Development and characterization of a three-dimensional radiochromic film stack dosimeter for megavoltage photon beam dosimetry
Purpose: Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, curren...
Saved in:
Published in | Medical physics (Lancaster) Vol. 41; no. 5; pp. 052104 - n/a |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
United States
American Association of Physicists in Medicine
01.05.2014
|
Subjects | |
Online Access | Get full text |
ISSN | 0094-2405 2473-4209 2473-4209 |
DOI | 10.1118/1.4871781 |
Cover
Abstract | Purpose:
Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry.
Methods:
A film stack dosimeter was developed using Gafchromic® EBT2 films. The dosimeter consists of 22 films separated by 1 mm-thick spacers. A Virtual Water™ phantom was created that maintains the radial film alignment within a maximum uncertainty of 0.3 mm. The film stack dosimeter was characterized using simulations and measurements of 6 MV fields. The absorbed-dose energy dependence and orientation dependence of the film stack dosimeter were investigated using Monte Carlo simulations. The water equivalence of the dosimeter was determined by comparing percentage-depth-dose (PDD) profiles measured with the film stack dosimeter and simulated using Monte Carlo methods. Film stack dosimeter measurements were verified with thermoluminescent dosimeter (TLD) microcube measurements. The film stack dosimeter was also used to verify the delivery of an intensity-modulated radiation therapy (IMRT) procedure.
Results:
The absorbed-dose energy response of EBT2 film differs less than 1.5% between the calibration and film stack dosimeter geometries for a 6 MV spectrum. Over a series of beam angles ranging from normal incidence to parallel incidence, the overall variation in the response of the film stack dosimeter is within a range of 2.5%. Relative to the response to a normally incident beam, the film stack dosimeter exhibits a 1% under-response when the beam axis is parallel to the film planes. Measured and simulated PDD profiles agree within a root-mean-square difference of 1.3%. In-field film stack dosimeter and TLD measurements agree within 5%, and measurements in the field penumbra agree within 0.5 mm. Film stack dosimeter and TLD measurements have expanded (k = 2) overall measurement uncertainties of 6.2% and 5.8%, respectively. Film stack dosimeter measurements of an IMRT dose distribution have 98% agreement with the treatment planning system dose calculation, using gamma criteria of 3% and 2 mm.
Conclusions:
The film stack dosimeter is capable of high-resolution, low-uncertainty 3D dose measurements, and can be readily incorporated into an existing film dosimetry program. |
---|---|
AbstractList | Purpose: Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry. Methods: A film stack dosimeter was developed using Gafchromic{sup ®} EBT2 films. The dosimeter consists of 22 films separated by 1 mm-thick spacers. A Virtual Water™ phantom was created that maintains the radial film alignment within a maximum uncertainty of 0.3 mm. The film stack dosimeter was characterized using simulations and measurements of 6 MV fields. The absorbed-dose energy dependence and orientation dependence of the film stack dosimeter were investigated using Monte Carlo simulations. The water equivalence of the dosimeter was determined by comparing percentage-depth-dose (PDD) profiles measured with the film stack dosimeter and simulated using Monte Carlo methods. Film stack dosimeter measurements were verified with thermoluminescent dosimeter (TLD) microcube measurements. The film stack dosimeter was also used to verify the delivery of an intensity-modulated radiation therapy (IMRT) procedure. Results: The absorbed-dose energy response of EBT2 film differs less than 1.5% between the calibration and film stack dosimeter geometries for a 6 MV spectrum. Over a series of beam angles ranging from normal incidence to parallel incidence, the overall variation in the response of the film stack dosimeter is within a range of 2.5%. Relative to the response to a normally incident beam, the film stack dosimeter exhibits a 1% under-response when the beam axis is parallel to the film planes. Measured and simulated PDD profiles agree within a root-mean-square difference of 1.3%. In-field film stack dosimeter and TLD measurements agree within 5%, and measurements in the field penumbra agree within 0.5 mm. Film stack dosimeter and TLD measurements have expanded (k = 2) overall measurement uncertainties of 6.2% and 5.8%, respectively. Film stack dosimeter measurements of an IMRT dose distribution have 98% agreement with the treatment planning system dose calculation, using gamma criteria of 3% and 2 mm. Conclusions: The film stack dosimeter is capable of high-resolution, low-uncertainty 3D dose measurements, and can be readily incorporated into an existing film dosimetry program. Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry. A film stack dosimeter was developed using Gafchromic(®) EBT2 films. The dosimeter consists of 22 films separated by 1 mm-thick spacers. A Virtual Water™ phantom was created that maintains the radial film alignment within a maximum uncertainty of 0.3 mm. The film stack dosimeter was characterized using simulations and measurements of 6 MV fields. The absorbed-dose energy dependence and orientation dependence of the film stack dosimeter were investigated using Monte Carlo simulations. The water equivalence of the dosimeter was determined by comparing percentage-depth-dose (PDD) profiles measured with the film stack dosimeter and simulated using Monte Carlo methods. Film stack dosimeter measurements were verified with thermoluminescent dosimeter (TLD) microcube measurements. The film stack dosimeter was also used to verify the delivery of an intensity-modulated radiation therapy (IMRT) procedure. The absorbed-dose energy response of EBT2 film differs less than 1.5% between the calibration and film stack dosimeter geometries for a 6 MV spectrum. Over a series of beam angles ranging from normal incidence to parallel incidence, the overall variation in the response of the film stack dosimeter is within a range of 2.5%. Relative to the response to a normally incident beam, the film stack dosimeter exhibits a 1% under-response when the beam axis is parallel to the film planes. Measured and simulated PDD profiles agree within a root-mean-square difference of 1.3%. In-field film stack dosimeter and TLD measurements agree within 5%, and measurements in the field penumbra agree within 0.5 mm. Film stack dosimeter and TLD measurements have expanded (k = 2) overall measurement uncertainties of 6.2% and 5.8%, respectively. Film stack dosimeter measurements of an IMRT dose distribution have 98% agreement with the treatment planning system dose calculation, using gamma criteria of 3% and 2 mm. The film stack dosimeter is capable of high-resolution, low-uncertainty 3D dose measurements, and can be readily incorporated into an existing film dosimetry program. Purpose: Three‐dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever‐increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry. Methods: A film stack dosimeter was developed using Gafchromic® EBT2 films. The dosimeter consists of 22 films separated by 1 mm‐thick spacers. A Virtual Water™ phantom was created that maintains the radial film alignment within a maximum uncertainty of 0.3 mm. The film stack dosimeter was characterized using simulations and measurements of 6 MV fields. The absorbed‐dose energy dependence and orientation dependence of the film stack dosimeter were investigated using Monte Carlo simulations. The water equivalence of the dosimeter was determined by comparing percentage‐depth‐dose (PDD) profiles measured with the film stack dosimeter and simulated using Monte Carlo methods. Film stack dosimeter measurements were verified with thermoluminescent dosimeter (TLD) microcube measurements. The film stack dosimeter was also used to verify the delivery of an intensity‐modulated radiation therapy (IMRT) procedure. Results: The absorbed‐dose energy response of EBT2 film differs less than 1.5% between the calibration and film stack dosimeter geometries for a 6 MV spectrum. Over a series of beam angles ranging from normal incidence to parallel incidence, the overall variation in the response of the film stack dosimeter is within a range of 2.5%. Relative to the response to a normally incident beam, the film stack dosimeter exhibits a 1% under‐response when the beam axis is parallel to the film planes. Measured and simulated PDD profiles agree within a root‐mean‐square difference of 1.3%. In‐field film stack dosimeter and TLD measurements agree within 5%, and measurements in the field penumbra agree within 0.5 mm. Film stack dosimeter and TLD measurements have expanded (k = 2) overall measurement uncertainties of 6.2% and 5.8%, respectively. Film stack dosimeter measurements of an IMRT dose distribution have 98% agreement with the treatment planning system dose calculation, using gamma criteria of 3% and 2 mm. Conclusions: The film stack dosimeter is capable of high‐resolution, low‐uncertainty 3D dose measurements, and can be readily incorporated into an existing film dosimetry program. Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry.PURPOSEThree-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry.A film stack dosimeter was developed using Gafchromic(®) EBT2 films. The dosimeter consists of 22 films separated by 1 mm-thick spacers. A Virtual Water™ phantom was created that maintains the radial film alignment within a maximum uncertainty of 0.3 mm. The film stack dosimeter was characterized using simulations and measurements of 6 MV fields. The absorbed-dose energy dependence and orientation dependence of the film stack dosimeter were investigated using Monte Carlo simulations. The water equivalence of the dosimeter was determined by comparing percentage-depth-dose (PDD) profiles measured with the film stack dosimeter and simulated using Monte Carlo methods. Film stack dosimeter measurements were verified with thermoluminescent dosimeter (TLD) microcube measurements. The film stack dosimeter was also used to verify the delivery of an intensity-modulated radiation therapy (IMRT) procedure.METHODSA film stack dosimeter was developed using Gafchromic(®) EBT2 films. The dosimeter consists of 22 films separated by 1 mm-thick spacers. A Virtual Water™ phantom was created that maintains the radial film alignment within a maximum uncertainty of 0.3 mm. The film stack dosimeter was characterized using simulations and measurements of 6 MV fields. The absorbed-dose energy dependence and orientation dependence of the film stack dosimeter were investigated using Monte Carlo simulations. The water equivalence of the dosimeter was determined by comparing percentage-depth-dose (PDD) profiles measured with the film stack dosimeter and simulated using Monte Carlo methods. Film stack dosimeter measurements were verified with thermoluminescent dosimeter (TLD) microcube measurements. The film stack dosimeter was also used to verify the delivery of an intensity-modulated radiation therapy (IMRT) procedure.The absorbed-dose energy response of EBT2 film differs less than 1.5% between the calibration and film stack dosimeter geometries for a 6 MV spectrum. Over a series of beam angles ranging from normal incidence to parallel incidence, the overall variation in the response of the film stack dosimeter is within a range of 2.5%. Relative to the response to a normally incident beam, the film stack dosimeter exhibits a 1% under-response when the beam axis is parallel to the film planes. Measured and simulated PDD profiles agree within a root-mean-square difference of 1.3%. In-field film stack dosimeter and TLD measurements agree within 5%, and measurements in the field penumbra agree within 0.5 mm. Film stack dosimeter and TLD measurements have expanded (k = 2) overall measurement uncertainties of 6.2% and 5.8%, respectively. Film stack dosimeter measurements of an IMRT dose distribution have 98% agreement with the treatment planning system dose calculation, using gamma criteria of 3% and 2 mm.RESULTSThe absorbed-dose energy response of EBT2 film differs less than 1.5% between the calibration and film stack dosimeter geometries for a 6 MV spectrum. Over a series of beam angles ranging from normal incidence to parallel incidence, the overall variation in the response of the film stack dosimeter is within a range of 2.5%. Relative to the response to a normally incident beam, the film stack dosimeter exhibits a 1% under-response when the beam axis is parallel to the film planes. Measured and simulated PDD profiles agree within a root-mean-square difference of 1.3%. In-field film stack dosimeter and TLD measurements agree within 5%, and measurements in the field penumbra agree within 0.5 mm. Film stack dosimeter and TLD measurements have expanded (k = 2) overall measurement uncertainties of 6.2% and 5.8%, respectively. Film stack dosimeter measurements of an IMRT dose distribution have 98% agreement with the treatment planning system dose calculation, using gamma criteria of 3% and 2 mm.The film stack dosimeter is capable of high-resolution, low-uncertainty 3D dose measurements, and can be readily incorporated into an existing film dosimetry program.CONCLUSIONSThe film stack dosimeter is capable of high-resolution, low-uncertainty 3D dose measurements, and can be readily incorporated into an existing film dosimetry program. |
Author | DeWerd, Larry A. Micka, John A. McCaw, Travis J. |
Author_xml | – sequence: 1 givenname: Travis J. surname: McCaw fullname: McCaw, Travis J. email: mccaw@wisc.edu organization: Department of Medical Physics, University of Wisconsin, Madison, Wisconsin 53705 – sequence: 2 givenname: John A. surname: Micka fullname: Micka, John A. organization: Department of Medical Physics, University of Wisconsin, Madison, Wisconsin 53705 – sequence: 3 givenname: Larry A. surname: DeWerd fullname: DeWerd, Larry A. organization: Department of Medical Physics, University of Wisconsin, Madison, Wisconsin 53705 |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/24784393$$D View this record in MEDLINE/PubMed https://www.osti.gov/biblio/22250658$$D View this record in Osti.gov |
BookMark | eNp90Utv1DAQAGALFdFt4cAfQJa4IKQUvxI7R9TykorgAGdr1pk0LkkcbO-i5cJfx8tuKyQkTiN5vpmxZs7IyRxmJOQpZxecc_OKXyijuTb8AVkJpWWlBGtPyIqxVlVCsfqUnKV0yxhrZM0ekdOCjJKtXJFfV7jFMSwTzpnC3FE3QASXMfqfkH2Yaegp0DxExKrzhaXyCCON0Pnghhgm72jvx4mmDO4b7UIqqtTTPkQ64Q1sw5jhBukyhFz6rRGmOxV3j8nDHsaET47xnHx9--bL5fvq-tO7D5evr6tbxSWv0GjdSC40MCGAKewR1gq56KRhoFpV96xtsOO1Eb00zVp3IHqteFszVTshz8nzQ9-QsrfJ-YxucGGe0WUrhKhZU5uiXhzUEsP3DaZsJ58cjiPMGDbJ8lpwKY3QbaHPjnSznrCzS_QTxJ29W20B1QH88CPu7vOc2f3NLLfHm9mPn_eh-JcHv__cn9Xf12xD_MsvXf8__M8A-RsF5qaJ |
CODEN | MPHYA6 |
ContentType | Journal Article |
Copyright | American Association of Physicists in Medicine 2014 American Association of Physicists in Medicine |
Copyright_xml | – notice: American Association of Physicists in Medicine – notice: 2014 American Association of Physicists in Medicine |
DBID | CGR CUY CVF ECM EIF NPM 7X8 OTOTI |
DOI | 10.1118/1.4871781 |
DatabaseName | Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed MEDLINE - Academic OSTI.GOV |
DatabaseTitle | MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) MEDLINE - Academic |
DatabaseTitleList | MEDLINE MEDLINE - Academic |
Database_xml | – sequence: 1 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database – sequence: 2 dbid: EIF name: MEDLINE url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search sourceTypes: Index Database |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Medicine Physics |
EISSN | 2473-4209 |
EndPage | n/a |
ExternalDocumentID | 22250658 24784393 MP1781 |
Genre | article Journal Article Comparative Study |
GroupedDBID | --- --Z -DZ .GJ 0R~ 1OB 1OC 29M 2WC 33P 36B 3O- 4.4 476 53G 5GY 5RE 5VS AAHHS AANLZ AAQQT AASGY AAXRX AAZKR ABCUV ABEFU ABFTF ABJNI ABLJU ABQWH ABTAH ABXGK ACAHQ ACBEA ACCFJ ACCZN ACGFO ACGFS ACGOF ACPOU ACSMX ACXBN ACXQS ADBBV ADBTR ADKYN ADOZA ADXAS ADZMN AEEZP AEGXH AEIGN AENEX AEQDE AEUYR AFBPY AFFPM AHBTC AIACR AIAGR AIURR AIWBW AJBDE ALMA_UNASSIGNED_HOLDINGS ALUQN AMYDB ASPBG BFHJK C45 CS3 DCZOG DRFUL DRMAN DRSTM DU5 EBD EBS EJD EMB EMOBN F5P G8K HDBZQ HGLYW I-F KBYEO LATKE LEEKS LOXES LUTES LYRES MEWTI O9- OVD P2P P2W PALCI PHY RJQFR RNS ROL SAMSI SUPJJ SV3 TEORI TN5 TWZ USG WOHZO WXSBR XJT ZGI ZVN ZXP ZY4 ZZTAW AAHQN AAIPD AAMNL AAYCA ABDPE AFWVQ AITYG ALVPJ ADMLS CGR CUY CVF ECM EIF NPM 7X8 AAMMB AEFGJ AEYWJ AGXDD AGYGG AIDQK AIDYY LH4 AAJUZ AAPBV ABCVL ABPTK ADDAD AEUQT OTOTI |
ID | FETCH-LOGICAL-j4131-e87763127a022a04efeab4e12d380a4945f096ed1582f386b7da2f74195045c23 |
ISSN | 0094-2405 2473-4209 |
IngestDate | Fri May 19 00:35:39 EDT 2023 Fri Sep 05 09:54:28 EDT 2025 Wed Feb 19 01:57:00 EST 2025 Wed Jan 22 16:48:22 EST 2025 Fri Jun 21 00:28:34 EDT 2024 Sun Jul 14 10:05:21 EDT 2019 |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 5 |
Keywords | three-dimensional dosimetry radiochromic film stack dosimeter Gafchromic® EBT2 film |
Language | English |
License | 0094-2405/2014/41(5)/052104/13/$30.00 |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-j4131-e87763127a022a04efeab4e12d380a4945f096ed1582f386b7da2f74195045c23 |
Notes | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 |
PMID | 24784393 |
PQID | 1521338279 |
PQPubID | 23479 |
PageCount | 13 |
ParticipantIDs | wiley_primary_10_1118_1_4871781_MP1781 osti_scitechconnect_22250658 proquest_miscellaneous_1521338279 pubmed_primary_24784393 scitation_primary_10_1118_1_4871781 |
PublicationCentury | 2000 |
PublicationDate | May 2014 |
PublicationDateYYYYMMDD | 2014-05-01 |
PublicationDate_xml | – month: 05 year: 2014 text: May 2014 |
PublicationDecade | 2010 |
PublicationPlace | United States |
PublicationPlace_xml | – name: United States |
PublicationTitle | Medical physics (Lancaster) |
PublicationTitleAlternate | Med Phys |
PublicationYear | 2014 |
Publisher | American Association of Physicists in Medicine |
Publisher_xml | – name: American Association of Physicists in Medicine |
References | Saur, Frengen (c62) 2008; 35 Ferreira, Lopes, Capela (c63) 2009; 54 Maryanski, Schulz, Ibbott, Gatenby, Xie, Horton, Gore (c10) 1994; 39 Hazle, Hefner, Nyerick, Wilson, Boyer (c4) 1991; 36 Nelms, Chan, Jarry, Lemire, Lowden, Hampton, Feygelman (c73) 2013; 40 Richter, Pawelke, Karsch, Woithe (c47) 2009; 36 Low, Harms, Mutic, Purdy (c55) 1998; 25 Lindsay, Rink, Ruschin, Jaffray (c70) 2010; 37 Olsson, Petersson, Ahlgren, Mattsson (c3) 1989; 34 Pedersen, Olsen, Skretting (c7) 1997; 42 Oldham, Sakhalkar, Guo, Adamovics (c23) 2008; 35 Doran, Koerkamp, Bero, Jenneson, Morton, Gilboy (c19) 2001; 46 Devic, Aldelaijan, Mohammed, Tomic, Liang, DeBlois, Seuntjens (c67) 2010; 37 Chiu-Tsao, Chan (c43) 2009; 36 Oldham, Kim (c17) 2004; 31 Chiu-Tsao, Duckworth, Patel, Pisch, Harrison (c42) 2004; 31 McEwen, Niven (c58) 2006; 33 De Deene, Pittomvils, Visalatchi (c13) 2007; 52 Thomas, Newton, Adamovics, Oldham (c26) 2011; 38 Maryanski, Gore, Kennan, Schulz (c9) 1993; 11 McCaw, Micka, DeWerd (c37) 2011; 38 Zhen, Nelms, Tomé (c72) 2011; 38 da Rosa, Regulla, Fill (c53) 1999; 50 Andrés, del Castillo, Tortosa, Alonso, Barquero (c35) 2010; 37 Sakhalkar, Adamovics, Ibbott, Oldham (c24) 2009; 36 Rogers, Faddegon, Ding, Ma, We, Mackie (c48) 1995; 22 Sakhalkar, Sterling, Adamovics, Ibbott, Oldham (c25) 2009; 36 Kelly, Jordan, Battista (c18) 1998; 25 Junell, DeWerd (c54) 2009; 36 Mourtada, Soares, Seltzer, Bergstrom, Fernandez-Verea, Asenjo, Lott (c41) 2003; 30 Mersseman, De Wagter (c65) 1998; 43 Cheung, Butson, Yu (c40) 2002; 47 Menegotti, Delana, Martignano (c61) 2008; 35 Thomas, Niebanck, Juang, Wang, Oldham (c27) 2013; 40 Devic, Seuntjens, Hegyi, Podgorsak, Soares, Kirov, Ali, Williamson, Elizondo (c45) 2004; 31 Kairn, Aland, Kenny (c64) 2010; 55 Gore, Ranade, Maryanski, Schulz (c14) 1996; 41 Kim, Yoon, Kim, Shin, Lee, Lim, Kim, Park (c44) 2012; 151 Sutherland, Rogers (c46) 2010; 37 Paelinck, De Neve, De Wagter (c60) 2007; 52 De Deene, De Wagter, Van Duyse, Derycke, De Neve, Achten (c11) 1998; 48 Arjomandy, Tailor, Anand, Sahoo, Gillin, Prado, Vicic (c51) 2010; 37 Niroomand-Rad, Blackwell, Coursey, Gall, Galvin, McLaughlin, Meigooni, Nath, Rodgers, Soares (c32) 1998; 25 Butson, Yu, Cheung, Alnawaf (c69) 2010; 45 De Deene, Hanselaer, De Wagter, Achten, De Neve (c12) 2000; 45 Cheung, Butson, Yu (c38) 2001; 46 Butson, Cheung, Yu (c39) 2001; 61 Schulz, deGuzman, Nguyen, Gore (c6) 1990; 35 Xu, Wuu, Maryanski (c16) 2003; 30 Xu, Wuu, Maryanski (c20) 2004; 31 Ezzell, Burmeister, Dogan, LoSasso, Mechalakos, Mihailidis, Molineu, Palta, Ramsey, Salter, Shi, Xia, Yue, Xiao (c56) 2009; 36 Gore, Kang, Schulz (c1) 1984; 29 Guo, Adamovics, Oldham (c21) 2006; 33 Nelms, Zhen, Tomé (c71) 2011; 38 Micke, Lewis, Yu (c68) 2011; 38 Martišíková, Ackermann, Jäkel (c66) 2008; 53 Rae, Willemse, Lötter, Engelbrecht, Swarts (c8) 1996; 23 Lynch, Kozelka, Ranade, Li, Simon, Dempsey (c33) 2006; 33 Klassen, van der Zwan, Cygler (c57) 1997; 24 Olsson, Fransson, Ericsson, Mattsson (c5) 1990; 35 Devic, Seuntjens, Sham, Podgorsak, Schmidtlein, Kirov, Soares (c34) 2005; 32 Murphy, Piper, Greenwood, Mitch, Lamperti, Seltzer, Bales, Phillips (c49) 2004; 31 Skyt, Balling, Petersen, Yates, Muren (c29) 2011; 38 Richley, John, Coomber, Fletcher (c36) 2010; 55 Devic, Wang, Tomic, Podgorsak (c59) 2006; 33 Doran, Brochard, Adamovics, Krstajic, Bräuer-Krisch (c22) 2010; 55 Almond, Biggs, Coursey, Hanson, Huq, Nath, Rogers (c52) 1999; 26 Appleby, Christman, Leghrouz (c2) 1987; 14 Oldham, Siewerdsen, Shetty, Jaffray (c15) 2001; 28 Baldock, De Deene, Doran, Ibbott, Jirasek, Lepage, McAuley, Oldham, Schreiner (c31) 2010; 55 Suchowerska, Hoban, Butson, Davison, Metcalfe (c50) 2001; 46 Guo, Adamovics, Oldham (c28) 2006; 33 Palmer, Di Pietro, Alobaidli, Issa, Doran, Bradley, Nisbet (c30) 2013; 40 1987; 14 1998; 48 2010; 55 2010; 37 1991; 36 1990; 35 1997; 42 2000; 45 2006; 33 2013; 40 1999; 26 1997; 24 2008; 35 1984; 29 2001; 28 2007; 52 2008; 53 2011; 38 2001; 46 1998; 43 2003; 30 1998; 25 2010; 45 2012; 151 2009; 36 2002; 47 2004; 31 2001; 61 1989; 34 2009; 54 1993; 11 1995; 22 1996; 41 2005; 32 1994; 39 1999; 50 1996; 23 |
References_xml | – volume: 14 start-page: 382 year: 1987 ident: c2 article-title: Imaging of spatial radiation dose distribution in agarose gels using magnetic resonance publication-title: Med. Phys. – volume: 35 start-page: 1611 year: 1990 ident: c6 article-title: Dose-response curves for Fricke-infused agarose gels as obtained by nuclear magnetic resonance publication-title: Phys. Med. Biol. – volume: 36 start-page: 1117 year: 1991 ident: c4 article-title: Dose-response characteristics of a ferrous-sulphate-doped gelatin system for determining radiation absorbed dose distributions by magnetic resonance imaging (Fe MRI) publication-title: Phys. Med. Biol. – volume: 31 start-page: 2392 year: 2004 ident: c45 article-title: Dosimetric properties of improved GafChromic films for seven different digitizers publication-title: Med. Phys. – volume: 31 start-page: 1529 year: 2004 ident: c49 article-title: Evaluation of the new cesium-131 seed for use in low-energy x-ray brachytherapy publication-title: Med. Phys. – volume: 36 start-page: 5506 year: 2009 ident: c47 article-title: Energy dependence of EBT-1 radiochromic film response for photon (10 kVp-15 MVp) and electron beams (6-18 MeV) readout by a flatbed scanner publication-title: Med. Phys. – volume: 45 start-page: 836 year: 2010 ident: c69 article-title: Energy response of the new EBT2 radiochromic film to x-ray radiation publication-title: Radiat. Meas. – volume: 38 start-page: 4846 year: 2011 ident: c26 article-title: Commissioning and benchmarking a 3D dosimetry system for clinical use publication-title: Med. Phys. – volume: 55 start-page: 2601 year: 2010 ident: c36 article-title: Evaluation and optimization of the new EBT2 radiochromic film dosimetry system for patient dose verification in radiotherapy publication-title: Phys. Med. Biol. – volume: 38 start-page: 2523 year: 2011 ident: c68 article-title: Multichannel film dosimetry with nonuniformity correction publication-title: Med. Phys. – volume: 41 start-page: 2695 year: 1996 ident: c14 article-title: Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: I. Development of an optical scanner publication-title: Phys. Med. Biol. – volume: 47 start-page: N31 year: 2002 ident: c40 article-title: Multilayer Gafchromic film detectors for breast skin dose determination publication-title: Phys. Med. Biol. – volume: 46 start-page: 1391 year: 2001 ident: c50 article-title: Directional dependence in film dosimetry: Radiographic and radiochromic film publication-title: Phys. Med. Biol. – volume: 31 start-page: 201 year: 2004 ident: c42 article-title: Verification of Ir-192 near source dosimetry using GAFCHROMIC film publication-title: Med. Phys. – volume: 52 start-page: 231 year: 2007 ident: c60 article-title: Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry publication-title: Phys. Med. Biol. – volume: 22 start-page: 503 year: 1995 ident: c48 article-title: BEAM: A Monte Carlo code to simulate radiotherapy treatment units publication-title: Med. Phys. – volume: 32 start-page: 2245 year: 2005 ident: c34 article-title: Precise radiochromic film dosimetry using a flat-bed document scanner publication-title: Med. Phys. – volume: 33 start-page: 876 year: 2006 ident: c58 article-title: Characterization of the phantom material Virtual Water™ in high-energy photon and electron beams publication-title: Med. Phys. – volume: 33 start-page: 3993 year: 2006 ident: c59 article-title: Sensitivity of linear CCD array based film scanners used for film dosimetry publication-title: Med. Phys. – volume: 55 start-page: 1531 year: 2010 ident: c22 article-title: An investigation of the potential of optical computed tomography for imaging of synchrotron-generated x-rays at high spatial resolution publication-title: Phys. Med. Biol. – volume: 151 start-page: 272 year: 2012 ident: c44 article-title: Three-dimensional radiochromic film dosimetry of proton clinical beams using a Gafchromic EBT2 film array publication-title: Radiat. Prot. Dosim. – volume: 55 start-page: L37 year: 2010 ident: c64 article-title: Local heterogeneities in early batches of EBT2 film: A suggested solution publication-title: Phys. Med. Biol. – volume: 36 start-page: 5359 year: 2009 ident: c56 article-title: IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 publication-title: Med. Phys. – volume: 35 start-page: 2072 year: 2008 ident: c23 article-title: An investigation of the accuracy of an IMRT dose distribution using two- and three-dimensional dosimetry techniques publication-title: Med. Phys. – volume: 35 start-page: 3094 year: 2008 ident: c62 article-title: GafChromic EBT film dosimetry with flatbed CCD scanner: A novel background correction method and full dose uncertainty analysis publication-title: Med. Phys. – volume: 25 start-page: 656 year: 1998 ident: c55 article-title: A technique for the quantitative evaluation of dose distributions publication-title: Med. Phys. – volume: 11 start-page: 253 year: 1993 ident: c9 article-title: NMR relaxation enhancement in gels polymerized and cross-linked by ionizing-radiation: A new approach to 3D dosimetry by MRI publication-title: Magn. Reson. Imaging – volume: 25 start-page: 2093 year: 1998 ident: c32 article-title: Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 publication-title: Med. Phys. – volume: 48 start-page: 283 year: 1998 ident: c11 article-title: Three-dimensional dosimetry using polymer gel and magnetic resonance imaging applied to the verification of conformal radiation therapy in head-and-neck cancer publication-title: Radiother. Oncol. – volume: 26 start-page: 1847 year: 1999 ident: c52 article-title: AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams publication-title: Med. Phys. – volume: 35 start-page: 3078 year: 2008 ident: c61 article-title: Radiochromic film dosimetry with flatbed scanners: A fast and accurate method for dose calibration and uniformity correction with single film exposure publication-title: Med. Phys. – volume: 53 start-page: 7013 year: 2008 ident: c66 article-title: Analysis of uncertainties in Gafchromic EBT film dosimetry of photon beams publication-title: Phys. Med. Biol. – volume: 52 start-page: 2719 year: 2007 ident: c13 article-title: The influence of cooling rate on the accuracy of normoxic polymer gel dosimeters publication-title: Phys. Med. Biol. – volume: 50 start-page: 573 year: 1999 ident: c53 article-title: Reproducibility study of TLD-100 micro-cubes at radiotherapy dose level publication-title: Appl. Radiat. Isot. – volume: 36 start-page: 3371 year: 2009 ident: c25 article-title: Investigation of the feasibility of relative 3D dosimetry in the Radiologic Physics Center Head and Neck IMRT phantom using Presage/optical-CT publication-title: Med. Phys. – volume: 54 start-page: 1073 year: 2009 ident: c63 article-title: Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetry publication-title: Phys. Med. Biol. – volume: 30 start-page: 959 year: 2003 ident: c41 article-title: Dosimetry characterization of a P source wire used for intravascular brachytherapy with automated stepping publication-title: Med. Phys. – volume: 34 start-page: 43 year: 1989 ident: c3 article-title: Ferrous sulphate gels for determination of absorbed dose distributions using MRI technique: Basic studies publication-title: Phys. Med. Biol. – volume: 36 start-page: 2594 year: 2009 ident: c54 article-title: Determination of the energy correction factor for TLD-100 in 6 and 10 MV photon beams relative to Co (abstract) publication-title: Med. Phys. – volume: 29 start-page: 1189 year: 1984 ident: c1 article-title: Measurement of radiation dose distributions by nuclear magnetic resonance (NMR) imaging publication-title: Phys. Med. Biol. – volume: 36 start-page: 71 year: 2009 ident: c24 article-title: A comprehensive evaluation of the PRESAGE/optical-CT 3D dosimetry system publication-title: Med. Phys. – volume: 23 start-page: 15 year: 1996 ident: c8 article-title: Chelator effect on ion diffusion in ferrous-sulfate-doped gelatin gel dosimeters as analyzed by MRI publication-title: Med. Phys. – volume: 45 start-page: 859 year: 2000 ident: c12 article-title: An investigation of the chemical stability of a monomer/polymer gel dosimeter publication-title: Phys. Med. Biol. – volume: 38 start-page: 5771 year: 2011 ident: c37 article-title: Characterizing the marker-dye correction for Gafchromic EBT2 film: A comparison of three analysis methods publication-title: Med. Phys. – volume: 36 start-page: 2074 year: 2009 ident: c43 article-title: Photon beam dosimetry in the superficial buildup region using radiochromic EBT film stack publication-title: Med. Phys. – volume: 43 start-page: 1803 year: 1998 ident: c65 article-title: Characteristics of a commercially available film digitizer and their significance for film dosimetry publication-title: Phys. Med. Biol. – volume: 33 start-page: 1338 year: 2006 ident: c21 article-title: Characterization of a new radiochromic three-dimensional dosimeter publication-title: Med. Phys. – volume: 37 start-page: 1110 year: 2010 ident: c46 article-title: Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film publication-title: Med. Phys. – volume: 39 start-page: 1437 year: 1994 ident: c10 article-title: Magnetic resonance imaging of radiation dose distributions using a polymer-gel dosimeter publication-title: Phys. Med. Biol. – volume: 31 start-page: 1093 year: 2004 ident: c17 article-title: Optical-CT gel-dosimetry. II: Optical artifacts and geometrical distortion publication-title: Med. Phys. – volume: 31 start-page: 3024 year: 2004 ident: c20 article-title: Performance of a commercial optical CT scanner and polymer gel dosimeters for 3-D dose verification publication-title: Med. Phys. – volume: 28 start-page: 1436 year: 2001 ident: c15 article-title: High resolution gel-dosimetry by optical-CT and MR scanning publication-title: Med. Phys. – volume: 40 start-page: 121725 year: 2013 ident: c27 article-title: A comprehensive investigation of the accuracy and reproducibility of a multitarget single isocenter VMAT radiosurgery technique publication-title: Med. Phys. – volume: 37 start-page: 571 year: 2010 ident: c70 article-title: Investigation of energy dependence of EBT and EBT-2 Gafchromic film publication-title: Med. Phys. – volume: 38 start-page: 5477 year: 2011 ident: c72 article-title: Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA publication-title: Med. Phys. – volume: 42 start-page: 1575 year: 1997 ident: c7 article-title: Measurement of the ferric diffusion coefficient in agarose and gelatine gels by utilization of the evolution of a radiation induced edge as reflected in relaxation rate images publication-title: Phys. Med. Biol. – volume: 25 start-page: 1741 year: 1998 ident: c18 article-title: Optical CT reconstruction of 3D dose distributions using the ferrous-benzoic-xylenol (FBX) gel dosimeter publication-title: Med. Phys. – volume: 33 start-page: 3962 year: 2006 ident: c28 article-title: A practical three-dimensional dosimetry system for radiation therapy publication-title: Med. Phys. – volume: 38 start-page: 1037 year: 2011 ident: c71 article-title: Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors publication-title: Med. Phys. – volume: 35 start-page: 1623 year: 1990 ident: c5 article-title: MR imaging of absorbed dose distributions for radiotherapy using ferrous sulphate gels publication-title: Phys. Med. Biol. – volume: 24 start-page: 1924 year: 1997 ident: c57 article-title: GafChromic MD-55: Investigated as a precision dosimeter publication-title: Med. Phys. – volume: 61 start-page: 211 year: 2001 ident: c39 article-title: Spatial resolution of a stacked radiochromic film dosimeter publication-title: Radiother. Oncol. – volume: 38 start-page: 2806 year: 2011 ident: c29 article-title: Temperature dependence of the dose response for a solid-state radiochromic dosimeter during irradiation and storage publication-title: Med. Phys. – volume: 40 start-page: 061707 year: 2013 ident: c30 article-title: Comparison of methods for the measurement of radiation dose distributions in high dose rate (HDR) brachytherapy: Ge-doped optical fiber, EBT3 Gafchromic film, and PRESAGE radiochromic plastic publication-title: Med. Phys. – volume: 37 start-page: 6271 year: 2010 ident: c35 article-title: A comprehensive study of the Gafchromic EBT2 radiochromic film: A comparison with EBT publication-title: Med. Phys. – volume: 46 start-page: N235 year: 2001 ident: c38 article-title: Use of multiple layers of Gafchromic film to increase sensitivity publication-title: Phys. Med. Biol. – volume: 30 start-page: 2257 year: 2003 ident: c16 article-title: Determining optimal gel sensitivity in optical CT scanning of gel dosimeters publication-title: Med. Phys. – volume: 46 start-page: 3191 year: 2001 ident: c19 article-title: A CCD-based optical CT scanner for high-resolution 3D imaging of radiation dose distributions: Equipment specifications, optical simulations and preliminary results publication-title: Phys. Med. Biol. – volume: 33 start-page: 4551 year: 2006 ident: c33 article-title: Important considerations for radiochromic film dosimetry with flatbed CCD scanners and EBT GAFCHROMIC film publication-title: Med. Phys. – volume: 55 start-page: R1 year: 2010 ident: c31 article-title: Polymer gel dosimetry publication-title: Phys. Med. Biol. – volume: 40 start-page: 111722 year: 2013 ident: c73 article-title: Evaluating IMRT and VMAT dose accuracy: Practical examples of failure to detect systematic errors when applying a commonly used metric and action levels publication-title: Med. Phys. – volume: 37 start-page: 1942 year: 2010 ident: c51 article-title: Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies publication-title: Med. Phys. – volume: 37 start-page: 2207 year: 2010 ident: c67 article-title: Absorption spectra time evolution of EBT-2 model GAFCHROMIC™ film publication-title: Med. Phys. – volume: 33 start-page: 1338 issue: 5 year: 2006 end-page: 1345 article-title: Characterization of a new radiochromic three‐dimensional dosimeter publication-title: Med. Phys. – volume: 33 start-page: 876 issue: 4 year: 2006 end-page: 887 article-title: Characterization of the phantom material Virtual Water™ in high‐energy photon and electron beams publication-title: Med. Phys. – volume: 33 start-page: 3993 issue: 11 year: 2006 end-page: 3996 article-title: Sensitivity of linear CCD array based film scanners used for film dosimetry publication-title: Med. Phys. – volume: 48 start-page: 283 issue: 3 year: 1998 end-page: 291 article-title: Three‐dimensional dosimetry using polymer gel and magnetic resonance imaging applied to the verification of conformal radiation therapy in head‐and‐neck cancer publication-title: Radiother. Oncol. – volume: 33 start-page: 3962 issue: 10 year: 2006 end-page: 3972 article-title: A practical three‐dimensional dosimetry system for radiation therapy publication-title: Med. Phys. – volume: 25 start-page: 656 issue: 5 year: 1998 end-page: 661 article-title: A technique for the quantitative evaluation of dose distributions publication-title: Med. Phys. – volume: 30 start-page: 959 issue: 5 year: 2003 end-page: 971 article-title: Dosimetry characterization of a P source wire used for intravascular brachytherapy with automated stepping publication-title: Med. Phys. – volume: 14 start-page: 382 issue: 3 year: 1987 end-page: 384 article-title: Imaging of spatial radiation dose distribution in agarose gels using magnetic resonance publication-title: Med. Phys. – volume: 25 start-page: 2093 issue: 11 year: 1998 end-page: 2115 article-title: Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 publication-title: Med. Phys. – volume: 40 start-page: 061707 issue: 6 year: 2013 article-title: Comparison of methods for the measurement of radiation dose distributions in high dose rate (HDR) brachytherapy: Ge‐doped optical fiber, EBT3 Gafchromic film, and PRESAGE radiochromic plastic publication-title: Med. Phys. – volume: 36 start-page: 2594 year: 2009 article-title: Determination of the energy correction factor for TLD‐100 in 6 and 10 MV photon beams relative to Co (abstract) publication-title: Med. Phys. – volume: 35 start-page: 1611 issue: 12 year: 1990 end-page: 1622 article-title: Dose‐response curves for Fricke‐infused agarose gels as obtained by nuclear magnetic resonance publication-title: Phys. Med. Biol. – volume: 46 start-page: N235 issue: 10 year: 2001 end-page: N240 article-title: Use of multiple layers of Gafchromic film to increase sensitivity publication-title: Phys. Med. Biol. – volume: 36 start-page: 1117 issue: 8 year: 1991 end-page: 1125 article-title: Dose‐response characteristics of a ferrous‐sulphate‐doped gelatin system for determining radiation absorbed dose distributions by magnetic resonance imaging (Fe MRI) publication-title: Phys. Med. Biol. – volume: 32 start-page: 2245 issue: 7 year: 2005 end-page: 2253 article-title: Precise radiochromic film dosimetry using a flat‐bed document scanner publication-title: Med. Phys. – volume: 39 start-page: 1437 issue: 9 year: 1994 end-page: 1455 article-title: Magnetic resonance imaging of radiation dose distributions using a polymer‐gel dosimeter publication-title: Phys. Med. Biol. – volume: 151 start-page: 272 issue: 2 year: 2012 end-page: 277 article-title: Three‐dimensional radiochromic film dosimetry of proton clinical beams using a Gafchromic EBT2 film array publication-title: Radiat. Prot. Dosim. – volume: 36 start-page: 5506 issue: 12 year: 2009 end-page: 5514 article-title: Energy dependence of EBT‐1 radiochromic film response for photon (10 kVp‐15 MVp) and electron beams (6‐18 MeV) readout by a flatbed scanner publication-title: Med. Phys. – volume: 35 start-page: 3094 issue: 7 year: 2008 end-page: 3101 article-title: GafChromic EBT film dosimetry with flatbed CCD scanner: A novel background correction method and full dose uncertainty analysis publication-title: Med. Phys. – volume: 55 start-page: L37 issue: 15 year: 2010 end-page: L42 article-title: Local heterogeneities in early batches of EBT2 film: A suggested solution publication-title: Phys. Med. Biol. – volume: 53 start-page: 7013 issue: 24 year: 2008 end-page: 7027 article-title: Analysis of uncertainties in Gafchromic EBT film dosimetry of photon beams publication-title: Phys. Med. Biol. – volume: 11 start-page: 253 issue: 2 year: 1993 end-page: 258 article-title: NMR relaxation enhancement in gels polymerized and cross‐linked by ionizing‐radiation: A new approach to 3D dosimetry by MRI publication-title: Magn. Reson. Imaging – volume: 36 start-page: 3371 issue: 7 year: 2009 end-page: 3377 article-title: Investigation of the feasibility of relative 3D dosimetry in the Radiologic Physics Center Head and Neck IMRT phantom using Presage/optical‐CT publication-title: Med. Phys. – volume: 38 start-page: 5477 issue: 10 year: 2011 end-page: 5489 article-title: Moving from gamma passing rates to patient DVH‐based QA metrics in pretreatment dose QA publication-title: Med. Phys. – volume: 37 start-page: 2207 issue: 5 year: 2010 end-page: 2214 article-title: Absorption spectra time evolution of EBT‐2 model GAFCHROMIC™ film publication-title: Med. Phys. – volume: 29 start-page: 1189 issue: 10 year: 1984 end-page: 1197 article-title: Measurement of radiation dose distributions by nuclear magnetic resonance (NMR) imaging publication-title: Phys. Med. Biol. – volume: 35 start-page: 2072 issue: 5 year: 2008 end-page: 2080 article-title: An investigation of the accuracy of an IMRT dose distribution using two‐ and three‐dimensional dosimetry techniques publication-title: Med. Phys. – volume: 40 start-page: 121725 issue: 12 year: 2013 article-title: A comprehensive investigation of the accuracy and reproducibility of a multitarget single isocenter VMAT radiosurgery technique publication-title: Med. Phys. – volume: 45 start-page: 836 issue: 7 year: 2010 end-page: 839 article-title: Energy response of the new EBT2 radiochromic film to x‐ray radiation publication-title: Radiat. Meas. – volume: 55 start-page: 1531 issue: 5 year: 2010 end-page: 1547 article-title: An investigation of the potential of optical computed tomography for imaging of synchrotron‐generated x‐rays at high spatial resolution publication-title: Phys. Med. Biol. – volume: 52 start-page: 231 issue: 1 year: 2007 end-page: 242 article-title: Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry publication-title: Phys. Med. Biol. – volume: 37 start-page: 6271 issue: 12 year: 2010 end-page: 6278 article-title: A comprehensive study of the Gafchromic EBT2 radiochromic film: A comparison with EBT publication-title: Med. Phys. – volume: 31 start-page: 2392 issue: 9 year: 2004 end-page: 2401 article-title: Dosimetric properties of improved GafChromic films for seven different digitizers publication-title: Med. Phys. – volume: 38 start-page: 1037 issue: 2 year: 2011 end-page: 1044 article-title: Per‐beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors publication-title: Med. Phys. – volume: 36 start-page: 2074 issue: 6 year: 2009 end-page: 2083 article-title: Photon beam dosimetry in the superficial buildup region using radiochromic EBT film stack publication-title: Med. Phys. – volume: 33 start-page: 4551 issue: 12 year: 2006 end-page: 4556 article-title: Important considerations for radiochromic film dosimetry with flatbed CCD scanners and EBT GAFCHROMIC film publication-title: Med. Phys. – volume: 47 start-page: N31 issue: 2 year: 2002 end-page: N37 article-title: Multilayer Gafchromic film detectors for breast skin dose determination publication-title: Phys. Med. Biol. – volume: 46 start-page: 3191 issue: 12 year: 2001 end-page: 3213 article-title: A CCD‐based optical CT scanner for high‐resolution 3D imaging of radiation dose distributions: Equipment specifications, optical simulations and preliminary results publication-title: Phys. Med. Biol. – volume: 35 start-page: 1623 issue: 12 year: 1990 end-page: 1631 article-title: MR imaging of absorbed dose distributions for radiotherapy using ferrous sulphate gels publication-title: Phys. Med. Biol. – volume: 36 start-page: 5359 issue: 11 year: 2009 end-page: 5373 article-title: IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 publication-title: Med. Phys. – volume: 43 start-page: 1803 issue: 6 year: 1998 end-page: 1812 article-title: Characteristics of a commercially available film digitizer and their significance for film dosimetry publication-title: Phys. Med. Biol. – volume: 46 start-page: 1391 issue: 5 year: 2001 end-page: 1397 article-title: Directional dependence in film dosimetry: Radiographic and radiochromic film publication-title: Phys. Med. Biol. – volume: 31 start-page: 201 issue: 2 year: 2004 end-page: 207 article-title: Verification of Ir‐192 near source dosimetry using GAFCHROMIC film publication-title: Med. Phys. – volume: 45 start-page: 859 issue: 4 year: 2000 end-page: 878 article-title: An investigation of the chemical stability of a monomer/polymer gel dosimeter publication-title: Phys. Med. Biol. – volume: 31 start-page: 1529 issue: 6 year: 2004 end-page: 1538 article-title: Evaluation of the new cesium‐131 seed for use in low‐energy x‐ray brachytherapy publication-title: Med. Phys. – volume: 61 start-page: 211 issue: 2 year: 2001 end-page: 213 article-title: Spatial resolution of a stacked radiochromic film dosimeter publication-title: Radiother. Oncol. – volume: 50 start-page: 573 issue: 3 year: 1999 end-page: 577 article-title: Reproducibility study of TLD‐100 micro‐cubes at radiotherapy dose level publication-title: Appl. Radiat. Isot. – volume: 28 start-page: 1436 issue: 7 year: 2001 end-page: 1445 article-title: High resolution gel‐dosimetry by optical‐CT and MR scanning publication-title: Med. Phys. – volume: 52 start-page: 2719 issue: 10 year: 2007 end-page: 2728 article-title: The influence of cooling rate on the accuracy of normoxic polymer gel dosimeters publication-title: Phys. Med. Biol. – volume: 35 start-page: 3078 issue: 7 year: 2008 end-page: 3085 article-title: Radiochromic film dosimetry with flatbed scanners: A fast and accurate method for dose calibration and uniformity correction with single film exposure publication-title: Med. Phys. – volume: 37 start-page: 571 issue: 2 year: 2010 end-page: 576 article-title: Investigation of energy dependence of EBT and EBT‐2 Gafchromic film publication-title: Med. Phys. – volume: 30 start-page: 2257 issue: 8 year: 2003 end-page: 2263 article-title: Determining optimal gel sensitivity in optical CT scanning of gel dosimeters publication-title: Med. Phys. – volume: 31 start-page: 3024 issue: 11 year: 2004 end-page: 3033 article-title: Performance of a commercial optical CT scanner and polymer gel dosimeters for 3‐D dose verification publication-title: Med. Phys. – volume: 24 start-page: 1924 issue: 12 year: 1997 end-page: 1934 article-title: GafChromic MD‐55: Investigated as a precision dosimeter publication-title: Med. Phys. – volume: 34 start-page: 43 issue: 1 year: 1989 end-page: 52 article-title: Ferrous sulphate gels for determination of absorbed dose distributions using MRI technique: Basic studies publication-title: Phys. Med. Biol. – volume: 36 start-page: 71 issue: 1 year: 2009 end-page: 82 article-title: A comprehensive evaluation of the PRESAGE/optical‐CT 3D dosimetry system publication-title: Med. Phys. – volume: 38 start-page: 5771 issue: 10 year: 2011 end-page: 5777 article-title: Characterizing the marker‐dye correction for Gafchromic EBT2 film: A comparison of three analysis methods publication-title: Med. Phys. – volume: 31 start-page: 1093 issue: 5 year: 2004 end-page: 1104 article-title: Optical‐CT gel‐dosimetry. II: Optical artifacts and geometrical distortion publication-title: Med. Phys. – volume: 55 start-page: R1 issue: 5 year: 2010 end-page: R63 article-title: Polymer gel dosimetry publication-title: Phys. Med. Biol. – volume: 23 start-page: 15 issue: 1 year: 1996 end-page: 23 article-title: Chelator effect on ion diffusion in ferrous‐sulfate‐doped gelatin gel dosimeters as analyzed by MRI publication-title: Med. Phys. – volume: 37 start-page: 1942 issue: 5 year: 2010 end-page: 1947 article-title: Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies publication-title: Med. Phys. – volume: 22 start-page: 503 issue: 5 year: 1995 end-page: 524 article-title: BEAM: A Monte Carlo code to simulate radiotherapy treatment units publication-title: Med. Phys. – volume: 54 start-page: 1073 issue: 4 year: 2009 end-page: 1085 article-title: Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetry publication-title: Phys. Med. Biol. – volume: 25 start-page: 1741 issue: 9 year: 1998 end-page: 1750 article-title: Optical CT reconstruction of 3D dose distributions using the ferrous‐benzoic‐xylenol (FBX) gel dosimeter publication-title: Med. Phys. – volume: 38 start-page: 2806 issue: 5 year: 2011 end-page: 2811 article-title: Temperature dependence of the dose response for a solid‐state radiochromic dosimeter during irradiation and storage publication-title: Med. Phys. – volume: 41 start-page: 2695 issue: 12 year: 1996 end-page: 2704 article-title: Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: I. Development of an optical scanner publication-title: Phys. Med. Biol. – volume: 42 start-page: 1575 issue: 8 year: 1997 end-page: 1585 article-title: Measurement of the ferric diffusion coefficient in agarose and gelatine gels by utilization of the evolution of a radiation induced edge as reflected in relaxation rate images publication-title: Phys. Med. Biol. – volume: 40 start-page: 111722 issue: 11 year: 2013 article-title: Evaluating IMRT and VMAT dose accuracy: Practical examples of failure to detect systematic errors when applying a commonly used metric and action levels publication-title: Med. Phys. – volume: 26 start-page: 1847 issue: 9 year: 1999 end-page: 1870 article-title: AAPMˈs TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beams publication-title: Med. Phys. – volume: 37 start-page: 1110 issue: 3 year: 2010 end-page: 1116 article-title: Monte Carlo calculated absorbed‐dose energy dependence of EBT and EBT2 film publication-title: Med. Phys. – volume: 38 start-page: 4846 issue: 8 year: 2011 end-page: 4857 article-title: Commissioning and benchmarking a 3D dosimetry system for clinical use publication-title: Med. Phys. – volume: 38 start-page: 2523 issue: 5 year: 2011 end-page: 2534 article-title: Multichannel film dosimetry with nonuniformity correction publication-title: Med. Phys. – volume: 55 start-page: 2601 issue: 9 year: 2010 end-page: 2617 article-title: Evaluation and optimization of the new EBT2 radiochromic film dosimetry system for patient dose verification in radiotherapy publication-title: Phys. Med. Biol. |
SSID | ssj0006350 |
Score | 2.1677163 |
Snippet | Purpose:
Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with... Purpose: Three‐dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with... Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the... Purpose: Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with... |
SourceID | osti proquest pubmed wiley scitation |
SourceType | Open Access Repository Aggregation Database Index Database Publisher Enrichment Source |
StartPage | 052104 |
SubjectTerms | ABSORBED RADIATION DOSES Algorithms biomedical equipment Calibrating of instruments or apparatus CALIBRATION COMMISSIONING Computer Simulation COMPUTERIZED SIMULATION Conformal radiation treatment DEPTH DOSE DISTRIBUTIONS Dose‐volume analysis dosimetry Dosimetry/exposure assessment ENERGY DEPENDENCE Equipment Design FILM DOSIMETRY Film Dosimetry - instrumentation Film Dosimetry - methods Gafchromic® EBT2 film Measuring radiation intensity MONTE CARLO METHOD Monte Carlo methods PHANTOMS Phantoms, Imaging PHOTON BEAMS Photons radiation therapy radiochromic film stack dosimeter RADIOLOGY AND NUCLEAR MEDICINE RADIOTHERAPY Radiotherapy Dosage Radiotherapy Planning, Computer-Assisted Radiotherapy, Intensity-Modulated - methods Scintigraphy Testing or calibrating of apparatus or arrangements provided for in groups G01D1/00 to G01D15/00 Therapeutic applications THERMOLUMINESCENT DOSEMETERS thermoluminescent dosimeters Thermoluminescent Dosimetry - methods three‐dimensional dosimetry Uncertainty Water |
Title | Development and characterization of a three-dimensional radiochromic film stack dosimeter for megavoltage photon beam dosimetry |
URI | http://dx.doi.org/10.1118/1.4871781 https://onlinelibrary.wiley.com/doi/abs/10.1118%2F1.4871781 https://www.ncbi.nlm.nih.gov/pubmed/24784393 https://www.proquest.com/docview/1521338279 https://www.osti.gov/biblio/22250658 |
Volume | 41 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3db9MwELegEwweEAwGhYGMQLxMKY3jfD1OG2hCC5rEJvYWOf6gZWpTpd0k9sK_zl2cOBl0EvCStrETR_n9er47350JeRuagHMmYs_4ifKw-p2XxgH3eKhkEGnpFyEmJ2efo8NT_uksPOuWC-rsklUxkldr80r-B1U4B7hiluw_IOtuCifgO-ALR0AYjn-FcS_ix-anuerLV04TFLgVj9aewjL-tgTHbiXUtJSTCjOSsTLTDD0K8nxXlUvohWUTMfhwpr8JEF4rjOpZTEqsv1FoMWt7Xc-lbhd8rKekduVierWwG384b0Mm90W9mARz5OV02S1KZVN5LtrY4M7BeqC_aht8fySq6kfT0HgpfN7FBI50Lc0YjwOPs3HaF7225lVDsbAnRzGl2G5LvEbEY9qCPwJLy4_tdi89qBezGmsYLQFVK-hmORd72DbdJhssjiI2IBt7B9nRFzd_gwo2bmpQwVjv3UhYN7q5FubvEiTwOqvkPtkE3cWGUVw3eGqN5eQhedCYGnTP8uYRuaXnW-Ru1gRTbJE7xxarx-Rnj0gUiER_JxItDRX0DyLRPpEoEonWRKKOSBSIRHtEopZIFIlEHZGekNOPH072D71mZw7vOyg9vqeTGOYln8UCVEAx5tpoUXDtMxUkY8FTHhowjbXyw4SZIImKWAlmQHlNQzAhJAu2yWBezvUzQtNIRZqPVZwmEpRZnaZGMPjlmyQwRWyGZAdfdY4vVcuJxPAvucrRTYHa85C8biHIQTDiapeY6_JimaNiGgQJi9MheWqxyRe2gkve4jgkbxxYrtGaxUnu5w3wa3tdllXXI18oeNB3Ndg33yfPjvHj-Y2P84Lc6_47O2Swqi70S9B5V8WrhqS_AFmtrOU |
linkProvider | EBSCOhost |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Development+and+characterization+of+a+three-dimensional+radiochromic+film+stack+dosimeter+for+megavoltage+photon+beam+dosimetry&rft.jtitle=Medical+physics+%28Lancaster%29&rft.au=McCaw%2C+Travis+J&rft.au=Micka%2C+John+A&rft.au=DeWerd%2C+Larry+A&rft.date=2014-05-01&rft.eissn=2473-4209&rft.volume=41&rft.issue=5&rft.spage=052104&rft_id=info:doi/10.1118%2F1.4871781&rft_id=info%3Apmid%2F24784393&rft.externalDocID=24784393 |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0094-2405&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0094-2405&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0094-2405&client=summon |