One‐year analysis of Elekta CBCT image quality using NPS and MTF
The image quality (IQ) of imaging systems must be sufficiently high for image‐guided radiation therapy (IGRT). Hence, users should implement a quality assurance program to maintain IQ. In our routine IQ tests of the kV cone‐beam CT system (Elekta XVI), image noise was quantified by noise standard de...
Saved in:
| Published in | Journal of Applied Clinical Medical Physics Vol. 17; no. 3; pp. 211 - 222 |
|---|---|
| Main Authors | , , |
| Format | Journal Article |
| Language | English |
| Published |
United States
Wiley
01.05.2016
John Wiley & Sons, Inc John Wiley and Sons Inc |
| Subjects | |
| Online Access | Get full text |
| ISSN | 1526-9914 1526-9914 |
| DOI | 10.1120/jacmp.v17i3.6047 |
Cover
| Abstract | The image quality (IQ) of imaging systems must be sufficiently high for image‐guided radiation therapy (IGRT). Hence, users should implement a quality assurance program to maintain IQ. In our routine IQ tests of the kV cone‐beam CT system (Elekta XVI), image noise was quantified by noise standard deviation (NSD), which was the standard deviation of CT numbers measured in a small area in an image of an IQ test phantom (Catphan), and the high spatial resolution (HSR) was evaluated by the number of line‐pairs (LPN) visually recognizable in the image. We also measured the image uniformity, the low contrast resolution, and the distances of two points for geometrical accuracy. For this study, we did an additional evaluation of the XVI data for 12 monthly IQ tests by using noise power spectrum (NPS) for noise, modulation transfer function (MTF) for HSR, and CT number‐to‐density relationship. NPS was obtained by applying Fourier analysis in a small area on the uniformity test section of Catphan. The MTF analysis was performed by applying the Droege‐Morin (D‐M) method to the line‐pair bar regions in the phantom. The CT number‐to‐density relationship was obtained for insert materials in the low‐contrast test section of the phantom. All the quantities showed a noticeable change over the one‐year period. Especially the noise level improved significantly after a repair of the imager. NPS was more sensitive to the IQ change than NSD. MTF could provide more quantitative and objective evaluation of HSR. The CT number was very different from the expected CT number, but the CT number‐to‐density curves were constant within 5% except for two months. Since the D‐M method is easy to implement, we recommend using MTF instead of LPN even for routine QA. The IQ of the imaging systems was constantly changing; hence, IQ tests should be periodically performed. Additionally, we found the importance of IQ tests after every service work, including detector calibration as well as preventive maintenance.
PACS number(s): 87.56Da, 87.57.C‐, 87.57.N‐, 8757.Q‐ |
|---|---|
| AbstractList | The image quality (IQ) of imaging systems must be sufficiently high for image‐guided radiation therapy (IGRT). Hence, users should implement a quality assurance program to maintain IQ. In our routine IQ tests of the kV cone‐beam CT system (Elekta XVI), image noise was quantified by noise standard deviation (NSD), which was the standard deviation of CT numbers measured in a small area in an image of an IQ test phantom (Catphan), and the high spatial resolution (HSR) was evaluated by the number of line‐pairs (LPN) visually recognizable in the image. We also measured the image uniformity, the low contrast resolution, and the distances of two points for geometrical accuracy. For this study, we did an additional evaluation of the XVI data for 12 monthly IQ tests by using noise power spectrum (NPS) for noise, modulation transfer function (MTF) for HSR, and CT number‐to‐density relationship. NPS was obtained by applying Fourier analysis in a small area on the uniformity test section of Catphan. The MTF analysis was performed by applying the Droege‐Morin (D‐M) method to the line‐pair bar regions in the phantom. The CT number‐to‐density relationship was obtained for insert materials in the low‐contrast test section of the phantom. All the quantities showed a noticeable change over the one‐year period. Especially the noise level improved significantly after a repair of the imager. NPS was more sensitive to the IQ change than NSD. MTF could provide more quantitative and objective evaluation of HSR. The CT number was very different from the expected CT number, but the CT number‐to‐density curves were constant within 5% except for two months. Since the D‐M method is easy to implement, we recommend using MTF instead of LPN even for routine QA. The IQ of the imaging systems was constantly changing; hence, IQ tests should be periodically performed. Additionally, we found the importance of IQ tests after every service work, including detector calibration as well as preventive maintenance. PACS number(s): 87.56Da, 87.57.C‐, 87.57.N‐, 8757.Q‐ The image quality (IQ) of imaging systems must be sufficiently high for image‐guided radiation therapy (IGRT). Hence, users should implement a quality assurance program to maintain IQ. In our routine IQ tests of the kV cone‐beam CT system (Elekta XVI), image noise was quantified by noise standard deviation (NSD), which was the standard deviation of CT numbers measured in a small area in an image of an IQ test phantom (Catphan), and the high spatial resolution (HSR) was evaluated by the number of line‐pairs (LPN) visually recognizable in the image. We also measured the image uniformity, the low contrast resolution, and the distances of two points for geometrical accuracy. For this study, we did an additional evaluation of the XVI data for 12 monthly IQ tests by using noise power spectrum (NPS) for noise, modulation transfer function (MTF) for HSR, and CT number‐to‐density relationship. NPS was obtained by applying Fourier analysis in a small area on the uniformity test section of Catphan. The MTF analysis was performed by applying the Droege‐Morin (D‐M) method to the line‐pair bar regions in the phantom. The CT number‐to‐density relationship was obtained for insert materials in the low‐contrast test section of the phantom. All the quantities showed a noticeable change over the one‐year period. Especially the noise level improved significantly after a repair of the imager. NPS was more sensitive to the IQ change than NSD. MTF could provide more quantitative and objective evaluation of HSR. The CT number was very different from the expected CT number, but the CT number‐to‐density curves were constant within 5% except for two months. Since the D‐M method is easy to implement, we recommend using MTF instead of LPN even for routine QA. The IQ of the imaging systems was constantly changing; hence, IQ tests should be periodically performed. Additionally, we found the importance of IQ tests after every service work, including detector calibration as well as preventive maintenance. PACS number(s): 87.56Da, 87.57.C‐, 87.57.N‐, 8757.Q‐ The image quality (IQ) of imaging systems must be sufficiently high for image-guided radiation therapy (IGRT). Hence, users should implement a quality assurance program to maintain IQ. In our routine IQ tests of the kV cone-beam CT system (Elekta XVI), image noise was quantified by noise standard deviation (NSD), which was the standard deviation of CT numbers measured in a small area in an image of an IQ test phantom (Catphan), and the high spatial resolution (HSR) was evaluated by the number of line-pairs (LPN) visually recognizable in the image. We also measured the image uniformity, the low contrast resolution, and the distances of two points for geometrical accuracy. For this study, we did an additional evaluation of the XVI data for 12 monthly IQ tests by using noise power spectrum (NPS) for noise, modulation transfer function (MTF) for HSR, and CT number-to-density relationship. NPS was obtained by applying Fourier analysis in a small area on the uniformity test section of Catphan. The MTF analysis was performed by applying the Droege-Morin (D-M) method to the line-pair bar regions in the phantom. The CT number-to-density relationship was obtained for insert materials in the low-contrast test section of the phantom. All the quantities showed a noticeable change over the one-year period. Especially the noise level improved significantly after a repair of the imager. NPS was more sensitive to the IQ change than NSD. MTF could provide more quantitative and objective evaluation of HSR. The CT number was very different from the expected CT number, but the CT number-to-density curves were constant within 5% except for two months. Since the D-M method is easy to implement, we recommend using MTF instead of LPN even for routine QA. The IQ of the imaging systems was constantly changing; hence, IQ tests should be periodically performed. Additionally, we found the importance of IQ tests after every service work, including detector calibration as well as preventive maintenance. |
| Author | Yoichi Watanabe Satomi Nakahara Masayuki Tachibana |
| AuthorAffiliation | 2 Department of Radiation Oncology University of Minnesota Minneapolis MN USA 1 Department of Diagnostic Radiology Hiroshima International University Higashi Hiroshima Hiroshima Japan |
| AuthorAffiliation_xml | – name: 2 Department of Radiation Oncology University of Minnesota Minneapolis MN USA – name: 1 Department of Diagnostic Radiology Hiroshima International University Higashi Hiroshima Hiroshima Japan |
| Author_xml | – sequence: 1 givenname: Satomi surname: Nakahara fullname: Nakahara, Satomi organization: Hiroshima International University – sequence: 2 givenname: Masayuki surname: Tachibana fullname: Tachibana, Masayuki organization: Hiroshima International University – sequence: 3 givenname: Yoichi surname: Watanabe fullname: Watanabe, Yoichi email: watan016@umn.edu organization: University of Minnesota |
| BackLink | https://cir.nii.ac.jp/crid/1873398392620873728$$DView record in CiNii https://www.ncbi.nlm.nih.gov/pubmed/27167279$$D View this record in MEDLINE/PubMed |
| BookMark | eNqNkc1u1DAUhS1UREvpnhWyBAs2M_jf8QapjVpAaikSw9ryODeDi8eZxkmr7HgEnpEnIekUKN3QjX0ln--ee4-fop3UJEDoOSVzShl5c-H8ejO_ojrwuSJCP0J7VDI1M4aKnTv1LjrIOSwJYVIQKvkTtMs0VZpps4eOzhP8_P5jANdil1wccsi4qfFxhG-dw-VRucBh7VaAL3sXQzfgPoe0wh8_fR71FT5bnDxDj2sXMxzc3vvoy8nxonw_Oz1_96E8PJ15aQSdAShQsjI1caL2RgpgwggDwviCVKyWldBMGqeo9L4mggJjy0IJzqXSlBO-j-i2b582brh2MdpNO87WDpYSO0VibyKxN5HYKZKRebtlNv1yDZWH1LXuL9e4YP99SeGrXTVXVipDDONjg9e3Ddrmsofc2XXIHmJ0CZo-W6oLQ7goyOT18p70ounbMdNsGTOEFFSoaYsXdyf6M8rvLxkFZCvwbZNzC_VDtlT3EB8614Vm2inEB4DXIcLwXyN7WJ4xwigdwVdbMIUwmk0nLTTnpuCGKUbGWrOC_wJvdc0D |
| CitedBy_id | crossref_primary_10_1007_s12194_023_00708_3 crossref_primary_10_1088_1742_6596_851_1_012029 crossref_primary_10_1002_acm2_13727 crossref_primary_10_1002_mp_16463 crossref_primary_10_1002_acm2_12476 crossref_primary_10_1088_2057_1976_abb2f8 crossref_primary_10_1002_mp_14992 crossref_primary_10_1097_IJG_0000000000001260 crossref_primary_10_32628_IJSRST52310632 crossref_primary_10_1088_1742_6596_1505_1_012039 crossref_primary_10_1007_s13246_020_00870_7 crossref_primary_10_5624_isd_2019_49_2_131 crossref_primary_10_1017_S1460396917000218 |
| Cites_doi | 10.1118/1.595124 10.1007/s00066-011-2236-4 10.1118/1.598243 10.1118/1.2900110 10.1118/1.599009 10.1120/jacmp.v12i2.3435 10.4103/0971-6203.41276 10.1118/1.3190392 10.1016/j.radonc.2009.07.012 |
| ContentType | Journal Article |
| Copyright | 2016 The Authors. 2016. This work is published under http://creativecommons.org/licenses/by/3.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. |
| Copyright_xml | – notice: 2016 The Authors. – notice: 2016. This work is published under http://creativecommons.org/licenses/by/3.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. |
| DBID | RYH 24P AAYXX CITATION CGR CUY CVF ECM EIF NPM 3V. 7X7 7XB 88I 8FI 8FJ 8FK ABUWG AFKRA AZQEC BENPR CCPQU DWQXO FYUFA GHDGH GNUQQ HCIFZ K9. M0S M2P PHGZM PHGZT PIMPY PJZUB PKEHL PPXIY PQEST PQQKQ PQUKI PRINS Q9U 7X8 5PM ADTOC UNPAY |
| DOI | 10.1120/jacmp.v17i3.6047 |
| DatabaseName | CiNii Complete Wiley Online Library Open Access CrossRef Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed ProQuest Central (Corporate) Health & Medical Collection ProQuest Central (purchase pre-March 2016) Science Database (Alumni Edition) Hospital Premium Collection Hospital Premium Collection (Alumni Edition) ProQuest Central (Alumni) (purchase pre-March 2016) ProQuest Central (Alumni) ProQuest Central ProQuest Central Essentials Proquest Central ProQuest One ProQuest Central Korea Health Research Premium Collection Health Research Premium Collection (Alumni) ProQuest Central Student SciTech Premium Collection ProQuest Health & Medical Complete (Alumni) ProQuest Health & Medical Collection Science Database ProQuest Central Premium ProQuest One Academic Publicly Available Content Database (ProQuest) ProQuest Health & Medical Research Collection ProQuest One Academic Middle East (New) ProQuest One Health & Nursing ProQuest One Academic Eastern Edition (DO NOT USE) ProQuest One Academic ProQuest One Academic UKI Edition ProQuest Central China ProQuest Central Basic MEDLINE - Academic PubMed Central (Full Participant titles) Unpaywall for CDI: Periodical Content Unpaywall |
| DatabaseTitle | CrossRef MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) Publicly Available Content Database ProQuest Central Student ProQuest One Academic Middle East (New) ProQuest Central Essentials ProQuest Health & Medical Complete (Alumni) ProQuest Central (Alumni Edition) SciTech Premium Collection ProQuest One Community College ProQuest One Health & Nursing ProQuest Central China ProQuest Central ProQuest Health & Medical Research Collection Health Research Premium Collection Health and Medicine Complete (Alumni Edition) ProQuest Central Korea ProQuest Central (New) ProQuest Science Journals (Alumni Edition) ProQuest Central Basic ProQuest Science Journals ProQuest One Academic Eastern Edition ProQuest Hospital Collection Health Research Premium Collection (Alumni) ProQuest Hospital Collection (Alumni) ProQuest Health & Medical Complete ProQuest One Academic UKI Edition ProQuest One Academic ProQuest One Academic (New) ProQuest Central (Alumni) MEDLINE - Academic |
| DatabaseTitleList | CrossRef MEDLINE - Academic Publicly Available Content Database MEDLINE |
| Database_xml | – sequence: 1 dbid: 24P name: Wiley Online Library Open Access url: https://authorservices.wiley.com/open-science/open-access/browse-journals.html sourceTypes: Publisher – sequence: 2 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database – sequence: 3 dbid: EIF name: MEDLINE url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search sourceTypes: Index Database – sequence: 4 dbid: UNPAY name: Unpaywall url: https://proxy.k.utb.cz/login?url=https://unpaywall.org/ sourceTypes: Open Access Repository – sequence: 5 dbid: BENPR name: ProQuest Central url: http://www.proquest.com/pqcentral?accountid=15518 sourceTypes: Aggregation Database |
| DeliveryMethod | fulltext_linktorsrc |
| Discipline | Medicine |
| EISSN | 1526-9914 |
| EndPage | 222 |
| ExternalDocumentID | 10.1120/jacmp.v17i3.6047 PMC5690923 27167279 10_1120_jacmp_v17i3_6047 ACM20211 |
| Genre | article Journal Article |
| GroupedDBID | 0R~ 1OC 24P 29J 2WC 53G 5GY 7X7 88I 8FI 8FJ AAHHS ABUWG ACCFJ ACCMX ACGFO ACXQS ADBBV ADKYN ADZMN ADZOD AEEZP AENEX AEQDE AFKRA AIWBW AJBDE ALIPV ALMA_UNASSIGNED_HOLDINGS ALUQN AOIJS AVUZU AZQEC BAWUL BCNDV BENPR BPHCQ BVXVI CCPQU CS3 DIK DU5 DWQXO E3Z EBS EJD FYUFA GNUQQ GROUPED_DOAJ GX1 HCIFZ HMCUK HYE IAO IHR KWQ M2P M~E OK1 OVT P6G PHGZM PHGZT PIMPY PQQKQ PROAC RNS RPM RYH TR2 UKHRP W2D XSB ADPDF EMOBN FRP H13 INH ITC OVD OVEED WIN AAMMB AAYXX AEFGJ AGXDD AIDQK AIDYY CITATION PUEGO CGR CUY CVF ECM EIF NPM 3V. 7XB 8FK K9. PJZUB PKEHL PPXIY PQEST PQUKI PRINS Q9U 7X8 5PM ADTOC AFFHD UNPAY |
| ID | FETCH-LOGICAL-c5941-ee6e65d9f0a4fc954e24949e49c80d2f5d47259a615ccf041e22b864335671303 |
| IEDL.DBID | UNPAY |
| ISSN | 1526-9914 |
| IngestDate | Wed Oct 29 12:17:07 EDT 2025 Tue Sep 30 16:44:58 EDT 2025 Thu Sep 04 18:50:56 EDT 2025 Tue Oct 07 07:27:46 EDT 2025 Mon Jul 21 05:58:10 EDT 2025 Wed Oct 01 02:06:40 EDT 2025 Thu Apr 24 23:06:07 EDT 2025 Wed Jan 22 16:46:51 EST 2025 Fri Jun 27 00:38:54 EDT 2025 |
| IsDoiOpenAccess | true |
| IsOpenAccess | true |
| IsPeerReviewed | true |
| IsScholarly | true |
| Issue | 3 |
| Language | English |
| License | Attribution http://creativecommons.org/licenses/by/3.0 This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. cc-by |
| LinkModel | DirectLink |
| MergedId | FETCHMERGED-LOGICAL-c5941-ee6e65d9f0a4fc954e24949e49c80d2f5d47259a615ccf041e22b864335671303 |
| Notes | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
| ORCID | 0000-0001-8086-6442 |
| OpenAccessLink | https://proxy.k.utb.cz/login?url=https://onlinelibrary.wiley.com/doi/pdfdirect/10.1120/jacmp.v17i3.6047 |
| PMID | 27167279 |
| PQID | 2290081460 |
| PQPubID | 4370306 |
| PageCount | 12 |
| ParticipantIDs | unpaywall_primary_10_1120_jacmp_v17i3_6047 pubmedcentral_primary_oai_pubmedcentral_nih_gov_5690923 proquest_miscellaneous_1789034807 proquest_journals_2290081460 pubmed_primary_27167279 crossref_primary_10_1120_jacmp_v17i3_6047 crossref_citationtrail_10_1120_jacmp_v17i3_6047 wiley_primary_10_1120_jacmp_v17i3_6047_ACM20211 nii_cinii_1873398392620873728 |
| ProviderPackageCode | CITATION AAYXX |
| PublicationCentury | 2000 |
| PublicationDate | May 2016 |
| PublicationDateYYYYMMDD | 2016-05-01 |
| PublicationDate_xml | – month: 05 year: 2016 text: May 2016 |
| PublicationDecade | 2010 |
| PublicationPlace | United States |
| PublicationPlace_xml | – name: United States – name: Malden Massachusetts – name: Hoboken |
| PublicationTitle | Journal of Applied Clinical Medical Physics |
| PublicationTitleAlternate | J Appl Clin Med Phys |
| PublicationYear | 2016 |
| Publisher | Wiley John Wiley & Sons, Inc John Wiley and Sons Inc |
| Publisher_xml | – name: Wiley – name: John Wiley & Sons, Inc – name: John Wiley and Sons Inc |
| References | 2009; 36 2000; 27 2009; 93 2009 2008; 25 1982; 9 2007 2008; 35 2008; 33 2011; 12 2003 2014 2011; 187 1998; 25 e_1_2_8_13_1 Ichikawa K (e_1_2_8_11_1) 2008; 25 e_1_2_8_15_1 Dunscombe PB (e_1_2_8_3_1) 2007 e_1_2_8_2_1 e_1_2_8_5_1 e_1_2_8_4_1 e_1_2_8_7_1 The Phantom Laboratory (e_1_2_8_14_1) 2014 e_1_2_8_6_1 e_1_2_8_9_1 e_1_2_8_8_1 Granfors PR (e_1_2_8_12_1) 2003 Ichikawa K (e_1_2_8_10_1) 2009 19893694 - J Med Phys. 2008 Apr;33(2):72-7 19810494 - Med Phys. 2009 Sep;36(9):4197-212 19695725 - Radiother Oncol. 2009 Oct;93(1):1-7 9608470 - Med Phys. 1998 May;25(5):614-28 21533757 - Strahlenther Onkol. 2011 May;187(5):284-91 10902561 - Med Phys. 2000 Jun;27(6):1311-23 18561655 - Med Phys. 2008 May;35(5):1807-15 21587192 - J Appl Clin Med Phys. 2011 Feb 04;12(2):3435 7155079 - Med Phys. 1982 Sep-Oct;9(5):758-60 |
| References_xml | – volume: 36 start-page: 4197 issue: 9 year: 2009 end-page: 212 article-title: Task Group 142 report: quality assurance of medical accelerators publication-title: Med Phys. – year: 2009 – volume: 93 start-page: 1 issue: 1 year: 2009 end-page: 7 article-title: Image quality and stability of image‐guided radiotherapy (IGRT) devices: a comparative study publication-title: Radiother Oncol. – volume: 187 start-page: 284 issue: 5 year: 2011 end-page: 91 article-title: kV cone‐beam CT‐based IGRT: a clinical review publication-title: Strahlenther Onkol. – year: 2003 article-title: DQE methodology—step by step – volume: 35 start-page: 1807 issue: 5 year: 2008 end-page: 15 article-title: A quality assurance program for image quality of cone‐beam CT guidance in radiation therapy publication-title: Med Phys. – year: 2007 article-title: Linac Integrated kV Imaging Systems and CBCT Simulators [draft report] – volume: 25 start-page: 614 issue: 5 year: 1998 end-page: 28 article-title: Signal, noise power spectrum, and detective quantum efficiency of indirect‐detection flat‐panel imagers for diagnostic radiology publication-title: Med Phys. – volume: 33 start-page: 72 issue: 2 year: 2008 end-page: 77 article-title: Commissioning and quality assurance of the X‐ray volume imaging system of an image‐guided radiotherapy capable linear accelerator publication-title: J Med Phys. – volume: 12 start-page: 3435 issue: 2 year: 2011 article-title: An image quality comparison study between XVI and OBI CBCT systems publication-title: J Appl Clin Med Phys. – year: 2014 – volume: 9 start-page: 758 issue: 5 year: 1982 end-page: 60 article-title: A practical method to measure the MTF of CT scanners publication-title: Med Phys. – volume: 25 start-page: 29 issue: 2 year: 2008 end-page: 34 article-title: Calculation methods for noise power spectrum measurement in computed tomography publication-title: Med Imaging Info Sci. – volume: 27 start-page: 1311 issue: 6 year: 2000 end-page: 23 article-title: Cone‐beam computed tomography with a flat‐panel imager: initial performance characterization publication-title: Med Phys. – volume: 25 start-page: 29 issue: 2 year: 2008 ident: e_1_2_8_11_1 article-title: Calculation methods for noise power spectrum measurement in computed tomography publication-title: Med Imaging Info Sci. – volume-title: CatPhan 500 and 600 manual year: 2014 ident: e_1_2_8_14_1 – ident: e_1_2_8_13_1 doi: 10.1118/1.595124 – ident: e_1_2_8_2_1 doi: 10.1007/s00066-011-2236-4 – ident: e_1_2_8_5_1 doi: 10.1118/1.598243 – ident: e_1_2_8_7_1 doi: 10.1118/1.2900110 – ident: e_1_2_8_6_1 doi: 10.1118/1.599009 – ident: e_1_2_8_15_1 doi: 10.1120/jacmp.v12i2.3435 – ident: e_1_2_8_8_1 doi: 10.4103/0971-6203.41276 – ident: e_1_2_8_4_1 doi: 10.1118/1.3190392 – ident: e_1_2_8_9_1 doi: 10.1016/j.radonc.2009.07.012 – volume-title: Standard X‐ray CT image measurement year: 2009 ident: e_1_2_8_10_1 – volume-title: Presented at the AAPM 45th Annual Meeting year: 2003 ident: e_1_2_8_12_1 – volume-title: The Canadian Organization of Medical Physicists (COMP) year: 2007 ident: e_1_2_8_3_1 – reference: 19893694 - J Med Phys. 2008 Apr;33(2):72-7 – reference: 18561655 - Med Phys. 2008 May;35(5):1807-15 – reference: 9608470 - Med Phys. 1998 May;25(5):614-28 – reference: 21587192 - J Appl Clin Med Phys. 2011 Feb 04;12(2):3435 – reference: 21533757 - Strahlenther Onkol. 2011 May;187(5):284-91 – reference: 10902561 - Med Phys. 2000 Jun;27(6):1311-23 – reference: 7155079 - Med Phys. 1982 Sep-Oct;9(5):758-60 – reference: 19695725 - Radiother Oncol. 2009 Oct;93(1):1-7 – reference: 19810494 - Med Phys. 2009 Sep;36(9):4197-212 |
| SSID | ssib002540153 ssj0011838 ssib045028836 ssib008799463 |
| Score | 2.1388445 |
| Snippet | The image quality (IQ) of imaging systems must be sufficiently high for image‐guided radiation therapy (IGRT). Hence, users should implement a quality... The image quality (IQ) of imaging systems must be sufficiently high for image-guided radiation therapy (IGRT). Hence, users should implement a quality... |
| SourceID | unpaywall pubmedcentral proquest pubmed crossref wiley nii |
| SourceType | Open Access Repository Aggregation Database Index Database Enrichment Source Publisher |
| StartPage | 211 |
| SubjectTerms | Accuracy Algorithms Cone-Beam Computed Tomography Cone-Beam Computed Tomography - instrumentation Cone-Beam Computed Tomography - methods cone‐beam CT Foreign students Humans Image Processing, Computer-Assisted Image Processing, Computer-Assisted - methods image quality International organizations MTF Noise NPS periodic QA Phantoms, Imaging Quality Control Radiation Oncology Physics Radiotherapy Dosage Radiotherapy Planning, Computer-Assisted Radiotherapy Planning, Computer-Assisted - methods Software Standard deviation |
| SummonAdditionalLinks | – databaseName: Proquest Central dbid: BENPR link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV1Lj9MwEB7tdiXggngT2EVG4gJSto6TOPEBoW3VaoXUsoKutLfI8WMJdNPCtqDe-An8Rn4JnrygAi1ccvEkij1j-7Nn5huAZzYJuTaJ9rmUwo-skL60JvQDpYXhYSBpxTM7mfLj0-j1WXy2A9M2FwbDKts1sVqo9ULhHXkfeckp3lfRV8tPPlaNQu9qW0JDNqUV9MuKYmwX9hgyY_VgbzCanrzt_ArOgNPWWclo_4NUF8vDL0FShIecYomV3zan3bIo_oY7_wyfvL4ul3LzVc7n2xC32qPGt-BmAy7JUW0Nt2HHlHfg2qRxn9-FwZvS_Pj2feOsm8iGjYQsLBnNzceVJMPBcEaKC7fCkDrXckMwLP6cTE_eOXlNJrPxPTgdj2bDY78pouCrWESBbww3PNbCUhlZJeLIMCSkMZFQKdXMxjpK3BFIOmSjlKVRYBjLU4dTwpgnuMHdh165KM1DIEkYKCN5rk1sImutMBqjwjhluaZhzjzot6OXqYZhHAtdzLPqpMFoVo13Vo13huPtwfPujWXNrnGF7IFTiPswPoM0CUOB4I4zmmKpndSD_VZVWTMPL7NfVuPB067ZzSB0i8jSLNaXWYC5wCGm1nvwoNZs9zPMHScdwhMeJFs67wSQnXu7pSzeVyzdMRfUoWcPXnTW8R997Ffm80_B7Gg4YQ6eBY-u7vVjuOEQHq8jNPeht_q8NgcORa3yJ83U-AkhWxnd priority: 102 providerName: ProQuest – databaseName: Wiley Online Library Open Access dbid: 24P link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV1Lb9QwEB7RIgEXxJtAi4zEBaR0_YqzPrarriqkLZXYSr1FTjyGiG12RXdBe-Mn8Bv5JdhONmIB8bjk4ocSz0z82TPzDcALlwtlMbepMkan0mmTGociZZXVqAQzNPLMTk7Vybl8fZFd_JAL0_JD9BduwTLi_zoYuCm7KiScRnL56nJx8InltThQVOY7cJ15OBO0nMuz3pPgVTamw2VcpR4LyY2rktPBzzNsbU07TV3_DnX-Gjx5c9UszPqzmc22AW7cocZ34HYHLclhqwt34Ro29-DGpHOe34ejNw1--_J17XWbmI6LhMwdOZ7hh6Uho6PRlNSX_v9C2kzLNQlB8e_I6dlb39-SyXT8AM7Hx9PRSdqVUEirTEuWIipUmdWOGukqnUnkgY4Gpa6G1HKXWZn7A5DxuKaqHJUMOS-HHqWITOVhe3sIu828wcdAcsEqNKq0mKF0zmm0ISZMUV5aKkqewGCzekXV8YuHMhezIp4zOC3iehdxvYuw3gm87EcsWm6NP_Td9wLxE4cnG-ZC6ADtFKfDUGhnmMDeRlRFZ4VXReCyp-GOkybwvG_29hOcIqbB-eqqYCETWITE-gQetZLtX4b7w6THdzqBfEvmfYfAzb3d0tTvI0d3pjT12DmBV712_MM3DqL6_LVjcTiacA_O2JP_HvEUbnnIp9qQzT3YXX5c4b6HVcvyWbSb7zrhGgU priority: 102 providerName: Wiley-Blackwell |
| Title | One‐year analysis of Elekta CBCT image quality using NPS and MTF |
| URI | https://cir.nii.ac.jp/crid/1873398392620873728 https://onlinelibrary.wiley.com/doi/abs/10.1120%2Fjacmp.v17i3.6047 https://www.ncbi.nlm.nih.gov/pubmed/27167279 https://www.proquest.com/docview/2290081460 https://www.proquest.com/docview/1789034807 https://pubmed.ncbi.nlm.nih.gov/PMC5690923 https://onlinelibrary.wiley.com/doi/pdfdirect/10.1120/jacmp.v17i3.6047 |
| UnpaywallVersion | publishedVersion |
| Volume | 17 |
| hasFullText | 1 |
| inHoldings | 1 |
| isFullTextHit | |
| isPrint | |
| journalDatabaseRights | – providerCode: PRVAON databaseName: DOAJ Directory of Open Access Journals customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: DOA dateStart: 20000101 isFulltext: true titleUrlDefault: https://www.doaj.org/ providerName: Directory of Open Access Journals – providerCode: PRVBFR databaseName: Free Medical Journals customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: DIK dateStart: 20000101 isFulltext: true titleUrlDefault: http://www.freemedicaljournals.com providerName: Flying Publisher – providerCode: PRVFQY databaseName: GFMER Free Medical Journals customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: GX1 dateStart: 20000101 isFulltext: true titleUrlDefault: http://www.gfmer.ch/Medical_journals/Free_medical.php providerName: Geneva Foundation for Medical Education and Research – providerCode: PRVHPJ databaseName: ROAD: Directory of Open Access Scholarly Resources customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: M~E dateStart: 20000101 isFulltext: true titleUrlDefault: https://road.issn.org providerName: ISSN International Centre – providerCode: PRVAQN databaseName: PubMed Central customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: RPM dateStart: 20000101 isFulltext: true titleUrlDefault: https://www.ncbi.nlm.nih.gov/pmc/ providerName: National Library of Medicine – providerCode: PRVOVD databaseName: Journals@Ovid LWW All Open Access Journal Collection Rolling customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssib045028836 issn: 1526-9914 databaseCode: OVEED dateStart: 20000101 isFulltext: true titleUrlDefault: http://ovidsp.ovid.com/ providerName: Ovid – providerCode: PRVPQU databaseName: Health & Medical Collection customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: 7X7 dateStart: 20000301 isFulltext: true titleUrlDefault: https://search.proquest.com/healthcomplete providerName: ProQuest – providerCode: PRVPQU databaseName: ProQuest Central customDbUrl: http://www.proquest.com/pqcentral?accountid=15518 eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: BENPR dateStart: 20000301 isFulltext: true titleUrlDefault: https://www.proquest.com/central providerName: ProQuest – providerCode: PRVWIB databaseName: KBPluse Wiley Online Library: Open Access customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: AVUZU dateStart: 20000101 isFulltext: true titleUrlDefault: https://www.kbplus.ac.uk/kbplus7/publicExport/pkg/559 providerName: Wiley-Blackwell – providerCode: PRVWIB databaseName: Wiley Online Library Open Access customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: 24P dateStart: 20000101 isFulltext: true titleUrlDefault: https://authorservices.wiley.com/open-science/open-access/browse-journals.html providerName: Wiley-Blackwell |
| link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV1Lj9MwEB5tWwm48H4EdisjcQEpreM4dn1sq5YVUksFrVROkRPbS9luWrEtqJz4CfxGfgl2kkYqrHiIiy8ZW7E9Y3-2Z74BeGZ4yJTmymdSCp8aIX1pdOgHqRKahYHEOc_saMxOZ_TVPJofwXAfC1PwQ1QXbs4y8vXaGfhamWKdL5MT4vYHmV6sW58CvghbDFNegwaLLCavQ2M2nnTf5WSphPkWBNH9G-UV1Q72pFq2WFwFN3_1mry-zdZy91kul4fINt-ahrfgbN-pwiPlvLXdJK30y098j__f69tws0SvqFuo2x040tlduDYq3-fvQe91pr9__baz5oNkSXeCVgYNlvp8I1G_15-ixYVdwlARzLlDzu_-DI0nb628QqPp8D7MhoNp_9QvszT4aSRo4GvNNIuUMFhSk4qIauIYbzQVaQcrYiJFuT1jSQud0tRgGmhCko4FQmHEuNtBH0A9W2X6ESAeBqmWLFE60tQYI7RybmcMk0ThMCEetPfzFKclhbnLpLGM86MMwXE-NHE-NLEbGg-eVzXWBX3Hb2RP7NTbhl0ZdHgYCoceGcEdl8un48HxXini0tAvY0eXj901KvbgafXZmqh7d5GZXm0v48AFG4cudt-Dh4UOVT9D7HnVQkjhAT_QrkrA0X8ffskW73Ma8IgJbOG5By8qPfyLPrZz7fqjYNztj4jFf8Hjf2n-CdywgJIVDqHHUN983OoTC9o2SRNqhE5syee8CY3eYDx508wvQGz5ch40S4v9AdicREg |
| linkProvider | Unpaywall |
| linkToHtml | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV3NbtNAEB71R6K9IP4xtLBIcADJzXp3vfYeKtSERCltQgWp1Ju78e6WQOoEklDlxiPwRDwMT8KuYxsiUOHSSy5eW_HMzs43nplvAJ6aiHKlI-VzKYXPjJC-NJr6QaqE5jSQOOeZ7XR5-5i9PglPVuB72QvjyirLMzE_qNUodd_Ia46XHLvvVfjl-JPvpka57Go5QkMWoxXUbk4xVjR2HOj5hQ3hJrv7r6y-nxHSavYabb-YMuCnoWCBrzXXPFTCYMlMKkKmiWNs0UykMVbEhIpFNkaQ1vWnqcEs0IT0Y-vIacgj5wHsc1dhnVEmbPC3Xm92j95WeQxrMHGZHCW49kGm5-OdL0E0oDscu5EuvznD1Www-BvO_bNcc2OWjeX8Qg6Hy5A694mtG3C9ALNob7H7bsKKzm7BtU6Rrr8N9TeZ_vH129yKCcmC_QSNDGoO9cepRI16o4cG5_ZEQ4vezjlyZfhnqHv0zq5XqNNr3YHjKxHnXVjLRpm-DyiiQaol7ysdamaMEVq5KjSOSV9h2ice1ErpJWnBaO4GawyTPLIhOMnlneTyTpy8PXhe3TFesHlcsnbbKsQ-2P0GcUSpcGCSExy70T6xB1ulqpLC7ifJr13qwZPqsrVYl4aRmR7NJkngeo-pa-X34N5Cs9WfITZ8tYhSeBAt6bxa4NjAl69kg_c5K3jIBbZo3YMX1e74j3es5dvnnwuTvUaHWDgYPLj8rR_DRrvXOUwO97sHD2HToku-qA7dgrXp55netghu2n9UmAmC06u2zJ_tSlTH |
| linkToPdf | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV3NbhMxEB61RSpcEP8stGAkOIC0jdfe9a4PCLVpo5aSUIlWym1x1jYE0k0gCVVuPALPw-PwJHj2DyJQ4dLLXuy1duf3G3s8A_DYxlxoE2tfKCX90ErlK2u4H2RaGsEDRYs6s92e2D8JX_aj_gp8r-_CYFplbRMLQ63HGe6Rt7AuOcX9KtqyVVrE0W7nxeSTjx2k8KS1bqdRisihWZy58G36_GDX8foJY5294_a-X3UY8LNIhoFvjDAi0tJSFdpMRqFhWK3FhDJLqGY20mHs4gPl3H6WWRoGhrFB4pw4j0SM1t-tuwqXYs4lphPG_SbYc7idJ_WxKKOtDyo7nWx9CeIh3xIUm7n85gZX8-Hwbwj3z0TNy_N8ohZnajRaBtOFN-xcg6sVjCXbpdxdhxWT34D1bnVQfxN2Xufmx9dvC0ckoqq6J2Rsyd7IfJwp0t5pH5PhqbNlpLzVuSCYgP-O9I7euPmadI87t-DkQoh5G9bycW7uAol5kBklBtpEJrTWSqMx_0xQNtCUD5gHrZp6aVbVMseWGqO0iGkYTQt6pwW9U6S3B0-bNyZlHY9z5m46hriF8RkkyF2EkYLRBJv6JB5s1KxKK42fpr_k04NHzbDTVTyAUbkZz6dpgLeOOV7i9-BOydnmY5gLXB2WlB7ESzxvJmAd8OWRfPi-qAceCUkdTvfgWSMd__GPrUJ8_jkx3W53mQOCwb3z__ohrDt9TF8d9A7vwxUHK0WZFroBa7PPc7PpoNts8KDQEQJvL1opfwKE7lJh |
| linkToUnpaywall | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV1LbxMxEB61qQRceD8WWmQkLiBt4rW93viYRo0qpIRKJFI5rbx-tEvTTUQTUDjxE_iN_BLsfUmBioe47MVja23P2J_tmW8AXtqEcm0SHXIpRciskKG0hoaR0sJwGklc8syOJ_x4xt6cxqc7MGpiYSp-iPbCzVtGuV57A19qW63zdXJC3Psg1eWy-ylKctrlmCW7sMdjh8k7sDebnAzel2SphIcOBLHmjfKaalt70m6R59fBzV-9Jm-ui6XcfJbz-TayLbem0R04azpVeaRcdNerrKu-_MT3-P-9vgu3a_SKBpW63YMdU9yHG-P6ff4BHL4tzPev3zbOfJCs6U7QwqKjublYSTQ8HE5RfumWMFQFc26Q97s_Q5OTd05eo_F09BBmo6Pp8DisszSEKhYsCo3hhsdaWCyZVSJmhnjGG8OE6mNNbKxZ4s5Y0kEnpSxmkSEk6zsgRGOe-B30EXSKRWGeAEpopIzkmTaxYdZaYbR3O-OYZBrTjATQa-YpVTWFuc-kMU_LowzBaTk0aTk0qR-aAF61NZYVfcdvZA_c1LuG_TfqJ5QKjx45wX2fy6cfwH6jFGlt6Fepp8vH_hoVB_CiLXYm6t9dZGEW66s08sHG1MfuB_C40qH2Z4g7rzoIKQJItrSrFfD039slRX5e0oDHXGAHzwN43erhX_SxV2rXHwXTwXBMHP6Lnv5L88_glgOUvHII3YfO6uPaHDjQtsqe1xb5A7OZPwo |
| openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=One%E2%80%90year+analysis+of+Elekta+CBCT+image+quality+using+NPS+and+MTF&rft.jtitle=Journal+of+applied+clinical+medical+physics&rft.au=Nakahara%2C+Satomi&rft.au=Tachibana%2C+Masayuki&rft.au=Watanabe%2C+Yoichi&rft.date=2016-05-01&rft.issn=1526-9914&rft.eissn=1526-9914&rft.volume=17&rft.issue=3&rft.spage=211&rft.epage=222&rft_id=info:doi/10.1120%2Fjacmp.v17i3.6047&rft.externalDBID=n%2Fa&rft.externalDocID=10_1120_jacmp_v17i3_6047 |
| thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1526-9914&client=summon |
| thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1526-9914&client=summon |
| thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1526-9914&client=summon |