Model predictive filtering MR thermometry: Effects of model inaccuracies, k-space reduction factor, and temperature increase rate
Purpose Evaluate effects of model parameter inaccuracies (thermal conductivity, k, and ultrasound power deposition density, Q), k‐space reduction factor (R), and rate of temperature increase ( T˙) in a thermal model‐based reconstruction for MR‐thermometry during focused‐ultrasound heating. Methods S...
Saved in:
| Published in | Magnetic resonance in medicine Vol. 75; no. 1; pp. 207 - 216 |
|---|---|
| Main Authors | , , , , |
| Format | Journal Article |
| Language | English |
| Published |
United States
Blackwell Publishing Ltd
01.01.2016
Wiley Subscription Services, Inc |
| Subjects | |
| Online Access | Get full text |
| ISSN | 0740-3194 1522-2594 1522-2594 |
| DOI | 10.1002/mrm.25622 |
Cover
| Summary: | Purpose
Evaluate effects of model parameter inaccuracies (thermal conductivity, k, and ultrasound power deposition density, Q), k‐space reduction factor (R), and rate of temperature increase (
T˙) in a thermal model‐based reconstruction for MR‐thermometry during focused‐ultrasound heating.
Methods
Simulations and ex vivo experiments were performed to investigate the accuracy of the thermal model and the model predictive filtering (MPF) algorithm for varying R and
T˙, and their sensitivity to errors in k and Q. Ex vivo data was acquired with a segmented EPI pulse sequence to achieve large field‐of‐view (192 × 162 × 96 mm) four‐dimensional temperature maps with high spatiotemporal resolution (1.5 × 1.5 × 2.0 mm, 1.7 s).
Results
In the simulations, 50% errors in k and Q resulted in maximum temperature root mean square errors (RMSE) of 6°C for model only and 3°C for MPF. Using recently developed methods, estimates of k and Q were accurate to within 3%. The RMSE between MPF and true temperature increased with R and
T˙. In the ex vivo study the RMSE remained below 0.7°C for R ranging from 4 to 12 and
T˙ of 0.28–0.75°C/s.
Conclusion
Errors in MPF temperatures occur due to errors in k and Q. These MPF temperature errors increase with increase in R and
T˙, but are smaller than those obtained using the thermal model alone. Magn Reson Med 75:207–216, 2016. © 2015 Wiley Periodicals, Inc. |
|---|---|
| Bibliography: | The Ben B. and Iris M. Margolis Foundation ark:/67375/WNG-95SQPB4B-L NIH - No. F32 EB012917-02; No. R01s EB013433; No. CA134599; No. CA172787 Siemens Healthcare The Focused Ultrasound Surgery Foundation istex:CDB843CCBFAF75D44F99A6CE1AA6936D73D767D5 ArticleID:MRM25622 ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
| ISSN: | 0740-3194 1522-2594 1522-2594 |
| DOI: | 10.1002/mrm.25622 |