Purpose Tracking of the internal anatomy by means of a motion model that uses the MR‐derived motion fields and noise covariance matrix (NCM) dynamic as a surrogate signal. Methods A 2D respiratory motion model was developed based on the MR‐derived motion fields and the NCM of a receive array used in...
        Saved in:
      
    
          | Published in | Magnetic resonance in medicine Vol. 79; no. 3; pp. 1730 - 1735 | 
|---|---|
| Main Authors | , , , , , | 
| Format | Journal Article | 
| Language | English | 
| Published | 
        United States
          Wiley Subscription Services, Inc
    
        01.03.2018
     Wiley  | 
| Subjects | |
| Online Access | Get full text | 
| ISSN | 0740-3194 1522-2594 1522-2594  | 
| DOI | 10.1002/mrm.26775 | 
Cover
| Summary: | Purpose
Tracking of the internal anatomy by means of a motion model that uses the MR‐derived motion fields and noise covariance matrix (NCM) dynamic as a surrogate signal.
Methods
A 2D respiratory motion model was developed based on the MR‐derived motion fields and the NCM of a receive array used in MRI. Temporal dynamics of the NCM were used as a motion surrogate for a linear correspondence motion model. The model performance was tested on five healthy volunteers with a liver as the target. The motion fields were calculated from the cineMR frames with an optical flow registration tool.
Results
The model estimated the liver motion with an average residual error of 2.3 mm (13% of the motion amplitude). The model formation takes 3 min and the model latency was 0.5 s in the current implementation. The limiting factor for the latency is the current update time of the NCM (0.48 s), which in principle can be reduced to 0.004 s with an alternative way to determine the NCM.
Conclusions
The 2D respiratory motion of the liver can be effectively estimated with the linear motion model that uses the temporal behavior of the NCM as motion surrogate. Magn Reson Med 79:1730–1735, 2018. © 2017 International Society for Magnetic Resonance in Medicine. | 
|---|---|