Dose Reduction and Low-Contrast Detectability Using Iterative CBCT Reconstruction Algorithm for Radiotherapy

Introduction: Several studies have reported the relation between the imaging dose and secondary cancer risk and have emphasized the need to minimize the additional imaging dose as low as reasonably achievable. The iterative cone-beam computed tomography (iCBCT) algorithm can improve the image qualit...

Full description

Saved in:
Bibliographic Details
Published inTechnology in Cancer Research & Treatment Vol. 21; p. 15330338211067312
Main Authors Washio, Hayate, Ohira, Shingo, Funama, Yoshinori, Ueda, Yoshihiro, Morimoto, Masahiro, Kanayama, Naoyuki, Isono, Masaru, Inui, Shoki, Nitta, Yuya, Miyazaki, Masayoshi, Teshima, Teruki
Format Journal Article
LanguageEnglish
Published Los Angeles, CA SAGE Publications 01.01.2022
Sage Publications Ltd
SAGE Publishing
Subjects
Online AccessGet full text
ISSN1533-0346
1533-0338
1533-0338
DOI10.1177/15330338211067312

Cover

More Information
Summary:Introduction: Several studies have reported the relation between the imaging dose and secondary cancer risk and have emphasized the need to minimize the additional imaging dose as low as reasonably achievable. The iterative cone-beam computed tomography (iCBCT) algorithm can improve the image quality by utilizing scatter correction and statistical reconstruction. We investigate the use of a novel iCBCT reconstruction algorithm to reduce the patient dose while maintaining low-contrast detectability and registration accuracy. Methods: Catphan and anthropomorphic phantoms were analyzed. All CBCT images were acquired with varying dose levels and reconstructed with a Feldkamp–Davis–Kress algorithm-based CBCT (FDK-CBCT) and iCBCT. The low-contrast detectability was subjectively assessed using a 9-point scale by 4 reviewers and objectively assessed using structure similarity index (SSIM). The soft tissue-based registration error was analyzed for each dose level and reconstruction technique. Results: The results of subjective low-contrast detectability found that the iCBCT acquired at two-thirds of a dose was superior to the FDK-CBCT acquired at a full dose (6.4 vs 5.4). Relative to FDK-CBCT acquired at full dose, SSIM was higher for iCBCT acquired at one-sixth dose in head and head and neck region while equivalent with iCBCT acquired at two-thirds dose in pelvis region. The soft tissue-based registration was 2.2 and 0.6 mm for FDK-CBCT and iCBCT, respectively. Conclusion: Use of iCBCT reconstruction algorithm can generally reduce the patient dose by approximately two-thirds compared to conventional reconstruction methods while maintaining low-contrast detectability and accuracy of registration.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
ISSN:1533-0346
1533-0338
1533-0338
DOI:10.1177/15330338211067312