Room temperature magnetoelectric sensor arrays for application of detecting iron profiles in organs

[Display omitted] Noninvasive measurement of liver iron concentration (LIC) is clinically important. Yet, at the present time, it can only be achieved with SQUID technology. However, SQUID based BLS suffers high costs and cumbersome cryogenic requirements that prevent SQUID BLS from being adopted by...

Full description

Saved in:
Bibliographic Details
Published inSensors and actuators. A. Physical. Vol. 311; p. 112064
Main Authors Xi, Hao, Lu, Meng-Chien, Yang, Qing X., Zhang, Q.M.
Format Journal Article
LanguageEnglish
Published Lausanne Elsevier B.V 15.08.2020
Elsevier BV
Subjects
Online AccessGet full text
ISSN0924-4247
1873-3069
DOI10.1016/j.sna.2020.112064

Cover

More Information
Summary:[Display omitted] Noninvasive measurement of liver iron concentration (LIC) is clinically important. Yet, at the present time, it can only be achieved with SQUID technology. However, SQUID based BLS suffers high costs and cumbersome cryogenic requirements that prevent SQUID BLS from being adopted by clinical applications. Recently, we demonstrated that a single channel ME sensor with piezo-single crystals could detect LIC from only 3 ml of mouse liver tissue without any magnetic field shielding. The results demonstrated not only the sensitivity of ME sensor system for LIC but also the feasibility for mapping LIC profiles spatially. This investigation further developed ME sensor arrays, exploiting the compact size and room temperature operation. A Dual-channel 1-D ME sensor array along the vertical, Z-direction, was developed and shown to be sensitive to the skin-liver distance change which can be utilized to calibrate and eliminate the inter-subject variability of the LIC measurement due to skin-liver distance. With phantom having spatially dependent iron concentrations, the 1-D ME sensor array was capable of mapping the one-dimensional profile of the iron concentration in the horizontal X- and Y-directions. The results of the prototype sensor devices show the feasibility of an array ME-sensors for imaging iron profile.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
Q.M.Z. initiated the study, along with Q.X.Y., Q.M.Z. directed the study. H.X., and M.-C.L carried out the study. Q.M.Z., H.X. and Q.X.Y. wrote the manuscript with comments from all authors.
Author_Statements
ISSN:0924-4247
1873-3069
DOI:10.1016/j.sna.2020.112064