Experimental evaluation of the GE NM/CT 870 CZT clinical SPECT system equipped with WEHR and MEHRS collimator

Purpose A high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium‐energy radionuclides....

Full description

Saved in:
Bibliographic Details
Published inJournal of applied clinical medical physics Vol. 22; no. 2; pp. 165 - 177
Main Authors Ito, Toshimune, Matsusaka, Yohji, Onoguchi, Masahisa, Ichikawa, Hajime, Okuda, Koichi, Shibutani, Takayuki, Shishido, Masaaki, Sato, Kozo
Format Journal Article
LanguageEnglish
Published United States John Wiley & Sons, Inc 01.02.2021
John Wiley and Sons Inc
Subjects
Online AccessGet full text
ISSN1526-9914
1526-9914
DOI10.1002/acm2.13158

Cover

Abstract Purpose A high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium‐energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C‐SPECT, and a NaI scintillation detector‐equipped Anger‐type SPECT (A‐SPECT) scanner, with a LEHR and LMEGP. Methods A line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99mTc and 123I. We also measured the system's sensitivity and high‐contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high‐contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction. Results The energy resolution with 99mTc was 5.6% in C‐SPECT with WEHR and 9.9% in A‐SPECT with LEHR. Using 123I, the results were 9.1% in C‐SPECT with WEHR, 5.5% in C‐SPECT with MEHRS, and 10.4% in A‐SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C‐SPECT performed better in SPECT acquisition. High‐contrast resolution was improved in C‐SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C‐SPECT. Conclusion C‐SPECT demonstrates excellent energy resolution and improved high‐contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C‐SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high‐contrast resolution.
AbstractList A high-energy-resolution whole-body SPECT-CT device (NM/CT 870 CZT; C-SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium-energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C-SPECT, and a NaI scintillation detector-equipped Anger-type SPECT (A-SPECT) scanner, with a LEHR and LMEGP. A line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using Tc and I. We also measured the system's sensitivity and high-contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high-contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction. The energy resolution with Tc was 5.6% in C-SPECT with WEHR and 9.9% in A-SPECT with LEHR. Using 123I, the results were 9.1% in C-SPECT with WEHR, 5.5% in C-SPECT with MEHRS, and 10.4% in A-SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C-SPECT performed better in SPECT acquisition. High-contrast resolution was improved in C-SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C-SPECT. C-SPECT demonstrates excellent energy resolution and improved high-contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C-SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high-contrast resolution.
Purpose A high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium‐energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C‐SPECT, and a NaI scintillation detector‐equipped Anger‐type SPECT (A‐SPECT) scanner, with a LEHR and LMEGP. Methods A line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99mTc and 123I. We also measured the system's sensitivity and high‐contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high‐contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction. Results The energy resolution with 99mTc was 5.6% in C‐SPECT with WEHR and 9.9% in A‐SPECT with LEHR. Using 123I, the results were 9.1% in C‐SPECT with WEHR, 5.5% in C‐SPECT with MEHRS, and 10.4% in A‐SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C‐SPECT performed better in SPECT acquisition. High‐contrast resolution was improved in C‐SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C‐SPECT. Conclusion C‐SPECT demonstrates excellent energy resolution and improved high‐contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C‐SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high‐contrast resolution.
PurposeA high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium‐energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C‐SPECT, and a NaI scintillation detector‐equipped Anger‐type SPECT (A‐SPECT) scanner, with a LEHR and LMEGP.MethodsA line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99mTc and 123I. We also measured the system's sensitivity and high‐contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high‐contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction.ResultsThe energy resolution with 99mTc was 5.6% in C‐SPECT with WEHR and 9.9% in A‐SPECT with LEHR. Using 123I, the results were 9.1% in C‐SPECT with WEHR, 5.5% in C‐SPECT with MEHRS, and 10.4% in A‐SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C‐SPECT performed better in SPECT acquisition. High‐contrast resolution was improved in C‐SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C‐SPECT.ConclusionC‐SPECT demonstrates excellent energy resolution and improved high‐contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C‐SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high‐contrast resolution.
A high-energy-resolution whole-body SPECT-CT device (NM/CT 870 CZT; C-SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium-energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C-SPECT, and a NaI scintillation detector-equipped Anger-type SPECT (A-SPECT) scanner, with a LEHR and LMEGP.PURPOSEA high-energy-resolution whole-body SPECT-CT device (NM/CT 870 CZT; C-SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium-energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C-SPECT, and a NaI scintillation detector-equipped Anger-type SPECT (A-SPECT) scanner, with a LEHR and LMEGP.A line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99m Tc and 123 I. We also measured the system's sensitivity and high-contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high-contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction.METHODSA line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99m Tc and 123 I. We also measured the system's sensitivity and high-contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high-contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction.The energy resolution with 99m Tc was 5.6% in C-SPECT with WEHR and 9.9% in A-SPECT with LEHR. Using 123I, the results were 9.1% in C-SPECT with WEHR, 5.5% in C-SPECT with MEHRS, and 10.4% in A-SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C-SPECT performed better in SPECT acquisition. High-contrast resolution was improved in C-SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C-SPECT.RESULTSThe energy resolution with 99m Tc was 5.6% in C-SPECT with WEHR and 9.9% in A-SPECT with LEHR. Using 123I, the results were 9.1% in C-SPECT with WEHR, 5.5% in C-SPECT with MEHRS, and 10.4% in A-SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C-SPECT performed better in SPECT acquisition. High-contrast resolution was improved in C-SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C-SPECT.C-SPECT demonstrates excellent energy resolution and improved high-contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C-SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high-contrast resolution.CONCLUSIONC-SPECT demonstrates excellent energy resolution and improved high-contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C-SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high-contrast resolution.
Author Sato, Kozo
Matsusaka, Yohji
Shishido, Masaaki
Shibutani, Takayuki
Ito, Toshimune
Onoguchi, Masahisa
Ichikawa, Hajime
Okuda, Koichi
AuthorAffiliation 1 Department of Radiology Saiseikai Yokohamashi Tobu Hospital Yokohama Japan
5 Department of Physics Kanazawa Medical University Kahoku, Ishikawa Japan
2 Department of Diagnostic Radiology School of Medicine Keio University Tokyo Japan
3 Department of Quantum Medical Technology Graduate School of Medical Sciences Kanazawa University Kanazawa Japan
6 Department of Diagnostic Radiology Saiseikai Yokohamashi Tobu Hospital Yokohama Kanagawa Japan
4 Department of Radiology Toyohashi Municipal Hospital Toyohashi Japan
AuthorAffiliation_xml – name: 1 Department of Radiology Saiseikai Yokohamashi Tobu Hospital Yokohama Japan
– name: 6 Department of Diagnostic Radiology Saiseikai Yokohamashi Tobu Hospital Yokohama Kanagawa Japan
– name: 4 Department of Radiology Toyohashi Municipal Hospital Toyohashi Japan
– name: 2 Department of Diagnostic Radiology School of Medicine Keio University Tokyo Japan
– name: 5 Department of Physics Kanazawa Medical University Kahoku, Ishikawa Japan
– name: 3 Department of Quantum Medical Technology Graduate School of Medical Sciences Kanazawa University Kanazawa Japan
Author_xml – sequence: 1
  givenname: Toshimune
  surname: Ito
  fullname: Ito, Toshimune
  email: toshimu2001@gmail.com
  organization: Saiseikai Yokohamashi Tobu Hospital
– sequence: 2
  givenname: Yohji
  surname: Matsusaka
  fullname: Matsusaka, Yohji
  organization: Keio University
– sequence: 3
  givenname: Masahisa
  surname: Onoguchi
  fullname: Onoguchi, Masahisa
  organization: Kanazawa University
– sequence: 4
  givenname: Hajime
  surname: Ichikawa
  fullname: Ichikawa, Hajime
  organization: Toyohashi Municipal Hospital
– sequence: 5
  givenname: Koichi
  surname: Okuda
  fullname: Okuda, Koichi
  organization: Kanazawa Medical University
– sequence: 6
  givenname: Takayuki
  surname: Shibutani
  fullname: Shibutani, Takayuki
  organization: Kanazawa University
– sequence: 7
  givenname: Masaaki
  surname: Shishido
  fullname: Shishido, Masaaki
  organization: Kanazawa University
– sequence: 8
  givenname: Kozo
  surname: Sato
  fullname: Sato, Kozo
  organization: Saiseikai Yokohamashi Tobu Hospital
BackLink https://www.ncbi.nlm.nih.gov/pubmed/33426812$$D View this record in MEDLINE/PubMed
BookMark eNp9kU1v1DAQhi1URD8v_ABkiQui2tZ27MS5VKqi0CJ1KaKLkLhYjuOwrhw7jZMu--_rbQotFerJY_uZd2be2QVbzjsNwFuMjjBC5FiqlhzhBDP-CuxgRtJZnmO69STeBrshXCOEMU_4G7CdJJSkHJMd0Ja_O92bVrtBWqhvpR3lYLyDvoHDUsOzEn6ZHxcLyDMEi58LqKxxRkX26msZn8M6DLqF-mY0XadruDLDEv4oz79B6Wo4j8EVVN5a08rB9_vgdSNt0AcP5x74_qlcFOezi8uzz8XpxUzRjPJZhXHCaMIQSnPOKVM8Xnilsap4I1nFScIqTEiCaoplSjJOMKnTlOWIEcWbZA8cTrqj6-R6Ja0VXRxS9muBkdiYJjamiXvTIn0y0d1YtbpW0YtePmZ4acS_P84sxS9_KzIeC-ONwIcHgd7fjDoMojVBaWul034MgtAs5ZRmDEX0_TP02o-9i2ZEiuco7geTSL172tHfVv7sLQJoAlTvQ-h1I5QZ7hcXGzT2_1N-fJbyoiV4glfG6vULpDgt5mTKuQNtoMOH
CitedBy_id crossref_primary_10_1053_j_semnuclmed_2023_10_007
crossref_primary_10_1186_s40658_024_00659_7
crossref_primary_10_3390_cancers13246183
crossref_primary_10_1053_j_semnuclmed_2022_01_004
crossref_primary_10_1186_s40658_024_00656_w
crossref_primary_10_1002_mp_17724
crossref_primary_10_1186_s40658_022_00463_1
crossref_primary_10_1186_s40658_022_00517_4
crossref_primary_10_1038_s41566_021_00909_5
crossref_primary_10_1007_s13246_023_01257_0
crossref_primary_10_1016_j_ijleo_2021_167681
crossref_primary_10_1007_s00247_023_05682_x
crossref_primary_10_6009_jjrt_2022_2024
crossref_primary_10_1007_s12194_023_00746_x
crossref_primary_10_2967_jnumed_124_267542
crossref_primary_10_4103_jmp_jmp_71_22
crossref_primary_10_1038_s41598_021_03853_0
crossref_primary_10_1055_a_1993_7420
crossref_primary_10_1007_s00259_023_06176_6
crossref_primary_10_1186_s13550_023_00973_8
crossref_primary_10_1186_s40658_024_00684_6
crossref_primary_10_1088_1361_6560_ac5671
crossref_primary_10_6009_jjrt_2022_2054
crossref_primary_10_1038_s41551_022_00866_6
crossref_primary_10_2967_jnmt_122_265328
crossref_primary_10_1186_s40658_023_00566_3
crossref_primary_10_1007_s13139_023_00825_2
crossref_primary_10_1055_a_1715_5305
Cites_doi 10.1186/2191-219X-3-45
10.1109/23.958705
10.1007/s12194-012-0168-z
10.1097/MNM.0000000000001084
10.1007/s13139-017-0474-9
10.1016/j.nima.2005.04.081
10.1007/s12149-015-0987-1
10.6009/jjrt.2012_JSRT_68.3.299
10.2967/jnumed.110.081323
10.1097/MNM.0000000000001028
10.1007/s12350-014-0008-0
10.1118/1.4934830
10.1007/s00259-013-2617-2
10.1053/j.semnuclmed.2016.01.002
10.1007/s12149-009-0293-x
10.1007/s12350-018-01585-9
10.2967/jnumed.108.060020
10.1088/0031-9155/60/8/3045
10.1109/TNS.2004.829437
10.1007/s12149-019-01382-5
10.1007/s00259-016-3415-4
10.1007/s12149-016-1144-1
ContentType Journal Article
Copyright 2021 The Authors. published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
2021. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Copyright_xml – notice: 2021 The Authors. published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
– notice: 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
– notice: 2021. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
DBID 24P
AAYXX
CITATION
NPM
3V.
7X7
7XB
88I
8FI
8FJ
8FK
ABUWG
AFKRA
AZQEC
BENPR
CCPQU
DWQXO
FYUFA
GHDGH
GNUQQ
HCIFZ
K9.
M0S
M2P
PHGZM
PHGZT
PIMPY
PJZUB
PKEHL
PPXIY
PQEST
PQQKQ
PQUKI
PRINS
Q9U
7X8
5PM
ADTOC
UNPAY
DOI 10.1002/acm2.13158
DatabaseName Wiley Online Library Open Access
CrossRef
PubMed
ProQuest Central (Corporate)
Health & Medical Collection
ProQuest Central (purchase pre-March 2016)
Science Database (Alumni Edition)
ProQuest Hospital Collection
Hospital Premium Collection (Alumni Edition)
ProQuest Central (Alumni) (purchase pre-March 2016)
ProQuest Central (Alumni)
ProQuest Central UK/Ireland
ProQuest Central Essentials
ProQuest Central
ProQuest One Community College
ProQuest Central
Health Research Premium Collection
Health Research Premium Collection (Alumni)
ProQuest Central Student
SciTech Premium Collection
ProQuest Health & Medical Complete (Alumni)
ProQuest Health & Medical Collection
Science Database
ProQuest Central Premium
ProQuest One Academic (New)
Publicly Available Content Database
ProQuest Health & Medical Research Collection
ProQuest One Academic Middle East (New)
ProQuest One Health & Nursing
ProQuest One Academic Eastern Edition (DO NOT USE)
ProQuest One Academic
ProQuest One Academic UKI Edition
ProQuest Central China
ProQuest Central Basic
MEDLINE - Academic
PubMed Central (Full Participant titles)
Unpaywall for CDI: Periodical Content
Unpaywall
DatabaseTitle CrossRef
PubMed
Publicly Available Content Database
ProQuest Central Student
ProQuest One Academic Middle East (New)
ProQuest Central Essentials
ProQuest Health & Medical Complete (Alumni)
ProQuest Central (Alumni Edition)
SciTech Premium Collection
ProQuest One Community College
ProQuest One Health & Nursing
ProQuest Central China
ProQuest Central
ProQuest Health & Medical Research Collection
Health Research Premium Collection
Health and Medicine Complete (Alumni Edition)
ProQuest Central Korea
ProQuest Central (New)
ProQuest Science Journals (Alumni Edition)
ProQuest Central Basic
ProQuest Science Journals
ProQuest One Academic Eastern Edition
ProQuest Hospital Collection
Health Research Premium Collection (Alumni)
ProQuest Hospital Collection (Alumni)
ProQuest Health & Medical Complete
ProQuest One Academic UKI Edition
ProQuest One Academic
ProQuest One Academic (New)
ProQuest Central (Alumni)
MEDLINE - Academic
DatabaseTitleList PubMed

Publicly Available Content Database
MEDLINE - Academic
Database_xml – sequence: 1
  dbid: 24P
  name: Wiley Online Library (Open Access Collection)
  url: https://authorservices.wiley.com/open-science/open-access/browse-journals.html
  sourceTypes: Publisher
– sequence: 2
  dbid: NPM
  name: PubMed
  url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 3
  dbid: UNPAY
  name: Unpaywall
  url: https://proxy.k.utb.cz/login?url=https://unpaywall.org/
  sourceTypes: Open Access Repository
– sequence: 4
  dbid: BENPR
  name: ProQuest Central
  url: http://www.proquest.com/pqcentral?accountid=15518
  sourceTypes: Aggregation Database
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
Physics
DocumentTitleAlternate Ito et al
EISSN 1526-9914
EndPage 177
ExternalDocumentID 10.1002/acm2.13158
PMC7882118
33426812
10_1002_acm2_13158
ACM213158
Genre article
Journal Article
GeographicLocations United States--US
Japan
GeographicLocations_xml – name: United States--US
– name: Japan
GroupedDBID 0R~
1OC
24P
29J
2WC
53G
5GY
7X7
88I
8FI
8FJ
AAHHS
ABUWG
ACCFJ
ACCMX
ACGFO
ACXQS
ADBBV
ADKYN
ADPDF
ADZMN
ADZOD
AEEZP
AENEX
AEQDE
AFKRA
AIWBW
AJBDE
ALIPV
ALMA_UNASSIGNED_HOLDINGS
ALUQN
AOIJS
AVUZU
AZQEC
BAWUL
BCNDV
BENPR
BPHCQ
BVXVI
CCPQU
CS3
DIK
DU5
DWQXO
E3Z
EBS
EJD
EMOBN
FRP
FYUFA
GNUQQ
GROUPED_DOAJ
GX1
H13
HCIFZ
HMCUK
HYE
IAO
IHR
INH
ITC
KWQ
M2P
M~E
OK1
OVD
OVEED
P6G
PIMPY
PQQKQ
PROAC
RNS
RPM
TR2
UKHRP
W2D
WIN
XSB
AAMMB
AAYXX
AEFGJ
AGXDD
AIDQK
AIDYY
CITATION
OVT
PHGZM
PHGZT
PUEGO
NPM
3V.
7XB
8FK
K9.
PJZUB
PKEHL
PPXIY
PQEST
PQUKI
PRINS
Q9U
7X8
5PM
ADTOC
UNPAY
ID FETCH-LOGICAL-c4748-b113543500698845c84358be1cb8fa5b8235b12230d41a6278212d6659052c8f3
IEDL.DBID BENPR
ISSN 1526-9914
IngestDate Sun Oct 26 03:58:02 EDT 2025
Tue Sep 30 16:37:47 EDT 2025
Fri Sep 05 11:07:52 EDT 2025
Tue Oct 07 07:08:01 EDT 2025
Thu Apr 03 07:04:47 EDT 2025
Wed Oct 01 02:06:46 EDT 2025
Thu Apr 24 22:51:25 EDT 2025
Wed Jan 22 16:30:00 EST 2025
IsDoiOpenAccess true
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Issue 2
Keywords medium-energy high-resolution sensitivity collimator
Discovery NM/CT 870 CZT ⇒ NM/CT 870 CZT
wide-energy high-resolution collimator
Language English
License Attribution
2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
cc-by
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c4748-b113543500698845c84358be1cb8fa5b8235b12230d41a6278212d6659052c8f3
Notes Funding
This research did not receive any specific grant from funding agencies in the public, commercial, or not‐for‐profit sectors.
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
OpenAccessLink https://www.proquest.com/docview/2489083812?pq-origsite=%requestingapplication%&accountid=15518
PMID 33426812
PQID 2489083812
PQPubID 4370306
PageCount 13
ParticipantIDs unpaywall_primary_10_1002_acm2_13158
pubmedcentral_primary_oai_pubmedcentral_nih_gov_7882118
proquest_miscellaneous_2476844750
proquest_journals_2489083812
pubmed_primary_33426812
crossref_citationtrail_10_1002_acm2_13158
crossref_primary_10_1002_acm2_13158
wiley_primary_10_1002_acm2_13158_ACM213158
ProviderPackageCode CITATION
AAYXX
PublicationCentury 2000
PublicationDate February 2021
PublicationDateYYYYMMDD 2021-02-01
PublicationDate_xml – month: 02
  year: 2021
  text: February 2021
PublicationDecade 2020
PublicationPlace United States
PublicationPlace_xml – name: United States
– name: Malden Massachusetts
– name: Hoboken
PublicationTitle Journal of applied clinical medical physics
PublicationTitleAlternate J Appl Clin Med Phys
PublicationYear 2021
Publisher John Wiley & Sons, Inc
John Wiley and Sons Inc
Publisher_xml – name: John Wiley & Sons, Inc
– name: John Wiley and Sons Inc
References 2009; 23
1998; 26
2013; 3
2010; 37
2005; 550
2019; 33
1991; 32
2011; 52
2007
2001; 48
2014; 41
2013; 6
2007; 35
2018; 46
2017; 51
2017; 31
1993; 34
2004; 51
2019; 40
2015; 29
2015; 60
2009; 50
2015; 42
2015; 22
2016; 43
2019
2012; 68
2016; 46
2003; 44
e_1_2_9_31_1
e_1_2_9_11_1
e_1_2_9_10_1
e_1_2_9_13_1
e_1_2_9_12_1
Bartel TB (e_1_2_9_16_1) 2018; 46
Ichihara T (e_1_2_9_23_1) 1993; 34
Association NEM (e_1_2_9_14_1) 2007
e_1_2_9_15_1
e_1_2_9_17_1
e_1_2_9_19_1
Mariani G (e_1_2_9_29_1) 2010; 37
Israel O (e_1_2_9_18_1) 1991; 32
e_1_2_9_22_1
e_1_2_9_21_1
e_1_2_9_8_1
e_1_2_9_7_1
Mueller B (e_1_2_9_24_1) 2003; 44
e_1_2_9_6_1
e_1_2_9_5_1
Takahashi Y (e_1_2_9_30_1) 2007; 35
e_1_2_9_4_1
e_1_2_9_3_1
e_1_2_9_2_1
e_1_2_9_9_1
e_1_2_9_26_1
e_1_2_9_25_1
e_1_2_9_28_1
Murase K (e_1_2_9_20_1) 1998; 26
e_1_2_9_27_1
References_xml – volume: 41
  start-page: 522
  year: 2014
  end-page: 528
  article-title: Factors affecting the myocardial activity acquired during exercise SPECT with a high‐sensitivity cardiac CZT camera as compared with conventional Anger camera
  publication-title: Eur J Nucl Med Mol Imaging
– volume: 51
  start-page: 1191
  year: 2004
  end-page: 1198
  article-title: CdTe and CdZnTe X‐ray and gamma‐ray detectors for imaging systems
  publication-title: IEEE Trans Nucl Sci
– volume: 40
  start-page: 693
  year: 2019
  end-page: 702
  article-title: Evaluation of edge‐preserving and noise‐reducing effects using the nonlinear diffusion method in bone single‐photon emission computed tomography
  publication-title: Nucl Med Commun
– volume: 42
  start-page: 6895
  year: 2015
  end-page: 6911
  article-title: Scatter and crosstalk corrections for (99m)Tc/(123)I dual‐radionuclide imaging using a CZT SPECT system with pinhole collimators
  publication-title: Med Phys
– volume: 31
  start-page: 172
  year: 2017
  end-page: 180
  article-title: Respiratory average CT for attenuation correction in myocardial perfusion SPECT/CT
  publication-title: Ann Nucl Med
– volume: 60
  start-page: 3045
  year: 2015
  end-page: 3063
  article-title: Model‐based correction for scatter and tailing effects in simultaneous 99mTc and 123I imaging for a CdZnTe cardiac SPECT camera
  publication-title: Phys Med Biol
– year: 2007
– volume: 32
  start-page: 1157
  year: 1991
  end-page: 1161
  article-title: In vivo SPECT quantitation of bone metabolism in hyperparathyroidism and thyrotoxicosis
  publication-title: J Nucl Med
– volume: 22
  start-page: 334
  year: 2015
  end-page: 343
  article-title: Scatter correction improves concordance in SPECT MPI with a dedicated cardiac SPECT solid‐state camera
  publication-title: J Nucl Cardiol
– volume: 37
  start-page: 1959
  year: 2010
  end-page: 1985
  article-title: A review on the clinical uses of SPECT/CT
  publication-title: EJNMMI
– volume: 52
  start-page: 210
  year: 2011
  end-page: 217
  article-title: Cardiac dedicated ultrafast SPECT cameras: new designs and clinical implications
  publication-title: J Nucl Med
– volume: 46
  start-page: 398
  year: 2018
  end-page: 404
  article-title: SNMMI procedure standard for bone scintigraphy 4.0
  publication-title: J Nucl Med Technol
– volume: 3
  start-page: 45
  year: 2013
  article-title: Quantification of 99mTc‐DPD concentration in the lumbar spine with SPECT/CT
  publication-title: EJNMMI Res
– volume: 51
  start-page: 331
  year: 2017
  end-page: 337
  article-title: Evaluation of cadmium‐zinc‐telluride detector‐based single‐photon emission computed tomography for nuclear cardiology: a comparison with conventional anger single‐photon emission computed tomography
  publication-title: Nucl Med Mol Imaging
– volume: 44
  start-page: 602
  year: 2003
  end-page: 609
  article-title: Evaluation of a small cadmium zinc telluride detector for scintimammography
  publication-title: J Nucl Med
– volume: 35
  start-page: 34
  year: 2007
  end-page: 38
  article-title: Simultaneous 3‐dimensional resolution correction in SPECT reconstruction with an ordered‐subsets expectation maximization algorithm
  publication-title: J Nucl Med Technol
– volume: 23
  start-page: 763
  year: 2009
  end-page: 770
  article-title: Development of an ultra‐high resolution SPECT system with a CdTe semiconductor detector
  publication-title: Ann Nucl Med
– volume: 33
  start-page: 725
  year: 2019
  end-page: 732
  article-title: Nuclear medicine practice in Japan: a report of the eighth nationwide survey in 2017
  publication-title: Ann Nucl Med
– volume: 50
  start-page: 635
  year: 2009
  end-page: 643
  article-title: A novel high‐sensitivity rapid‐acquisition single‐photon cardiac imaging camera
  publication-title: J Nucl Med
– volume: 46
  start-page: 273
  year: 2016
  end-page: 285
  article-title: Novel cadmium‐zinc telluride devices for myocardial perfusion imaging‐technological aspects and clinical applications
  publication-title: Semin Nucl Med
– volume: 34
  start-page: 2216
  year: 1993
  end-page: 2221
  article-title: Compton scatter compensation using the triple‐energy window method for single‐ and dual‐isotope SPECT
  publication-title: J Nucl Med
– volume: 6
  start-page: 42
  year: 2013
  end-page: 53
  article-title: Quantitative performance of advanced resolution recovery strategies on SPECT images: evaluation with use of digital phantom models
  publication-title: Radio Phys Technol
– volume: 43
  start-page: 1723
  year: 2016
  end-page: 1738
  article-title: The EANM practice guidelines for bone scintigraphy
  publication-title: Eur J Nucl Med Mol Imaging
– volume: 48
  start-page: 950
  year: 2001
  end-page: 959
  article-title: Recent progress in CdTe and CdZnTe detectors
  publication-title: IEEE Trans Nucl Sci
– volume: 550
  start-page: 267
  year: 2005
  end-page: 277
  article-title: A photon counting CdTe gamma‐ and X‐ray camera
  publication-title: Nucl Inst Methods Phys Res
– volume: 26
  start-page: 191
  year: 1998
  end-page: 195
  article-title: Measurement of blood radioactivity for quantification of cerebral blood flow using a gamma camera
  publication-title: J Nucl Med Technol
– volume: 68
  start-page: 299
  year: 2012
  end-page: 306
  article-title: Development of the software package of the nuclear medicine data processor for education and research
  publication-title: Nihon Hoshasen Gijutsu Gakkai Zasshi
– volume: 29
  start-page: 636
  year: 2015
  end-page: 642
  article-title: Optimization of iterative reconstruction parameters with 3‐dimensional resolution recovery, scatter and attenuation correction in I‐FP‐CIT SPECT
  publication-title: Ann Nucl Med
– volume: 40
  start-page: 1130
  year: 2019
  end-page: 1137
  article-title: Shortened acquisition time in simultaneous 99mTc‐tetrofosmin and 123I‐β‐methyl‐p‐iodophenyl pentadecanoic acid dual‐tracer imaging with cadmium‐zinc‐telluride detectors in patients undergoing primary coronary intervention for acute myocardial infarction
  publication-title: Nucl Med Commun
– year: 2019
  article-title: Feasibility of simultaneous 99mTc‐tetrofosmin and 123I‐BMIPP dual‐tracer imaging with cadmium‐zinc‐telluride detectors in patients undergoing primary coronary intervention for acute myocardial infarction
  publication-title: J Nucl Cardiol
– ident: e_1_2_9_19_1
  doi: 10.1186/2191-219X-3-45
– ident: e_1_2_9_26_1
  doi: 10.1109/23.958705
– volume: 32
  start-page: 1157
  year: 1991
  ident: e_1_2_9_18_1
  article-title: In vivo SPECT quantitation of bone metabolism in hyperparathyroidism and thyrotoxicosis
  publication-title: J Nucl Med
– ident: e_1_2_9_31_1
  doi: 10.1007/s12194-012-0168-z
– ident: e_1_2_9_11_1
  doi: 10.1097/MNM.0000000000001084
– ident: e_1_2_9_9_1
  doi: 10.1007/s13139-017-0474-9
– volume-title: Nema Standards Publication NU‐1: Performance Measurements of Gamma Cameras
  year: 2007
  ident: e_1_2_9_14_1
– ident: e_1_2_9_3_1
  doi: 10.1016/j.nima.2005.04.081
– volume: 37
  start-page: 1959
  year: 2010
  ident: e_1_2_9_29_1
  article-title: A review on the clinical uses of SPECT/CT
  publication-title: EJNMMI
– ident: e_1_2_9_21_1
  doi: 10.1007/s12149-015-0987-1
– volume: 46
  start-page: 398
  year: 2018
  ident: e_1_2_9_16_1
  article-title: SNMMI procedure standard for bone scintigraphy 4.0
  publication-title: J Nucl Med Technol
– ident: e_1_2_9_13_1
  doi: 10.6009/jjrt.2012_JSRT_68.3.299
– volume: 34
  start-page: 2216
  year: 1993
  ident: e_1_2_9_23_1
  article-title: Compton scatter compensation using the triple‐energy window method for single‐ and dual‐isotope SPECT
  publication-title: J Nucl Med
– volume: 44
  start-page: 602
  year: 2003
  ident: e_1_2_9_24_1
  article-title: Evaluation of a small cadmium zinc telluride detector for scintimammography
  publication-title: J Nucl Med
– ident: e_1_2_9_6_1
  doi: 10.2967/jnumed.110.081323
– ident: e_1_2_9_17_1
  doi: 10.1097/MNM.0000000000001028
– ident: e_1_2_9_22_1
  doi: 10.1007/s12350-014-0008-0
– ident: e_1_2_9_27_1
  doi: 10.1118/1.4934830
– volume: 26
  start-page: 191
  year: 1998
  ident: e_1_2_9_20_1
  article-title: Measurement of blood radioactivity for quantification of cerebral blood flow using a gamma camera
  publication-title: J Nucl Med Technol
– volume: 35
  start-page: 34
  year: 2007
  ident: e_1_2_9_30_1
  article-title: Simultaneous 3‐dimensional resolution correction in SPECT reconstruction with an ordered‐subsets expectation maximization algorithm
  publication-title: J Nucl Med Technol
– ident: e_1_2_9_7_1
  doi: 10.1007/s00259-013-2617-2
– ident: e_1_2_9_8_1
  doi: 10.1053/j.semnuclmed.2016.01.002
– ident: e_1_2_9_4_1
  doi: 10.1007/s12149-009-0293-x
– ident: e_1_2_9_10_1
  doi: 10.1007/s12350-018-01585-9
– ident: e_1_2_9_5_1
  doi: 10.2967/jnumed.108.060020
– ident: e_1_2_9_25_1
  doi: 10.1088/0031-9155/60/8/3045
– ident: e_1_2_9_2_1
  doi: 10.1109/TNS.2004.829437
– ident: e_1_2_9_12_1
  doi: 10.1007/s12149-019-01382-5
– ident: e_1_2_9_15_1
  doi: 10.1007/s00259-016-3415-4
– ident: e_1_2_9_28_1
  doi: 10.1007/s12149-016-1144-1
SSID ssj0011838
Score 2.3974695
Snippet Purpose A high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used...
A high-energy-resolution whole-body SPECT-CT device (NM/CT 870 CZT; C-SPECT) equipped with a CZT detector has been developed and is being used clinically. A...
PurposeA high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used...
SourceID unpaywall
pubmedcentral
proquest
pubmed
crossref
wiley
SourceType Open Access Repository
Aggregation Database
Index Database
Enrichment Source
Publisher
StartPage 165
SubjectTerms Clinical medicine
Discovery NM/CT 870 CZT ⇒ NM/CT 870 CZT
Energy
Medical Imaging
medium‐energy high‐resolution sensitivity collimator
Physics
Sensors
Tomography
wide‐energy high‐resolution collimator
SummonAdditionalLinks – databaseName: Unpaywall
  dbid: UNPAY
  link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV1Lj9MwEB4tXQm48H4EFmRgLyClWzt26h6rKEuFlGrFtmLhEtmOIyratEAjBL-ecV5sWVRxc-Jx4tgz8jeZ8WeAY4PLuM5o6JuRkD63WeCPMiN8yVUYWEMHumLbT6bhZM7fXYiLA3jZ7oXZjd-zE2VWrE8DKuQ1OAwF4u0eHM6nZ-OPFREqC30EOLzjHb3cYHeluQIfr2ZB3iiLjfr5Qy2Xu0i1WmpOb0PUdrLOMPnSL7e6b379xd-4_yvuwK0GaZJxrRp34cAW9-B60sTS78MqvkTuT_6wfpN1ThAVkrcxmSYn0YygrZHo04y0myjJ-VmMt2sOaGK_lovNxmbE_dElH-LJe6KKjCRYOCdOyxYr59c_gPlpPIsmfnP4gm_4kEtfUxoIxFKOylhKLozEC6ktNVrmSmjJAqEpgotBxqkKGSINyrIQp2ggmJF58BB6xbqwj4EgBqPa8jzX6PuNFHo0wlId6lwhVgm08eB1O1GpaZjJ3QEZy7TmVGapG7-0Gj8PXnWym5qP459SR-18p41Nfk8ZlyMEnIhoPHjRVaM1uRCJKuy6dDIuMMkRRnnwqFaP7jVBgGimaj3cUZxOwDF179YUi88VY_cQ_Rj05Dw47lRsb-_fVNq3RyQdRwmrSk_-75lP4SZz2ThVvvkR9LbfSvsM4dRWP2_s6Td3IBcy
  priority: 102
  providerName: Unpaywall
– databaseName: Wiley Online Library Open Access
  dbid: 24P
  link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlZ3fa9RAEMeHWkF9EX8brTJiXxTi3W52cxvwpRyph5BS7BWLL2F3s8GDu9xpe4j_vbObXM6jUvAtyU5I2JnJfvbXNwCHlppxU7E0tplUsXBVEmeVlbESOk2cZUMT1PaLk3RyLj5fyIs9-LjZC9PqQ_QDbj4zwvfaJ7g2l4OtaKi2C_6BJUyqW3CbEcj4-ObitJ9DoGANG-EkT2OiINGLk_LB9t7d5ugaY15fKnl33az07196Pt_F2dAeHT-A-x1I4lHr-Yew55pHcKfopsofwyL_S7sft6LeuKyRoA8_5XhSDMZTpFTC8bcpbvZI4tlpTpdbiWd0P9az1cpV6Ads8Ws--YK6qbCggzP0QTRb-G77Ezg_zqfjSdz9WyG2YiRUbKjGJKGSVypWSkir6EQZx6xRtZZG8UQaRuwwrATTKSeQYLxKU5kNJbeqTp7CfrNs3HNAQixmnKhrQ127TFOHRTpmUlNrQpHE2Ajebaq4tJ3wuP__xbxsJZN56d1RBndE8La3XbVyG_-0Oth4quxS7rLkQmXEkwQsEbzpiylZ_AyIbtxy7W38vKMgSorgWevY_jFJQrAS7h7tuLw38ELcuyXN7HsQ5B5RN4XCLYLDPjhufPv3IW5uMCmPxgUPRy_-x_gl3ON-yU1YVH4A-1c_1-4VMdOVeR1S4w-_SAyG
  priority: 102
  providerName: Wiley-Blackwell
Title Experimental evaluation of the GE NM/CT 870 CZT clinical SPECT system equipped with WEHR and MEHRS collimator
URI https://onlinelibrary.wiley.com/doi/abs/10.1002%2Facm2.13158
https://www.ncbi.nlm.nih.gov/pubmed/33426812
https://www.proquest.com/docview/2489083812
https://www.proquest.com/docview/2476844750
https://pubmed.ncbi.nlm.nih.gov/PMC7882118
https://doi.org/10.1002/acm2.13158
UnpaywallVersion publishedVersion
Volume 22
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
journalDatabaseRights – providerCode: PRVAON
  databaseName: en accès libre
  customDbUrl:
  eissn: 1526-9914
  dateEnd: 99991231
  omitProxy: true
  ssIdentifier: ssj0011838
  issn: 1526-9914
  databaseCode: DOA
  dateStart: 20000101
  isFulltext: true
  titleUrlDefault: https://www.doaj.org/
  providerName: Directory of Open Access Journals
– providerCode: PRVBFR
  databaseName: Free Medical Journals
  customDbUrl:
  eissn: 1526-9914
  dateEnd: 99991231
  omitProxy: true
  ssIdentifier: ssj0011838
  issn: 1526-9914
  databaseCode: DIK
  dateStart: 20000101
  isFulltext: true
  titleUrlDefault: http://www.freemedicaljournals.com
  providerName: Flying Publisher
– providerCode: PRVFQY
  databaseName: GFMER Free Medical Journals
  customDbUrl:
  eissn: 1526-9914
  dateEnd: 99991231
  omitProxy: true
  ssIdentifier: ssj0011838
  issn: 1526-9914
  databaseCode: GX1
  dateStart: 20000101
  isFulltext: true
  titleUrlDefault: http://www.gfmer.ch/Medical_journals/Free_medical.php
  providerName: Geneva Foundation for Medical Education and Research
– providerCode: PRVHPJ
  databaseName: ROAD: Directory of Open Access Scholarly Resources
  customDbUrl:
  eissn: 1526-9914
  dateEnd: 99991231
  omitProxy: true
  ssIdentifier: ssj0011838
  issn: 1526-9914
  databaseCode: M~E
  dateStart: 20000101
  isFulltext: true
  titleUrlDefault: https://road.issn.org
  providerName: ISSN International Centre
– providerCode: PRVAQN
  databaseName: PubMed Central
  customDbUrl:
  eissn: 1526-9914
  dateEnd: 99991231
  omitProxy: true
  ssIdentifier: ssj0011838
  issn: 1526-9914
  databaseCode: RPM
  dateStart: 20000101
  isFulltext: true
  titleUrlDefault: https://www.ncbi.nlm.nih.gov/pmc/
  providerName: National Library of Medicine
– providerCode: PRVPQU
  databaseName: Health & Medical Collection
  customDbUrl:
  eissn: 1526-9914
  dateEnd: 99991231
  omitProxy: true
  ssIdentifier: ssj0011838
  issn: 1526-9914
  databaseCode: 7X7
  dateStart: 20000301
  isFulltext: true
  titleUrlDefault: https://search.proquest.com/healthcomplete
  providerName: ProQuest
– providerCode: PRVPQU
  databaseName: ProQuest Central
  customDbUrl: http://www.proquest.com/pqcentral?accountid=15518
  eissn: 1526-9914
  dateEnd: 99991231
  omitProxy: true
  ssIdentifier: ssj0011838
  issn: 1526-9914
  databaseCode: BENPR
  dateStart: 20000301
  isFulltext: true
  titleUrlDefault: https://www.proquest.com/central
  providerName: ProQuest
– providerCode: PRVWIB
  databaseName: KBPluse Wiley Online Library: Open Access
  customDbUrl:
  eissn: 1526-9914
  dateEnd: 99991231
  omitProxy: true
  ssIdentifier: ssj0011838
  issn: 1526-9914
  databaseCode: AVUZU
  dateStart: 20000101
  isFulltext: true
  titleUrlDefault: https://www.kbplus.ac.uk/kbplus7/publicExport/pkg/559
  providerName: Wiley-Blackwell
– providerCode: PRVWIB
  databaseName: Wiley Online Library (Open Access Collection)
  customDbUrl:
  eissn: 1526-9914
  dateEnd: 99991231
  omitProxy: true
  ssIdentifier: ssj0011838
  issn: 1526-9914
  databaseCode: 24P
  dateStart: 20000101
  isFulltext: true
  titleUrlDefault: https://authorservices.wiley.com/open-science/open-access/browse-journals.html
  providerName: Wiley-Blackwell
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwfV3da9swED_alH28jK37aLYuaKwvG3iJZNlRHsbIgtswSAhtwrK-GEmWaSBx3LVh7L_fSf5oQ0dehG2d8cedpJ90p98BnGgcxlVCQ0_3AuFxk_heL9GBJ7gMfaNpRzm2_dE4HM74j3kw34NxtRfGhlVWfaLrqJO1tmvkbcZFD-ECjkff8mvPZo2y3tUqhYYsUyskXx3F2D4cMMuM1YCD79F4cl77FdCARU1SytpSr9gX6lOb8P3-sPQAaz4MmXyyyXL5949cLrdhrRuXTp_DsxJQkn5hAS9gz2SH8HhUuswP4ZGL8dQ3L2EV3WPzJ3c032SdEoSB5Cwi41F7MCXYuMjgckqqXZPkYhLh5YL0mZjrzSLPTULsEi75GQ3PicwSMsKDC2LNarGyE_lXMDuNpoOhV2Zb8DTvcuEpSv0AwZPlLhaCB1rgiVCGaiVSGSjB_EBRRBOdhFMZMoQWlCVhGPQ6AdMi9V9DI1tn5ggIgi6qDE9ThZO9nsQpTGCoClUqEZz4SjfhU_WzY11SkduMGMu4IFFmsVVM7BTThI-1bF4QcPxX6rjSWVw2wpv4zmSa8KGuxuZjfSIyM-uNlbGeSI64qQlvChXXj_F9hC_u7u6W8msBS829XZMtrhxFdxcnLmhsTTipzWTn2392FrRDJO4PRswdvd39qe_gKbNhNy6w_Bgat7835j3iplvVgn3GJ1h2591W2TBabg0Cy7M5xWuz8aT_6x9kERkH
linkProvider ProQuest
linkToHtml http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV3fb9MwED6NTTBeEIxfhQFGjAeQQmvHSZ2HCY2S0bGlmrZOTLwE23FEpTbt6Kpp_xx_G2c3yVYN9W1vaXNt2txd7jvf-TuALY1hXGU09HQUCI-bzPeiTAee4DL0jaYt5dj2k17YPeHfT4PTFfhb7YWxbZXVM9E9qLOxtmvkTcZFhHAB49HnyZlnp0bZ6mo1QkOWoxWybUcxVm7s2DeXF5jCTbf3vqK-3zO2G_c7Xa-cMuBp3ubCU5T6AYIGy9krBA-0wBdCGaqVyGWgBPMDRTGKtjJOZcgwpFKWhWEQtQKmRe7j996BNe7zCJO_tS9x7_CormOgw4iaFJU1pR6xT9SndsD89TB4A9vebNFcnxUTeXkhh8NFGO3i4O5DeFACWLIzt7hHsGKKDbiXlCX6Dbjrekr19DGM4mvTA8gVrTgZ5wRhJ_kWk17S7PQJOjPp_OyTapcmOT6M8e05yTQxZ7PBZGIyYpeMyY-4e0RkkZEED46JNePByC4cPIGTW7nvT2G1GBfmORAEeVQZnucKk8tIYsoUGKpClUsEQ77SDfhQ3exUl9TndgLHMJ2TNrPUKiZ1imnAu1p2Mif8-K_UZqWztHT6aXplog14W59Gd7U1GFmY8czK2MonR5zWgGdzFdeX8X2ES-7T7QXl1wKWCnzxTDH47SjB25goobE1YKs2k6W__qOzoCUi6U4nYe7oxfK_-gbWu_3kID3Y6-2_hPvMtvy4pvZNWD3_MzOvELOdq9elYxD4ddu--A-VbkyJ
linkToPdf http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lb9NAEB6VIgoXBOVlKLCIcgDJJLvrx_qAUJUmpJREFU1FxMV412sRKXFc0qjqX-PXMbt-tFFRbr35MX7OjOeb3fE3ALsKw7hMaeCqyBeup1PuRqnyXeElAdeKtqVl2x8Mg_6J93Xsjzfgb_0vjCmrrL-J9kOdzpUZI28xT0QIFzAetbKqLOJov_e5OHVNBykz01q30yhN5FBfnGP6tvh0sI-6fsdYrzvq9N2qw4CrvNATrqSU-wgYDF-vEJ6vBK4IqamSIkt8KRj3JcUI2k49mgQMwyllaRD4UdtnSmQcz3sLboecR6acMBw3yR7idi4aOlTWStSMfaScmtbyVwPgNVR7vTjz7jIvkovzZDpdBdA2AvYewP0KupK90tYewobOt2FrUE3Ob8MdW02qFo9g1r3SN4BcEoqTeUYQcJIvXTIctDojgm5MOj9HpP4_kxwfdXFzSS9N9OlyUhQ6JWawmPzo9r-TJE_JABeOiTHgycwMGTyGkxt5609gM5_n-hkQhHdUai_LJKaVUYLJkq-pDGSWIAziUjnwvn7ZsapIz03vjWlc0jWz2Cgmtopx4G0jW5RUH_-V2ql1FlfuvogvjdOBN81udFQz-5Lker40MmbO00OE5sDTUsXNZThHoGSPDleU3wgYEvDVPfnktyUDDzFFQmNzYLcxk7V3_8Fa0BqReK8zYHbp-fpHfQ1b6IHxt4Ph4Qu4x0ytj61m34HNsz9L_RLB2pl8Zb2CwK-bdsN_WrNKIw
linkToUnpaywall http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV1Lj9MwEB4tXQm48H4EFmRgLyClWzt26h6rKEuFlGrFtmLhEtmOIyratEAjBL-ecV5sWVRxc-Jx4tgz8jeZ8WeAY4PLuM5o6JuRkD63WeCPMiN8yVUYWEMHumLbT6bhZM7fXYiLA3jZ7oXZjd-zE2VWrE8DKuQ1OAwF4u0eHM6nZ-OPFREqC30EOLzjHb3cYHeluQIfr2ZB3iiLjfr5Qy2Xu0i1WmpOb0PUdrLOMPnSL7e6b379xd-4_yvuwK0GaZJxrRp34cAW9-B60sTS78MqvkTuT_6wfpN1ThAVkrcxmSYn0YygrZHo04y0myjJ-VmMt2sOaGK_lovNxmbE_dElH-LJe6KKjCRYOCdOyxYr59c_gPlpPIsmfnP4gm_4kEtfUxoIxFKOylhKLozEC6ktNVrmSmjJAqEpgotBxqkKGSINyrIQp2ggmJF58BB6xbqwj4EgBqPa8jzX6PuNFHo0wlId6lwhVgm08eB1O1GpaZjJ3QEZy7TmVGapG7-0Gj8PXnWym5qP459SR-18p41Nfk8ZlyMEnIhoPHjRVaM1uRCJKuy6dDIuMMkRRnnwqFaP7jVBgGimaj3cUZxOwDF179YUi88VY_cQ_Rj05Dw47lRsb-_fVNq3RyQdRwmrSk_-75lP4SZz2ThVvvkR9LbfSvsM4dRWP2_s6Td3IBcy
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Experimental+evaluation+of+the+GE+NM%2FCT+870+CZT+clinical+SPECT+system+equipped+with+WEHR+and+MEHRS+collimator&rft.jtitle=Journal+of+applied+clinical+medical+physics&rft.au=Ito%2C+Toshimune&rft.au=Matsusaka%2C+Yohji&rft.au=Onoguchi%2C+Masahisa&rft.au=Ichikawa%2C+Hajime&rft.date=2021-02-01&rft.pub=John+Wiley+%26+Sons%2C+Inc&rft.eissn=1526-9914&rft.volume=22&rft.issue=2&rft.spage=165&rft.epage=177&rft_id=info:doi/10.1002%2Facm2.13158&rft.externalDBID=HAS_PDF_LINK
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1526-9914&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1526-9914&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1526-9914&client=summon