螺旋流血液ポンプのモーター改良による差圧流量推定精度の向上

Sensor less estimation of pressure and pump flow has been an important subject for developing artificial hearts. However, estimated values are often incorrect influenced by the change in blood viscosity that changes with various factors. Therefore, viscosity estimation is essential for accurate esti...

Full description

Saved in:
Bibliographic Details
Published inTransactions of Japanese Society for Medical and Biological Engineering Vol. 54Annual; no. 26AM-Abstract; p. S55
Main Authors 斎藤, 逸郎, 圦本, 晃海, 原, 伸太郎, 村上, 遥, 磯山, 隆, 阿部, 裕輔, 李, 欣陽, 羽合, 佳範
Format Journal Article
LanguageJapanese
Published 公益社団法人 日本生体医工学会 2016
Japanese Society for Medical and Biological Engineering
Online AccessGet full text
ISSN1347-443X
1881-4379
DOI10.11239/jsmbe.54Annual.S55

Cover

More Information
Summary:Sensor less estimation of pressure and pump flow has been an important subject for developing artificial hearts. However, estimated values are often incorrect influenced by the change in blood viscosity that changes with various factors. Therefore, viscosity estimation is essential for accurate estimation of pressure and flow. In previous study, we proposed viscosity estimation with table estimation method and in vitro study was performed using helical flow pump. However, dispersion was found with nonlinear characteristic of motor. In this study, we tried to improve viscosity estimation with modification of motor having linear characteristics. As a result, standard deviations of estimated viscosity reduced from 0.51 to 0.20. Consequently, dispersion of estimated pressure and flow could be reduced.
ISSN:1347-443X
1881-4379
DOI:10.11239/jsmbe.54Annual.S55