Evaluation of the VIDAS Anti-HCV Assay for Detection of Hepatitis C Virus Infection

Anti-hepatitis C virus antibody (anti-HCV) assays are recommended for screening HCV-infected persons. The VIDAS Anti-HCV Assay (bioMérieux, France), based on the enzyme-linked fluorescence test principle, was recently introduced in Korea. We evaluated the clinical performance of the VIDAS assay. One...

Full description

Saved in:
Bibliographic Details
Published inAnnals of laboratory medicine Vol. 36; no. 6; pp. 550 - 554
Main Authors Hyun, Jungwon, Ko, Dae-Hyun, Kang, Hee Jung, Whang, Dong Hee, Cha, Young Joo, Kim, Hyun Soo
Format Journal Article
LanguageEnglish
Published Korea (South) The Korean Society for Laboratory Medicine 01.11.2016
대한진단검사의학회
Subjects
Online AccessGet full text
ISSN2234-3806
2234-3814
2234-3814
DOI10.3343/alm.2016.36.6.550

Cover

More Information
Summary:Anti-hepatitis C virus antibody (anti-HCV) assays are recommended for screening HCV-infected persons. The VIDAS Anti-HCV Assay (bioMérieux, France), based on the enzyme-linked fluorescence test principle, was recently introduced in Korea. We evaluated the clinical performance of the VIDAS assay. One hundred HCV-positive and 1,002 HCV-negative blood samples confirmed by Architect anti-HCV (Abbott Laboratories, USA) and COBAS TaqMan HCV real-time PCR (Roche Diagnostics, USA) or the Procleix Ultrio Plus Assay (Gen-Probe Incorporated, USA) were obtained from the Human Serum Bank (HSB) and tested by VIDAS. In case of discrepant results, we conducted a recombinant immunoblot assay (RIBA). The agreement rates for known HCV-positive and HCV-negative samples between the VIDAS assay and the HSB testing were 100% (95% confidence interval [CI]: 96.4-100%) and 99.5% (95% CI: 98.8-99.8%), respectively. One of the five discrepant samples was positive for Core 2+ and NS3-2 2+ reactivity, two samples were negative, and the other two were indeterminate regarding NS4 2+ reactivity in RIBA. We observed a significant but weak positive correlation between the titers of VIDAS and Architect assays (r=0.315, P<0.001). The VIDAS anti-HCV assay, developed on the VIDAS automated immunoassay platform based on the ready-to-use, single-sample test concept may be useful in small-to-medium-sized laboratories. It showed good agreement with Architect anti-HCV and COBAS PCR assays and is therefore useful for detection of HCV infection. Weakly test-positive (ambiguous) samples require additional testing by another anti-HCV, RIBA, or HCV RNA assay.
Bibliography:ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Undefined-1
ObjectType-Feature-3
content type line 23
G704-000327.2016.36.6.012
ISSN:2234-3806
2234-3814
2234-3814
DOI:10.3343/alm.2016.36.6.550