CT attenuation measurements are valuable to discriminate pledgets used in prosthetic heart valve implantation from paravalvular leakage

Sutures with polytetrafluorethylene (PTFE) felt pledgets are commonly used in prosthetic heart valve (PHV) implantation. Paravalvular leakage can be difficult to distinguish from PTFE felt pledgets on multislice CT because both present as hyperdense structures. We assessed whether pledgets can be di...

Full description

Saved in:
Bibliographic Details
Published inBritish journal of radiology Vol. 85; no. 1017; pp. e616 - e621
Main Authors Habets, J, Meijer, T S, Meijer, R C A, Mali, W P Th M, Vonken, E-J P A, Budde, R P J
Format Journal Article
LanguageEnglish
Published England The British Institute of Radiology 01.09.2012
Subjects
Online AccessGet full text
ISSN0007-1285
1748-880X
1748-880X
DOI10.1259/bjr/29602784

Cover

More Information
Summary:Sutures with polytetrafluorethylene (PTFE) felt pledgets are commonly used in prosthetic heart valve (PHV) implantation. Paravalvular leakage can be difficult to distinguish from PTFE felt pledgets on multislice CT because both present as hyperdense structures. We assessed whether pledgets can be discriminated from contrast-enhanced solutions (blood/saline) on CT images based on attenuation difference in an ex vivo experiment and under in vivo conditions. PTFE felt pledgets were sutured to the suture ring of a mechanical PHV and porcine aortic annulus, and immersed and scanned in four different contrast-enhanced (Ultravist®; 300 mg jopromide ml(-1)) saline concentrations (10.0, 12.0, 13.6 and 15.0 mg ml(-1)). Scanning was performed on a 256-slice scanner with eight different scan protocols with various tube voltage (100 kV, 120 kV) and tube current (400 mAs, 600 mAs, 800 mAs, 1000 mAs) settings. Attenuation of the pledgets and surrounding contrast-enhanced saline were measured. Additionally, the attenuation of pledgets and contrast-enhanced blood was measured on electrocardiography (ECG)-gated CTA scans of 19 patients with 22 PHVs. Ex vivo CT attenuation differences between the pledgets and contrast-enhanced solutions were larger by using higher tube voltages. CT attenuation values of the pledgets were higher than contrast-enhanced blood in patients: 420±26 Hounsfield units (mean±SD, range 383-494) and 288±41 Hounsfield units (range 202-367), respectively. PTFE felt pledgets have consistently higher attenuation than surrounding contrast-enhanced blood. CT attenuation measurements therefore may help to differentiate pledgets from paravalvular leakage, and detect paravalvular leakage in patients with suspected PHV dysfunction.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0007-1285
1748-880X
1748-880X
DOI:10.1259/bjr/29602784