Test-retest reproducibility and validity of the back-leg-chest strength measurements
BACKGROUND: A single measure to characterize overall muscle strength is advantageous because it saves time and costs of evaluation. For this reason, the back-leg-chest (BLC) strength might be an appropriate single measure in characterizing total body strength. OBJECTIVE: To assess the test-retest re...
Saved in:
Published in | Isokinetics and exercise science Vol. 24; no. 3; pp. 209 - 216 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
London, England
SAGE Publications
01.08.2016
|
Subjects | |
Online Access | Get full text |
ISSN | 0959-3020 1878-5913 |
DOI | 10.3233/IES-160619 |
Cover
Summary: | BACKGROUND:
A single measure to characterize overall muscle strength is advantageous because it saves time and costs of evaluation. For this reason, the back-leg-chest (BLC) strength might be an appropriate single measure in characterizing total body strength.
OBJECTIVE:
To assess the test-retest reproducibility and smallest real difference (SRD) of the BLC dynamometer in healthy adults and adolescents and to examine whether handgrip, knee-extensor and knee-flexor strength predict BLC strength in healthy adults.
METHODS:
Forty-five adults and 58 adolescents were recruited. In a first session back-leg-chest strength, handgrip strength, and additionally, in adults, isometric knee-extensor strength, and knee-flexor strength were measured. In a second session, 2-5 days later, BLC strength was measured again for test-retest reproducibility.
RESULTS:
Inter-session correlations of BLC strength were high (all r's and ICC's > 0.92). Bland-Altman-plots showed high agreement. The SRD and SRD% were between 14-26, and 19% and 26% respectively. Strength variables (handgrip, knee-extensor, and knee-flexor strength) explained 87% of the variance in BLC strength. A stepwise linear regression showed that dominant knee extensor and flexor strength were the most important significant predictors of BLC strength (r2 = 0.86).
CONCLUSIONS:
This study demonstrated that the BLC dynamometer has reasonably high test-retest reproducibility and hence may serve in some pertinent situations to be an appropriate tool for clinical, basic and applied research. |
---|---|