Accuracy of a CGM Sensor in Pediatric Subjects With Type 1 Diabetes. Comparison of Three Insertion Sites: Arm, Abdomen, and Gluteus

Background: Patients with diabetes, especially pediatric ones, sometimes use continuous glucose monitoring (CGM) sensor in different positions from the approved ones. Here we compare the accuracy of Dexcom® G5 CGM sensor in three different sites: abdomen, gluteus (both approved) and arm (off-label)....

Full description

Saved in:
Bibliographic Details
Published inJournal of diabetes science and technology Vol. 11; no. 6; pp. 1147 - 1154
Main Authors Faccioli, Simone, Del Favero, Simone, Visentin, Roberto, Bonfanti, Riccardo, Iafusco, Dario, Rabbone, Ivana, Marigliano, Marco, Schiaffini, Riccardo, Bruttomesso, Daniela, Cobelli, Claudio
Format Journal Article
LanguageEnglish
Published Los Angeles, CA SAGE Publications 01.11.2017
Subjects
Online AccessGet full text
ISSN1932-2968
1932-3107
1932-3107
DOI10.1177/1932296817706377

Cover

More Information
Summary:Background: Patients with diabetes, especially pediatric ones, sometimes use continuous glucose monitoring (CGM) sensor in different positions from the approved ones. Here we compare the accuracy of Dexcom® G5 CGM sensor in three different sites: abdomen, gluteus (both approved) and arm (off-label). Method: Thirty youths, 5-9 years old, with type 1 diabetes (T1D) wore the sensor during a clinical trial where frequent self-monitoring of blood glucose (SMBG) measurements were obtained. Sensor was inserted in different sites according to the patient habit. Accuracy metrics include absolute relative difference (ARD) and absolute difference (AD) of CGM with respect to SMBG. The three sites were compared with ANOVA. If the test detected a difference, an additional pair-wise comparison was performed. Results: Overall, no accuracy difference was detected: the mean ARD was 13.3% (SD = 13.5%) for abdomen, 13.4% (12.9%) for arm and 12.9% (20.2%) for gluteus (P value = .83); the mean AD was 17.0 mg/dl (17.2 mg/dl) for abdomen, 17.2 mg/dl (17.1 mg/dl) for arm and 18.3 mg/dl (18.5 mg/dl) for gluteus (P value = .30). In hypo- and euglycemia ARD (P value = .87 and .15, respectively), and AD (P value = .68 and .37, respectively) were not statistically different. At variance, in hyperglycemia, a significant difference was detected between the two approved sites, abdomen and gluteus (ΔARD = −2.2% [CI = −4.2%, −0.1%], P value = .04), whereas the comparisons with the off-label location, arm-abdomen, and arm-gluteus were not significant. Conclusions: These results suggest that the accuracy of the sensor placed on the arm was not significantly different with respect to the two approved insertion sites (abdomen and gluteus). Larger, randomized trials are needed to draw final conclusions.
Bibliography:ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Feature-1
content type line 23
ISSN:1932-2968
1932-3107
1932-3107
DOI:10.1177/1932296817706377