Validation for tremor quantification of an electromagnetic tracking device

An electromagnetic tracking system was used to record arm motion in subjects with Parkinson's disease (n = 23), essential tremor (n = 28) or without neurological disease (n = 4). Tremor magnitude was calculated by averaging the three‐dimensional displacement of individual tremor bursts. Tremor...

Full description

Saved in:
Bibliographic Details
Published inMovement disorders Vol. 16; no. 2; pp. 265 - 271
Main Authors O'Suilleabhain, Padraig E., Dewey Jr, Richard B.
Format Journal Article
LanguageEnglish
Published New York John Wiley & Sons, Inc 01.03.2001
Wiley
Subjects
Online AccessGet full text
ISSN0885-3185
1531-8257
DOI10.1002/mds.1064

Cover

More Information
Summary:An electromagnetic tracking system was used to record arm motion in subjects with Parkinson's disease (n = 23), essential tremor (n = 28) or without neurological disease (n = 4). Tremor magnitude was calculated by averaging the three‐dimensional displacement of individual tremor bursts. Tremor magnitude calculated in this manner was quite closely correlated with a clinician's estimate (r = 0.88 and 0.86 for Parkinsonian and essential tremors, respectively) and was reproducible (r = 0.93 for repeated recordings). The accuracy of the device and algorithm was confirmed by mechanically generating oscillations of known magnitudes and frequencies. This device is adaptable for quantifying different types of tremors and its accuracy is easy to verify. Because position rather than acceleration is tracked, tremor amplitude can be stated in readily comprehensible units. © 2001 Movement Disorder Society.
Bibliography:ArticleID:MDS1064
ark:/67375/WNG-5GZ3BLLF-6
istex:79A712FF1F4170B3DB8BC3ACA8EFD437DAA22156
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ObjectType-Article-2
ObjectType-Undefined-1
ObjectType-Feature-3
ObjectType-Feature-1
ISSN:0885-3185
1531-8257
DOI:10.1002/mds.1064