Significant Insulin Dose Errors May Occur if Blood Glucose Results are Obtained from Miscoded Meters

The objective of this study was to determine inaccuracies of miscoded blood glucose (BG) meters and potential errors in insulin dose based on values from these meters. Fasting diabetic subjects at three clinical centers participated in a 2-hour meal tolerance test. At various times subjects' bl...

Full description

Saved in:
Bibliographic Details
Published inJournal of diabetes science and technology Vol. 1; no. 2; pp. 205 - 210
Main Authors Raine, Charles H., Schrock, Linda E., Edelman, Steven V., Mudaliar, Sunder Raj D., Zhong, Weiping, Proud, Lois J., Parkes, Joan Lee
Format Journal Article
LanguageEnglish
Published United States Diabetes Technology Society 01.03.2007
Subjects
Online AccessGet full text
ISSN1932-2968
1932-3107
1932-3107
DOI10.1177/193229680700100211

Cover

More Information
Summary:The objective of this study was to determine inaccuracies of miscoded blood glucose (BG) meters and potential errors in insulin dose based on values from these meters. Fasting diabetic subjects at three clinical centers participated in a 2-hour meal tolerance test. At various times subjects' blood was tested on five BG meters and on a Yellow Springs Instruments laboratory glucose analyzer. Some meters were purposely miscoded. Using the BG values from these meters, along with three insulin dose algorithms, Monte Carlo simulations were conducted to generate ideal and simulated-meter glucose values and subsequent probability of insulin dose errors based on normal and empirical distribution assumptions. Maximal median percentage biases of miscoded meters were +29% and -37%, while maximal median percentage biases of correctly coded meters were only +0.64% and -10.45% (p = 0.000, chi(2) test, df = 1). Using the low-dose algorithm and the normal distribution assumption, the combined data showed that the probability of insulin error of +/-1U, +/-2, +/-3, +/-4, and +/-5U for miscoded meters could be as high as 49.6, 50.0, 22.3, 1.4, and 0.04%, respectively. This is compared to manually, correctly coded meters where the probability of error of +/-1, +/-2, and +/-3U could be as high as 44.6, 7.1, and 0.49%, respectively. There was no instance of a +/-4 or +/-5U insulin dose error with a manually, correctly coded meter. For autocoded meters, the probability of +/-1 and +/-2U could be as high as 35.4 and 1.4%, respectively. For autocoded meters there were no calculated insulin dose errors above +/-2U. The probability of insulin misdosing with either manually, correctly coded or autocoded meters was significantly lower than that with miscoded meters. Results using empirical distributions showed similar trends of insulin dose errors. Blood glucose meter coding errors may result in significant insulin dosing errors. To avoid error, patients should be instructed to code their meters correctly or be advised to use an autocoded meter that showed superior performance over manually, correctly coded meters in this study.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
Disclosures: Weiping Zhong is a full-time employee of Bayer HealthCareLLC; Lois J Proud was a full-time employee of Bayer HealthCareLLC until retirement on November 30, 2006; and Joan Lee Parkes is a full-time employee of Bayer HealthCareLLC.
Funding: This work was funded by a grant from Bayer HealthCare LLC
ISSN:1932-2968
1932-3107
1932-3107
DOI:10.1177/193229680700100211