Real-time adaptive sequential design for optimal acquisition of arterial spin labeling MRI data

An optimal sampling schedule strategy based on the Fisher information matrix and the D‐optimality criterion has previously been proposed as a formal framework for optimizing inversion time scheduling for multi‐inversion‐time arterial spin labeling experiments. Optimal sampling schedule possesses the...

Full description

Saved in:
Bibliographic Details
Published inMagnetic resonance in medicine Vol. 64; no. 1; pp. 203 - 210
Main Authors Xie, Jingyi, Clare, Stuart, Gallichan, Daniel, Gunn, Roger N., Jezzard, Peter
Format Journal Article
LanguageEnglish
Published Hoboken Wiley Subscription Services, Inc., A Wiley Company 01.07.2010
Subjects
Online AccessGet full text
ISSN0740-3194
1522-2594
1522-2594
DOI10.1002/mrm.22398

Cover

More Information
Summary:An optimal sampling schedule strategy based on the Fisher information matrix and the D‐optimality criterion has previously been proposed as a formal framework for optimizing inversion time scheduling for multi‐inversion‐time arterial spin labeling experiments. Optimal sampling schedule possesses the primary advantage of improving parameter estimation precision but requires a priori estimation of plausible parameter distributions that may not be available in all situations. An adaptive sequential design approach addresses this issue by incorporating the optimal sampling schedule strategy into an adaptive process that iteratively updates the parameter estimates and adjusts the optimal sampling schedule accordingly as data are acquired. In this study, the adaptive sequential design method was experimentally implemented with a real‐time feedback scheme on a clinical MRI scanner and was tested in six normal volunteers. Adapted schedules were found to accommodate the intrinsically prolonged arterial transit times in the occipital lobe of the brain. Simulation of applying the adaptive sequential design approach on subjects with pathologically reduced perfusion was also implemented. Simulation results show that the adaptive sequential design approach is capable of incorporating pathologic parameter information into an optimal arterial spin labeling scheduling design within a clinically useful experimental time. Magn Reson Med, 2010. © 2010 Wiley‐Liss, Inc.
Bibliography:Medical Research Council for additional research
ArticleID:MRM22398
M: male. F: female. yr: year.
istex:96BF3784FE7A7608DAAA477AA902B937B7B5BC70
ark:/67375/WNG-275RRR0L-S
Dr. Xie and Dr. Clare share joint first authorship for this paper.
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ObjectType-Article-2
ObjectType-Feature-1
ISSN:0740-3194
1522-2594
1522-2594
DOI:10.1002/mrm.22398