Variation in target volume and centroid position due to breath holding during four‐dimensional computed tomography scanning: A phantom study

This study investigated the effects of respiratory motion, including unwanted breath holding, on the target volume and centroid position on four‐dimensional computed tomography (4DCT) imaging. Cine 4DCT images were reconstructed based on a time‐based sorting algorithm, and helical 4DCT images were r...

Full description

Saved in:
Bibliographic Details
Published inJournal of applied clinical medical physics Vol. 21; no. 1; pp. 11 - 17
Main Authors Miyamae, Yuta, Akimoto, Mami, Sasaki, Makoto, Fujimoto, Takahiro, Yano, Shinsuke, Nakamura, Mitsuhiro
Format Journal Article
LanguageEnglish
Published United States John Wiley & Sons, Inc 01.01.2020
John Wiley and Sons Inc
Subjects
Online AccessGet full text
ISSN1526-9914
1526-9914
DOI10.1002/acm2.12692

Cover

More Information
Summary:This study investigated the effects of respiratory motion, including unwanted breath holding, on the target volume and centroid position on four‐dimensional computed tomography (4DCT) imaging. Cine 4DCT images were reconstructed based on a time‐based sorting algorithm, and helical 4DCT images were reconstructed based on both the time‐based sorting algorithm and an amplitude‐based sorting algorithm. A spherical object 20 mm in diameter was moved according to several simulated respiratory motions, with a motion period of 4.0 s and maximum amplitude of 5 mm. The object was extracted automatically, and the target volume and centroid position in the craniocaudal direction were measured using a treatment planning system. When the respiratory motion included unwanted breath‐holding times shorter than the breathing cycle, the root mean square errors (RSME) between the reference and imaged target volumes were 18.8%, 14.0%, and 5.5% in time‐based images in cine mode, time‐based images in helical mode, and amplitude‐based images in helical mode, respectively. In helical mode, the RSME between the reference and imaged centroid position was reduced from 1.42 to 0.50 mm by changing the reconstruction method from time‐ to amplitude‐based sorting. When the respiratory motion included unwanted breath‐holding times equal to the breathing cycle, the RSME between the reference and imaged target volumes were 19.1%, 24.3%, and 15.6% in time‐based images in cine mode, time‐based images in helical mode, and amplitude‐based images in helical mode, respectively. In helical mode, the RSME between the reference and imaged centroid position was reduced from 1.61 to 0.83 mm by changing the reconstruction method from time‐ to amplitude‐based sorting. With respiratory motion including breath holding of shorter duration than the breathing cycle, the accuracies of the target volume and centroid position were improved by amplitude‐based sorting, particularly in helical 4DCT.
Bibliography:Funding information
This work was supported in part by JSPS KAKENHI (Grant No. 18H02766).
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
ISSN:1526-9914
1526-9914
DOI:10.1002/acm2.12692