Calibration System for Zener Voltage Standards Using a 10 V Programmable Josephson Voltage Standard at NMIJ
We have developed a calibration system for the Zener voltage standards using a programmable Josephson voltage standard (PJVS) for liquid-helium consumption reduction and improvement in the calibration performance. The PJVS array is composed of 524288 niobium nitride (NbN)-based overdamped Josephson...
        Saved in:
      
    
          | Published in | IEEE transactions on instrumentation and measurement Vol. 64; no. 6; pp. 1606 - 1612 | 
|---|---|
| Main Authors | , , , , , | 
| Format | Journal Article | 
| Language | English | 
| Published | 
            IEEE
    
        01.06.2015
     | 
| Subjects | |
| Online Access | Get full text | 
| ISSN | 0018-9456 1557-9662  | 
| DOI | 10.1109/TIM.2015.2411998 | 
Cover
| Summary: | We have developed a calibration system for the Zener voltage standards using a programmable Josephson voltage standard (PJVS) for liquid-helium consumption reduction and improvement in the calibration performance. The PJVS array is composed of 524288 niobium nitride (NbN)-based overdamped Josephson junctions and is cooled by a 10-K compact cooler. To verify the validity of the system, indirect and direct comparisons of the PJVS system and our conventional Josephson voltage standard system have been carried out. As a result, an agreement within a relative expanded uncertainty of 8.4 × 10 -10 was obtained for a nominal voltage of 10 V. We also discuss the uncertainty sources in the calibration of the Zener voltage standards using our PJVS system. | 
|---|---|
| ISSN: | 0018-9456 1557-9662  | 
| DOI: | 10.1109/TIM.2015.2411998 |