Toward real-time digital pulse process algorithms for CsI(Tl) detector array at external target facility in HIRFL-CSR
A fully digital data acquisition system based on a field-programmable gate array (FPGA) was developed for a CsI(Tl) array at the external target facility (ETF) in the Heavy Ion Research Facility in Lanzhou (HIRFL). To process the CsI(Tl) signals generated by γ -rays and light-charged ions, a scheme...
        Saved in:
      
    
          | Published in | Nuclear science and techniques Vol. 34; no. 9 | 
|---|---|
| Main Authors | , , , , , , , , , , , , , , , | 
| Format | Journal Article | 
| Language | English | 
| Published | 
        Singapore
          Springer Nature Singapore
    
        01.09.2023
     | 
| Subjects | |
| Online Access | Get full text | 
| ISSN | 1001-8042 2210-3147  | 
| DOI | 10.1007/s41365-023-01272-6 | 
Cover
| Summary: | A fully digital data acquisition system based on a field-programmable gate array (FPGA) was developed for a CsI(Tl) array at the external target facility (ETF) in the Heavy Ion Research Facility in Lanzhou (HIRFL). To process the CsI(Tl) signals generated by
γ
-rays and light-charged ions, a scheme for digital pulse processing algorithms is proposed. Every step in the algorithms was benchmarked using standard
γ
and
α
sources. The scheme, which included a moving average filter, baseline restoration, leading-edge discrimination, moving window deconvolution, and digital charge comparison, was subsequently implemented on the FPGA. A good energy resolution of 5.7% for 1.33-MeV
γ
-rays and excellent
α
-
γ
identification using the digital charge comparison method were achieved, which satisfies CsI(Tl) array performance requirements. | 
|---|---|
| ISSN: | 1001-8042 2210-3147  | 
| DOI: | 10.1007/s41365-023-01272-6 |