The Successful Measurement of Clitoral Pulse Amplitude Using a New Clitoral Photoplethysmograph: A Pilot Study

Although the clitoris is more sensitive to stimulation and its innervation more conducive to sensory feedback than the vagina, the field of sexual psychophysiology, which uses psychophysiological methods including genital response measures to study sexual arousal, relies heavily on the measurement o...

Full description

Saved in:
Bibliographic Details
Published inJournal of sexual medicine Vol. 17; no. 6; pp. 1118 - 1125
Main Authors Mechelmans, Daisy J., Sachtler, Wendelin L., von Wiegand, Thomas E., Goodrich, David, Heiman, Julia R., Janssen, Erick
Format Journal Article
LanguageEnglish
Published Netherlands Elsevier Inc 01.06.2020
Subjects
Online AccessGet full text
ISSN1743-6095
1743-6109
1743-6109
DOI10.1016/j.jsxm.2020.02.017

Cover

More Information
Summary:Although the clitoris is more sensitive to stimulation and its innervation more conducive to sensory feedback than the vagina, the field of sexual psychophysiology, which uses psychophysiological methods including genital response measures to study sexual arousal, relies heavily on the measurement of vaginal, rather than clitoral, pulse amplitude. To develop and test a new clitoral photoplethysmograph for the measurement of clitoral pulse amplitude (CPA). 2 versions of the new device, which differed in the orientation of the sensor and light source (parallel vs angled), were tested in 15 premenopausal, sexually functional women. Vaginal pulse amplitude (VPA) was assessed simultaneously. The women viewed a 3-minute erotic and an anxiety-inducing film segment with each clitoral sensor, interspersed with neutral video excerpts. In addition, they were asked to indicate their subjective level of sexual arousal during and after erotic video presentations. The main outcome measures are CPA, VPA, and subjective sexual arousal. The clitoral photoplethysmograph successfully detected CPA. The quality of the signals was best for the angled sensor. Main effects of the film and the interaction between the film and epoch were stronger for the clitoral than for the vaginal device. In addition, CPA followed more closely changes in intensity of sexual films than VPA. Within- and between-subject correlations between genital response and subjective sexual arousal were higher for the clitoral than for the vaginal device. Comparison of CPA with other genital blood flow measures in clinical samples is indicated and may contribute to improved physiological assessments of sexual response in women. Our sample was small and consisted of healthy volunteers. Future research could examine test-retest reliability, by including multiple recording sessions, and further explore the specificity of CPA by comparing sexual and non-sexual stimuli with positive valence. This study presents the first instrument to successfully measure CPA. Mechelmans DJ, Sachtler WL, von Wiegand TE, et al. The Successful Measurement of Clitoral Pulse Amplitude Using a New Clitoral Photoplethysmograph: A Pilot Study. J Sex Med 2020;17:1118–1125.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
STATEMENT OF AUTHORSHIP
Category 3
Category 2
(a) Conception and Design
(b) Revising It for Intellectual Content
Daisy J. Mechelmans; Wendelin. L. Sachtler; Thomas E. von Wiegand; David Goodrich; Julia R. Heiman; Erick Janssen
(a) Final Approval of the Completed Article
Wendelin. L. Sachtler; Thomas E. von Wiegand; David Goodrich; Julia R. Heiman; Erick Janssen
David Goodrich; Erick Janssen
Category 1
(a) Drafting the Article
(b) Acquisition of Data
Daisy J. Mechelmans; Erick Janssen
(c) Analysis and Interpretation of Data
ISSN:1743-6095
1743-6109
1743-6109
DOI:10.1016/j.jsxm.2020.02.017