ECG Triggering in Ultra-High Field Cardiovascular MRI
Cardiac magnetic resonance imaging at ultra-high field (B0 ≥ 7 T) potentially provides improved resolution and new opportunities for tissue characterization. Although an accurate synchronization of the acquisition to the cardiac cycle is essential, electrocardiogram (ECG) triggering at ultra-high fi...
Saved in:
| Published in | Tomography (Ann Arbor) Vol. 2; no. 3; pp. 167 - 174 |
|---|---|
| Main Authors | , , , , |
| Format | Journal Article |
| Language | English |
| Published |
United States
Grapho Publications, LLC
01.09.2016
|
| Subjects | |
| Online Access | Get full text |
| ISSN | 2379-139X 2379-1381 2379-139X |
| DOI | 10.18383/j.tom.2016.00193 |
Cover
| Summary: | Cardiac magnetic resonance imaging at ultra-high field (B0 ≥ 7 T) potentially provides improved resolution and new opportunities for tissue characterization. Although an accurate synchronization of the acquisition to the cardiac cycle is essential, electrocardiogram (ECG) triggering at ultra-high field can be significantly impacted by the magnetohydrodynamic (MHD) effect. Blood flow within a static magnetic field induces a voltage, which superimposes the ECG and often affects the recognition of the R-wave. The MHD effect scales with B0 and is particularly pronounced at ultra-high field creating triggering-related image artifacts. Here, we investigated the performance of a conventional 3-lead ECG trigger device and a state-of-the-art trigger algorithm for cardiac ECG synchronization at 7 T. We show that by appropriate subject preparation and by including a learning phase for the R-wave detection outside of the magnetic field, reliable ECG triggering is feasible in healthy subjects at 7 T without additional equipment. Ultra-high field cardiac imaging was performed with the ECG signal and the trigger events recorded in 8 healthy subjects. Despite severe ECG signal distortions, synchronized imaging was successfully performed. Recorded ECG signals, vectorcardiograms, and large consistency in trigger event spacing indicate high accuracy for R-wave detection. |
|---|---|
| Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
| ISSN: | 2379-139X 2379-1381 2379-139X |
| DOI: | 10.18383/j.tom.2016.00193 |