CEST MR‐Fingerprinting: Practical considerations and insights for acquisition schedule design and improved reconstruction
Purpose To understand the influence of various acquisition parameters on the ability of CEST MR‐Fingerprinting (MRF) to discriminate different chemical exchange parameters and to provide tools for optimal acquisition schedule design and parameter map reconstruction. Methods Numerical simulations wer...
Saved in:
| Published in | Magnetic resonance in medicine Vol. 83; no. 2; pp. 462 - 478 |
|---|---|
| Main Authors | , , , , , |
| Format | Journal Article |
| Language | English |
| Published |
United States
Wiley Subscription Services, Inc
01.02.2020
|
| Subjects | |
| Online Access | Get full text |
| ISSN | 0740-3194 1522-2594 1522-2594 |
| DOI | 10.1002/mrm.27937 |
Cover
| Summary: | Purpose
To understand the influence of various acquisition parameters on the ability of CEST MR‐Fingerprinting (MRF) to discriminate different chemical exchange parameters and to provide tools for optimal acquisition schedule design and parameter map reconstruction.
Methods
Numerical simulations were conducted using a parallel computing implementation of the Bloch‐McConnell equations, examining the effect of TR, TE, flip‐angle, water T1 and T2, saturation‐pulse duration, power, and frequency on the discrimination ability of CEST‐MRF. A modified Euclidean distance matching metric was evaluated and compared to traditional dot product matching. L‐Arginine phantoms of various concentrations and pH were scanned at 4.7T and the results compared to numerical findings.
Results
Simulations for dot product matching demonstrated that the optimal flip‐angle and saturation times are 30∘ and 1100 ms, respectively. The optimal maximal saturation power was 3.4 μT for concentrated solutes with a slow exchange rate, and 5.2 μT for dilute solutes with medium‐to‐fast exchange rates. Using the Euclidean distance matching metric, much lower maximum saturation powers were required (1.6 and 2.4 μT, respectively), with a slightly longer saturation time (1500 ms) and 90∘ flip‐angle. For both matching metrics, the discrimination ability increased with the repetition time. The experimental results were in agreement with simulations, demonstrating that more than a 50% reduction in scan‐time can be achieved by Euclidean distance‐based matching.
Conclusions
Optimization of the CEST‐MRF acquisition schedule is critical for obtaining the best exchange parameter accuracy. The use of Euclidean distance‐based matching of signal trajectories simultaneously improved the discrimination ability and reduced the scan time and maximal saturation power required. |
|---|---|
| Bibliography: | Christian T. Farrar and Matthew S. Rosen contributed equally to this work. ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
| ISSN: | 0740-3194 1522-2594 1522-2594 |
| DOI: | 10.1002/mrm.27937 |