of 470 pulmonary nodules, the sensitivity readings for radiologists without CAD and radiologist with CAD, were 71.9% (95% CI: 66.0%, 77.0%) and 80.3% (95% CI: 75.2%, 85.0%) (p < 0.01), with average FP rate of 0.11 and 0.16 per CT scan, respectively. Accuracy and kappa of CAD for classifying solid vs sub-solid nodules was 94.2% and 0.77, respectively. Average inter-reader Dice coefficient for nodule segmentation was 0.83 (95% CI: 0.39, 0.96) and 0.86 (95% CI: 0.51, 0.95) for CAD versus readers. Mean growth percentage discrepancy of readers and CAD alone was 1.30 (95% CI: 1.02, 2.21) and 1.35 (95% CI: 1.01, 4.99), respectively. The applied CAD significantly increased radiologist's detection of actionable nodules yet also minimally increasing the false positive rate. The CAD can automatically classify and quantify nodules and calculate nodule growth rate in a cohort of a routine clinical population. Results suggest this Deep Learning software has the potential to assist chest radiologists in the tasks of pulmonary nodule detection and management within their routine clinical practice.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
Competing Interests: JTM, GR, and DS declare no competing interests. EJRvB declares ownership of QCTIS Ltd and serves on the medical advisory boards of Aidence BV and Imbio LLC. EJRvB received a restricted research grant from Siemens Healthineers and speaker fees from AstraZeneca and Roche Diagnostics. EJRvB’s non-financial competing interest is serving as an expert witness on medicolegal advice on imaging based cases. JHN and GvV are full time, paid employees of Aidence and have aided in part of the raw data analysis (GvV) or re-submitting this manuscript to PLOS ONE (JHN). These interests do not alter our adherence to PLOS ONE policies on sharing data and materials.
ISSN:1932-6203
1932-6203
DOI:10.1371/journal.pone.0266799