A Simple Procedure for Correcting Loading Effects of Aethalometer Data
A simple method for correcting for the loading effects of aethalometer data is presented. The formula BC(CORRECTED) = (1 + k x ATN) x BC(NONCORRECTED), where ATN is the attenuation and BC is black carbon, was used for correcting aethalometer data obtained from measurements at three different sites:...
        Saved in:
      
    
          | Published in | Journal of the Air & Waste Management Association (1995) Vol. 57; no. 10; pp. 1214 - 1222 | 
|---|---|
| Main Authors | , , , , , , | 
| Format | Journal Article | 
| Language | English | 
| Published | 
        Pittsburgh, PA
          Air & Waste Management Association
    
        01.10.2007
     Air and Waste Management Association Taylor & Francis Ltd  | 
| Subjects | |
| Online Access | Get full text | 
| ISSN | 1096-2247 2162-2906 2162-2906  | 
| DOI | 10.3155/1047-3289.57.10.1214 | 
Cover
| Summary: | A simple method for correcting for the loading effects of aethalometer data is presented. The formula BC(CORRECTED) = (1 + k x ATN) x BC(NONCORRECTED), where ATN is the attenuation and BC is black carbon, was used for correcting aethalometer data obtained from measurements at three different sites: a subway station in Helsinki, an urban background measurement station in Helsinki, and a rural station in Hyytiälä in central Finland. The BC data were compared with simultaneously measured aerosol volume concentrations (V). After the correction algorithm, the BC-to-V ratio remained relatively stable between consequent filter spots, which can be regarded as indirect evidence that the correction algorithm works. The k value calculated from the outdoor sites had a clear seasonal cycle that could be explained by darker aerosol in winter than in summer. When the contribution of BC to the total aerosol volume was high, the k factor was high and vice versa. In winter, the k values at all wavelengths were very close to that obtained from the subway station data. In summer, the k value was wavelength dependent and often negative. When the k value is negative, the noncorrected BC concentrations overestimated the true concentrations. | 
|---|---|
| Bibliography: | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 14 ObjectType-Article-1 ObjectType-Feature-2 content type line 23  | 
| ISSN: | 1096-2247 2162-2906 2162-2906  | 
| DOI: | 10.3155/1047-3289.57.10.1214 |