Evaluation of von Willebrand factor with a fully magnetically levitated centrifugal continuous-flow left ventricular assist device in advanced heart failure

Contemporary continuous-flow left ventricular assist devices (CF-LVADs) are associated with degradation of von Willebrand factor (vWF) high-molecular-weight multimers (HMWMs), a critical factor supporting platelet function. We hypothesized that the HeartMate 3 fully magnetically levitated LVAD, desi...

Full description

Saved in:
Bibliographic Details
Published inThe Journal of heart and lung transplantation Vol. 35; no. 7; pp. 860 - 867
Main Authors Netuka, Ivan, Kvasnička, Tomáš, Kvasnička, Jan, Hrachovinová, Ingrid, Ivák, Peter, Mareček, František, Bílková, Jana, Malíková, Ivana, Jančová, Michaela, Malý, Jiří, Sood, Poornima, Sundareswaran, Kartik S., Connors, Jean M., Mehra, Mandeep R.
Format Journal Article
LanguageEnglish
Published United States Elsevier Inc 01.07.2016
Subjects
Online AccessGet full text
ISSN1053-2498
1557-3117
1557-3117
DOI10.1016/j.healun.2016.05.019

Cover

More Information
Summary:Contemporary continuous-flow left ventricular assist devices (CF-LVADs) are associated with degradation of von Willebrand factor (vWF) high-molecular-weight multimers (HMWMs), a critical factor supporting platelet function. We hypothesized that the HeartMate 3 fully magnetically levitated LVAD, designed to reduce circulatory shear stress, favorably influences these hemostatic parameters. Fifteen consecutive HeartMate 3 LVAD patients were compared with 11 consecutive HeartMate II controls. Serial plasma samples were collected pre-implant and on Days 2, 7, 30 and 45 post-operatively. Changes in vWF HMWMs were evaluated by 2 independent, study-blind hematologists and confirmed using densitometry-based computerized software. Ristocetin cofactor (RiCO) and vWF antigen (vWF Ag) were measured using standard protocols with enzyme-linked immunosorbent assay. HeartMate 3 patients and HeartMate II controls had a mean age of 67.3 ± 1.4 and 52.8 ± 2.5 years, respectively (INTERMACS Profiles 2 to 4 in 93.3% and 91%, respectively). HeartMate 3 group demonstrated a significantly greater preservation of HMWMs compared with the HeartMate II group, with the most prominent decrease occurring by Day 2 post-operatively and sustained through 45 days (71.94% vs 31.16%, p = 0.001). Laboratory values (normalized to baseline) for RiCO activity, vWF Ag and RiCO:vWF Ag ratio remained in the functional range with no statistically significant differences observed between groups. The HeartMate 3 LVAD is associated with enhanced hemocompatibility compared with the HeartMate II LVAD, as demonstrated by the improved preservation of vWF HMWMs, In contrast, effects on HMWM degradation appeared to be dissociated from functional attributes. Further confirmation of these findings in randomized clinical trials is warranted.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1053-2498
1557-3117
1557-3117
DOI:10.1016/j.healun.2016.05.019