Calibration System for Zener Voltage Standards Using a 10 V Programmable Josephson Voltage Standard at NMIJ

We have developed a calibration system for the Zener voltage standards using a programmable Josephson voltage standard (PJVS) for liquid-helium consumption reduction and improvement in the calibration performance. The PJVS array is composed of 524288 niobium nitride (NbN)-based overdamped Josephson...

Full description

Saved in:
Bibliographic Details
Published inIEEE transactions on instrumentation and measurement Vol. 64; no. 6; pp. 1606 - 1612
Main Authors Maruyama, Michitaka, Iwasa, Akio, Yamamori, Hirotake, Shih-Fang Chen, Urano, Chiharu, Kaneko, Nobu-hisa
Format Journal Article
LanguageEnglish
Published IEEE 01.06.2015
Subjects
Online AccessGet full text
ISSN0018-9456
1557-9662
DOI10.1109/TIM.2015.2411998

Cover

More Information
Summary:We have developed a calibration system for the Zener voltage standards using a programmable Josephson voltage standard (PJVS) for liquid-helium consumption reduction and improvement in the calibration performance. The PJVS array is composed of 524288 niobium nitride (NbN)-based overdamped Josephson junctions and is cooled by a 10-K compact cooler. To verify the validity of the system, indirect and direct comparisons of the PJVS system and our conventional Josephson voltage standard system have been carried out. As a result, an agreement within a relative expanded uncertainty of 8.4 × 10 -10 was obtained for a nominal voltage of 10 V. We also discuss the uncertainty sources in the calibration of the Zener voltage standards using our PJVS system.
ISSN:0018-9456
1557-9662
DOI:10.1109/TIM.2015.2411998