A Correction of Measurement for Complex Magnetic Permeability by Direct Contact Probe
A method for obtaining magnetic permeability of thin film by the direct contact probe was developed and the measurement error of permeability was reduced. We analyzed the frequency dependence of a current path inside the magnetic thin film using by finite element method. The magnetic permeability wa...
Saved in:
Published in | Journal of the Magnetics Society of Japan Vol. 46; no. 4 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
Tokyo
Japan Science and Technology Agency
01.07.2022
|
Subjects | |
Online Access | Get full text |
ISSN | 1882-2924 1882-2932 |
DOI | 10.3379/msjmag.2207R002 |
Cover
Summary: | A method for obtaining magnetic permeability of thin film by the direct contact probe was developed and the measurement error of permeability was reduced. We analyzed the frequency dependence of a current path inside the magnetic thin film using by finite element method. The magnetic permeability was corrected in consideration of the frequency dependence of the high-frequency current inside the magnetic thin film. The permeability of CoNbZr film (5 mm x 5 mm, 2 μm in thickness) was optimized by the method, and measured permeability was roughly agreed with the theoretical value up to 4 GHz. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 |