Visualization of Anterior Chamber Angle Structures With Scattering- and Polarization-Sensitive Anterior Segment Optical Coherence Tomography
The purpose of this study was to compare three optical coherence tomography (OCT) modalities in the observation of anterior chamber angle structures; trabecular meshwork (TM), Schlemm's canal (SC), and band of extracanalicular limbal lamina (BELL). Three OCT modalities were used: (1) 2 × 2 Jone...
Saved in:
Published in | Translational vision science & technology Vol. 10; no. 14; p. 29 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
United States
The Association for Research in Vision and Ophthalmology
01.12.2021
|
Subjects | |
Online Access | Get full text |
ISSN | 2164-2591 2164-2591 |
DOI | 10.1167/tvst.10.14.29 |
Cover
Summary: | The purpose of this study was to compare three optical coherence tomography (OCT) modalities in the observation of anterior chamber angle structures; trabecular meshwork (TM), Schlemm's canal (SC), and band of extracanalicular limbal lamina (BELL).
Three OCT modalities were used: (1) 2 × 2 Jones-matrix scattering OCT (S-OCT) representing conventional intensity OCT, (2) polarization-diverse S-OCT that was calculated as summation of all elements of the Jones-matrix to eliminate the influence of artifacts caused by sample birefringence, and (3) polarization-sensitive OCT (PS-OCT) to assess depth-resolved phase retardation.
In a total of 97 eyes of 55 subjects, nasal and temporal angles were scanned. The detection rate of TM and BELL was significantly different among modalities; highest with PS-OCT (95.1% and 99.2%), followed by 2 × 2 Jones-matrix S-OCT (71.1% and 88.7%) and polarization-diverse S-OCT (33.2% and 25.0%), indicating the influence of artifacts on 2 × 2 Jones-matrix S-OCT measurements. SC was visible with 2 × 2 Jones-matrix S-OCT, polarization-diverse S-OCT, and PS-OCT in 14.2%, 14.9%, and 0.3% of images, respectively. The intergrader agreement as evaluated with the prevalence-adjusted bias-adjusted κ value was higher with PS-OCT than with other S-OCTs.
Visibility of anterior chamber angle structures was assessed with three OCT modalities. For TM and BELL that are rich in collagen fibers, PS-OCT provides significantly better visibility than S-OCT without the influence of artifacts arising from polarization or birefringence. Visualization of SC was more difficult with any OCT modalities.
PS-OCT is a useful tool to investigate the anterior chamber angle structures which are difficult to observe with conventional OCT. |
---|---|