Verification of Qfix Encompass™ couch modeling using the Acuros XB algorithm and HypeArc™ using a high-spatial-resolution two-dimensional diode array

We modeled the Qfix Encompass™ immobilization system and further verified the calculated dose distribution of the AcurosXB (AXB) dose calculation algorithm using SRS MapCHECK (SRSMC) in the HyperArc™ (HA) clinical plan. An Encompass system with a StereoPHAN™ QA phantom was scanned by SOMATOM go.Sim...

Full description

Saved in:
Bibliographic Details
Published inMedical dosimetry : official journal of the American Association of Medical Dosimetrists
Main Authors Yamada, Takumi, Nakano, Hisashi, Tanabe, Satoshi, Sakai, Tatsuya, Tanabe, Shunpei, Oka, Tetsuya, Sakai, Hironori, Oshikane, Tomoya, Nakano, Toshimichi, Ohta, Atsushi, Kanazawa, Tsutomu, Kaidu, Motoki, Ishikawa, Hiroyuki
Format Journal Article
LanguageEnglish
Published United States 01.01.2023
Subjects
Online AccessGet more information
ISSN1873-4022
DOI10.1016/j.meddos.2023.06.002

Cover

More Information
Summary:We modeled the Qfix Encompass™ immobilization system and further verified the calculated dose distribution of the AcurosXB (AXB) dose calculation algorithm using SRS MapCHECK (SRSMC) in the HyperArc™ (HA) clinical plan. An Encompass system with a StereoPHAN™ QA phantom was scanned by SOMATOM go.Sim and imported to an Eclipse™ treatment planning system to create a treatment plan for Encompass modeling. The Encompass modeling was performed in the StereoPHAN with a pinpoint ion chamber for 6 MV and 6 MV flattening filter free (6 MV FFF), and 2 × 2 cm , 4 × 4 cm , and 6 × 6 cm irradiation field sizes. The dose calculation algorithm used was AXB ver. 15.5 with a 1.0 mm calculation grid size. The Hounsfield unit (HU) values of the Encompass modeling were set to 400, -100, -200, and -300 for Encompass, and -400, -600, -700, and -800 for the Encompass base. We evaluated the dose distribution after Encompass modeling by SRSMC using gamma analysis in 12 patients. We adopted HU values of -200 for Encompass, -800 for Encompass base for 6 MV, and -200 for Encompass and -700 for Encompass. Base for 6 MV FFF was adopted as the HU values for the Encompass modeling based on the measurement results. The proposed Encompass modeling resulted in a mean pass rate evaluation >98% for both 6 MV and 6 MV FFF when the 1%/1 mm criterion was used, demonstrating that the proposed HU value can be adopted to calculate more accurate dose distributions.