Novel implant design of the proximal interphalangeal joint using an optimized rolling contact joint mechanism

Background The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact joint (RCJ) mechanism and to derive an optimal implant design based on human PIPJ kinematics. Methods In total, 10 participants with normal PIPJs...

Full description

Saved in:
Bibliographic Details
Published inJournal of orthopaedic surgery and research Vol. 14; no. 1; pp. 212 - 13
Main Authors Hong, Seok Woo, Yoon, Junsuk, Kim, Yong-Jae, Gong, Hyun Sik
Format Journal Article
LanguageEnglish
Published London BioMed Central 12.07.2019
BioMed Central Ltd
BMC
Subjects
Online AccessGet full text
ISSN1749-799X
1749-799X
DOI10.1186/s13018-019-1234-6

Cover

Abstract Background The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact joint (RCJ) mechanism and to derive an optimal implant design based on human PIPJ kinematics. Methods In total, 10 participants with normal PIPJs were enrolled in this study. True lateral finger radiographs were obtained in 10° increments from 0º (full extension) to 120° flexion of PIPJ. Radiographs were used to determine the average center of rotation, which formed the basis of a mathematical expression of the PIPJ kinematics. The variations in extensor tendon excursions in relation to the range of motion of PIPJ were determined using results from previous cadaveric studies. As the next step, a PIPJ implant design using an RCJ mechanism that was most consistent with the mathematically expressed PIPJ kinematics and tendon excursions was determined using a constrained optimization algorithm. Results The final proposed PIPJ implant had a relatively constant center of rotation over the entire PIPJ range of motion among the participants. In addition, the extensor tendon excursions of the proposed implant as applied to the phalangeal bones were similar to those of the human tendon. The proposed PIPJ implant achieved an acceptable position of the RCJ surface on the proximal and middle phalanges, which was derived from the constrained optimization algorithm. Conclusions A novel PIPJ implant design using an RCJ mechanism demonstrated acceptable outcomes in terms of PIPJ human kinematics and tendon excursions.
AbstractList Background The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact joint (RCJ) mechanism and to derive an optimal implant design based on human PIPJ kinematics. Methods In total, 10 participants with normal PIPJs were enrolled in this study. True lateral finger radiographs were obtained in 10[degrees] increments from 0º (full extension) to 120[degrees] flexion of PIPJ. Radiographs were used to determine the average center of rotation, which formed the basis of a mathematical expression of the PIPJ kinematics. The variations in extensor tendon excursions in relation to the range of motion of PIPJ were determined using results from previous cadaveric studies. As the next step, a PIPJ implant design using an RCJ mechanism that was most consistent with the mathematically expressed PIPJ kinematics and tendon excursions was determined using a constrained optimization algorithm. Results The final proposed PIPJ implant had a relatively constant center of rotation over the entire PIPJ range of motion among the participants. In addition, the extensor tendon excursions of the proposed implant as applied to the phalangeal bones were similar to those of the human tendon. The proposed PIPJ implant achieved an acceptable position of the RCJ surface on the proximal and middle phalanges, which was derived from the constrained optimization algorithm. Conclusions A novel PIPJ implant design using an RCJ mechanism demonstrated acceptable outcomes in terms of PIPJ human kinematics and tendon excursions. Keywords: Proximal interphalangeal joint implant, Rolling contact joint mechanism, Average center of rotation, Tendon excursion, Constrained optimization
Abstract Background The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact joint (RCJ) mechanism and to derive an optimal implant design based on human PIPJ kinematics. Methods In total, 10 participants with normal PIPJs were enrolled in this study. True lateral finger radiographs were obtained in 10° increments from 0º (full extension) to 120° flexion of PIPJ. Radiographs were used to determine the average center of rotation, which formed the basis of a mathematical expression of the PIPJ kinematics. The variations in extensor tendon excursions in relation to the range of motion of PIPJ were determined using results from previous cadaveric studies. As the next step, a PIPJ implant design using an RCJ mechanism that was most consistent with the mathematically expressed PIPJ kinematics and tendon excursions was determined using a constrained optimization algorithm. Results The final proposed PIPJ implant had a relatively constant center of rotation over the entire PIPJ range of motion among the participants. In addition, the extensor tendon excursions of the proposed implant as applied to the phalangeal bones were similar to those of the human tendon. The proposed PIPJ implant achieved an acceptable position of the RCJ surface on the proximal and middle phalanges, which was derived from the constrained optimization algorithm. Conclusions A novel PIPJ implant design using an RCJ mechanism demonstrated acceptable outcomes in terms of PIPJ human kinematics and tendon excursions.
The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact joint (RCJ) mechanism and to derive an optimal implant design based on human PIPJ kinematics. In total, 10 participants with normal PIPJs were enrolled in this study. True lateral finger radiographs were obtained in 10° increments from 0º (full extension) to 120° flexion of PIPJ. Radiographs were used to determine the average center of rotation, which formed the basis of a mathematical expression of the PIPJ kinematics. The variations in extensor tendon excursions in relation to the range of motion of PIPJ were determined using results from previous cadaveric studies. As the next step, a PIPJ implant design using an RCJ mechanism that was most consistent with the mathematically expressed PIPJ kinematics and tendon excursions was determined using a constrained optimization algorithm. The final proposed PIPJ implant had a relatively constant center of rotation over the entire PIPJ range of motion among the participants. In addition, the extensor tendon excursions of the proposed implant as applied to the phalangeal bones were similar to those of the human tendon. The proposed PIPJ implant achieved an acceptable position of the RCJ surface on the proximal and middle phalanges, which was derived from the constrained optimization algorithm. A novel PIPJ implant design using an RCJ mechanism demonstrated acceptable outcomes in terms of PIPJ human kinematics and tendon excursions.
The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact joint (RCJ) mechanism and to derive an optimal implant design based on human PIPJ kinematics.BACKGROUNDThe aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact joint (RCJ) mechanism and to derive an optimal implant design based on human PIPJ kinematics.In total, 10 participants with normal PIPJs were enrolled in this study. True lateral finger radiographs were obtained in 10° increments from 0º (full extension) to 120° flexion of PIPJ. Radiographs were used to determine the average center of rotation, which formed the basis of a mathematical expression of the PIPJ kinematics. The variations in extensor tendon excursions in relation to the range of motion of PIPJ were determined using results from previous cadaveric studies. As the next step, a PIPJ implant design using an RCJ mechanism that was most consistent with the mathematically expressed PIPJ kinematics and tendon excursions was determined using a constrained optimization algorithm.METHODSIn total, 10 participants with normal PIPJs were enrolled in this study. True lateral finger radiographs were obtained in 10° increments from 0º (full extension) to 120° flexion of PIPJ. Radiographs were used to determine the average center of rotation, which formed the basis of a mathematical expression of the PIPJ kinematics. The variations in extensor tendon excursions in relation to the range of motion of PIPJ were determined using results from previous cadaveric studies. As the next step, a PIPJ implant design using an RCJ mechanism that was most consistent with the mathematically expressed PIPJ kinematics and tendon excursions was determined using a constrained optimization algorithm.The final proposed PIPJ implant had a relatively constant center of rotation over the entire PIPJ range of motion among the participants. In addition, the extensor tendon excursions of the proposed implant as applied to the phalangeal bones were similar to those of the human tendon. The proposed PIPJ implant achieved an acceptable position of the RCJ surface on the proximal and middle phalanges, which was derived from the constrained optimization algorithm.RESULTSThe final proposed PIPJ implant had a relatively constant center of rotation over the entire PIPJ range of motion among the participants. In addition, the extensor tendon excursions of the proposed implant as applied to the phalangeal bones were similar to those of the human tendon. The proposed PIPJ implant achieved an acceptable position of the RCJ surface on the proximal and middle phalanges, which was derived from the constrained optimization algorithm.A novel PIPJ implant design using an RCJ mechanism demonstrated acceptable outcomes in terms of PIPJ human kinematics and tendon excursions.CONCLUSIONSA novel PIPJ implant design using an RCJ mechanism demonstrated acceptable outcomes in terms of PIPJ human kinematics and tendon excursions.
Background The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact joint (RCJ) mechanism and to derive an optimal implant design based on human PIPJ kinematics. Methods In total, 10 participants with normal PIPJs were enrolled in this study. True lateral finger radiographs were obtained in 10° increments from 0º (full extension) to 120° flexion of PIPJ. Radiographs were used to determine the average center of rotation, which formed the basis of a mathematical expression of the PIPJ kinematics. The variations in extensor tendon excursions in relation to the range of motion of PIPJ were determined using results from previous cadaveric studies. As the next step, a PIPJ implant design using an RCJ mechanism that was most consistent with the mathematically expressed PIPJ kinematics and tendon excursions was determined using a constrained optimization algorithm. Results The final proposed PIPJ implant had a relatively constant center of rotation over the entire PIPJ range of motion among the participants. In addition, the extensor tendon excursions of the proposed implant as applied to the phalangeal bones were similar to those of the human tendon. The proposed PIPJ implant achieved an acceptable position of the RCJ surface on the proximal and middle phalanges, which was derived from the constrained optimization algorithm. Conclusions A novel PIPJ implant design using an RCJ mechanism demonstrated acceptable outcomes in terms of PIPJ human kinematics and tendon excursions.
The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact joint (RCJ) mechanism and to derive an optimal implant design based on human PIPJ kinematics. In total, 10 participants with normal PIPJs were enrolled in this study. True lateral finger radiographs were obtained in 10[degrees] increments from 0º (full extension) to 120[degrees] flexion of PIPJ. Radiographs were used to determine the average center of rotation, which formed the basis of a mathematical expression of the PIPJ kinematics. The variations in extensor tendon excursions in relation to the range of motion of PIPJ were determined using results from previous cadaveric studies. As the next step, a PIPJ implant design using an RCJ mechanism that was most consistent with the mathematically expressed PIPJ kinematics and tendon excursions was determined using a constrained optimization algorithm. The final proposed PIPJ implant had a relatively constant center of rotation over the entire PIPJ range of motion among the participants. In addition, the extensor tendon excursions of the proposed implant as applied to the phalangeal bones were similar to those of the human tendon. The proposed PIPJ implant achieved an acceptable position of the RCJ surface on the proximal and middle phalanges, which was derived from the constrained optimization algorithm. A novel PIPJ implant design using an RCJ mechanism demonstrated acceptable outcomes in terms of PIPJ human kinematics and tendon excursions.
ArticleNumber 212
Audience Academic
Author Gong, Hyun Sik
Hong, Seok Woo
Kim, Yong-Jae
Yoon, Junsuk
Author_xml – sequence: 1
  givenname: Seok Woo
  surname: Hong
  fullname: Hong, Seok Woo
  organization: Department of Orthopedic Surgery, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine
– sequence: 2
  givenname: Junsuk
  surname: Yoon
  fullname: Yoon, Junsuk
  organization: Department of Electrical Engineering, Korea University of Technology and Education
– sequence: 3
  givenname: Yong-Jae
  surname: Kim
  fullname: Kim, Yong-Jae
  organization: Department of Electrical Engineering, Korea University of Technology and Education
– sequence: 4
  givenname: Hyun Sik
  surname: Gong
  fullname: Gong, Hyun Sik
  email: hsgong@snu.ac.kr
  organization: Department of Orthopedic Surgery, Seoul National University Bundang Hospital, Seoul National University College of Medicine
BackLink https://www.ncbi.nlm.nih.gov/pubmed/31299978$$D View this record in MEDLINE/PubMed
BookMark eNp9kktv1DAUhSNURB_wA9igSGzYpPgZ2xukquJRqYJNF-wsx3EyHiV2sD1V4ddzh0xRi1Dlhe3r7xzda53T6ijE4KrqNUbnGMv2fcYUYdkgrBpMKGvaZ9UJFkw1QqnvRw_Ox9VpzluEOOKSvaiOKSZKKSFPqvlrvHVT7edlMqHUvct-DHUc6rJx9ZLinZ8NPIfi0rIxwIwO7tsIlXqXfRhrA_hS_Ox_ub5OcZr2RRtDMbYcwNnZjQk-zy-r54OZsnt12M-qm08fby6_NNffPl9dXlw3lgteGqsMbikTsjOYSsRb1na2F8oazJETxtKOt8QQI6lkiCqulO06RwnqjLKMnlVXq20fzVYvCWZIP3U0Xv8pxDRqk4q3k9O9lYyjTiAsWkaRkIhY2fcDEo7RniPw-rB6Lbtudr11oSQzPTJ9_BL8Ro_xVrctYaolYPDuYJDij53LRc8-WzfBX7q4y5oQLgQSgglA367oaKA1H4YIjnaP6wuuqBScSwXU-X8oWL2bPXy8GzzUHwnePBzhb-_3KQBArIBNMefkBm19McXH_UR-0hjpfd70mjcNedP7vOkWlPgf5b35UxqyajKwkKekt3GXAuThCdFv-Lfn7Q
CitedBy_id crossref_primary_10_1109_TRO_2022_3190790
crossref_primary_10_3390_prosthesis6030040
crossref_primary_10_1109_LRA_2023_3243439
crossref_primary_10_1186_s13018_023_04477_y
crossref_primary_10_18019_1028_4427_2024_30_2_191_199
crossref_primary_10_7746_jkros_2022_17_3_322
Cites_doi 10.1115/DETC2004-57264
10.1177/0278364913518998
10.1016/j.hcl.2017.12.011
10.5435/00124635-200703000-00009
10.1016/j.hcl.2017.12.014
10.1016/j.mechmachtheory.2018.04.013
10.1177/1753193414533754
10.1016/j.hcl.2017.12.002
10.1016/j.hcl.2017.12.003
10.1016/J.JHSB.2004.01.010
10.1155/2013/138063
10.1016/0377-0427(88)90405-0
10.1016/j.medengphy.2012.06.016
10.5312/wjo.v6.i2.172
10.1109/BIOROB.2016.7523795
10.1007/s11704-016-5556-9
10.1016/S0363-5023(98)80054-3
10.1016/j.medengphy.2004.08.010
10.3109/17453677608989718
10.1016/0266-7681(86)90019-7
10.1016/j.jbiomech.2011.08.007
10.1016/j.hcl.2013.08.012
10.1007/s13675-015-0039-6
10.1016/S1350-4533(01)00043-1
10.1016/j.hcl.2017.12.009
10.1016/j.hcl.2017.12.008
ContentType Journal Article
Copyright The Author(s). 2019
COPYRIGHT 2019 BioMed Central Ltd.
Copyright_xml – notice: The Author(s). 2019
– notice: COPYRIGHT 2019 BioMed Central Ltd.
DBID C6C
AAYXX
CITATION
CGR
CUY
CVF
ECM
EIF
NPM
7X8
5PM
DOA
DOI 10.1186/s13018-019-1234-6
DatabaseName Springer Nature OA Free Journals
CrossRef
Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
MEDLINE - Academic
PubMed Central (Full Participant titles)
DOAJ Directory of Open Access Journals
DatabaseTitle CrossRef
MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
MEDLINE - Academic
DatabaseTitleList

MEDLINE
MEDLINE - Academic


Database_xml – sequence: 1
  dbid: C6C
  name: Springer Nature OA Free Journals
  url: http://www.springeropen.com/
  sourceTypes: Publisher
– sequence: 2
  dbid: DOA
  name: DOAJ Directory of Open Access Journals
  url: https://www.doaj.org/
  sourceTypes: Open Website
– sequence: 3
  dbid: NPM
  name: PubMed
  url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 4
  dbid: EIF
  name: MEDLINE
  url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search
  sourceTypes: Index Database
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
EISSN 1749-799X
EndPage 13
ExternalDocumentID oai_doaj_org_article_dc8450b701764307802c8ddf07e43d50
PMC6624962
A593875589
31299978
10_1186_s13018_019_1234_6
Genre Journal Article
GeographicLocations South Korea
GeographicLocations_xml – name: South Korea
GrantInformation_xml – fundername: Seoul National University Bundang Hospital
  grantid: 14-2017-001
– fundername: National Research Foundation of Korea
  grantid: 2015M3C1B2052817
– fundername: ;
  grantid: 14-2017-001
– fundername: ;
  grantid: 2015M3C1B2052817
GroupedDBID ---
0R~
29L
2WC
53G
5GY
5VS
7X7
88E
8FI
8FJ
AAFWJ
AAJSJ
AASML
AAWTL
ABDBF
ABUWG
ACGFO
ACGFS
ACPRK
ACUHS
ADBBV
ADRAZ
ADUKV
AENEX
AFKRA
AFPKN
AHBYD
AHMBA
AHYZX
ALMA_UNASSIGNED_HOLDINGS
AMKLP
AMTXH
AOIJS
BAPOH
BAWUL
BCNDV
BENPR
BFQNJ
BMC
BPHCQ
BVXVI
C6C
CCPQU
CS3
DIK
E3Z
EBD
EBLON
EBS
EJD
EMOBN
ESX
F5P
FYUFA
GROUPED_DOAJ
GX1
HMCUK
HYE
IAO
IHR
INH
INR
IPT
ITC
KQ8
M1P
M48
M~E
O5R
O5S
OK1
OVT
P2P
PGMZT
PHGZM
PHGZT
PIMPY
PJZUB
PPXIY
PQQKQ
PROAC
PSQYO
PUEGO
RBZ
RNS
ROL
RPM
RSV
SMD
SOJ
SV3
TUS
UKHRP
WOQ
WOW
~8M
AAYXX
ALIPV
CITATION
-5E
-5G
-A0
-BR
3V.
ACRMQ
ADINQ
C24
CGR
CUY
CVF
ECM
EIF
NPM
PMFND
7X8
5PM
ID FETCH-LOGICAL-c575t-c9a163478ba13805646bcd79ca150e7ac3b562a2a8384039599cbbe320ba9c43
IEDL.DBID M48
ISSN 1749-799X
IngestDate Wed Aug 27 01:27:18 EDT 2025
Thu Aug 21 18:32:14 EDT 2025
Thu Sep 04 16:39:34 EDT 2025
Tue Jun 17 20:58:15 EDT 2025
Tue Jun 10 20:32:45 EDT 2025
Thu Jan 02 23:00:45 EST 2025
Tue Jul 01 02:17:21 EDT 2025
Thu Apr 24 23:00:12 EDT 2025
Sat Sep 06 07:24:19 EDT 2025
IsDoiOpenAccess true
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Issue 1
Keywords Proximal interphalangeal joint implant
Tendon excursion
Average center of rotation
Rolling contact joint mechanism
Constrained optimization
Language English
License Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c575t-c9a163478ba13805646bcd79ca150e7ac3b562a2a8384039599cbbe320ba9c43
Notes ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
OpenAccessLink https://doi.org/10.1186/s13018-019-1234-6
PMID 31299978
PQID 2257707747
PQPubID 23479
PageCount 13
ParticipantIDs doaj_primary_oai_doaj_org_article_dc8450b701764307802c8ddf07e43d50
pubmedcentral_primary_oai_pubmedcentral_nih_gov_6624962
proquest_miscellaneous_2257707747
gale_infotracmisc_A593875589
gale_infotracacademiconefile_A593875589
pubmed_primary_31299978
crossref_citationtrail_10_1186_s13018_019_1234_6
crossref_primary_10_1186_s13018_019_1234_6
springer_journals_10_1186_s13018_019_1234_6
ProviderPackageCode CITATION
AAYXX
PublicationCentury 2000
PublicationDate 2019-07-12
PublicationDateYYYYMMDD 2019-07-12
PublicationDate_xml – month: 07
  year: 2019
  text: 2019-07-12
  day: 12
PublicationDecade 2010
PublicationPlace London
PublicationPlace_xml – name: London
– name: England
PublicationTitle Journal of orthopaedic surgery and research
PublicationTitleAbbrev J Orthop Surg Res
PublicationTitleAlternate J Orthop Surg Res
PublicationYear 2019
Publisher BioMed Central
BioMed Central Ltd
BMC
Publisher_xml – name: BioMed Central
– name: BioMed Central Ltd
– name: BMC
References D Elliot (1234_CR19) 1986; 11
W Zhao (1234_CR20) 2018; 12
MG Catalano (1234_CR8) 2014; 33
1234_CR2
1234_CR7
M Schmidt (1234_CR14) 2015; 3
AD Markiewitz (1234_CR29) 2013; 29
TG Nelson (1234_CR9) 2018; 126
S Kamnerdnakta (1234_CR3) 2018; 34
CM DeDeugd (1234_CR5) 2018; 34
MA Elloy (1234_CR22) 1976; 47
J Vahey (1234_CR30) 1998; 23
AH Slocum Jr (1234_CR17) 2013
AP Sanders (1234_CR21) 2011; 44
FJ Aversano (1234_CR6) 2018; 34
PE Gill (1234_CR13) 1988; 22
M Rizzo (1234_CR1) 2007; 15
T Lawrence (1234_CR18) 2004; 29
D Bitar (1234_CR11) 2015; 6
AH Slocum Jr (1234_CR12) 2012; 34
B McCane (1234_CR24) 2005; 27
M Yamamoto (1234_CR28) 2018; 34
F Hess (1234_CR25) 2013; 2013
AF Zhu (1234_CR4) 2018; 34
GI Bain (1234_CR27) 2015; 40
1234_CR15
S Kim (1234_CR10) 2016
1234_CR16
EQ Pang (1234_CR26) 2018; 34
JH Challis (1234_CR23) 2001; 23
References_xml – ident: 1234_CR7
  doi: 10.1115/DETC2004-57264
– ident: 1234_CR15
– volume: 33
  start-page: 768
  issue: 5
  year: 2014
  ident: 1234_CR8
  publication-title: Int J Rob Res.
  doi: 10.1177/0278364913518998
– volume: 34
  start-page: 217
  issue: 2
  year: 2018
  ident: 1234_CR6
  publication-title: Hand Clin.
  doi: 10.1016/j.hcl.2017.12.011
– volume: 15
  start-page: 189
  issue: 3
  year: 2007
  ident: 1234_CR1
  publication-title: J Am Acad Orthop Surg
  doi: 10.5435/00124635-200703000-00009
– volume: 34
  start-page: 267
  issue: 2
  year: 2018
  ident: 1234_CR3
  publication-title: prevention and treatment. Hand Clin
  doi: 10.1016/j.hcl.2017.12.014
– volume: 126
  start-page: 225
  year: 2018
  ident: 1234_CR9
  publication-title: Mech Mach Theo.
  doi: 10.1016/j.mechmachtheory.2018.04.013
– volume: 40
  start-page: 406
  issue: 4
  year: 2015
  ident: 1234_CR27
  publication-title: J Hand Surg Eur Vol.
  doi: 10.1177/1753193414533754
– volume: 34
  start-page: 121
  issue: 2
  year: 2018
  ident: 1234_CR26
  publication-title: Hand Clin.
  doi: 10.1016/j.hcl.2017.12.002
– volume: 34
  start-page: 127
  issue: 2
  year: 2018
  ident: 1234_CR5
  publication-title: Hand Clin.
  doi: 10.1016/j.hcl.2017.12.003
– volume: 29
  start-page: 244
  issue: 3
  year: 2004
  ident: 1234_CR18
  publication-title: J Hand Surg Br.
  doi: 10.1016/J.JHSB.2004.01.010
– volume: 2013
  start-page: 138063
  year: 2013
  ident: 1234_CR25
  publication-title: Comput Math Methods Med.
  doi: 10.1155/2013/138063
– volume: 22
  start-page: 257
  issue: 2
  year: 1988
  ident: 1234_CR13
  publication-title: J Comput Appl Math.
  doi: 10.1016/0377-0427(88)90405-0
– volume-title: Rolling contact orthopedic joint design [PhD]
  year: 2013
  ident: 1234_CR17
– volume: 34
  start-page: 1009
  issue: 7
  year: 2012
  ident: 1234_CR12
  publication-title: Med Eng Phys.
  doi: 10.1016/j.medengphy.2012.06.016
– volume: 6
  start-page: 172
  issue: 2
  year: 2015
  ident: 1234_CR11
  publication-title: World J Orthop.
  doi: 10.5312/wjo.v6.i2.172
– ident: 1234_CR2
– start-page: 1207
  volume-title: Force characteristics of rolling contact joint for compact structure. 2016 6th IEEE International Conference on Biomedical Robotics and Biomechatronics (BioRob); 26-29 June
  year: 2016
  ident: 1234_CR10
  doi: 10.1109/BIOROB.2016.7523795
– ident: 1234_CR16
– volume: 12
  start-page: 316
  issue: 2
  year: 2018
  ident: 1234_CR20
  publication-title: Frontiers of Computer Science.
  doi: 10.1007/s11704-016-5556-9
– volume: 23
  start-page: 673
  issue: 4
  year: 1998
  ident: 1234_CR30
  publication-title: J Hand Surg Am.
  doi: 10.1016/S0363-5023(98)80054-3
– volume: 27
  start-page: 75
  issue: 1
  year: 2005
  ident: 1234_CR24
  publication-title: Med Eng Phys.
  doi: 10.1016/j.medengphy.2004.08.010
– volume: 47
  start-page: 193
  issue: 2
  year: 1976
  ident: 1234_CR22
  publication-title: Acta Orthop Scand.
  doi: 10.3109/17453677608989718
– volume: 11
  start-page: 77
  issue: 1
  year: 1986
  ident: 1234_CR19
  publication-title: J Hand Surg Br.
  doi: 10.1016/0266-7681(86)90019-7
– volume: 44
  start-page: 2802
  issue: 16
  year: 2011
  ident: 1234_CR21
  publication-title: J Biomech.
  doi: 10.1016/j.jbiomech.2011.08.007
– volume: 29
  start-page: 601
  issue: 4
  year: 2013
  ident: 1234_CR29
  publication-title: Hand Clin.
  doi: 10.1016/j.hcl.2013.08.012
– volume: 3
  start-page: 309
  issue: 4
  year: 2015
  ident: 1234_CR14
  publication-title: EURO J Comput Opt.
  doi: 10.1007/s13675-015-0039-6
– volume: 23
  start-page: 227
  issue: 3
  year: 2001
  ident: 1234_CR23
  publication-title: Med Eng Phys.
  doi: 10.1016/S1350-4533(01)00043-1
– volume: 34
  start-page: 195
  issue: 2
  year: 2018
  ident: 1234_CR28
  publication-title: Hand Clin.
  doi: 10.1016/j.hcl.2017.12.009
– volume: 34
  start-page: 185
  issue: 2
  year: 2018
  ident: 1234_CR4
  publication-title: Hand Clin.
  doi: 10.1016/j.hcl.2017.12.008
SSID ssj0050584
Score 2.2070549
Snippet Background The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact...
The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact joint (RCJ)...
Background The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling contact...
Abstract Background The aims of this study were to propose a novel implant design for the proximal interphalangeal joint (PIPJ) of the hand using a rolling...
SourceID doaj
pubmedcentral
proquest
gale
pubmed
crossref
springer
SourceType Open Website
Open Access Repository
Aggregation Database
Index Database
Enrichment Source
Publisher
StartPage 212
SubjectTerms Adult
Algorithms
Arthritis
Average center of rotation
Care and treatment
Constrained optimization
Design and construction
Female
Finger Joint - diagnostic imaging
Finger Joint - physiology
Finger Joint - surgery
Fingers
Humans
Joint prostheses
Joint Prosthesis - trends
Kinematics
Male
Medicine
Medicine & Public Health
Middle Aged
Novels
Optimization theory
Orthopedics
Prosthesis Design - methods
Prosthesis Design - trends
Proximal interphalangeal joint implant
Radius Fractures - diagnostic imaging
Radius Fractures - surgery
Range of Motion, Articular - physiology
Research Article
Rolling contact joint mechanism
Surgical Orthopedics
Tendon excursion
Young Adult
SummonAdditionalLinks – databaseName: DOAJ Directory of Open Access Journals
  dbid: DOA
  link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwrV1Nb9QwELVQT1wQqLQEWmSkSkhFUb2x449jQVQVh56K1JtlO04btEmq7rYHfj0zdnbVFAEXrvFEcjzjmTfK8zMhRyyga01d-kUrSqFaU3rlVBm08k10qq3bxPK9kOffxber-urRVV_ICcvywHnhTpqgRc28gsiB4gkFjVVBN03LVBS8yd06M2zTTOUcDGVdi-kf5kLLkxVk6gWStlD6n4tSzqpQEuv_PSU_qklP-ZJPfpqmWnT2kryYQCQ9zZN_RZ7FYZf0F-NDXNKuv13CatEmUTPo2FKAeBTZKl0P73SJZHiDhMZrwIj0xwhPKLLfr6kDc8ggffczNvQuq3VT5LK7sJ4M-4gnhbtV_5pcnn29_HJeTpcplAEQ2boMxgH0Ekp7t-AaYI-QPjTKBAeQMCoXuAco5CqnOfR84D1jgveRV8w7EwTfIzvDOMQ3hEbHGw-wgzMWRCuMl7WP2qiIYmh15QvCNmtrwyQ0jvddLG1qOLS02R0W3GHRHVYW5Hj7ym1W2fib8Wd02NYQBbLTAwgbO4WN_VfYFOQjutviNobJBTedRoBPREEse1obDq1crU1BDmaWsP3CbPjDJmAsDiFnbYjj_cpCplSKAbxWBdnPAbSdMweYZaCBL4iahdbso-YjQ3eT1L-lhI5ZVgX5tAlCO6Wd1Z_X7O3_WLN35HnF0pk32EUHZGd9dx8PAZKt_fu0-34BXukxHg
  priority: 102
  providerName: Directory of Open Access Journals
– databaseName: Springer Nature OA Free Journals
  dbid: C6C
  link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV1baxQxFA5aQXwR745WiSAIymB2ksnlsS6WItinCn0LSSbTjuzMlO7WB3-952SyS6dewNfkBGZyLvlCvnwh5C0L6FpTl37RilKo1pReOVUGrXwTnWrrNrF8j-XRN_HltD7NYtF4F-b6-f1Cy49rqLELpFuhaD8XpbxN7tTQhsG8lMtt0YV1XIt8aPnHYbNlJ6nz_16Dry1CNwmSN05J0-Jz-IDcz6iRHkxufkhuxeERufs1n4s_Jv3x-COuaNdfrGCmaJNoGXRsKcA7ikyVrofhXSIYniOZ8QzwIf0-QgtF5vsZdWAO1aPvfsaGXk5K3RR57C5ssmEf8ZZwt-6fkJPDzyfLozI_pFAGQGObMhgHsEso7d2Ca4A8QvrQKBMcwMGoXOAeYJCrnOaw3wPPGRO8j7xi3pkg-FOyN4xDfE5odLzxADk4Y0G0wnhZ-6iNiiiEVle-IGw7zTZkkXF862Jl02ZDSzt5xoJnLHrGyoK83w25mBQ2_mX8CX23M0Rx7NQAMWNzrtkmaFEzr6DYAN4CDMSqoJumZSoK3tSsIO_Q8xZTGD4uuHwTAX4RxbDsQW04bONqbQqyP7OE1Auz7jfb2LHYhXy1IY5XawtVUikG0FoV5NkUS7tv5gCxDGzeC6JmUTb7qXnP0J0n5W8pYbcsq4J82MajzSVn_fc5e_Ff1i_JvYqli22QOftkb3N5FV8B7tr41ynjfgFI8SVU
  priority: 102
  providerName: Springer Nature
Title Novel implant design of the proximal interphalangeal joint using an optimized rolling contact joint mechanism
URI https://link.springer.com/article/10.1186/s13018-019-1234-6
https://www.ncbi.nlm.nih.gov/pubmed/31299978
https://www.proquest.com/docview/2257707747
https://pubmed.ncbi.nlm.nih.gov/PMC6624962
https://doaj.org/article/dc8450b701764307802c8ddf07e43d50
Volume 14
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwfV1ba9swFBa9wOjL2NjNWxc0GAw2vCmWrMvDGGloKXkIY2shb0KS5dQljrskHdt-_Y4UOzRdu6eAdASyzkWfok_nIPSWuKBalae2X7KUiVKlVhiROils4Y0o8zKyfMf89JyNJvlkB3XlrdoFXN55tAv1pM4Xs4-_fvz-Ag7_OTq85J-WEIf7gZIVEvtTlvJdtB-viwKTj20uFWCvl6y92Lxz2AF6QGH7U7Hm2o1dKibz_zdk39izbvMpb12qxr3q5BF62IJMPFhbxWO04-dPUD1ufvoZruqrGawmLiJ1AzclBgiIA5ulqmFMFUmIF4HwOAUMiS8baMGBHT_FBsQhwtTVH1_gxTqbNw5cd-NWrWDtw0vialk_RWcnx2fD07QttpA6QGyr1CkD0IwJaU2fSoBFjFtXCOUMQEYvjKMWoJLJjKRwJgTtKuWs9TQj1ijH6DO0N2_m_gXC3tDCAiyhhDhWMmV5br1UwodkaXlmE0S6tdWuTUQe6mHMdDyQSK7XmtGgGR00o3mC3m-GXK2zcPxP-CgobCMYEmjHhmYx1a0_6sJJlhMrICABJgOcRDIni6IkwjNa5CRB74K6dTA8mJwz7WsF-MSQMEsPckXhqJdLlaDDLUlwT7fV_aYzGB26Aqdt7pvrpYZIKgQB-C0S9HxtQJs5d3aYILFlWlsftd0zry5idnDO4UTNswR96IxQd151_5q9vHcGr9BBRuJDN_CSQ7S3Wlz714DDVraHdsVE9ND-YDD6PoLfo-Px12_QOuTDXvxvoxf97y_zRjJz
linkProvider Scholars Portal
linkToHtml http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV1Lb9QwELagSMAF8SZQwEhISKCo3tjx41hWVAu0e1qk3izbcdqgTVJ1tz301zOTZFdNeUhc47GUeB7-rPn8hZD3LKBrTZ76SSlSoUqTeuVUGrTyRXSqzMuO5TuXsx_i23F-PIhF412Y6_37iZZ7K6ixE6RboWg_F6m8Te5g4xJl8qdyuim6sI9rMTQt_zhttO106vy_1-Brm9BNguSNLmm3-Rw8JA8G1Ej3ezc_Irdi85jcPRr64k9IPW8v45JW9dkSVooWHS2DtiUFeEeRqVLVML3qCIanSGY8AXxIf7bwhCLz_YQ6MIfqUVdXsaDnvVI3RR67C-vBsI54S7ha1U_J4uDLYjpLhx8ppAHQ2DoNxgHsEkp7N-EaII-QPhTKBAdwMCoXuAcY5DKnOZz3wHPGBO8jz5h3Jgj-jOw0bRNfEBodLzxADs5YEKUwXuY-aqMiCqHlmU8I2yyzDYPIOP7rYmm7w4aWtveMBc9Y9IyVCfm4nXLWK2z8y_gz-m5riOLY3QOIGTvkmi2CFjnzCooN4C3AQCwLuihKpqLgRc4S8gE9bzGF4eWCG24iwCeiGJbdzw2HY1yuTUJ2R5aQemE0_G4TOxaHkK_WxPZiZaFKKsUAWquEPO9jafvOHCCWgcN7QtQoykYfNR5pqtNO-VtKOC3LLCGfNvFoh5Kz-vuavfwv67fk3mxxdGgPv86_vyL3M9ZdcoMs2iU76_OL-Bow2Nq_6bLvFzysKEM
linkToPdf http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV1bi9QwFA66C4svsl63umoEQVDKZpq0SR7Hy7COOgiusG8hSdPZyrQdprP74K_3pE0Hu17A1-YE0pyck--QL18QekGsd61MYzMpWMx4IWPDNY-t4CZ3mhdp0bF8F9npNzY_T8_DO6ftwHYfjiT7Ow1epanenqzzog9xkZ20kHknnoTlpfwpi7ObaF-kUkL1tT-dzr_Oh2QM-7tg4TDzjx1H21Gn2v97bv5lc7pOnLx2etptSrNDdDugSTzt3X8H3XD1XXTwOZyX30PVorlyK1xW6xXMIM47ugZuCgywD3sGS1lB97IjHl54kuMScCP-3sAX7BnxS6zBHLJKVf5wOd70Ct7YT5a222BYOX97uGyr--hs9v7s7WkcHliILaC0bWylBjjGuDB6QgVAIZYZm3NpNcBEx7WlBuCRTrSgUAeCR6W0xjiaEKOlZfQB2qub2h0h7DTNDUARSohlBZMmS40TkjsvkJYmJkJkmGZlg_i4fwNjpboiRGSq94wCzyjvGZVF6NWuy7pX3viX8Rvvu52hF83uPjSbpQoxqHIrWEoMhyQEOAywEUmsyPOCcMdonpIIvfSeVz60YXBWhxsK8IteJEtNU0mhvEuFjNDxyBJC0o6anw9rR_kmz2OrXXPZKsienBOA3DxCD_u1tBszBegFS1hEiI9W2einxi11edEpgmcZVNFZEqHXw3pUIRW1f5-zR_9l_QwdfHk3U58-LD4-RrcS0t19gyA6RnvbzaV7AtBsa56G8PsJ6vcyAQ
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Novel+implant+design+of+the+proximal+interphalangeal+joint+using+an+optimized+rolling+contact+joint+mechanism&rft.jtitle=Journal+of+orthopaedic+surgery+and+research&rft.au=Hong%2C+Seok+Woo&rft.au=Yoon%2C+Junsuk&rft.au=Kim%2C+Yong-Jae&rft.au=Gong%2C+Hyun+Sik&rft.date=2019-07-12&rft.eissn=1749-799X&rft.volume=14&rft.issue=1&rft.spage=212&rft_id=info:doi/10.1186%2Fs13018-019-1234-6&rft_id=info%3Apmid%2F31299978&rft.externalDocID=31299978
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1749-799X&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1749-799X&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1749-799X&client=summon