Experimental evaluation of the GE NM/CT 870 CZT clinical SPECT system equipped with WEHR and MEHRS collimator
Purpose A high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium‐energy radionuclides....
Saved in:
| Published in | Journal of applied clinical medical physics Vol. 22; no. 2; pp. 165 - 177 |
|---|---|
| Main Authors | , , , , , , , |
| Format | Journal Article |
| Language | English |
| Published |
United States
John Wiley & Sons, Inc
01.02.2021
John Wiley and Sons Inc |
| Subjects | |
| Online Access | Get full text |
| ISSN | 1526-9914 1526-9914 |
| DOI | 10.1002/acm2.13158 |
Cover
| Abstract | Purpose
A high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium‐energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C‐SPECT, and a NaI scintillation detector‐equipped Anger‐type SPECT (A‐SPECT) scanner, with a LEHR and LMEGP.
Methods
A line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99mTc and 123I. We also measured the system's sensitivity and high‐contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high‐contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction.
Results
The energy resolution with 99mTc was 5.6% in C‐SPECT with WEHR and 9.9% in A‐SPECT with LEHR. Using 123I, the results were 9.1% in C‐SPECT with WEHR, 5.5% in C‐SPECT with MEHRS, and 10.4% in A‐SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C‐SPECT performed better in SPECT acquisition. High‐contrast resolution was improved in C‐SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C‐SPECT.
Conclusion
C‐SPECT demonstrates excellent energy resolution and improved high‐contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C‐SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high‐contrast resolution. |
|---|---|
| AbstractList | A high-energy-resolution whole-body SPECT-CT device (NM/CT 870 CZT; C-SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium-energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C-SPECT, and a NaI scintillation detector-equipped Anger-type SPECT (A-SPECT) scanner, with a LEHR and LMEGP.
A line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using
Tc and
I. We also measured the system's sensitivity and high-contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high-contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction.
The energy resolution with
Tc was 5.6% in C-SPECT with WEHR and 9.9% in A-SPECT with LEHR. Using 123I, the results were 9.1% in C-SPECT with WEHR, 5.5% in C-SPECT with MEHRS, and 10.4% in A-SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C-SPECT performed better in SPECT acquisition. High-contrast resolution was improved in C-SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C-SPECT.
C-SPECT demonstrates excellent energy resolution and improved high-contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C-SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high-contrast resolution. Purpose A high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium‐energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C‐SPECT, and a NaI scintillation detector‐equipped Anger‐type SPECT (A‐SPECT) scanner, with a LEHR and LMEGP. Methods A line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99mTc and 123I. We also measured the system's sensitivity and high‐contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high‐contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction. Results The energy resolution with 99mTc was 5.6% in C‐SPECT with WEHR and 9.9% in A‐SPECT with LEHR. Using 123I, the results were 9.1% in C‐SPECT with WEHR, 5.5% in C‐SPECT with MEHRS, and 10.4% in A‐SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C‐SPECT performed better in SPECT acquisition. High‐contrast resolution was improved in C‐SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C‐SPECT. Conclusion C‐SPECT demonstrates excellent energy resolution and improved high‐contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C‐SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high‐contrast resolution. PurposeA high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium‐energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C‐SPECT, and a NaI scintillation detector‐equipped Anger‐type SPECT (A‐SPECT) scanner, with a LEHR and LMEGP.MethodsA line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99mTc and 123I. We also measured the system's sensitivity and high‐contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high‐contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction.ResultsThe energy resolution with 99mTc was 5.6% in C‐SPECT with WEHR and 9.9% in A‐SPECT with LEHR. Using 123I, the results were 9.1% in C‐SPECT with WEHR, 5.5% in C‐SPECT with MEHRS, and 10.4% in A‐SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C‐SPECT performed better in SPECT acquisition. High‐contrast resolution was improved in C‐SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C‐SPECT.ConclusionC‐SPECT demonstrates excellent energy resolution and improved high‐contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C‐SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high‐contrast resolution. A high-energy-resolution whole-body SPECT-CT device (NM/CT 870 CZT; C-SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium-energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C-SPECT, and a NaI scintillation detector-equipped Anger-type SPECT (A-SPECT) scanner, with a LEHR and LMEGP.PURPOSEA high-energy-resolution whole-body SPECT-CT device (NM/CT 870 CZT; C-SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium-energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C-SPECT, and a NaI scintillation detector-equipped Anger-type SPECT (A-SPECT) scanner, with a LEHR and LMEGP.A line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99m Tc and 123 I. We also measured the system's sensitivity and high-contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high-contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction.METHODSA line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99m Tc and 123 I. We also measured the system's sensitivity and high-contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high-contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction.The energy resolution with 99m Tc was 5.6% in C-SPECT with WEHR and 9.9% in A-SPECT with LEHR. Using 123I, the results were 9.1% in C-SPECT with WEHR, 5.5% in C-SPECT with MEHRS, and 10.4% in A-SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C-SPECT performed better in SPECT acquisition. High-contrast resolution was improved in C-SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C-SPECT.RESULTSThe energy resolution with 99m Tc was 5.6% in C-SPECT with WEHR and 9.9% in A-SPECT with LEHR. Using 123I, the results were 9.1% in C-SPECT with WEHR, 5.5% in C-SPECT with MEHRS, and 10.4% in A-SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C-SPECT performed better in SPECT acquisition. High-contrast resolution was improved in C-SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C-SPECT.C-SPECT demonstrates excellent energy resolution and improved high-contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C-SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high-contrast resolution.CONCLUSIONC-SPECT demonstrates excellent energy resolution and improved high-contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C-SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high-contrast resolution. |
| Author | Sato, Kozo Matsusaka, Yohji Shishido, Masaaki Shibutani, Takayuki Ito, Toshimune Onoguchi, Masahisa Ichikawa, Hajime Okuda, Koichi |
| AuthorAffiliation | 1 Department of Radiology Saiseikai Yokohamashi Tobu Hospital Yokohama Japan 5 Department of Physics Kanazawa Medical University Kahoku, Ishikawa Japan 2 Department of Diagnostic Radiology School of Medicine Keio University Tokyo Japan 3 Department of Quantum Medical Technology Graduate School of Medical Sciences Kanazawa University Kanazawa Japan 6 Department of Diagnostic Radiology Saiseikai Yokohamashi Tobu Hospital Yokohama Kanagawa Japan 4 Department of Radiology Toyohashi Municipal Hospital Toyohashi Japan |
| AuthorAffiliation_xml | – name: 1 Department of Radiology Saiseikai Yokohamashi Tobu Hospital Yokohama Japan – name: 6 Department of Diagnostic Radiology Saiseikai Yokohamashi Tobu Hospital Yokohama Kanagawa Japan – name: 4 Department of Radiology Toyohashi Municipal Hospital Toyohashi Japan – name: 2 Department of Diagnostic Radiology School of Medicine Keio University Tokyo Japan – name: 5 Department of Physics Kanazawa Medical University Kahoku, Ishikawa Japan – name: 3 Department of Quantum Medical Technology Graduate School of Medical Sciences Kanazawa University Kanazawa Japan |
| Author_xml | – sequence: 1 givenname: Toshimune surname: Ito fullname: Ito, Toshimune email: toshimu2001@gmail.com organization: Saiseikai Yokohamashi Tobu Hospital – sequence: 2 givenname: Yohji surname: Matsusaka fullname: Matsusaka, Yohji organization: Keio University – sequence: 3 givenname: Masahisa surname: Onoguchi fullname: Onoguchi, Masahisa organization: Kanazawa University – sequence: 4 givenname: Hajime surname: Ichikawa fullname: Ichikawa, Hajime organization: Toyohashi Municipal Hospital – sequence: 5 givenname: Koichi surname: Okuda fullname: Okuda, Koichi organization: Kanazawa Medical University – sequence: 6 givenname: Takayuki surname: Shibutani fullname: Shibutani, Takayuki organization: Kanazawa University – sequence: 7 givenname: Masaaki surname: Shishido fullname: Shishido, Masaaki organization: Kanazawa University – sequence: 8 givenname: Kozo surname: Sato fullname: Sato, Kozo organization: Saiseikai Yokohamashi Tobu Hospital |
| BackLink | https://www.ncbi.nlm.nih.gov/pubmed/33426812$$D View this record in MEDLINE/PubMed |
| BookMark | eNp9kU1v1DAQhi1URD8v_ABkiQui2tZ27MS5VKqi0CJ1KaKLkLhYjuOwrhw7jZMu--_rbQotFerJY_uZd2be2QVbzjsNwFuMjjBC5FiqlhzhBDP-CuxgRtJZnmO69STeBrshXCOEMU_4G7CdJJSkHJMd0Ja_O92bVrtBWqhvpR3lYLyDvoHDUsOzEn6ZHxcLyDMEi58LqKxxRkX26msZn8M6DLqF-mY0XadruDLDEv4oz79B6Wo4j8EVVN5a08rB9_vgdSNt0AcP5x74_qlcFOezi8uzz8XpxUzRjPJZhXHCaMIQSnPOKVM8Xnilsap4I1nFScIqTEiCaoplSjJOMKnTlOWIEcWbZA8cTrqj6-R6Ja0VXRxS9muBkdiYJjamiXvTIn0y0d1YtbpW0YtePmZ4acS_P84sxS9_KzIeC-ONwIcHgd7fjDoMojVBaWul034MgtAs5ZRmDEX0_TP02o-9i2ZEiuco7geTSL172tHfVv7sLQJoAlTvQ-h1I5QZ7hcXGzT2_1N-fJbyoiV4glfG6vULpDgt5mTKuQNtoMOH |
| CitedBy_id | crossref_primary_10_1053_j_semnuclmed_2023_10_007 crossref_primary_10_1186_s40658_024_00659_7 crossref_primary_10_3390_cancers13246183 crossref_primary_10_1053_j_semnuclmed_2022_01_004 crossref_primary_10_1186_s40658_024_00656_w crossref_primary_10_1002_mp_17724 crossref_primary_10_1186_s40658_022_00463_1 crossref_primary_10_1186_s40658_022_00517_4 crossref_primary_10_1038_s41566_021_00909_5 crossref_primary_10_1007_s13246_023_01257_0 crossref_primary_10_1016_j_ijleo_2021_167681 crossref_primary_10_1007_s00247_023_05682_x crossref_primary_10_6009_jjrt_2022_2024 crossref_primary_10_1007_s12194_023_00746_x crossref_primary_10_2967_jnumed_124_267542 crossref_primary_10_4103_jmp_jmp_71_22 crossref_primary_10_1038_s41598_021_03853_0 crossref_primary_10_1055_a_1993_7420 crossref_primary_10_1007_s00259_023_06176_6 crossref_primary_10_1186_s13550_023_00973_8 crossref_primary_10_1186_s40658_024_00684_6 crossref_primary_10_1088_1361_6560_ac5671 crossref_primary_10_6009_jjrt_2022_2054 crossref_primary_10_1038_s41551_022_00866_6 crossref_primary_10_2967_jnmt_122_265328 crossref_primary_10_1186_s40658_023_00566_3 crossref_primary_10_1007_s13139_023_00825_2 crossref_primary_10_1055_a_1715_5305 |
| Cites_doi | 10.1186/2191-219X-3-45 10.1109/23.958705 10.1007/s12194-012-0168-z 10.1097/MNM.0000000000001084 10.1007/s13139-017-0474-9 10.1016/j.nima.2005.04.081 10.1007/s12149-015-0987-1 10.6009/jjrt.2012_JSRT_68.3.299 10.2967/jnumed.110.081323 10.1097/MNM.0000000000001028 10.1007/s12350-014-0008-0 10.1118/1.4934830 10.1007/s00259-013-2617-2 10.1053/j.semnuclmed.2016.01.002 10.1007/s12149-009-0293-x 10.1007/s12350-018-01585-9 10.2967/jnumed.108.060020 10.1088/0031-9155/60/8/3045 10.1109/TNS.2004.829437 10.1007/s12149-019-01382-5 10.1007/s00259-016-3415-4 10.1007/s12149-016-1144-1 |
| ContentType | Journal Article |
| Copyright | 2021 The Authors. published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. 2021. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. |
| Copyright_xml | – notice: 2021 The Authors. published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. – notice: 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. – notice: 2021. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. |
| DBID | 24P AAYXX CITATION NPM 3V. 7X7 7XB 88I 8FI 8FJ 8FK ABUWG AFKRA AZQEC BENPR CCPQU DWQXO FYUFA GHDGH GNUQQ HCIFZ K9. M0S M2P PHGZM PHGZT PIMPY PJZUB PKEHL PPXIY PQEST PQQKQ PQUKI PRINS Q9U 7X8 5PM ADTOC UNPAY |
| DOI | 10.1002/acm2.13158 |
| DatabaseName | Wiley Online Library Open Access CrossRef PubMed ProQuest Central (Corporate) Health & Medical Collection ProQuest Central (purchase pre-March 2016) Science Database (Alumni Edition) ProQuest Hospital Collection Hospital Premium Collection (Alumni Edition) ProQuest Central (Alumni) (purchase pre-March 2016) ProQuest Central (Alumni) ProQuest Central UK/Ireland ProQuest Central Essentials ProQuest Central ProQuest One Community College ProQuest Central Health Research Premium Collection Health Research Premium Collection (Alumni) ProQuest Central Student SciTech Premium Collection ProQuest Health & Medical Complete (Alumni) ProQuest Health & Medical Collection Science Database ProQuest Central Premium ProQuest One Academic (New) Publicly Available Content Database ProQuest Health & Medical Research Collection ProQuest One Academic Middle East (New) ProQuest One Health & Nursing ProQuest One Academic Eastern Edition (DO NOT USE) ProQuest One Academic ProQuest One Academic UKI Edition ProQuest Central China ProQuest Central Basic MEDLINE - Academic PubMed Central (Full Participant titles) Unpaywall for CDI: Periodical Content Unpaywall |
| DatabaseTitle | CrossRef PubMed Publicly Available Content Database ProQuest Central Student ProQuest One Academic Middle East (New) ProQuest Central Essentials ProQuest Health & Medical Complete (Alumni) ProQuest Central (Alumni Edition) SciTech Premium Collection ProQuest One Community College ProQuest One Health & Nursing ProQuest Central China ProQuest Central ProQuest Health & Medical Research Collection Health Research Premium Collection Health and Medicine Complete (Alumni Edition) ProQuest Central Korea ProQuest Central (New) ProQuest Science Journals (Alumni Edition) ProQuest Central Basic ProQuest Science Journals ProQuest One Academic Eastern Edition ProQuest Hospital Collection Health Research Premium Collection (Alumni) ProQuest Hospital Collection (Alumni) ProQuest Health & Medical Complete ProQuest One Academic UKI Edition ProQuest One Academic ProQuest One Academic (New) ProQuest Central (Alumni) MEDLINE - Academic |
| DatabaseTitleList | PubMed Publicly Available Content Database MEDLINE - Academic |
| Database_xml | – sequence: 1 dbid: 24P name: Wiley Online Library (Open Access Collection) url: https://authorservices.wiley.com/open-science/open-access/browse-journals.html sourceTypes: Publisher – sequence: 2 dbid: NPM name: PubMed url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database – sequence: 3 dbid: UNPAY name: Unpaywall url: https://proxy.k.utb.cz/login?url=https://unpaywall.org/ sourceTypes: Open Access Repository – sequence: 4 dbid: BENPR name: ProQuest Central url: http://www.proquest.com/pqcentral?accountid=15518 sourceTypes: Aggregation Database |
| DeliveryMethod | fulltext_linktorsrc |
| Discipline | Medicine Physics |
| DocumentTitleAlternate | Ito et al |
| EISSN | 1526-9914 |
| EndPage | 177 |
| ExternalDocumentID | 10.1002/acm2.13158 PMC7882118 33426812 10_1002_acm2_13158 ACM213158 |
| Genre | article Journal Article |
| GeographicLocations | United States--US Japan |
| GeographicLocations_xml | – name: United States--US – name: Japan |
| GroupedDBID | 0R~ 1OC 24P 29J 2WC 53G 5GY 7X7 88I 8FI 8FJ AAHHS ABUWG ACCFJ ACCMX ACGFO ACXQS ADBBV ADKYN ADPDF ADZMN ADZOD AEEZP AENEX AEQDE AFKRA AIWBW AJBDE ALIPV ALMA_UNASSIGNED_HOLDINGS ALUQN AOIJS AVUZU AZQEC BAWUL BCNDV BENPR BPHCQ BVXVI CCPQU CS3 DIK DU5 DWQXO E3Z EBS EJD EMOBN FRP FYUFA GNUQQ GROUPED_DOAJ GX1 H13 HCIFZ HMCUK HYE IAO IHR INH ITC KWQ M2P M~E OK1 OVD OVEED P6G PIMPY PQQKQ PROAC RNS RPM TR2 UKHRP W2D WIN XSB AAMMB AAYXX AEFGJ AGXDD AIDQK AIDYY CITATION OVT PHGZM PHGZT PUEGO NPM 3V. 7XB 8FK K9. PJZUB PKEHL PPXIY PQEST PQUKI PRINS Q9U 7X8 5PM ADTOC UNPAY |
| ID | FETCH-LOGICAL-c4748-b113543500698845c84358be1cb8fa5b8235b12230d41a6278212d6659052c8f3 |
| IEDL.DBID | BENPR |
| ISSN | 1526-9914 |
| IngestDate | Sun Oct 26 03:58:02 EDT 2025 Tue Sep 30 16:37:47 EDT 2025 Fri Sep 05 11:07:52 EDT 2025 Tue Oct 07 07:08:01 EDT 2025 Thu Apr 03 07:04:47 EDT 2025 Wed Oct 01 02:06:46 EDT 2025 Thu Apr 24 22:51:25 EDT 2025 Wed Jan 22 16:30:00 EST 2025 |
| IsDoiOpenAccess | true |
| IsOpenAccess | true |
| IsPeerReviewed | true |
| IsScholarly | true |
| Issue | 2 |
| Keywords | medium-energy high-resolution sensitivity collimator Discovery NM/CT 870 CZT ⇒ NM/CT 870 CZT wide-energy high-resolution collimator |
| Language | English |
| License | Attribution 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. cc-by |
| LinkModel | DirectLink |
| MergedId | FETCHMERGED-LOGICAL-c4748-b113543500698845c84358be1cb8fa5b8235b12230d41a6278212d6659052c8f3 |
| Notes | Funding This research did not receive any specific grant from funding agencies in the public, commercial, or not‐for‐profit sectors. ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
| OpenAccessLink | https://www.proquest.com/docview/2489083812?pq-origsite=%requestingapplication%&accountid=15518 |
| PMID | 33426812 |
| PQID | 2489083812 |
| PQPubID | 4370306 |
| PageCount | 13 |
| ParticipantIDs | unpaywall_primary_10_1002_acm2_13158 pubmedcentral_primary_oai_pubmedcentral_nih_gov_7882118 proquest_miscellaneous_2476844750 proquest_journals_2489083812 pubmed_primary_33426812 crossref_citationtrail_10_1002_acm2_13158 crossref_primary_10_1002_acm2_13158 wiley_primary_10_1002_acm2_13158_ACM213158 |
| ProviderPackageCode | CITATION AAYXX |
| PublicationCentury | 2000 |
| PublicationDate | February 2021 |
| PublicationDateYYYYMMDD | 2021-02-01 |
| PublicationDate_xml | – month: 02 year: 2021 text: February 2021 |
| PublicationDecade | 2020 |
| PublicationPlace | United States |
| PublicationPlace_xml | – name: United States – name: Malden Massachusetts – name: Hoboken |
| PublicationTitle | Journal of applied clinical medical physics |
| PublicationTitleAlternate | J Appl Clin Med Phys |
| PublicationYear | 2021 |
| Publisher | John Wiley & Sons, Inc John Wiley and Sons Inc |
| Publisher_xml | – name: John Wiley & Sons, Inc – name: John Wiley and Sons Inc |
| References | 2009; 23 1998; 26 2013; 3 2010; 37 2005; 550 2019; 33 1991; 32 2011; 52 2007 2001; 48 2014; 41 2013; 6 2007; 35 2018; 46 2017; 51 2017; 31 1993; 34 2004; 51 2019; 40 2015; 29 2015; 60 2009; 50 2015; 42 2015; 22 2016; 43 2019 2012; 68 2016; 46 2003; 44 e_1_2_9_31_1 e_1_2_9_11_1 e_1_2_9_10_1 e_1_2_9_13_1 e_1_2_9_12_1 Bartel TB (e_1_2_9_16_1) 2018; 46 Ichihara T (e_1_2_9_23_1) 1993; 34 Association NEM (e_1_2_9_14_1) 2007 e_1_2_9_15_1 e_1_2_9_17_1 e_1_2_9_19_1 Mariani G (e_1_2_9_29_1) 2010; 37 Israel O (e_1_2_9_18_1) 1991; 32 e_1_2_9_22_1 e_1_2_9_21_1 e_1_2_9_8_1 e_1_2_9_7_1 Mueller B (e_1_2_9_24_1) 2003; 44 e_1_2_9_6_1 e_1_2_9_5_1 Takahashi Y (e_1_2_9_30_1) 2007; 35 e_1_2_9_4_1 e_1_2_9_3_1 e_1_2_9_2_1 e_1_2_9_9_1 e_1_2_9_26_1 e_1_2_9_25_1 e_1_2_9_28_1 Murase K (e_1_2_9_20_1) 1998; 26 e_1_2_9_27_1 |
| References_xml | – volume: 41 start-page: 522 year: 2014 end-page: 528 article-title: Factors affecting the myocardial activity acquired during exercise SPECT with a high‐sensitivity cardiac CZT camera as compared with conventional Anger camera publication-title: Eur J Nucl Med Mol Imaging – volume: 51 start-page: 1191 year: 2004 end-page: 1198 article-title: CdTe and CdZnTe X‐ray and gamma‐ray detectors for imaging systems publication-title: IEEE Trans Nucl Sci – volume: 40 start-page: 693 year: 2019 end-page: 702 article-title: Evaluation of edge‐preserving and noise‐reducing effects using the nonlinear diffusion method in bone single‐photon emission computed tomography publication-title: Nucl Med Commun – volume: 42 start-page: 6895 year: 2015 end-page: 6911 article-title: Scatter and crosstalk corrections for (99m)Tc/(123)I dual‐radionuclide imaging using a CZT SPECT system with pinhole collimators publication-title: Med Phys – volume: 31 start-page: 172 year: 2017 end-page: 180 article-title: Respiratory average CT for attenuation correction in myocardial perfusion SPECT/CT publication-title: Ann Nucl Med – volume: 60 start-page: 3045 year: 2015 end-page: 3063 article-title: Model‐based correction for scatter and tailing effects in simultaneous 99mTc and 123I imaging for a CdZnTe cardiac SPECT camera publication-title: Phys Med Biol – year: 2007 – volume: 32 start-page: 1157 year: 1991 end-page: 1161 article-title: In vivo SPECT quantitation of bone metabolism in hyperparathyroidism and thyrotoxicosis publication-title: J Nucl Med – volume: 22 start-page: 334 year: 2015 end-page: 343 article-title: Scatter correction improves concordance in SPECT MPI with a dedicated cardiac SPECT solid‐state camera publication-title: J Nucl Cardiol – volume: 37 start-page: 1959 year: 2010 end-page: 1985 article-title: A review on the clinical uses of SPECT/CT publication-title: EJNMMI – volume: 52 start-page: 210 year: 2011 end-page: 217 article-title: Cardiac dedicated ultrafast SPECT cameras: new designs and clinical implications publication-title: J Nucl Med – volume: 46 start-page: 398 year: 2018 end-page: 404 article-title: SNMMI procedure standard for bone scintigraphy 4.0 publication-title: J Nucl Med Technol – volume: 3 start-page: 45 year: 2013 article-title: Quantification of 99mTc‐DPD concentration in the lumbar spine with SPECT/CT publication-title: EJNMMI Res – volume: 51 start-page: 331 year: 2017 end-page: 337 article-title: Evaluation of cadmium‐zinc‐telluride detector‐based single‐photon emission computed tomography for nuclear cardiology: a comparison with conventional anger single‐photon emission computed tomography publication-title: Nucl Med Mol Imaging – volume: 44 start-page: 602 year: 2003 end-page: 609 article-title: Evaluation of a small cadmium zinc telluride detector for scintimammography publication-title: J Nucl Med – volume: 35 start-page: 34 year: 2007 end-page: 38 article-title: Simultaneous 3‐dimensional resolution correction in SPECT reconstruction with an ordered‐subsets expectation maximization algorithm publication-title: J Nucl Med Technol – volume: 23 start-page: 763 year: 2009 end-page: 770 article-title: Development of an ultra‐high resolution SPECT system with a CdTe semiconductor detector publication-title: Ann Nucl Med – volume: 33 start-page: 725 year: 2019 end-page: 732 article-title: Nuclear medicine practice in Japan: a report of the eighth nationwide survey in 2017 publication-title: Ann Nucl Med – volume: 50 start-page: 635 year: 2009 end-page: 643 article-title: A novel high‐sensitivity rapid‐acquisition single‐photon cardiac imaging camera publication-title: J Nucl Med – volume: 46 start-page: 273 year: 2016 end-page: 285 article-title: Novel cadmium‐zinc telluride devices for myocardial perfusion imaging‐technological aspects and clinical applications publication-title: Semin Nucl Med – volume: 34 start-page: 2216 year: 1993 end-page: 2221 article-title: Compton scatter compensation using the triple‐energy window method for single‐ and dual‐isotope SPECT publication-title: J Nucl Med – volume: 6 start-page: 42 year: 2013 end-page: 53 article-title: Quantitative performance of advanced resolution recovery strategies on SPECT images: evaluation with use of digital phantom models publication-title: Radio Phys Technol – volume: 43 start-page: 1723 year: 2016 end-page: 1738 article-title: The EANM practice guidelines for bone scintigraphy publication-title: Eur J Nucl Med Mol Imaging – volume: 48 start-page: 950 year: 2001 end-page: 959 article-title: Recent progress in CdTe and CdZnTe detectors publication-title: IEEE Trans Nucl Sci – volume: 550 start-page: 267 year: 2005 end-page: 277 article-title: A photon counting CdTe gamma‐ and X‐ray camera publication-title: Nucl Inst Methods Phys Res – volume: 26 start-page: 191 year: 1998 end-page: 195 article-title: Measurement of blood radioactivity for quantification of cerebral blood flow using a gamma camera publication-title: J Nucl Med Technol – volume: 68 start-page: 299 year: 2012 end-page: 306 article-title: Development of the software package of the nuclear medicine data processor for education and research publication-title: Nihon Hoshasen Gijutsu Gakkai Zasshi – volume: 29 start-page: 636 year: 2015 end-page: 642 article-title: Optimization of iterative reconstruction parameters with 3‐dimensional resolution recovery, scatter and attenuation correction in I‐FP‐CIT SPECT publication-title: Ann Nucl Med – volume: 40 start-page: 1130 year: 2019 end-page: 1137 article-title: Shortened acquisition time in simultaneous 99mTc‐tetrofosmin and 123I‐β‐methyl‐p‐iodophenyl pentadecanoic acid dual‐tracer imaging with cadmium‐zinc‐telluride detectors in patients undergoing primary coronary intervention for acute myocardial infarction publication-title: Nucl Med Commun – year: 2019 article-title: Feasibility of simultaneous 99mTc‐tetrofosmin and 123I‐BMIPP dual‐tracer imaging with cadmium‐zinc‐telluride detectors in patients undergoing primary coronary intervention for acute myocardial infarction publication-title: J Nucl Cardiol – ident: e_1_2_9_19_1 doi: 10.1186/2191-219X-3-45 – ident: e_1_2_9_26_1 doi: 10.1109/23.958705 – volume: 32 start-page: 1157 year: 1991 ident: e_1_2_9_18_1 article-title: In vivo SPECT quantitation of bone metabolism in hyperparathyroidism and thyrotoxicosis publication-title: J Nucl Med – ident: e_1_2_9_31_1 doi: 10.1007/s12194-012-0168-z – ident: e_1_2_9_11_1 doi: 10.1097/MNM.0000000000001084 – ident: e_1_2_9_9_1 doi: 10.1007/s13139-017-0474-9 – volume-title: Nema Standards Publication NU‐1: Performance Measurements of Gamma Cameras year: 2007 ident: e_1_2_9_14_1 – ident: e_1_2_9_3_1 doi: 10.1016/j.nima.2005.04.081 – volume: 37 start-page: 1959 year: 2010 ident: e_1_2_9_29_1 article-title: A review on the clinical uses of SPECT/CT publication-title: EJNMMI – ident: e_1_2_9_21_1 doi: 10.1007/s12149-015-0987-1 – volume: 46 start-page: 398 year: 2018 ident: e_1_2_9_16_1 article-title: SNMMI procedure standard for bone scintigraphy 4.0 publication-title: J Nucl Med Technol – ident: e_1_2_9_13_1 doi: 10.6009/jjrt.2012_JSRT_68.3.299 – volume: 34 start-page: 2216 year: 1993 ident: e_1_2_9_23_1 article-title: Compton scatter compensation using the triple‐energy window method for single‐ and dual‐isotope SPECT publication-title: J Nucl Med – volume: 44 start-page: 602 year: 2003 ident: e_1_2_9_24_1 article-title: Evaluation of a small cadmium zinc telluride detector for scintimammography publication-title: J Nucl Med – ident: e_1_2_9_6_1 doi: 10.2967/jnumed.110.081323 – ident: e_1_2_9_17_1 doi: 10.1097/MNM.0000000000001028 – ident: e_1_2_9_22_1 doi: 10.1007/s12350-014-0008-0 – ident: e_1_2_9_27_1 doi: 10.1118/1.4934830 – volume: 26 start-page: 191 year: 1998 ident: e_1_2_9_20_1 article-title: Measurement of blood radioactivity for quantification of cerebral blood flow using a gamma camera publication-title: J Nucl Med Technol – volume: 35 start-page: 34 year: 2007 ident: e_1_2_9_30_1 article-title: Simultaneous 3‐dimensional resolution correction in SPECT reconstruction with an ordered‐subsets expectation maximization algorithm publication-title: J Nucl Med Technol – ident: e_1_2_9_7_1 doi: 10.1007/s00259-013-2617-2 – ident: e_1_2_9_8_1 doi: 10.1053/j.semnuclmed.2016.01.002 – ident: e_1_2_9_4_1 doi: 10.1007/s12149-009-0293-x – ident: e_1_2_9_10_1 doi: 10.1007/s12350-018-01585-9 – ident: e_1_2_9_5_1 doi: 10.2967/jnumed.108.060020 – ident: e_1_2_9_25_1 doi: 10.1088/0031-9155/60/8/3045 – ident: e_1_2_9_2_1 doi: 10.1109/TNS.2004.829437 – ident: e_1_2_9_12_1 doi: 10.1007/s12149-019-01382-5 – ident: e_1_2_9_15_1 doi: 10.1007/s00259-016-3415-4 – ident: e_1_2_9_28_1 doi: 10.1007/s12149-016-1144-1 |
| SSID | ssj0011838 |
| Score | 2.3974695 |
| Snippet | Purpose
A high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used... A high-energy-resolution whole-body SPECT-CT device (NM/CT 870 CZT; C-SPECT) equipped with a CZT detector has been developed and is being used clinically. A... PurposeA high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used... |
| SourceID | unpaywall pubmedcentral proquest pubmed crossref wiley |
| SourceType | Open Access Repository Aggregation Database Index Database Enrichment Source Publisher |
| StartPage | 165 |
| SubjectTerms | Clinical medicine Discovery NM/CT 870 CZT ⇒ NM/CT 870 CZT Energy Medical Imaging medium‐energy high‐resolution sensitivity collimator Physics Sensors Tomography wide‐energy high‐resolution collimator |
| SummonAdditionalLinks | – databaseName: Unpaywall dbid: UNPAY link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV1Lj9MwEB4tXQm48H4EFmRgLyClWzt26h6rKEuFlGrFtmLhEtmOIyratEAjBL-ecV5sWVRxc-Jx4tgz8jeZ8WeAY4PLuM5o6JuRkD63WeCPMiN8yVUYWEMHumLbT6bhZM7fXYiLA3jZ7oXZjd-zE2VWrE8DKuQ1OAwF4u0eHM6nZ-OPFREqC30EOLzjHb3cYHeluQIfr2ZB3iiLjfr5Qy2Xu0i1WmpOb0PUdrLOMPnSL7e6b379xd-4_yvuwK0GaZJxrRp34cAW9-B60sTS78MqvkTuT_6wfpN1ThAVkrcxmSYn0YygrZHo04y0myjJ-VmMt2sOaGK_lovNxmbE_dElH-LJe6KKjCRYOCdOyxYr59c_gPlpPIsmfnP4gm_4kEtfUxoIxFKOylhKLozEC6ktNVrmSmjJAqEpgotBxqkKGSINyrIQp2ggmJF58BB6xbqwj4EgBqPa8jzX6PuNFHo0wlId6lwhVgm08eB1O1GpaZjJ3QEZy7TmVGapG7-0Gj8PXnWym5qP459SR-18p41Nfk8ZlyMEnIhoPHjRVaM1uRCJKuy6dDIuMMkRRnnwqFaP7jVBgGimaj3cUZxOwDF179YUi88VY_cQ_Rj05Dw47lRsb-_fVNq3RyQdRwmrSk_-75lP4SZz2ThVvvkR9LbfSvsM4dRWP2_s6Td3IBcy priority: 102 providerName: Unpaywall – databaseName: Wiley Online Library Open Access dbid: 24P link: http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlZ3fa9RAEMeHWkF9EX8brTJiXxTi3W52cxvwpRyph5BS7BWLL2F3s8GDu9xpe4j_vbObXM6jUvAtyU5I2JnJfvbXNwCHlppxU7E0tplUsXBVEmeVlbESOk2cZUMT1PaLk3RyLj5fyIs9-LjZC9PqQ_QDbj4zwvfaJ7g2l4OtaKi2C_6BJUyqW3CbEcj4-ObitJ9DoGANG-EkT2OiINGLk_LB9t7d5ugaY15fKnl33az07196Pt_F2dAeHT-A-x1I4lHr-Yew55pHcKfopsofwyL_S7sft6LeuKyRoA8_5XhSDMZTpFTC8bcpbvZI4tlpTpdbiWd0P9az1cpV6Ads8Ws--YK6qbCggzP0QTRb-G77Ezg_zqfjSdz9WyG2YiRUbKjGJKGSVypWSkir6EQZx6xRtZZG8UQaRuwwrATTKSeQYLxKU5kNJbeqTp7CfrNs3HNAQixmnKhrQ127TFOHRTpmUlNrQpHE2Ajebaq4tJ3wuP__xbxsJZN56d1RBndE8La3XbVyG_-0Oth4quxS7rLkQmXEkwQsEbzpiylZ_AyIbtxy7W38vKMgSorgWevY_jFJQrAS7h7tuLw38ELcuyXN7HsQ5B5RN4XCLYLDPjhufPv3IW5uMCmPxgUPRy_-x_gl3ON-yU1YVH4A-1c_1-4VMdOVeR1S4w-_SAyG priority: 102 providerName: Wiley-Blackwell |
| Title | Experimental evaluation of the GE NM/CT 870 CZT clinical SPECT system equipped with WEHR and MEHRS collimator |
| URI | https://onlinelibrary.wiley.com/doi/abs/10.1002%2Facm2.13158 https://www.ncbi.nlm.nih.gov/pubmed/33426812 https://www.proquest.com/docview/2489083812 https://www.proquest.com/docview/2476844750 https://pubmed.ncbi.nlm.nih.gov/PMC7882118 https://doi.org/10.1002/acm2.13158 |
| UnpaywallVersion | publishedVersion |
| Volume | 22 |
| hasFullText | 1 |
| inHoldings | 1 |
| isFullTextHit | |
| isPrint | |
| journalDatabaseRights | – providerCode: PRVAON databaseName: en accès libre customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: DOA dateStart: 20000101 isFulltext: true titleUrlDefault: https://www.doaj.org/ providerName: Directory of Open Access Journals – providerCode: PRVBFR databaseName: Free Medical Journals customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: DIK dateStart: 20000101 isFulltext: true titleUrlDefault: http://www.freemedicaljournals.com providerName: Flying Publisher – providerCode: PRVFQY databaseName: GFMER Free Medical Journals customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: GX1 dateStart: 20000101 isFulltext: true titleUrlDefault: http://www.gfmer.ch/Medical_journals/Free_medical.php providerName: Geneva Foundation for Medical Education and Research – providerCode: PRVHPJ databaseName: ROAD: Directory of Open Access Scholarly Resources customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: M~E dateStart: 20000101 isFulltext: true titleUrlDefault: https://road.issn.org providerName: ISSN International Centre – providerCode: PRVAQN databaseName: PubMed Central customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: RPM dateStart: 20000101 isFulltext: true titleUrlDefault: https://www.ncbi.nlm.nih.gov/pmc/ providerName: National Library of Medicine – providerCode: PRVPQU databaseName: Health & Medical Collection customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: 7X7 dateStart: 20000301 isFulltext: true titleUrlDefault: https://search.proquest.com/healthcomplete providerName: ProQuest – providerCode: PRVPQU databaseName: ProQuest Central customDbUrl: http://www.proquest.com/pqcentral?accountid=15518 eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: BENPR dateStart: 20000301 isFulltext: true titleUrlDefault: https://www.proquest.com/central providerName: ProQuest – providerCode: PRVWIB databaseName: KBPluse Wiley Online Library: Open Access customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: AVUZU dateStart: 20000101 isFulltext: true titleUrlDefault: https://www.kbplus.ac.uk/kbplus7/publicExport/pkg/559 providerName: Wiley-Blackwell – providerCode: PRVWIB databaseName: Wiley Online Library (Open Access Collection) customDbUrl: eissn: 1526-9914 dateEnd: 99991231 omitProxy: true ssIdentifier: ssj0011838 issn: 1526-9914 databaseCode: 24P dateStart: 20000101 isFulltext: true titleUrlDefault: https://authorservices.wiley.com/open-science/open-access/browse-journals.html providerName: Wiley-Blackwell |
| link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwfV3da9swED_alH28jK37aLYuaKwvG3iJZNlRHsbIgtswSAhtwrK-GEmWaSBx3LVh7L_fSf5oQ0dehG2d8cedpJ90p98BnGgcxlVCQ0_3AuFxk_heL9GBJ7gMfaNpRzm2_dE4HM74j3kw34NxtRfGhlVWfaLrqJO1tmvkbcZFD-ECjkff8mvPZo2y3tUqhYYsUyskXx3F2D4cMMuM1YCD79F4cl77FdCARU1SytpSr9gX6lOb8P3-sPQAaz4MmXyyyXL5949cLrdhrRuXTp_DsxJQkn5hAS9gz2SH8HhUuswP4ZGL8dQ3L2EV3WPzJ3c032SdEoSB5Cwi41F7MCXYuMjgckqqXZPkYhLh5YL0mZjrzSLPTULsEi75GQ3PicwSMsKDC2LNarGyE_lXMDuNpoOhV2Zb8DTvcuEpSv0AwZPlLhaCB1rgiVCGaiVSGSjB_EBRRBOdhFMZMoQWlCVhGPQ6AdMi9V9DI1tn5ggIgi6qDE9ThZO9nsQpTGCoClUqEZz4SjfhU_WzY11SkduMGMu4IFFmsVVM7BTThI-1bF4QcPxX6rjSWVw2wpv4zmSa8KGuxuZjfSIyM-uNlbGeSI64qQlvChXXj_F9hC_u7u6W8msBS829XZMtrhxFdxcnLmhsTTipzWTn2392FrRDJO4PRswdvd39qe_gKbNhNy6w_Bgat7835j3iplvVgn3GJ1h2591W2TBabg0Cy7M5xWuz8aT_6x9kERkH |
| linkProvider | ProQuest |
| linkToHtml | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV3fb9MwED6NTTBeEIxfhQFGjAeQQmvHSZ2HCY2S0bGlmrZOTLwE23FEpTbt6Kpp_xx_G2c3yVYN9W1vaXNt2txd7jvf-TuALY1hXGU09HQUCI-bzPeiTAee4DL0jaYt5dj2k17YPeHfT4PTFfhb7YWxbZXVM9E9qLOxtmvkTcZFhHAB49HnyZlnp0bZ6mo1QkOWoxWybUcxVm7s2DeXF5jCTbf3vqK-3zO2G_c7Xa-cMuBp3ubCU5T6AYIGy9krBA-0wBdCGaqVyGWgBPMDRTGKtjJOZcgwpFKWhWEQtQKmRe7j996BNe7zCJO_tS9x7_CormOgw4iaFJU1pR6xT9SndsD89TB4A9vebNFcnxUTeXkhh8NFGO3i4O5DeFACWLIzt7hHsGKKDbiXlCX6Dbjrekr19DGM4mvTA8gVrTgZ5wRhJ_kWk17S7PQJOjPp_OyTapcmOT6M8e05yTQxZ7PBZGIyYpeMyY-4e0RkkZEED46JNePByC4cPIGTW7nvT2G1GBfmORAEeVQZnucKk8tIYsoUGKpClUsEQ77SDfhQ3exUl9TndgLHMJ2TNrPUKiZ1imnAu1p2Mif8-K_UZqWztHT6aXplog14W59Gd7U1GFmY8czK2MonR5zWgGdzFdeX8X2ES-7T7QXl1wKWCnzxTDH47SjB25goobE1YKs2k6W__qOzoCUi6U4nYe7oxfK_-gbWu_3kID3Y6-2_hPvMtvy4pvZNWD3_MzOvELOdq9elYxD4ddu--A-VbkyJ |
| linkToPdf | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lb9NAEB6VIgoXBOVlKLCIcgDJJLvrx_qAUJUmpJREFU1FxMV412sRKXFc0qjqX-PXMbt-tFFRbr35MX7OjOeb3fE3ALsKw7hMaeCqyBeup1PuRqnyXeElAdeKtqVl2x8Mg_6J93Xsjzfgb_0vjCmrrL-J9kOdzpUZI28xT0QIFzAetbKqLOJov_e5OHVNBykz01q30yhN5FBfnGP6tvh0sI-6fsdYrzvq9N2qw4CrvNATrqSU-wgYDF-vEJ6vBK4IqamSIkt8KRj3JcUI2k49mgQMwyllaRD4UdtnSmQcz3sLboecR6acMBw3yR7idi4aOlTWStSMfaScmtbyVwPgNVR7vTjz7jIvkovzZDpdBdA2AvYewP0KupK90tYewobOt2FrUE3Ob8MdW02qFo9g1r3SN4BcEoqTeUYQcJIvXTIctDojgm5MOj9HpP4_kxwfdXFzSS9N9OlyUhQ6JWawmPzo9r-TJE_JABeOiTHgycwMGTyGkxt5609gM5_n-hkQhHdUai_LJKaVUYLJkq-pDGSWIAziUjnwvn7ZsapIz03vjWlc0jWz2Cgmtopx4G0jW5RUH_-V2ql1FlfuvogvjdOBN81udFQz-5Lker40MmbO00OE5sDTUsXNZThHoGSPDleU3wgYEvDVPfnktyUDDzFFQmNzYLcxk7V3_8Fa0BqReK8zYHbp-fpHfQ1b6IHxt4Ph4Qu4x0ytj61m34HNsz9L_RLB2pl8Zb2CwK-bdsN_WrNKIw |
| linkToUnpaywall | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwjV1Lj9MwEB4tXQm48H4EFmRgLyClWzt26h6rKEuFlGrFtmLhEtmOIyratEAjBL-ecV5sWVRxc-Jx4tgz8jeZ8WeAY4PLuM5o6JuRkD63WeCPMiN8yVUYWEMHumLbT6bhZM7fXYiLA3jZ7oXZjd-zE2VWrE8DKuQ1OAwF4u0eHM6nZ-OPFREqC30EOLzjHb3cYHeluQIfr2ZB3iiLjfr5Qy2Xu0i1WmpOb0PUdrLOMPnSL7e6b379xd-4_yvuwK0GaZJxrRp34cAW9-B60sTS78MqvkTuT_6wfpN1ThAVkrcxmSYn0YygrZHo04y0myjJ-VmMt2sOaGK_lovNxmbE_dElH-LJe6KKjCRYOCdOyxYr59c_gPlpPIsmfnP4gm_4kEtfUxoIxFKOylhKLozEC6ktNVrmSmjJAqEpgotBxqkKGSINyrIQp2ggmJF58BB6xbqwj4EgBqPa8jzX6PuNFHo0wlId6lwhVgm08eB1O1GpaZjJ3QEZy7TmVGapG7-0Gj8PXnWym5qP459SR-18p41Nfk8ZlyMEnIhoPHjRVaM1uRCJKuy6dDIuMMkRRnnwqFaP7jVBgGimaj3cUZxOwDF179YUi88VY_cQ_Rj05Dw47lRsb-_fVNq3RyQdRwmrSk_-75lP4SZz2ThVvvkR9LbfSvsM4dRWP2_s6Td3IBcy |
| openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Experimental+evaluation+of+the+GE+NM%2FCT+870+CZT+clinical+SPECT+system+equipped+with+WEHR+and+MEHRS+collimator&rft.jtitle=Journal+of+applied+clinical+medical+physics&rft.au=Ito%2C+Toshimune&rft.au=Matsusaka%2C+Yohji&rft.au=Onoguchi%2C+Masahisa&rft.au=Ichikawa%2C+Hajime&rft.date=2021-02-01&rft.pub=John+Wiley+%26+Sons%2C+Inc&rft.eissn=1526-9914&rft.volume=22&rft.issue=2&rft.spage=165&rft.epage=177&rft_id=info:doi/10.1002%2Facm2.13158&rft.externalDBID=HAS_PDF_LINK |
| thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1526-9914&client=summon |
| thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1526-9914&client=summon |
| thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1526-9914&client=summon |