分光光度计法测量氨基聚醚的专属性
氨基聚醚(K2.2.2)含量是18F-FDG质控中的关键指标,比较了两种常用测量K2.2.2方法的专属性。分别采用分光光度计和半定量TLC碘显色法测量了14个样品,其中9个阴性样品、2个阳性样品和3个供试品,并与LC-MS/MS测量对比。结果显示:9个阴性样品经分光光度计法测量均为阳性,K2.2.2的测量结果在6.7-470.0μg/mL;2个阳性样品结果偏高(53,73μg/mL),3个供试品的K2.2.2含量在14.3-19.2μg/mL;半定量TLC碘显色法测量9个阴性样品结果为阴性,2个阳性样品半定量结果与实际一致,3个供试品的K2.2.2含量低于10μg/mL;LC-MS/MS法测量...
        Saved in:
      
    
          | Published in | 同位素 Vol. 28; no. 3; pp. 129 - 134 | 
|---|---|
| Main Author | |
| Format | Journal Article | 
| Language | Chinese | 
| Published | 
            解放军总医院 核医学科,北京,100853%总后勤部卫生部药品器材局,北京,100842%解放军总后药品检验所,北京,100071%北京肿瘤医院 核医学科,北京,100036
    
        2015
     | 
| Subjects | |
| Online Access | Get full text | 
| ISSN | 1000-7512 | 
| DOI | 10.7538/tws.2015.28.03.0129 | 
Cover
| Summary: | 氨基聚醚(K2.2.2)含量是18F-FDG质控中的关键指标,比较了两种常用测量K2.2.2方法的专属性。分别采用分光光度计和半定量TLC碘显色法测量了14个样品,其中9个阴性样品、2个阳性样品和3个供试品,并与LC-MS/MS测量对比。结果显示:9个阴性样品经分光光度计法测量均为阳性,K2.2.2的测量结果在6.7-470.0μg/mL;2个阳性样品结果偏高(53,73μg/mL),3个供试品的K2.2.2含量在14.3-19.2μg/mL;半定量TLC碘显色法测量9个阴性样品结果为阴性,2个阳性样品半定量结果与实际一致,3个供试品的K2.2.2含量低于10μg/mL;LC-MS/MS法测量的14个样品的结果与半定量TLC碘显色法的结果一致。以上结果表明,半定量TLC碘显色法测量K2.2.2的专属性较好,适用于测量18F-FDG溶液中K2.2.2含量。 | 
|---|---|
| Bibliography: | 11-2566/TL ZHANG Jin-ming , CHEN Zheng-yu , WU Xiang-feng , ZHANG Xiao-jun, ZHU Hua, ZHOU Ming, YANG Zhi4(1. Dept of Nucl Med , The PLA General Hops pital , Beij ing 100853, China ;2. Drug and Medical Equipment Administration Bureau of Health Department, Beij ing 100842, China 3. Drug and Medical Equipment Inspection Instritue of Health Department, Beijing 100071, China; 4. Dept of Nucl IVied, Peking University Cancer Hospital, Beijing 100036, China) The The spec 14 s The residue of K2.2.2 was one of the key metric in ^18F-FDG quality control. ificity of the spectrophotometry for K2.2.2 was compared with others. Methods: amples were determined for the K2.2.2 by the spectrophotometry and iodinevapor TLC stain method, respectively. The results were compared with LC-MS/MS. The spectrophotometry showed that the 9 of negative samples were false positive , 2 of positive samples were higher than its actual K2.2.2 contents, 3 of ^18F-FDG injections contained 14.3-19.2 μg/mL for K2.2.2. The iodine-vapor TLC stain method  | 
| ISSN: | 1000-7512 | 
| DOI: | 10.7538/tws.2015.28.03.0129 |