Non-Invasive Measurement of Hemoglobin: Assessment of Two Different Point-of-Care Technologies
Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for the patient and allows time saving in patient care. The aims of the present study were to assess the concordance of Hb levels obtained non-invasively w...
Saved in:
| Published in | PloS one Vol. 7; no. 1; p. e30065 |
|---|---|
| Main Authors | , , , , |
| Format | Journal Article |
| Language | English |
| Published |
United States
Public Library of Science
06.01.2012
Public Library of Science (PLoS) |
| Subjects | |
| Online Access | Get full text |
| ISSN | 1932-6203 1932-6203 |
| DOI | 10.1371/journal.pone.0030065 |
Cover
| Abstract | Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for the patient and allows time saving in patient care. The aims of the present study were to assess the concordance of Hb levels obtained non-invasively with the Pronto-7 monitor (version 2.1.9, Masimo Corporation, Irvine, USA) or with the NBM-200MP monitor (Orsense, Nes Ziona, Israel) and the values obtained from the usual colorimetric method using blood samples and to determine the source of discordance.
We conducted two consecutive prospective open trials enrolling patients presenting in the emergency department of a university hospital. The first was designed to assess Pronto-7™ and the second NBM-200MP™. In each study, the main outcome measure was the agreement between both methods. Independent factors associated with the bias were determined using multiple linear regression. Three hundred patients were prospectively enrolled in each study. For Pronto-7™, the absolute mean difference was 0.56 g.L(-1) (95% confidence interval [CI] 0.41 to 0.69) with an upper agreement limit at 2.94 g.L(-1) (95% CI [2.70;3.19]), a lower agreement limit at -1.84 g.L(-1) (95% CI [-2.08;-1.58]) and an intra-class correlation coefficient at 0.80 (95% CI [0.74;0.84]). The corresponding values for the NBM-200MP™ were 0.21 [0.02;0.39], 3.42 [3.10;3.74], -3.01 [-3.32;-2.69] and 0.69 [0.62;0.75]. Multivariate analysis showed that age and laboratory values of hemoglobin were independently associated with the bias when using Pronto-7™, while perfusion index and laboratory value of hemoglobin were independently associated with the bias when using NBM-200MP™.
Despite a relatively limited bias in both cases, the large limits of agreement found in both cases render the clinical usefulness of such devices debatable. For both devices, the bias is independently and inversely associated with the true value of hemoglobin.
ClinicalTrials.gov NCT01321580 and NCT01321593. |
|---|---|
| AbstractList | Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for the patient and allows time saving in patient care. The aims of the present study were to assess the concordance of Hb levels obtained non-invasively with the Pronto-7 monitor (version 2.1.9, Masimo Corporation, Irvine, USA) or with the NBM-200MP monitor (Orsense, Nes Ziona, Israel) and the values obtained from the usual colorimetric method using blood samples and to determine the source of discordance.
We conducted two consecutive prospective open trials enrolling patients presenting in the emergency department of a university hospital. The first was designed to assess Pronto-7™ and the second NBM-200MP™. In each study, the main outcome measure was the agreement between both methods. Independent factors associated with the bias were determined using multiple linear regression. Three hundred patients were prospectively enrolled in each study. For Pronto-7™, the absolute mean difference was 0.56 g.L(-1) (95% confidence interval [CI] 0.41 to 0.69) with an upper agreement limit at 2.94 g.L(-1) (95% CI [2.70;3.19]), a lower agreement limit at -1.84 g.L(-1) (95% CI [-2.08;-1.58]) and an intra-class correlation coefficient at 0.80 (95% CI [0.74;0.84]). The corresponding values for the NBM-200MP™ were 0.21 [0.02;0.39], 3.42 [3.10;3.74], -3.01 [-3.32;-2.69] and 0.69 [0.62;0.75]. Multivariate analysis showed that age and laboratory values of hemoglobin were independently associated with the bias when using Pronto-7™, while perfusion index and laboratory value of hemoglobin were independently associated with the bias when using NBM-200MP™.
Despite a relatively limited bias in both cases, the large limits of agreement found in both cases render the clinical usefulness of such devices debatable. For both devices, the bias is independently and inversely associated with the true value of hemoglobin.
ClinicalTrials.gov NCT01321580 and NCT01321593. Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for the patient and allows time saving in patient care. The aims of the present study were to assess the concordance of Hb levels obtained non-invasively with the Pronto-7 monitor (version 2.1.9, Masimo Corporation, Irvine, USA) or with the NBM-200MP monitor (Orsense, Nes Ziona, Israel) and the values obtained from the usual colorimetric method using blood samples and to determine the source of discordance.We conducted two consecutive prospective open trials enrolling patients presenting in the emergency department of a university hospital. The first was designed to assess Pronto-7™ and the second NBM-200MP™. In each study, the main outcome measure was the agreement between both methods. Independent factors associated with the bias were determined using multiple linear regression. Three hundred patients were prospectively enrolled in each study. For Pronto-7™, the absolute mean difference was 0.56 g.L(-1) (95% confidence interval [CI] 0.41 to 0.69) with an upper agreement limit at 2.94 g.L(-1) (95% CI [2.70;3.19]), a lower agreement limit at -1.84 g.L(-1) (95% CI [-2.08;-1.58]) and an intra-class correlation coefficient at 0.80 (95% CI [0.74;0.84]). The corresponding values for the NBM-200MP™ were 0.21 [0.02;0.39], 3.42 [3.10;3.74], -3.01 [-3.32;-2.69] and 0.69 [0.62;0.75]. Multivariate analysis showed that age and laboratory values of hemoglobin were independently associated with the bias when using Pronto-7™, while perfusion index and laboratory value of hemoglobin were independently associated with the bias when using NBM-200MP™.Despite a relatively limited bias in both cases, the large limits of agreement found in both cases render the clinical usefulness of such devices debatable. For both devices, the bias is independently and inversely associated with the true value of hemoglobin.ClinicalTrials.gov NCT01321580 and NCT01321593. Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for the patient and allows time saving in patient care. The aims of the present study were to assess the concordance of Hb levels obtained non-invasively with the Pronto-7 monitor (version 2.1.9, Masimo Corporation, Irvine, USA) or with the NBM-200MP monitor (Orsense, Nes Ziona, Israel) and the values obtained from the usual colorimetric method using blood samples and to determine the source of discordance. We conducted two consecutive prospective open trials enrolling patients presenting in the emergency department of a university hospital. The first was designed to assess Pronto-7[TM] and the second NBM-200MP[TM]. In each study, the main outcome measure was the agreement between both methods. Independent factors associated with the bias were determined using multiple linear regression. Three hundred patients were prospectively enrolled in each study. For Pronto-7[TM], the absolute mean difference was 0.56 g.L.sup.-1 (95% confidence interval [CI] 0.41 to 0.69) with an upper agreement limit at 2.94 g.L.sup.-1 (95% CI [2.70;3.19]), a lower agreement limit at -1.84 g.L.sup.-1 (95% CI [-2.08;-1.58]) and an intra-class correlation coefficient at 0.80 (95% CI [0.74;0.84]). The corresponding values for the NBM-200MP[TM] were 0.21 [0.02;0.39], 3.42 [3.10;3.74], -3.01 [-3.32;-2.69] and 0.69 [0.62;0.75]. Multivariate analysis showed that age and laboratory values of hemoglobin were independently associated with the bias when using Pronto-7[TM], while perfusion index and laboratory value of hemoglobin were independently associated with the bias when using NBM-200MP[TM]. Despite a relatively limited bias in both cases, the large limits of agreement found in both cases render the clinical usefulness of such devices debatable. For both devices, the bias is independently and inversely associated with the true value of hemoglobin. Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for the patient and allows time saving in patient care. The aims of the present study were to assess the concordance of Hb levels obtained non-invasively with the Pronto-7 monitor (version 2.1.9, Masimo Corporation, Irvine, USA) or with the NBM-200MP monitor (Orsense, Nes Ziona, Israel) and the values obtained from the usual colorimetric method using blood samples and to determine the source of discordance.BACKGROUNDMeasurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for the patient and allows time saving in patient care. The aims of the present study were to assess the concordance of Hb levels obtained non-invasively with the Pronto-7 monitor (version 2.1.9, Masimo Corporation, Irvine, USA) or with the NBM-200MP monitor (Orsense, Nes Ziona, Israel) and the values obtained from the usual colorimetric method using blood samples and to determine the source of discordance.We conducted two consecutive prospective open trials enrolling patients presenting in the emergency department of a university hospital. The first was designed to assess Pronto-7™ and the second NBM-200MP™. In each study, the main outcome measure was the agreement between both methods. Independent factors associated with the bias were determined using multiple linear regression. Three hundred patients were prospectively enrolled in each study. For Pronto-7™, the absolute mean difference was 0.56 g.L(-1) (95% confidence interval [CI] 0.41 to 0.69) with an upper agreement limit at 2.94 g.L(-1) (95% CI [2.70;3.19]), a lower agreement limit at -1.84 g.L(-1) (95% CI [-2.08;-1.58]) and an intra-class correlation coefficient at 0.80 (95% CI [0.74;0.84]). The corresponding values for the NBM-200MP™ were 0.21 [0.02;0.39], 3.42 [3.10;3.74], -3.01 [-3.32;-2.69] and 0.69 [0.62;0.75]. Multivariate analysis showed that age and laboratory values of hemoglobin were independently associated with the bias when using Pronto-7™, while perfusion index and laboratory value of hemoglobin were independently associated with the bias when using NBM-200MP™.METHODS AND FINDINGSWe conducted two consecutive prospective open trials enrolling patients presenting in the emergency department of a university hospital. The first was designed to assess Pronto-7™ and the second NBM-200MP™. In each study, the main outcome measure was the agreement between both methods. Independent factors associated with the bias were determined using multiple linear regression. Three hundred patients were prospectively enrolled in each study. For Pronto-7™, the absolute mean difference was 0.56 g.L(-1) (95% confidence interval [CI] 0.41 to 0.69) with an upper agreement limit at 2.94 g.L(-1) (95% CI [2.70;3.19]), a lower agreement limit at -1.84 g.L(-1) (95% CI [-2.08;-1.58]) and an intra-class correlation coefficient at 0.80 (95% CI [0.74;0.84]). The corresponding values for the NBM-200MP™ were 0.21 [0.02;0.39], 3.42 [3.10;3.74], -3.01 [-3.32;-2.69] and 0.69 [0.62;0.75]. Multivariate analysis showed that age and laboratory values of hemoglobin were independently associated with the bias when using Pronto-7™, while perfusion index and laboratory value of hemoglobin were independently associated with the bias when using NBM-200MP™.Despite a relatively limited bias in both cases, the large limits of agreement found in both cases render the clinical usefulness of such devices debatable. For both devices, the bias is independently and inversely associated with the true value of hemoglobin.CONCLUSIONDespite a relatively limited bias in both cases, the large limits of agreement found in both cases render the clinical usefulness of such devices debatable. For both devices, the bias is independently and inversely associated with the true value of hemoglobin.ClinicalTrials.gov NCT01321580 and NCT01321593.TRIAL REGISTRATIONClinicalTrials.gov NCT01321580 and NCT01321593. Background Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for the patient and allows time saving in patient care. The aims of the present study were to assess the concordance of Hb levels obtained non-invasively with the Pronto-7 monitor (version 2.1.9, Masimo Corporation, Irvine, USA) or with the NBM-200MP monitor (Orsense, Nes Ziona, Israel) and the values obtained from the usual colorimetric method using blood samples and to determine the source of discordance. Methods and Findings We conducted two consecutive prospective open trials enrolling patients presenting in the emergency department of a university hospital. The first was designed to assess Pronto-7[TM] and the second NBM-200MP[TM]. In each study, the main outcome measure was the agreement between both methods. Independent factors associated with the bias were determined using multiple linear regression. Three hundred patients were prospectively enrolled in each study. For Pronto-7[TM], the absolute mean difference was 0.56 g.L.sup.-1 (95% confidence interval [CI] 0.41 to 0.69) with an upper agreement limit at 2.94 g.L.sup.-1 (95% CI [2.70;3.19]), a lower agreement limit at -1.84 g.L.sup.-1 (95% CI [-2.08;-1.58]) and an intra-class correlation coefficient at 0.80 (95% CI [0.74;0.84]). The corresponding values for the NBM-200MP[TM] were 0.21 [0.02;0.39], 3.42 [3.10;3.74], -3.01 [-3.32;-2.69] and 0.69 [0.62;0.75]. Multivariate analysis showed that age and laboratory values of hemoglobin were independently associated with the bias when using Pronto-7[TM], while perfusion index and laboratory value of hemoglobin were independently associated with the bias when using NBM-200MP[TM]. Conclusion Despite a relatively limited bias in both cases, the large limits of agreement found in both cases render the clinical usefulness of such devices debatable. For both devices, the bias is independently and inversely associated with the true value of hemoglobin. Trial Registration ClinicalTrials.gov NCT01321580 NCT01321593 Background Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for the patient and allows time saving in patient care. The aims of the present study were to assess the concordance of Hb levels obtained non-invasively with the Pronto-7 monitor (version 2.1.9, Masimo Corporation, Irvine, USA) or with the NBM-200MP monitor (Orsense, Nes Ziona, Israel) and the values obtained from the usual colorimetric method using blood samples and to determine the source of discordance. Methods and Findings We conducted two consecutive prospective open trials enrolling patients presenting in the emergency department of a university hospital. The first was designed to assess Pronto-7™ and the second NBM-200MP™. In each study, the main outcome measure was the agreement between both methods. Independent factors associated with the bias were determined using multiple linear regression. Three hundred patients were prospectively enrolled in each study. For Pronto-7™, the absolute mean difference was 0.56 g.L−1 (95% confidence interval [CI] 0.41 to 0.69) with an upper agreement limit at 2.94 g.L−1 (95% CI [2.70;3.19]), a lower agreement limit at -1.84 g.L−1 (95% CI [-2.08;-1.58]) and an intra-class correlation coefficient at 0.80 (95% CI [0.74;0.84]). The corresponding values for the NBM-200MP™ were 0.21 [0.02;0.39], 3.42 [3.10;3.74], -3.01 [-3.32;-2.69] and 0.69 [0.62;0.75]. Multivariate analysis showed that age and laboratory values of hemoglobin were independently associated with the bias when using Pronto-7™, while perfusion index and laboratory value of hemoglobin were independently associated with the bias when using NBM-200MP™. Conclusion Despite a relatively limited bias in both cases, the large limits of agreement found in both cases render the clinical usefulness of such devices debatable. For both devices, the bias is independently and inversely associated with the true value of hemoglobin. Trial Registration ClinicalTrials.gov NCT01321580 NCT01321593 Background Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for the patient and allows time saving in patient care. The aims of the present study were to assess the concordance of Hb levels obtained non-invasively with the Pronto-7 monitor (version 2.1.9, Masimo Corporation, Irvine, USA) or with the NBM-200MP monitor (Orsense, Nes Ziona, Israel) and the values obtained from the usual colorimetric method using blood samples and to determine the source of discordance. Methods and Findings We conducted two consecutive prospective open trials enrolling patients presenting in the emergency department of a university hospital. The first was designed to assess Pronto-7™ and the second NBM-200MP™. In each study, the main outcome measure was the agreement between both methods. Independent factors associated with the bias were determined using multiple linear regression. Three hundred patients were prospectively enrolled in each study. For Pronto-7™, the absolute mean difference was 0.56 g.L −1 (95% confidence interval [CI] 0.41 to 0.69) with an upper agreement limit at 2.94 g.L −1 (95% CI [2.70;3.19]), a lower agreement limit at -1.84 g.L −1 (95% CI [-2.08;-1.58]) and an intra-class correlation coefficient at 0.80 (95% CI [0.74;0.84]). The corresponding values for the NBM-200MP™ were 0.21 [0.02;0.39], 3.42 [3.10;3.74], -3.01 [-3.32;-2.69] and 0.69 [0.62;0.75]. Multivariate analysis showed that age and laboratory values of hemoglobin were independently associated with the bias when using Pronto-7™, while perfusion index and laboratory value of hemoglobin were independently associated with the bias when using NBM-200MP™. Conclusion Despite a relatively limited bias in both cases, the large limits of agreement found in both cases render the clinical usefulness of such devices debatable. For both devices, the bias is independently and inversely associated with the true value of hemoglobin. Trial Registration ClinicalTrials.gov NCT01321580 NCT01321593 |
| Audience | Academic |
| Author | Fischler, Marc Gayat, Etienne Boisson, Mireille Aulagnier, Jérôme Matthieu, Emmanuel |
| AuthorAffiliation | University of Colorado Denver, United States of America 2 Clinical Epidemiology and Biostatistics, INSERM U717, Saint-Louis Hospital, Paris, France 3 Emergency room, Foch Hospital, Suresnes, France 1 Department of Anesthesia, Foch Hospital, Suresnes, France 4 Department of Biology, Foch Hospital, Suresnes, France |
| AuthorAffiliation_xml | – name: 4 Department of Biology, Foch Hospital, Suresnes, France – name: 3 Emergency room, Foch Hospital, Suresnes, France – name: University of Colorado Denver, United States of America – name: 2 Clinical Epidemiology and Biostatistics, INSERM U717, Saint-Louis Hospital, Paris, France – name: 1 Department of Anesthesia, Foch Hospital, Suresnes, France |
| Author_xml | – sequence: 1 givenname: Etienne surname: Gayat fullname: Gayat, Etienne – sequence: 2 givenname: Jérôme surname: Aulagnier fullname: Aulagnier, Jérôme – sequence: 3 givenname: Emmanuel surname: Matthieu fullname: Matthieu, Emmanuel – sequence: 4 givenname: Mireille surname: Boisson fullname: Boisson, Mireille – sequence: 5 givenname: Marc surname: Fischler fullname: Fischler, Marc |
| BackLink | https://www.ncbi.nlm.nih.gov/pubmed/22238693$$D View this record in MEDLINE/PubMed |
| BookMark | eNqNkttv0zAUxiM0xC7wHyCohATiIcWXXJw9IFXlskqDISg8YrnOcevKtYudbOy_x1nbqZkmNOUhzvHv-3z8nRwnB9ZZSJLnGA0xLfG7pWu9FWa4juUhQhShIn-UHOGKkrQgiB7srQ-T4xCWCOWUFcWT5JAQElcVPUp-f3U2ndhLEfQlDL6ACK2HFdhm4NTgDFZubtxM29PBKAQIYbczvXKDD1op8F3hm9O2SZ1Kx8LDYApyYZ1xcw3hafJYCRPg2fZ9kvz89HE6PkvPLz5PxqPzVBYVblJGWYVmIqe0RqTMoZRUxisRDFIBLYsiVzJjqI6ftazFDGU5yiCnBWKMRQU9SV5ufNfGBb6NJnBMCclKlmUsEpMNUTux5GuvV8Jfcyc0vyk4P-fCN1oa4JJVNValglqRrK4xq3M5w5RJpXJGVeeVb7xauxbXV8KYW0OMeDedXQu8mw7fTifq3m-7bGcrqGXMzgvTa6a_Y_WCz90lpySnmBTR4M3WwLs_LYSGr3SQYIyw4NrAK1zgrMxuyFd3yPtD2VJzEe-trXLxWNl58lFWlrioWEYiNbyHik8NKy3jBZWO9Z7gbU8QmQb-NnPRhsAnP74_nL341Wdf77ELEKZZBGfaRjsb-uCL_aBvE9799xHINoD0LgQP6qEDPL0jk7oR3fExEW3-L_4HDXwqjQ |
| CitedBy_id | crossref_primary_10_1371_journal_pgph_0001541 crossref_primary_10_1016_j_blre_2013_10_001 crossref_primary_10_3390_jcm12247517 crossref_primary_10_1186_s12884_023_05783_3 crossref_primary_10_1186_s12968_018_0443_1 crossref_primary_10_1007_s11739_014_1129_9 crossref_primary_10_1109_JBHI_2019_2954553 crossref_primary_10_1371_journal_pone_0239931 crossref_primary_10_3343_alm_2013_33_4_261 crossref_primary_10_1111_nyas_14003 crossref_primary_10_1111_nyas_14124 crossref_primary_10_1007_s40140_014_0070_9 crossref_primary_10_1016_j_hybadv_2023_100039 crossref_primary_10_1016_j_bjane_2020_06_001 crossref_primary_10_1038_s41598_022_13285_z crossref_primary_10_3390_jcm12175733 crossref_primary_10_1093_bja_aeu403 crossref_primary_10_1097_MCC_0000000000000197 crossref_primary_10_1111_trf_13241 crossref_primary_10_1111_anae_14358 crossref_primary_10_1109_JSEN_2021_3070971 crossref_primary_10_3109_14767058_2012_704453 crossref_primary_10_17945_kjbt_2015_26_3_273 crossref_primary_10_1016_j_heliyon_2024_e35777 crossref_primary_10_1088_1361_6579_ac41b7 crossref_primary_10_1213_ANE_0000000000000272 crossref_primary_10_4103_jcn_JCN_89_19 crossref_primary_10_1016_j_bjan_2019_05_006 crossref_primary_10_1186_s13613_022_01012_w crossref_primary_10_1111_ijlh_12056 crossref_primary_10_1002_ajhb_23670 crossref_primary_10_1155_2020_3034260 crossref_primary_10_1213_ANE_0000000000000198 crossref_primary_10_1016_j_arth_2014_06_029 crossref_primary_10_1016_j_transci_2013_10_007 crossref_primary_10_1182_blood_2012_09_453522 crossref_primary_10_3390_bios13080816 crossref_primary_10_1111_trf_17237 crossref_primary_10_1016_j_jgyn_2014_10_001 crossref_primary_10_1109_JSEN_2017_2752371 crossref_primary_10_1177_0310057X1504300310 crossref_primary_10_56126_71_3_4 crossref_primary_10_1111_anae_15312 crossref_primary_10_1186_1749_8090_8_159 crossref_primary_10_1371_journal_pmed_1003739 crossref_primary_10_15406_htij_2015_01_00016 crossref_primary_10_1111_trf_12819 crossref_primary_10_1136_bmjinnov_2015_000085 crossref_primary_10_1159_000526100 crossref_primary_10_1186_s12916_020_01775_8 crossref_primary_10_1016_j_amjsurg_2013_08_012 crossref_primary_10_1016_j_tacc_2014_07_002 crossref_primary_10_1039_D0LC01235A crossref_primary_10_1089_whr_2024_0028 crossref_primary_10_1371_journal_pone_0187663 crossref_primary_10_1007_s10877_022_00813_5 crossref_primary_10_1097_TA_0000000000000326 crossref_primary_10_1155_2013_701529 crossref_primary_10_1097_BOT_0000000000001677 crossref_primary_10_1111_jgs_12623 crossref_primary_10_4187_respcare_06364 crossref_primary_10_1371_journal_pone_0254629 crossref_primary_10_1053_j_jvca_2016_03_148 crossref_primary_10_1186_s12917_023_03804_z crossref_primary_10_1213_ANE_0000000000000927 crossref_primary_10_1002_ijgo_15385 crossref_primary_10_4103_ijmr_IJMR_301_17 crossref_primary_10_21518_2079_701X_2021_12_398_404 crossref_primary_10_1007_s00431_015_2602_9 crossref_primary_10_1093_bja_aet252 crossref_primary_10_1186_s12916_015_0320_9 crossref_primary_10_1177_2050312118796065 crossref_primary_10_4274_anatoljmed_2024_61587 crossref_primary_10_1213_ANE_0000000000000605 crossref_primary_10_1016_j_transci_2020_102770 crossref_primary_10_1097_MPH_0000000000001038 crossref_primary_10_1364_BOE_549671 crossref_primary_10_1038_s43856_023_00267_z crossref_primary_10_1016_j_pratan_2013_04_004 crossref_primary_10_1111_pan_12007 crossref_primary_10_1111_vox_12033 crossref_primary_10_1136_bmjopen_2023_083356 crossref_primary_10_1007_s40846_022_00750_3 crossref_primary_10_3810_pgm_2012_07_2584 crossref_primary_10_1111_ijlh_12118 crossref_primary_10_1038_s41598_023_42572_6 crossref_primary_10_2174_1573405617666210823100316 crossref_primary_10_1016_j_injms_2018_06_008 crossref_primary_10_1097_CCM_0b013e31828ca59c crossref_primary_10_1515_bmt_2018_0066 |
| Cites_doi | 10.1191/0962280204sm365ra 10.1093/clinchem/32.3.526 10.1007/BF02851177 10.1111/j.1399-6576.2011.02469.x 10.1080/00365519950185229 10.1213/ANE.0b013e3181fc74b9 10.1046/j.1365-2044.1998.00328.x 10.1016/j.annemergmed.2010.11.032 10.1034/j.1399-6576.2002.460809.x 10.1136/bmj.326.7379.41 10.1093/jn/129.9.1669 10.1111/j.1399-6576.2010.02321.x 10.1213/ANE.0b013e31820eecd1 10.1016/S0140-6736(86)90837-8 |
| ContentType | Journal Article |
| Copyright | COPYRIGHT 2012 Public Library of Science 2012 Gayat et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License: https://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. Gayat et al. 2012 |
| Copyright_xml | – notice: COPYRIGHT 2012 Public Library of Science – notice: 2012 Gayat et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License: https://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. – notice: Gayat et al. 2012 |
| DBID | AAYXX CITATION CGR CUY CVF ECM EIF NPM IOV ISR 3V. 7QG 7QL 7QO 7RV 7SN 7SS 7T5 7TG 7TM 7U9 7X2 7X7 7XB 88E 8AO 8C1 8FD 8FE 8FG 8FH 8FI 8FJ 8FK ABJCF ABUWG AEUYN AFKRA ARAPS ATCPS AZQEC BBNVY BENPR BGLVJ BHPHI C1K CCPQU D1I DWQXO FR3 FYUFA GHDGH GNUQQ H94 HCIFZ K9. KB. KB0 KL. L6V LK8 M0K M0S M1P M7N M7P M7S NAPCQ P5Z P62 P64 PATMY PDBOC PHGZM PHGZT PIMPY PJZUB PKEHL PPXIY PQEST PQGLB PQQKQ PQUKI PRINS PTHSS PYCSY RC3 7X8 5PM ADTOC UNPAY DOA |
| DOI | 10.1371/journal.pone.0030065 |
| DatabaseName | CrossRef Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed Gale In Context: Opposing Viewpoints Gale In Context: Science ProQuest Central (Corporate) Animal Behavior Abstracts Bacteriology Abstracts (Microbiology B) Biotechnology Research Abstracts Nursing & Allied Health Database Ecology Abstracts Entomology Abstracts (Full archive) Immunology Abstracts Meteorological & Geoastrophysical Abstracts Nucleic Acids Abstracts Virology and AIDS Abstracts Agricultural Science Collection Health & Medical Collection ProQuest Central (purchase pre-March 2016) Medical Database (Alumni Edition) ProQuest Pharma Collection Public Health Database Technology Research Database ProQuest SciTech Collection ProQuest Technology Collection ProQuest Natural Science Journals ProQuest Hospital Collection Hospital Premium Collection (Alumni Edition) ProQuest Central (Alumni) (purchase pre-March 2016) ProQuest Materials Science & Engineering Collection ProQuest Central (Alumni) ProQuest One Sustainability ProQuest Central UK/Ireland Advanced Technologies & Computer Science Collection ProQuest Agricultural & Environmental Science Collection ProQuest Central Essentials Biological Science Collection ProQuest Central Technology Collection Natural Science Collection Environmental Sciences and Pollution Management ProQuest One Community College ProQuest Materials Science Collection ProQuest Central Engineering Research Database Health Research Premium Collection Health Research Premium Collection (Alumni) ProQuest Central Student AIDS and Cancer Research Abstracts SciTech Premium Collection ProQuest Health & Medical Complete (Alumni) Materials Science Database Nursing & Allied Health Database (Alumni Edition) Meteorological & Geoastrophysical Abstracts - Academic ProQuest Engineering Collection Biological Sciences Agriculture Science Database Health & Medical Collection (Alumni Edition) Medical Database Algology Mycology and Protozoology Abstracts (Microbiology C) Biological Science Database Engineering Database Nursing & Allied Health Premium Advanced Technologies & Aerospace Database ProQuest Advanced Technologies & Aerospace Collection Biotechnology and BioEngineering Abstracts Environmental Science Database Materials Science Collection ProQuest Central Premium ProQuest One Academic Publicly Available Content Database ProQuest Health & Medical Research Collection ProQuest One Academic Middle East (New) ProQuest One Health & Nursing ProQuest One Academic Eastern Edition (DO NOT USE) ProQuest One Applied & Life Sciences ProQuest One Academic ProQuest One Academic UKI Edition ProQuest Central China Engineering Collection Environmental Science Collection Genetics Abstracts MEDLINE - Academic PubMed Central (Full Participant titles) Unpaywall for CDI: Periodical Content Unpaywall DOAJ Directory of Open Access Journals |
| DatabaseTitle | CrossRef MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) Agricultural Science Database Publicly Available Content Database ProQuest Central Student ProQuest Advanced Technologies & Aerospace Collection ProQuest Central Essentials Nucleic Acids Abstracts SciTech Premium Collection ProQuest Central China Environmental Sciences and Pollution Management ProQuest One Applied & Life Sciences ProQuest One Sustainability Health Research Premium Collection Meteorological & Geoastrophysical Abstracts Natural Science Collection Health & Medical Research Collection Biological Science Collection ProQuest Central (New) ProQuest Medical Library (Alumni) Engineering Collection Advanced Technologies & Aerospace Collection Engineering Database Virology and AIDS Abstracts ProQuest Biological Science Collection ProQuest One Academic Eastern Edition Agricultural Science Collection ProQuest Hospital Collection ProQuest Technology Collection Health Research Premium Collection (Alumni) Biological Science Database Ecology Abstracts ProQuest Hospital Collection (Alumni) Biotechnology and BioEngineering Abstracts Environmental Science Collection Entomology Abstracts Nursing & Allied Health Premium ProQuest Health & Medical Complete ProQuest One Academic UKI Edition Environmental Science Database ProQuest Nursing & Allied Health Source (Alumni) Engineering Research Database ProQuest One Academic Meteorological & Geoastrophysical Abstracts - Academic ProQuest One Academic (New) Technology Collection Technology Research Database ProQuest One Academic Middle East (New) Materials Science Collection ProQuest Health & Medical Complete (Alumni) ProQuest Central (Alumni Edition) ProQuest One Community College ProQuest One Health & Nursing ProQuest Natural Science Collection ProQuest Pharma Collection ProQuest Central ProQuest Health & Medical Research Collection Genetics Abstracts ProQuest Engineering Collection Biotechnology Research Abstracts Health and Medicine Complete (Alumni Edition) ProQuest Central Korea Bacteriology Abstracts (Microbiology B) Algology Mycology and Protozoology Abstracts (Microbiology C) Agricultural & Environmental Science Collection AIDS and Cancer Research Abstracts Materials Science Database ProQuest Materials Science Collection ProQuest Public Health ProQuest Nursing & Allied Health Source ProQuest SciTech Collection Advanced Technologies & Aerospace Database ProQuest Medical Library Animal Behavior Abstracts Materials Science & Engineering Collection Immunology Abstracts ProQuest Central (Alumni) MEDLINE - Academic |
| DatabaseTitleList | MEDLINE MEDLINE - Academic Agricultural Science Database |
| Database_xml | – sequence: 1 dbid: DOA name: DOAJ Directory of Open Access Journals url: https://www.doaj.org/ sourceTypes: Open Website – sequence: 2 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database – sequence: 3 dbid: EIF name: MEDLINE url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search sourceTypes: Index Database – sequence: 4 dbid: UNPAY name: Unpaywall url: https://proxy.k.utb.cz/login?url=https://unpaywall.org/ sourceTypes: Open Access Repository – sequence: 5 dbid: 8FG name: ProQuest Technology Collection url: https://search.proquest.com/technologycollection1 sourceTypes: Aggregation Database |
| DeliveryMethod | fulltext_linktorsrc |
| Discipline | Sciences (General) Medicine Chemistry Biology |
| DocumentTitleAlternate | Non-Invasive Measurement of Hemoglobin |
| EISSN | 1932-6203 |
| ExternalDocumentID | 1322478448 oai_doaj_org_article_c89d1f7fedf24dd18d5cb138cff583f8 10.1371/journal.pone.0030065 PMC3253126 2933376921 A477169842 22238693 10_1371_journal_pone_0030065 |
| Genre | Clinical Trial Journal Article Comparative Study |
| GeographicLocations | France United States--US Israel |
| GeographicLocations_xml | – name: Israel – name: France – name: United States--US |
| GroupedDBID | --- 123 29O 2WC 53G 5VS 7RV 7X2 7X7 7XC 88E 8AO 8C1 8CJ 8FE 8FG 8FH 8FI 8FJ A8Z AAFWJ AAUCC AAWOE AAYXX ABDBF ABIVO ABJCF ABUWG ACGFO ACIHN ACIWK ACPRK ACUHS ADBBV ADRAZ AEAQA AENEX AEUYN AFKRA AFPKN AFRAH AHMBA ALMA_UNASSIGNED_HOLDINGS AOIJS APEBS ARAPS ATCPS BAWUL BBNVY BCNDV BENPR BGLVJ BHPHI BKEYQ BPHCQ BVXVI BWKFM CCPQU CITATION CS3 D1I D1J D1K DIK DU5 E3Z EAP EAS EBD EMOBN ESTFP ESX EX3 F5P FPL FYUFA GROUPED_DOAJ GX1 HCIFZ HH5 HMCUK HYE IAO IEA IGS IHR IHW INH INR IOV IPNFZ IPY ISE ISR ITC K6- KB. KQ8 L6V LK5 LK8 M0K M1P M48 M7P M7R M7S M~E NAPCQ O5R O5S OK1 OVT P2P P62 PATMY PDBOC PHGZM PHGZT PIMPY PJZUB PPXIY PQGLB PQQKQ PROAC PSQYO PTHSS PUEGO PYCSY RIG RNS RPM SV3 TR2 UKHRP WOQ WOW ~02 ~KM ALIPV BBORY CGR CUY CVF ECM EIF NPM PV9 RZL 3V. 7QG 7QL 7QO 7SN 7SS 7T5 7TG 7TM 7U9 7XB 8FD 8FK AZQEC C1K DWQXO FR3 GNUQQ H94 K9. KL. M7N P64 PKEHL PQEST PQUKI PRINS RC3 7X8 5PM ADTOC UNPAY - 02 AAPBV ABPTK ADACO BBAFP KM |
| ID | FETCH-LOGICAL-c691t-83890ba533d0275e7c3c00321ecfe37665fc480decfdcdab04504e53608880273 |
| IEDL.DBID | M48 |
| ISSN | 1932-6203 |
| IngestDate | Fri Nov 26 17:13:53 EST 2021 Fri Oct 03 12:50:51 EDT 2025 Sun Oct 26 03:47:14 EDT 2025 Tue Sep 30 16:35:14 EDT 2025 Thu Oct 02 06:30:53 EDT 2025 Tue Oct 07 06:42:09 EDT 2025 Mon Oct 20 21:47:32 EDT 2025 Mon Oct 20 16:11:21 EDT 2025 Thu Oct 16 15:20:21 EDT 2025 Thu Oct 16 15:12:04 EDT 2025 Thu May 22 21:22:11 EDT 2025 Thu Apr 03 07:01:55 EDT 2025 Thu Apr 24 23:13:56 EDT 2025 Wed Oct 01 03:43:39 EDT 2025 |
| IsDoiOpenAccess | true |
| IsOpenAccess | true |
| IsPeerReviewed | true |
| IsScholarly | true |
| Issue | 1 |
| Language | English |
| License | This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. cc-by Creative Commons Attribution License |
| LinkModel | DirectLink |
| MergedId | FETCHMERGED-LOGICAL-c691t-83890ba533d0275e7c3c00321ecfe37665fc480decfdcdab04504e53608880273 |
| Notes | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 ObjectType-Article-2 ObjectType-Feature-1 content type line 23 Conceived and designed the experiments: EG MF. Performed the experiments: JA EM MB. Analyzed the data: EG. Contributed reagents/materials/analysis tools: MB MF. Wrote the paper: EG MF. |
| OpenAccessLink | http://journals.scholarsportal.info/openUrl.xqy?doi=10.1371/journal.pone.0030065 |
| PMID | 22238693 |
| PQID | 1322478448 |
| PQPubID | 1436336 |
| ParticipantIDs | plos_journals_1322478448 doaj_primary_oai_doaj_org_article_c89d1f7fedf24dd18d5cb138cff583f8 unpaywall_primary_10_1371_journal_pone_0030065 pubmedcentral_primary_oai_pubmedcentral_nih_gov_3253126 proquest_miscellaneous_916147426 proquest_journals_1322478448 gale_infotracmisc_A477169842 gale_infotracacademiconefile_A477169842 gale_incontextgauss_ISR_A477169842 gale_incontextgauss_IOV_A477169842 gale_healthsolutions_A477169842 pubmed_primary_22238693 crossref_primary_10_1371_journal_pone_0030065 crossref_citationtrail_10_1371_journal_pone_0030065 |
| ProviderPackageCode | CITATION AAYXX |
| PublicationCentury | 2000 |
| PublicationDate | 2012-01-06 |
| PublicationDateYYYYMMDD | 2012-01-06 |
| PublicationDate_xml | – month: 01 year: 2012 text: 2012-01-06 day: 06 |
| PublicationDecade | 2010 |
| PublicationPlace | United States |
| PublicationPlace_xml | – name: United States – name: San Francisco – name: San Francisco, USA |
| PublicationTitle | PloS one |
| PublicationTitleAlternate | PLoS One |
| PublicationYear | 2012 |
| Publisher | Public Library of Science Public Library of Science (PLoS) |
| Publisher_xml | – name: Public Library of Science – name: Public Library of Science (PLoS) |
| References | H Khusun (ref3) 1999; 129 MR Macknet (ref6) 2010; 111 E Gayat (ref4) 2011 H von Schenck (ref12) 1986; 32 H Gehring (ref14) 2002; 46 C Ricos (ref10) 1999; 59 (ref2) 1994 JM Bland (ref9) 1986; 1 RG Hahn (ref5) 2010; 54 RD Miller (ref7) 2011; 112 PM Bossuyt (ref8) 2003; 326 JW Severinghaus (ref1) 1986; 2 E Gayat (ref13) 2011 MM Shoukri (ref11) 2004; 13 AM Lardi (ref15) 1998; 53 10460202 - J Nutr. 1999 Sep;129(9):1669-74 3948400 - Clin Chem. 1986 Mar;32(3):526-9 12190799 - Acta Anaesthesiol Scand. 2002 Sep;46(8):980-6 21385985 - Anesth Analg. 2011 Apr;112(4):858-63 21039345 - Acta Anaesthesiol Scand. 2010 Nov;54(10):1233-40 21658016 - Acta Anaesthesiol Scand. 2011 Aug;55(7):902 12511463 - BMJ. 2003 Jan 4;326(7379):41-4 10667686 - Scand J Clin Lab Invest. 1999 Nov;59(7):491-500 21048100 - Anesth Analg. 2010 Dec;111(6):1424-6 2868172 - Lancet. 1986 Feb 8;1(8476):307-10 3537215 - J Clin Monit. 1986 Oct;2(4):270-88 9613300 - Anaesthesia. 1998 Apr;53(4):349-52 21251729 - Ann Emerg Med. 2011 Apr;57(4):330-3 |
| References_xml | – volume: 13 start-page: 251 year: 2004 ident: ref11 article-title: Sample size requirements for the design of reliability study: review and new results. publication-title: Statistical Methods in Medical Research doi: 10.1191/0962280204sm365ra – volume: 32 start-page: 526 year: 1986 ident: ref12 article-title: Evaluation of “HemoCue,” a new device for determining hemoglobin. publication-title: Clin Chem doi: 10.1093/clinchem/32.3.526 – volume: 2 start-page: 270 year: 1986 ident: ref1 article-title: History of blood gas analysis. VI. Oximetry. publication-title: J Clin Monit doi: 10.1007/BF02851177 – year: 2011 ident: ref13 article-title: Instability in non-invasive haemoglobin measurement: a possible influence of oxygen administration. doi: 10.1111/j.1399-6576.2011.02469.x – volume: 59 start-page: 491 year: 1999 ident: ref10 article-title: Current databases on biological variation: pros, cons and progress. publication-title: Scand J Clin Lab Invest doi: 10.1080/00365519950185229 – volume: 111 start-page: 1424 year: 2010 ident: ref6 article-title: The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution. publication-title: Anesth Analg doi: 10.1213/ANE.0b013e3181fc74b9 – volume: 53 start-page: 349 year: 1998 ident: ref15 article-title: Evaluation of the HemoCue for measuring intra-operative haemoglobin concentrations: a comparison with the Coulter Max-M. publication-title: Anaesthesia doi: 10.1046/j.1365-2044.1998.00328.x – year: 2011 ident: ref4 article-title: Performance Evaluation of a Noninvasive Hemoglobin Monitoring Device. doi: 10.1016/j.annemergmed.2010.11.032 – volume: 46 start-page: 980 year: 2002 ident: ref14 article-title: Accuracy of point-of-care-testing (POCT) for determining hemoglobin concentrations. publication-title: Acta Anaesthesiol Scand doi: 10.1034/j.1399-6576.2002.460809.x – volume: 326 start-page: 41 year: 2003 ident: ref8 article-title: Towards complete and accurate reporting of studies of diagnostic accuracy: the STARD initiative. publication-title: Bmj doi: 10.1136/bmj.326.7379.41 – year: 1994 ident: ref2 article-title: Indicators and Strategies for Iron Deficiency and Anemia Programmes. – volume: 129 start-page: 1669 year: 1999 ident: ref3 article-title: World Health Organization hemoglobin cut-off points for the detection of anemia are valid for an Indonesian population. publication-title: J Nutr doi: 10.1093/jn/129.9.1669 – volume: 54 start-page: 1233 year: 2010 ident: ref5 article-title: Non-invasive monitoring of blood haemoglobin for analysis of fluid volume kinetics. publication-title: Acta Anaesthesiol Scand doi: 10.1111/j.1399-6576.2010.02321.x – volume: 112 start-page: 858 year: 2011 ident: ref7 article-title: A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. publication-title: Anesth Analg doi: 10.1213/ANE.0b013e31820eecd1 – volume: 1 start-page: 307 year: 1986 ident: ref9 article-title: Statistical methods for assessing agreement between two methods of clinical measurement. publication-title: Lancet doi: 10.1016/S0140-6736(86)90837-8 – reference: 3948400 - Clin Chem. 1986 Mar;32(3):526-9 – reference: 2868172 - Lancet. 1986 Feb 8;1(8476):307-10 – reference: 21039345 - Acta Anaesthesiol Scand. 2010 Nov;54(10):1233-40 – reference: 10667686 - Scand J Clin Lab Invest. 1999 Nov;59(7):491-500 – reference: 12511463 - BMJ. 2003 Jan 4;326(7379):41-4 – reference: 21658016 - Acta Anaesthesiol Scand. 2011 Aug;55(7):902 – reference: 10460202 - J Nutr. 1999 Sep;129(9):1669-74 – reference: 21251729 - Ann Emerg Med. 2011 Apr;57(4):330-3 – reference: 21385985 - Anesth Analg. 2011 Apr;112(4):858-63 – reference: 9613300 - Anaesthesia. 1998 Apr;53(4):349-52 – reference: 12190799 - Acta Anaesthesiol Scand. 2002 Sep;46(8):980-6 – reference: 3537215 - J Clin Monit. 1986 Oct;2(4):270-88 – reference: 21048100 - Anesth Analg. 2010 Dec;111(6):1424-6 |
| SSID | ssj0053866 |
| Score | 2.3716972 |
| Snippet | Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for the... Background Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for... Background Measurement of blood hemoglobin (Hb) concentration is a routine procedure. Using a non-invasive point-of-care device reduces pain and discomfort for... |
| SourceID | plos doaj unpaywall pubmedcentral proquest gale pubmed crossref |
| SourceType | Open Website Open Access Repository Aggregation Database Index Database Enrichment Source |
| StartPage | e30065 |
| SubjectTerms | Accuracy Adult Aged Algorithms Analysis Anemia Anxiety - diagnosis Anxiety - etiology Bias Biological variation Biology Blood Blood levels Blood Specimen Collection - adverse effects Blood Specimen Collection - methods Blood Specimen Collection - psychology Blood Specimen Collection - statistics & numerical data Chemistry Clinical trials Colorimetry Confidence intervals Correlation coefficient Correlation coefficients Discordance Emergency medical care Emergency medical services Female Glycosylated hemoglobin Hemoglobin Hemoglobins - analysis Hospitalization - statistics & numerical data Human subjects Humans Laboratories Male Measurement Medical research Medicine Middle Aged Minimally Invasive Surgical Procedures - adverse effects Minimally Invasive Surgical Procedures - psychology Minimally Invasive Surgical Procedures - statistics & numerical data Monitoring, Physiologic - methods Multivariate analysis Pain Pain Measurement Patient care Patients Perfusion Point-of-Care Systems Political aspects Regression analysis Statistical analysis Statistical methods Studies |
| SummonAdditionalLinks | – databaseName: DOAJ Directory of Open Access Journals dbid: DOA link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwrV1Lb9QwELbQXuCCaHk00IKFkICDt3k4icOtIKqCRJGAop6wHD_albbOimSp-PfM5MVGVGoPHBNPLGVenrFnPhPyQqQaT7804xCMMp4mlikTF6w0XMF6EpeiRdf_dJwdnfCPp-npxlVfWBPWwQN3jNvXojCRy501LubGRMKkuowSoZ1LReLaNt9QFEMy1flgsOIs6xvlkjza7-UyX1XezlGvQ1xMNhaiFq9_9Mqz1bKqrwo5_62cvL32K_X7Ui2XG8vS4T1yt48n6UH3H1vklvXbZKu32Jq-6mGlX98nP44rzxb-l8KCdXrxd2-QVo6e24sKsUEW_g1VI1gnjjSXFR1uUWnoqlr4hlWOYcUYbYZ9eUi3H5CTw_ff3h2x_nYFprMiapgAGYWlgnDP4NGlzXWigTVxZLWz4Hay1GkuQgOPRhtVQuwXcpsmGfglgSg4D8nMAz93CFVtx2qYKETjKy0vYgOzu7DQeVRoqwKSDKyWuocexxswlrI9T8shBem4JVFAshdQQNj41aqD3riG_i1KcaRF4Oz2BaiT7NVJXqdOAXmGOiC7LtTR_OUBzxFXSPA4IM9bCgTP8Fidc6bWdS0_fP5-A6KvXyZEL3siVwE7tOo7IuCfEJRrQrk7oQQXoCfDO6ixA1dqiVsMIBFIveHLQYuvHqbjME6KFXfeVutaQtoQ8Rzit4A86nR-ZCyGlCIrkoDkE2uYcH464hfnLXR5EoPPxznno93cSLaP_4dsn5A7EO7G7QZatktmzc-13YOQsimftt7jD42fd0A priority: 102 providerName: Directory of Open Access Journals – databaseName: ProQuest Central dbid: BENPR link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV1Lb9QwELbK9gAXRHk1UCBCSMDBbR5O4iAhtIVWW6QuVWlRT40cP8pKS7w0u1T8e2ayTtqICnpMPLGUeXlsz3xDyCueSLz9kpRBMEpZEmsqVJTTUjEB60lU8gZdf3-cjo7Z55PkZIWM21oYTKtsfWLjqJWVeEa-hbsmlnHYTXyY_aTYNQpvV9sWGsK1VlDvG4ixW2Q1QmSsAVnd3hkfHLa-Gaw7TV0BXZyFW05emzNb6U3U9wAXmSsLVIPj33nrwWxq6-tC0b8zKm8vqpn4fSGm0yvL1e49ctfFmf5wqRhrZEVX98mas-Taf-Pgpt8-IKdjW9G96pfARHZ___LM0LfGH-kfFjFDJtU7f9iBeOLI0YX1P7nuKnP_wE6qObWGYj2T353Xwzb8ITne3Tn6OKKu6wKVaR7OKQfZBaWAMFDhlabOZCyBNVGopdHgjtLESMYDBY9KKlFCTBgwncQp-CuO6DiPyKACfq4TXzSVrEEsEKWv1CyPFMxuglxmYS618EjcsrqQDpIcO2NMi-aeLYOtyZJbBQqocALyCO2-mi0hOf5Dv41S7GgRULt5Yc_PCmefheS5Ck1mtDIRUyrkKpFlGHNpTMJjwz3yAnWgWFandm6hGLIM8YY4izzysqFAUI0Ks3bOxKKui70v325A9PWwR_TaERkL7JDCVUrAPyFYV49yo0cJrkH2htdRY1uu1MWlEcGXrRZfP-x3wzgpZuJV2i7qArYTIcsgrvPI46XOd4zFUJOneeyRrGcNPc73R6rJ9wbSPI5gLcA5Nzu7uZFsn_z7N56SOxDgRs2RWbpBBvPzhX4GQeS8fO48wx8AjXJd priority: 102 providerName: ProQuest – databaseName: Unpaywall dbid: UNPAY link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lb9QwELbK9gAXoLwaWCBCiMchaR5O4nBbHtUWqUsFLWoPEDl-wIolWZEsFRz47cwkTiBQRDlw211PnPXn8Xjs8Xwm5C6LBEa_hEPBGXVoFCqHyyB1ckk5zCdBzhp2_d1ZPD2gzw-jwzXypsuFMQjCGnFRVk0kHz-UhdoySG4hX1EbPXX9MPG7J9wlCLmoszCp3msYh3BnrMYEpDNkPY7AVR-R9YPZ3uSojTQHThx4oUmn-1NNg-mqYfXvbfcI_9lJjunv5yvProol_3LMF4ufJq_tC-Rb1-z2zMoHd1Xnrvj6CyPkf8PlIjlv3F570tayQdZUcYlsGMNS2Q8M-_XDy-TtrCycneIzx3P19u6PLUy71PZUfSyRwmRePLInPacoluwfl_ZTc9lLbe-V8G6n1A6mV9l9-GCuqivkYPvZ_pOpYy6BcESc-rXDQJW8nINXKjHCqhIRCmhN4CuhFVjHONKCMk_CVykkz8FF9aiKwhjMJ0OynqtkVAAEm8TmTWKtF3IkDcwVTQMJtWsvFYmfCsUtEnZ9nQnDkI4XdSyyJuyXwEqpRStDTDODqUWc_qllyxDyF_nHqEa9LPJ7Nz9Ap2amMzPBUunrRCupAyqlz2Qkcj9kQuuIhZpZ5DYqYdYmy_ZWKpvQBOmPGA0scqeRQI6PAg8RveOrqsp2Xrw-hdCrlwOh-0ZIlwCH4CZxA9qEOjeQHA8kwVKJQfEmKm2HSpXhTgj0CKXQnnE3jE4utvtirBQPBhaqXFUZrG58moCbaZFr7aDrgUXPl8VpaJFkMBwHyA9Livn7hmE9DGBqwjrdfuCeqm-v_-sDN8g58MCDZk8vHpNR_WmlboKXW-e3jK36Drpuq3E priority: 102 providerName: Unpaywall |
| Title | Non-Invasive Measurement of Hemoglobin: Assessment of Two Different Point-of-Care Technologies |
| URI | https://www.ncbi.nlm.nih.gov/pubmed/22238693 https://www.proquest.com/docview/1322478448 https://www.proquest.com/docview/916147426 https://pubmed.ncbi.nlm.nih.gov/PMC3253126 https://journals.plos.org/plosone/article/file?id=10.1371/journal.pone.0030065&type=printable https://doaj.org/article/c89d1f7fedf24dd18d5cb138cff583f8 http://dx.doi.org/10.1371/journal.pone.0030065 |
| UnpaywallVersion | publishedVersion |
| Volume | 7 |
| hasFullText | 1 |
| inHoldings | 1 |
| isFullTextHit | |
| isPrint | |
| journalDatabaseRights | – providerCode: PRVFSB databaseName: Free Full-Text Journals in Chemistry customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: HH5 dateStart: 20060101 isFulltext: true titleUrlDefault: http://abc-chemistry.org/ providerName: ABC ChemistRy – providerCode: PRVAFT databaseName: Open Access Digital Library customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: KQ8 dateStart: 20060101 isFulltext: true titleUrlDefault: http://grweb.coalliance.org/oadl/oadl.html providerName: Colorado Alliance of Research Libraries – providerCode: PRVAFT databaseName: Open Access Digital Library customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: KQ8 dateStart: 20061001 isFulltext: true titleUrlDefault: http://grweb.coalliance.org/oadl/oadl.html providerName: Colorado Alliance of Research Libraries – providerCode: PRVAON databaseName: DOAJ Directory of Open Access Journals customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: DOA dateStart: 20060101 isFulltext: true titleUrlDefault: https://www.doaj.org/ providerName: Directory of Open Access Journals – providerCode: PRVEBS databaseName: EBSCOhost Academic Search Ultimate customDbUrl: https://search.ebscohost.com/login.aspx?authtype=ip,shib&custid=s3936755&profile=ehost&defaultdb=asn eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: ABDBF dateStart: 20080101 isFulltext: true titleUrlDefault: https://search.ebscohost.com/direct.asp?db=asn providerName: EBSCOhost – providerCode: PRVEBS databaseName: EBSCOhost Food Science Source customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: false ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: A8Z dateStart: 20080101 isFulltext: true titleUrlDefault: https://search.ebscohost.com/login.aspx?authtype=ip,uid&profile=ehost&defaultdb=fsr providerName: EBSCOhost – providerCode: PRVBFR databaseName: Free Medical Journals customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: DIK dateStart: 20060101 isFulltext: true titleUrlDefault: http://www.freemedicaljournals.com providerName: Flying Publisher – providerCode: PRVFQY databaseName: GFMER Free Medical Journals customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: GX1 dateStart: 20060101 isFulltext: true titleUrlDefault: http://www.gfmer.ch/Medical_journals/Free_medical.php providerName: Geneva Foundation for Medical Education and Research – providerCode: PRVHPJ databaseName: ROAD: Directory of Open Access Scholarly Resources customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: M~E dateStart: 20060101 isFulltext: true titleUrlDefault: https://road.issn.org providerName: ISSN International Centre – providerCode: PRVAQN databaseName: PubMed Central customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: RPM dateStart: 20060101 isFulltext: true titleUrlDefault: https://www.ncbi.nlm.nih.gov/pmc/ providerName: National Library of Medicine – providerCode: PRVPQU databaseName: Health & Medical Collection customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: 7X7 dateStart: 20061201 isFulltext: true titleUrlDefault: https://search.proquest.com/healthcomplete providerName: ProQuest – providerCode: PRVPQU databaseName: ProQuest Central customDbUrl: http://www.proquest.com/pqcentral?accountid=15518 eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: BENPR dateStart: 20061201 isFulltext: true titleUrlDefault: https://www.proquest.com/central providerName: ProQuest – providerCode: PRVPQU databaseName: ProQuest Technology Collection customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: 8FG dateStart: 20061201 isFulltext: true titleUrlDefault: https://search.proquest.com/technologycollection1 providerName: ProQuest – providerCode: PRVPQU databaseName: Public Health Database customDbUrl: eissn: 1932-6203 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: 8C1 dateStart: 20061201 isFulltext: true titleUrlDefault: https://search.proquest.com/publichealth providerName: ProQuest – providerCode: PRVFZP databaseName: Scholars Portal Journals: Open Access customDbUrl: eissn: 1932-6203 dateEnd: 20250930 omitProxy: true ssIdentifier: ssj0053866 issn: 1932-6203 databaseCode: M48 dateStart: 20061201 isFulltext: true titleUrlDefault: http://journals.scholarsportal.info providerName: Scholars Portal |
| link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV1bb9MwFLa27gFeEOO2wCgRQgIeXOXiJA4SQt1Y2ZBaqrGi7oUosZ1RqSSlaRn795zjuoGIou0lUuJjSzl3375DyAseCNz9EpRBMkpZ4CuaSi-mmWQpxBMv4xpdvz8Ij0fs4zgYb5F1zVbDwGrj1A7rSY3m086vH1fvwODf6qoNkbvu1JmVheqg1kJY3SY7EKtiLObQZ_W-Ali33r3ErIWGnuOby3T_G6URrDSmf-25W7NpWW1KS_89XXlrWczSq8t0Ov0rdPXukjsm57S7KyXZJVuquEd2jVVX9isDPf36Pvk6KAt6UvxM8VC73f-zfmiXuX2svpeIHzIp3tjdGtATW84uS_u9qbSysIflpFjQMqd4t8mu1-5hSv6AjHpHZ4fH1FRgoCKM3QXlIEcnSyEllLi9qSLhC2CN5yqRK3BNYZALxh0Jr1LINIP80GEq8EPwXRyRch6SVgH83CN2qm-1On6KiH2ZYrEnYfTciUXkxkKlFvHXrE6EgSfHKhnTRO-5RTBNWXErQQElRkAWoXWv2Qqe4xr6A5RiTYvg2vpDOb9IjK0mgsfSzaNcydxjUrpcBiJzfS7yPOB-zi3yDHUgWd1UrV1E0mURYg9x5lnkuaZAgI0CT_BcpMuqSk4-fbkB0efTBtFLQ5SXwA6RmlsT8E8I3NWg3G9QgpsQjeY91Ng1V6oElyFAIjA9h55rLd7cbNfNOCieyitUuawSmFq4LIIczyKPVjpfMxbTTh7GvkWihjU0ON9sKSbfNLy570FcwDE7td3cSLaPrxXLE3Ib8l1Pr6CF-6S1mC_VU8gpF1mbbEfjCJ780MVn70Ob7BwcDYanbb1K09ZuBL6NBsPu-W8J3Hty |
| linkProvider | Scholars Portal |
| linkToHtml | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lb9NAEF6VcCgXRHnVUKiFQMDBqR9re42EUKBUCW0CgrTKCWPvo0QK3lAnRP1T_EZm_GotKuilx3jHq3h2nrs73xDylPkcT7-4RSEYtajvSSsRbmSlgibgT9yUFej6w1HQP6QfJv5kjfyua2HwWmVtEwtDLTTHPfIdzJpoyCCbeDP_aWHXKDxdrVtolGKxL09XkLLlrwe7sL7PXHfv_fhd36q6Clg8iJyFxeC_2WkCYY7AIzsZco-DaLuO5EqCugW-4pTZAn4KLpIUYh6bSt8LQB8Zor_AvNfIdeqBLQH9CSdNgge2Iwiq8jwvdHYqaejOdSa7qE02urBz7q_oEtD4gs58pvOLAt2_72uuL7N5crpKZrNzznDvFrlZRbFmrxS7DbIms9tko7ITufmiArN-eYd8HenMGmS_Erwmbw7PdiRNrcy-_KERkWSavTJ7DUQojoxX2tytercszE96mi0srSysljKb0wBI8u-Swyvh_j3SyYCfm8RMijpZ20sQAzCVNHIFzK7siIdOxGViEK9mdcwrwHPsuzGLi1O8EBKfklsxLlBcLZBBrOateQn48R_6t7iKDS3CdRcP9MlxXGl_zFkkHBUqKZRLhXCY8HnqeIwr5TNPMYNsowzEZe1rY3TiHg0RzYhR1yBPCgqE7MjwTtBxsszzePDx6BJEXz63iJ5XREoDO3hS1WHANyEUWItyq0UJhoe3hjdRYmuu5PGZisKbtRRfPGw2wzgp3vPLpF7mMSQrDg0hajTI_VLmG8ZiIMuCyDNI2NKGFufbI9n0ewGY7rngaXDObqM3l1rbB__-jG2y3h8PD-KDwWj_IbkBobRbbM4FW6SzOFnKRxCuLtLHhY0wyberNkp_ACBgp6Y |
| linkToPdf | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV3db9NADD-NIgEviPG1wGARAgEPafOdCxJChTK1jJWJbahPhMt9jEolKUtLtX-Nvw47uWSLmGAve2zOOTU-22ff2T8T8pQGHG-_uOWDM2r5gSctJtzYSoXPYD9xU1qi6--Ow-Gh_2ESTNbI77oWBtMqa5tYGmqRczwj72HU5EcUoome0mkRe4PtN_OfFnaQwpvWup1GJSI78mQF4VvxejSAtX7mutvvD94NLd1hwOJh7CwsCv_TThm4PAKv72TEPQ5i7jqSKwmqFwaK-9QW8FNwwVLwf2xfBl4IukkRCQbmvUKuRp4XYzphNGmCPbAjYahL9bzI6WnJ6M7zTHZRs2zczs5shWXHgGZf6MxneXGe0_t37ub1ZTZnJys2m53ZGLdvkZvaozX7lQiukzWZ3Sbr2mYU5gsNbP3yDvk6zjNrlP1imDJv7p6eTpq5MofyR47oJNPsldlv4EJx5GCVmwPdx2Vh7uXTbGHlysLKKbO5GYCA_y45vBTu3yOdDPi5QUxW1szaHkM8wFT6sStgdmXHPHJiLplBvJrVCdfg59iDY5aUN3oRBEEVtxJcoEQvkEGs5q15Bf7xH_q3uIoNLUJ3lw_y46NEW4KE01g4KlJSKNcXwqEi4KnjUa5UQD1FDbKFMpBUdbCNAUr6foTIRtR3DfKkpED4jgwV4YgtiyIZffpyAaL9zy2i55pI5cAOznRNBnwTwoK1KDdblGCEeGt4AyW25kqRnKorvFlL8fnDZjOMk2LOXybzZZFA4OL4EXiQBrlfyXzDWHRqaRh7Bola2tDifHskm34vwdM9F3YdnLPb6M2F1vbBvz9ji1wDc5R8HI13HpIb4FW75TlduEk6i-OlfASe6yJ9XJoIk3y7bJv0B04eq-k |
| linkToUnpaywall | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lb9QwELbK9gAXoLwaWCBCiMchaR5O4nBbHtUWqUsFLWoPEDl-wIolWZEsFRz47cwkTiBQRDlw211PnPXn8Xjs8Xwm5C6LBEa_hEPBGXVoFCqHyyB1ckk5zCdBzhp2_d1ZPD2gzw-jwzXypsuFMQjCGnFRVk0kHz-UhdoySG4hX1EbPXX9MPG7J9wlCLmoszCp3msYh3BnrMYEpDNkPY7AVR-R9YPZ3uSojTQHThx4oUmn-1NNg-mqYfXvbfcI_9lJjunv5yvProol_3LMF4ufJq_tC-Rb1-z2zMoHd1Xnrvj6CyPkf8PlIjlv3F570tayQdZUcYlsGMNS2Q8M-_XDy-TtrCycneIzx3P19u6PLUy71PZUfSyRwmRePLInPacoluwfl_ZTc9lLbe-V8G6n1A6mV9l9-GCuqivkYPvZ_pOpYy6BcESc-rXDQJW8nINXKjHCqhIRCmhN4CuhFVjHONKCMk_CVykkz8FF9aiKwhjMJ0OynqtkVAAEm8TmTWKtF3IkDcwVTQMJtWsvFYmfCsUtEnZ9nQnDkI4XdSyyJuyXwEqpRStDTDODqUWc_qllyxDyF_nHqEa9LPJ7Nz9Ap2amMzPBUunrRCupAyqlz2Qkcj9kQuuIhZpZ5DYqYdYmy_ZWKpvQBOmPGA0scqeRQI6PAg8RveOrqsp2Xrw-hdCrlwOh-0ZIlwCH4CZxA9qEOjeQHA8kwVKJQfEmKm2HSpXhTgj0CKXQnnE3jE4utvtirBQPBhaqXFUZrG58moCbaZFr7aDrgUXPl8VpaJFkMBwHyA9Livn7hmE9DGBqwjrdfuCeqm-v_-sDN8g58MCDZk8vHpNR_WmlboKXW-e3jK36Drpuq3E |
| openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Non-Invasive+Measurement+of+Hemoglobin%3A+Assessment+of+Two+Different+Point-of-Care+Technologies&rft.jtitle=PloS+one&rft.au=Boisson%2C+Mireille&rft.au=Gayat%2C+Etienne&rft.au=Aulagnier%2C+J%C3%A9r%C3%B4me&rft.au=Matthieu%2C+Emmanuel&rft.date=2012-01-06&rft.pub=Public+Library+of+Science&rft.issn=1932-6203&rft.eissn=1932-6203&rft.volume=7&rft.issue=1&rft_id=info:doi/10.1371%2Fjournal.pone.0030065&rft.externalDBID=n%2Fa&rft.externalDocID=A477169842 |
| thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1932-6203&client=summon |
| thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1932-6203&client=summon |
| thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1932-6203&client=summon |