Comparison of Keratometry Measurement Values between HTG-1 and Pentacam According to ISO19980
Purpose: This study compared the keratometry values obtained from the HTG-1 (Huvitz, Anyang, Korea) and the Scheimpflug-based Pentacam HR (Oculus, Wetzlar, Germany) in accordance with ISO19980:2012. The study aimed to assess the repeatability, reproducibility, and agreement between the devices, ther...
Saved in:
Published in | Annals of Optometry and Contact Lens Vol. 24; no. 2; pp. 57 - 66 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | English |
Published |
25.06.2025
|
Online Access | Get full text |
ISSN | 2384-0919 2384-0927 |
DOI | 10.52725/aocl.2025.24.2.57 |
Cover
Abstract | Purpose: This study compared the keratometry values obtained from the HTG-1 (Huvitz, Anyang, Korea) and the Scheimpflug-based Pentacam HR (Oculus, Wetzlar, Germany) in accordance with ISO19980:2012. The study aimed to assess the repeatability, reproducibility, and agreement between the devices, thereby evaluating the clinical validity of the HTG-1 as a corneal topographer.Methods: A total of 75 eyes from 75 subjects were examined for corneal curvature measurements following the ISO19980:2012 standards. The measurements were taken in the central (3 mm zone), middle (5 mm zone), outer (7 mm zone), and simulated K areas. These included steep K, flat K, mean K, astigmatism, and axis, as measured by both the HTG-1 and Pentacam HR. Astigmatism and axis were further analyzed using Jackson Cross Cylinder values (J0 and J45) through power vector transformation. Additional parameters such as corneal asphericity, white-to-white measurements, Keratoconus index, and Keratoconus prediction index were also compared. To investigate how astigmatism might influence these comparisons, subjects were divided into two groups: Group 1 with less than 1.5 D of astigmatism, and Group 2 with greater than or equal to 1.5 D, followed by subgroup analysis.Results: In Group 1 (n = 38) and Group 2 (n = 37), correlation analysis revealed statistically significant differences (p < 0.001) between HTG-1 and Pentacam HR for all corneal curvature and astigmatism measurements. However, the 95% limits of agreement from the Bland-Altman plots generally fell within the repeatability limits for all measurement zones, demonstrating a high degree of agreement between the two devices.Conclusions: Despite employing different measurement principles, the HTG-1 and Pentacam HR exhibit a strong repeatability and agreement across various keratometry values. While there is a need for some adjustment, particularly in the flat K values, the HTG-1 measurements may potentially substitute Pentacam readings in clinical practice. |
---|---|
AbstractList | Purpose: This study compared the keratometry values obtained from the HTG-1 (Huvitz, Anyang, Korea) and the Scheimpflug-based Pentacam HR (Oculus, Wetzlar, Germany) in accordance with ISO19980:2012. The study aimed to assess the repeatability, reproducibility, and agreement between the devices, thereby evaluating the clinical validity of the HTG-1 as a corneal topographer.Methods: A total of 75 eyes from 75 subjects were examined for corneal curvature measurements following the ISO19980:2012 standards. The measurements were taken in the central (3 mm zone), middle (5 mm zone), outer (7 mm zone), and simulated K areas. These included steep K, flat K, mean K, astigmatism, and axis, as measured by both the HTG-1 and Pentacam HR. Astigmatism and axis were further analyzed using Jackson Cross Cylinder values (J0 and J45) through power vector transformation. Additional parameters such as corneal asphericity, white-to-white measurements, Keratoconus index, and Keratoconus prediction index were also compared. To investigate how astigmatism might influence these comparisons, subjects were divided into two groups: Group 1 with less than 1.5 D of astigmatism, and Group 2 with greater than or equal to 1.5 D, followed by subgroup analysis.Results: In Group 1 (n = 38) and Group 2 (n = 37), correlation analysis revealed statistically significant differences (p < 0.001) between HTG-1 and Pentacam HR for all corneal curvature and astigmatism measurements. However, the 95% limits of agreement from the Bland-Altman plots generally fell within the repeatability limits for all measurement zones, demonstrating a high degree of agreement between the two devices.Conclusions: Despite employing different measurement principles, the HTG-1 and Pentacam HR exhibit a strong repeatability and agreement across various keratometry values. While there is a need for some adjustment, particularly in the flat K values, the HTG-1 measurements may potentially substitute Pentacam readings in clinical practice. |
Author | Choi, Yun Seok Choi, Seung Yong |
Author_xml | – sequence: 1 givenname: Yun Seok surname: Choi fullname: Choi, Yun Seok – sequence: 2 givenname: Seung Yong surname: Choi fullname: Choi, Seung Yong |
BookMark | eNo90N9KwzAYBfAgE5xzL-BVXqA1_74muRxD53AyweGdhDT9KoW1GUmH7O11Kl6dAwfOxe-aTIY4ICG3nJUgtIA7H8O-FExAKVQpStAXZCqkUQWzQk_-O7dXZJ5zVzMlZAUGYErel7E_-NTlONDY0idMfow9julEn9HnY8Ieh5G--f0RM61x_EQc6ONuVXDqh4a-fK8--J4uQoip6YYPOka6ft1yaw27IZet32ec_-WM7B7ud8vHYrNdrZeLTRGM1gVvrQWoawjaKq9Qm4pLr2WNsuUNk1UTuFFWB1ZX3EioECQqVAxBIJNKzoj4vQ0p5pywdYfU9T6dHGfuh8ididyZyAnlhAMtvwBqmVsK |
Cites_doi | 10.1016/s1607-551x(09)70336-5 10.1016/s0002-9394(02)01528-3 10.1016/j.jcrs.2010.02.025 10.1038/s41598-019-42968-3 10.3341/kjo.2018.0008 10.3928/1081597x-20090512-07 10.1016/j.jcrs.2007.10.031 10.3341/jkos.2009.50.11.1730 10.1590/s0004-27492010000400013 10.1097/01.opx.0000180817.46312.0a 10.1016/j.jcrs.2015.03.019 10.1016/j.ajo.2013.01.029 10.1097/opx.0000000000000378 10.1177/1352458510371640 10.1186/s12886-017-0528-x 10.1016/j.ajo.2016.12.003 10.1016/j.ophtha.2006.05.053 10.1016/j.ajo.2006.08.031 10.3928/1081597x-20150224-01 10.1016/s0886-3350(13)80917-0 |
ContentType | Journal Article |
DBID | AAYXX CITATION |
DOI | 10.52725/aocl.2025.24.2.57 |
DatabaseName | CrossRef |
DatabaseTitle | CrossRef |
DatabaseTitleList | CrossRef |
DeliveryMethod | fulltext_linktorsrc |
EISSN | 2384-0927 |
EndPage | 66 |
ExternalDocumentID | 10_52725_aocl_2025_24_2_57 |
GroupedDBID | AAYXX ALMA_UNASSIGNED_HOLDINGS CITATION P5Y |
ID | FETCH-LOGICAL-c877-1f9955bb5c794a4e78613a73be3f1d036dc18497c0b618356e53e4e40e52e0343 |
ISSN | 2384-0919 |
IngestDate | Wed Oct 01 05:52:26 EDT 2025 |
IsDoiOpenAccess | false |
IsOpenAccess | true |
IsPeerReviewed | false |
IsScholarly | true |
Issue | 2 |
Language | English |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-c877-1f9955bb5c794a4e78613a73be3f1d036dc18497c0b618356e53e4e40e52e0343 |
OpenAccessLink | https://www.annocl.org/upload/pdf/aocl-2025-24.pdf |
PageCount | 10 |
ParticipantIDs | crossref_primary_10_52725_aocl_2025_24_2_57 |
ProviderPackageCode | CITATION AAYXX |
PublicationCentury | 2000 |
PublicationDate | 2025-06-25 |
PublicationDateYYYYMMDD | 2025-06-25 |
PublicationDate_xml | – month: 06 year: 2025 text: 2025-06-25 day: 25 |
PublicationDecade | 2020 |
PublicationTitle | Annals of Optometry and Contact Lens |
PublicationYear | 2025 |
References | ref13 ref12 ref23 ref15 ref14 ref20 Muscat (ref16) 2002 ref11 ref22 ref10 ref21 ref2 ref1 ref17 ref18 ref8 Karunaratne (ref19) 2013 Park (ref24) 2012 ref9 ref4 Wang (ref7) 2006 ref3 ref6 ref5 |
References_xml | – ident: ref14 doi: 10.1016/s1607-551x(09)70336-5 – ident: ref18 doi: 10.1016/s0002-9394(02)01528-3 – ident: ref4 doi: 10.1016/j.jcrs.2010.02.025 – year: 2006 ident: ref7 – ident: ref9 doi: 10.1038/s41598-019-42968-3 – ident: ref6 doi: 10.3341/kjo.2018.0008 – ident: ref12 doi: 10.3928/1081597x-20090512-07 – ident: ref2 doi: 10.1016/j.jcrs.2007.10.031 – ident: ref3 doi: 10.3341/jkos.2009.50.11.1730 – ident: ref13 doi: 10.1590/s0004-27492010000400013 – ident: ref23 doi: 10.1097/01.opx.0000180817.46312.0a – ident: ref8 doi: 10.1016/j.jcrs.2015.03.019 – ident: ref21 doi: 10.1016/j.ajo.2013.01.029 – ident: ref10 doi: 10.1097/opx.0000000000000378 – ident: ref17 doi: 10.1177/1352458510371640 – ident: ref5 doi: 10.1186/s12886-017-0528-x – ident: ref22 doi: 10.1016/j.ajo.2016.12.003 – start-page: 490 volume-title: Repeatability and reproducibility of macular thickness measurements with the Humphrey OCT system year: 2002 ident: ref16 – ident: ref15 doi: 10.1016/j.ophtha.2006.05.053 – start-page: 77 volume-title: The prevalence of cataract and glaucoma in Korean year: 2012 ident: ref24 – ident: ref20 doi: 10.1016/j.ajo.2006.08.031 – ident: ref11 doi: 10.3928/1081597x-20150224-01 – ident: ref1 doi: 10.1016/s0886-3350(13)80917-0 – start-page: 825 volume-title: Comparison of the Pentacam equivalent keratometry reading and IOL Master keratometry measurement in intraocular lens power calculations year: 2013 ident: ref19 |
SSID | ssib042365855 ssib025320965 |
Score | 2.2991974 |
Snippet | Purpose: This study compared the keratometry values obtained from the HTG-1 (Huvitz, Anyang, Korea) and the Scheimpflug-based Pentacam HR (Oculus, Wetzlar,... |
SourceID | crossref |
SourceType | Index Database |
StartPage | 57 |
Title | Comparison of Keratometry Measurement Values between HTG-1 and Pentacam According to ISO19980 |
Volume | 24 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
journalDatabaseRights | – providerCode: PRVHPJ databaseName: ROAD: Directory of Open Access Scholarly Resources customDbUrl: eissn: 2384-0927 dateEnd: 99991231 omitProxy: true ssIdentifier: ssib025320965 issn: 2384-0919 databaseCode: M~E dateStart: 20150101 isFulltext: true titleUrlDefault: https://road.issn.org providerName: ISSN International Centre |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV1Nb9QwELWgXLggECCgBfnQW5Qlcew4OVar0m1h20MX1B5QFCcTIUE31So90AO_nRk7H9YCEu0lWlmb2Wzm6fmN7ZlhbL-GJo0aZD_ImjyURqrQZGkUxhXNPxJiU9M65PI0XXyWJxfqwtsxpeySzsyq27_mldzHqziGfqUs2Tt4djSKA_gZ_YtX9DBe_8vHc7-JYPCR6iO3V9BtfgbLaekv-FL-QO4fT2QtVkdh7JIE6Nx4VV7ZlhEbm92CSvT4_Iwy4iJftk5lls-uh58gC1TbipIsP8F61Obzb609IXB5Q0zUft8ePwfkl-Cy7afMfsVBKDoZ5bKTHTHhLC9D1BmO6sAfc3n-A7O67OgeQcKjSVeTup9wXdeVbSpXQtvGTGVb0Q6RUDMhZ2I23OnXzd6az8ZThhjfWCsF2SjIRiFkIQqlH7JHQqcpdbxY_joc-EdQk4x82vVFpYn6zDbOHf-yy7uyZt__8WietvFEyuope9JHF_zAQeUZewDr5-zrBBPeNtyDCfdgwh1MeA8TbmHC0cl8gAkfYcK7lg8wecFWHw5X80XYd9UIq0zrMG7yXCljVIVMXErQGQq6UicGkiauUc_UFQb9ua4ikyLdqxRUAhJkBEpAlMjkJdtZt2t4xTjUGLxnUQ0kqrXKDQZqqHoanUhoMDJ4zYLhdRTXrnZK8W-fvLnTt3fZ4wmae2yn29zAW5SHnXlnffobsF5fYA |
linkProvider | ISSN International Centre |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Comparison+of+Keratometry+Measurement+Values+between+HTG-1+and+Pentacam+According+to+ISO19980&rft.jtitle=Annals+of+Optometry+and+Contact+Lens&rft.au=Choi%2C+Yun+Seok&rft.au=Choi%2C+Seung+Yong&rft.date=2025-06-25&rft.issn=2384-0919&rft.eissn=2384-0927&rft.volume=24&rft.issue=2&rft.spage=57&rft.epage=66&rft_id=info:doi/10.52725%2Faocl.2025.24.2.57&rft.externalDBID=n%2Fa&rft.externalDocID=10_52725_aocl_2025_24_2_57 |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=2384-0919&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=2384-0919&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=2384-0919&client=summon |