Radiation exposure in cone beam CT measured using a MOSFET and RPLGD dosimeter and Monte Carlo simulation in phantom

BACKGROUND: Due to the wide application of the cone beam computed tomography (CBCT) in clinical practice, it is important to assess radiation dose of CBCT more accurately and efficiently in different clinical applications. OBJECTIVE: This study aims to calculate effective and absorbed doses in CBCT...

Full description

Saved in:
Bibliographic Details
Published inJournal of X-ray science and technology Vol. 31; no. 4; pp. 837 - 852
Main Author Kweon, Dae Cheol
Format Journal Article
LanguageEnglish
Published London, England SAGE Publications 01.01.2023
Subjects
Online AccessGet full text
ISSN0895-3996
1095-9114
1095-9114
DOI10.3233/XST-230026

Cover

Abstract BACKGROUND: Due to the wide application of the cone beam computed tomography (CBCT) in clinical practice, it is important to assess radiation dose of CBCT more accurately and efficiently in different clinical applications. OBJECTIVE: This study aims to calculate effective and absorbed doses in CBCT measured in an anthropomorphic phantom using computer-based Monte Carlo (PCXMC) software, and to conduct comparative evaluations of MOSFET (metal– oxide– semiconductor field-effect transistor) and radiophotoluminescence glass dosimeters (RPLGD). METHODS: Effective and absorbed organ doses are compared with those obtained using MOSFET and RPLGD dosimetry in an anthropomorphic phantom given the same exposure settings. Effective and absorbed organ doses from CBCT during scout and main projections are calculated using PCXMC and PCXMCRotation software, respectively. RESULTS: The mean effective dose from CBCT calculated using PCXMC software is 233.8μSv, while the doses calculated using dosimetry (MOSFET and RPLGD) are 266.67μSv and 268.78μSv, respectively. The X-ray source variation is 0.79%. The prescription limits based on the Friedman test for MOSFET and RPLGD pre-points (i.e., in an analytical analysis of diagnostic names in CBCT) are not statistically significant. The calculated correlation coefficient between MOSFET- and RPLGD-derived absorbed dose values with respect to a field of view CBCT parameter of 17×13.5 mm is r = 0.8623. CONCLUSIONS: The study demonstrates that the PCXMC software may be used as an alternative to MOSFET and RPLGD dosimetry for effective and absorbed organ dose estimation in CBCT conducted with a large FOV in an anthropomorphic phantom.
AbstractList BACKGROUND: Due to the wide application of the cone beam computed tomography (CBCT) in clinical practice, it is important to assess radiation dose of CBCT more accurately and efficiently in different clinical applications. OBJECTIVE: This study aims to calculate effective and absorbed doses in CBCT measured in an anthropomorphic phantom using computer-based Monte Carlo (PCXMC) software, and to conduct comparative evaluations of MOSFET (metal– oxide– semiconductor field-effect transistor) and radiophotoluminescence glass dosimeters (RPLGD). METHODS: Effective and absorbed organ doses are compared with those obtained using MOSFET and RPLGD dosimetry in an anthropomorphic phantom given the same exposure settings. Effective and absorbed organ doses from CBCT during scout and main projections are calculated using PCXMC and PCXMCRotation software, respectively. RESULTS: The mean effective dose from CBCT calculated using PCXMC software is 233.8μSv, while the doses calculated using dosimetry (MOSFET and RPLGD) are 266.67μSv and 268.78μSv, respectively. The X-ray source variation is 0.79%. The prescription limits based on the Friedman test for MOSFET and RPLGD pre-points (i.e., in an analytical analysis of diagnostic names in CBCT) are not statistically significant. The calculated correlation coefficient between MOSFET- and RPLGD-derived absorbed dose values with respect to a field of view CBCT parameter of 17×13.5 mm is r = 0.8623. CONCLUSIONS: The study demonstrates that the PCXMC software may be used as an alternative to MOSFET and RPLGD dosimetry for effective and absorbed organ dose estimation in CBCT conducted with a large FOV in an anthropomorphic phantom.
BACKGROUND: Due to the wide application of the cone beam computed tomography (CBCT) in clinical practice, it is important to assess radiation dose of CBCT more accurately and efficiently in different clinical applications. OBJECTIVE: This study aims to calculate effective and absorbed doses in CBCT measured in an anthropomorphic phantom using computer-based Monte Carlo (PCXMC) software, and to conduct comparative evaluations of MOSFET (metal– oxide– semiconductor field-effect transistor) and radiophotoluminescence glass dosimeters (RPLGD). METHODS: Effective and absorbed organ doses are compared with those obtained using MOSFET and RPLGD dosimetry in an anthropomorphic phantom given the same exposure settings. Effective and absorbed organ doses from CBCT during scout and main projections are calculated using PCXMC and PCXMCRotation software, respectively. RESULTS: The mean effective dose from CBCT calculated using PCXMC software is 233.8μSv, while the doses calculated using dosimetry (MOSFET and RPLGD) are 266.67μSv and 268.78μSv, respectively. The X-ray source variation is 0.79%. The prescription limits based on the Friedman test for MOSFET and RPLGD pre-points (i.e., in an analytical analysis of diagnostic names in CBCT) are not statistically significant. The calculated correlation coefficient between MOSFET- and RPLGD-derived absorbed dose values with respect to a field of view CBCT parameter of 17×13.5 mm is r = 0.8623. CONCLUSIONS: The study demonstrates that the PCXMC software may be used as an alternative to MOSFET and RPLGD dosimetry for effective and absorbed organ dose estimation in CBCT conducted with a large FOV in an anthropomorphic phantom.
Due to the wide application of the cone beam computed tomography (CBCT) in clinical practice, it is important to assess radiation dose of CBCT more accurately and efficiently in different clinical applications.BACKGROUNDDue to the wide application of the cone beam computed tomography (CBCT) in clinical practice, it is important to assess radiation dose of CBCT more accurately and efficiently in different clinical applications.This study aims to calculate effective and absorbed doses in CBCT measured in an anthropomorphic phantom using computer-based Monte Carlo (PCXMC) software, and to conduct comparative evaluations of MOSFET (metal- oxide- semiconductor field-effect transistor) and radiophotoluminescence glass dosimeters (RPLGD).OBJECTIVEThis study aims to calculate effective and absorbed doses in CBCT measured in an anthropomorphic phantom using computer-based Monte Carlo (PCXMC) software, and to conduct comparative evaluations of MOSFET (metal- oxide- semiconductor field-effect transistor) and radiophotoluminescence glass dosimeters (RPLGD).Effective and absorbed organ doses are compared with those obtained using MOSFET and RPLGD dosimetry in an anthropomorphic phantom given the same exposure settings. Effective and absorbed organ doses from CBCT during scout and main projections are calculated using PCXMC and PCXMCRotation software, respectively.METHODSEffective and absorbed organ doses are compared with those obtained using MOSFET and RPLGD dosimetry in an anthropomorphic phantom given the same exposure settings. Effective and absorbed organ doses from CBCT during scout and main projections are calculated using PCXMC and PCXMCRotation software, respectively.The mean effective dose from CBCT calculated using PCXMC software is 233.8μSv, while the doses calculated using dosimetry (MOSFET and RPLGD) are 266.67μSv and 268.78μSv, respectively. The X-ray source variation is 0.79%. The prescription limits based on the Friedman test for MOSFET and RPLGD pre-points (i.e., in an analytical analysis of diagnostic names in CBCT) are not statistically significant. The calculated correlation coefficient between MOSFET- and RPLGD-derived absorbed dose values with respect to a field of view CBCT parameter of 17×13.5 mm is r = 0.8623.RESULTSThe mean effective dose from CBCT calculated using PCXMC software is 233.8μSv, while the doses calculated using dosimetry (MOSFET and RPLGD) are 266.67μSv and 268.78μSv, respectively. The X-ray source variation is 0.79%. The prescription limits based on the Friedman test for MOSFET and RPLGD pre-points (i.e., in an analytical analysis of diagnostic names in CBCT) are not statistically significant. The calculated correlation coefficient between MOSFET- and RPLGD-derived absorbed dose values with respect to a field of view CBCT parameter of 17×13.5 mm is r = 0.8623.The study demonstrates that the PCXMC software may be used as an alternative to MOSFET and RPLGD dosimetry for effective and absorbed organ dose estimation in CBCT conducted with a large FOV in an anthropomorphic phantom.CONCLUSIONSThe study demonstrates that the PCXMC software may be used as an alternative to MOSFET and RPLGD dosimetry for effective and absorbed organ dose estimation in CBCT conducted with a large FOV in an anthropomorphic phantom.
Due to the wide application of the cone beam computed tomography (CBCT) in clinical practice, it is important to assess radiation dose of CBCT more accurately and efficiently in different clinical applications. This study aims to calculate effective and absorbed doses in CBCT measured in an anthropomorphic phantom using computer-based Monte Carlo (PCXMC) software, and to conduct comparative evaluations of MOSFET (metal- oxide- semiconductor field-effect transistor) and radiophotoluminescence glass dosimeters (RPLGD). Effective and absorbed organ doses are compared with those obtained using MOSFET and RPLGD dosimetry in an anthropomorphic phantom given the same exposure settings. Effective and absorbed organ doses from CBCT during scout and main projections are calculated using PCXMC and PCXMCRotation software, respectively. The mean effective dose from CBCT calculated using PCXMC software is 233.8μSv, while the doses calculated using dosimetry (MOSFET and RPLGD) are 266.67μSv and 268.78μSv, respectively. The X-ray source variation is 0.79%. The prescription limits based on the Friedman test for MOSFET and RPLGD pre-points (i.e., in an analytical analysis of diagnostic names in CBCT) are not statistically significant. The calculated correlation coefficient between MOSFET- and RPLGD-derived absorbed dose values with respect to a field of view CBCT parameter of 17×13.5 mm is r = 0.8623. The study demonstrates that the PCXMC software may be used as an alternative to MOSFET and RPLGD dosimetry for effective and absorbed organ dose estimation in CBCT conducted with a large FOV in an anthropomorphic phantom.
Author Kweon, Dae Cheol
Author_xml – sequence: 1
  givenname: Dae Cheol
  surname: Kweon
  fullname: Kweon, Dae Cheol
  email: dckweon@shinhan.ac.kr
  organization: Department of Radiological Science
BackLink https://www.ncbi.nlm.nih.gov/pubmed/37248944$$D View this record in MEDLINE/PubMed
BookMark eNptkU9LxDAQxYMouv65-AEkN0WoJk23TY5SdRV2WdEVvJVpO10rbVKTFvTbG-3qydMMb368Yebtk21tNBJyzNmFCIW4fHlaBaFgLIy3yIQzNQ0U59E2mTDpe6FUvEf2nXtjjPOplLtkTyRhJFUUTUj_CGUNfW00xY_OuMEirTUt_AaaI7Q0XdEW4Vsv6eBqvaZAF8un25sVBV3Sx4f57JqWxtUt9mh_tIXRPdIUbGOo14dm9Pe23Svo3rSHZKeCxuHRph6QZ--X3gXz5ew-vZoHRZjIPuCCA8-hjJRkcR5zKTGCWJVTriTKqGDJtKrCKk9iUflBBRIVFzkXAEkJeS4OyNno21nzPqDrs7Z2BTYNaDSDy0IZMhVLliiPnmzQIW-xzDpbt2A_s99PeeB8BAprnLNY_SGcZd8xZD6GbIzBw6cj7GCN2ZsZrPZ3_kd-AQuBhV0
Cites_doi 10.1093/rpd/ncr369
10.5516/NET.08.2012.080
10.1259/dmfr/15779208
10.1259/dmfr.20120260
10.1118/1.597314
10.1118/1.2047827
10.1016/S0360-3016(96)00600-1
10.3938/jkps.64.923
10.1007/s00247-009-1161-5
10.1080/10420150.2020.1768391
10.1259/dmfr.20120417
10.1002/acm2.12799
10.5624/isd.2019.49.2.139
10.3938/jkps.73.1377
10.1016/j.oooo.2019.11.012
10.1259/dmfr.20140202
10.1016/j.oooo.2012.06.003
10.1088/0952-4746/35/1/153
10.1118/1.2804555
10.2214/AJR.08.2168
10.5624/isd.2018.48.1.21
10.1259/bjr/31419627
10.1016/j.oooo.2015.10.030
10.2214/AJR.06.0742
10.1118/1.1481520
10.1118/1.3602456
10.1016/j.ejrad.2010.11.028
10.1007/s13246-022-01103-9
10.1259/dmfr.20140287
10.22974/jkda.2020.58.7.001
ContentType Journal Article
Copyright 2023 – IOS Press. All rights reserved
Copyright_xml – notice: 2023 – IOS Press. All rights reserved
DBID AAYXX
CITATION
CGR
CUY
CVF
ECM
EIF
NPM
7X8
DOI 10.3233/XST-230026
DatabaseName CrossRef
Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
MEDLINE - Academic
DatabaseTitle CrossRef
MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
MEDLINE - Academic
DatabaseTitleList CrossRef

MEDLINE - Academic
MEDLINE
Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: https://proxy.k.utb.cz/login?url=http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: EIF
  name: MEDLINE
  url: https://proxy.k.utb.cz/login?url=https://www.webofscience.com/wos/medline/basic-search
  sourceTypes: Index Database
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
Engineering
Physics
EISSN 1095-9114
EndPage 852
ExternalDocumentID 37248944
10_3233_XST_230026
10.3233_XST-230026
Genre Journal Article
GroupedDBID ---
0R~
1~5
4.4
4G.
5GY
7-5
AAGLT
AAQXI
ABDBF
ABJNI
ABJZC
ABUJY
ACGFS
ACPQW
ACUHS
ADZMO
AENEX
AFRHK
AHDMH
AJNRN
AKRWK
ALMA_UNASSIGNED_HOLDINGS
CS3
DU5
EAD
EAP
EBD
EBS
EMK
EMOBN
EPL
EST
ESX
F5P
FDB
HZ~
I-F
IOS
J8X
MET
MIO
MV1
NGNOM
O-L
O9-
P2P
SAUOL
SCNPE
SFC
SV3
TUS
AAYXX
AJGYC
CITATION
--K
1B1
1RT
29L
53G
5VS
AACTN
AAEDT
AAFNC
AALRI
AAQFI
AAQXK
AAXUO
ABEFU
ABMAC
ABUBZ
ABWVN
ACRPL
ADBBV
ADMUD
ADNMO
AFFNX
AGIAB
AHHHB
AITUG
CAG
CGR
COF
CUY
CVF
DM4
ECM
EIF
EJD
FEDTE
FGOYB
FIRID
G-2
HEI
HMK
HMO
HVGLF
IHE
IL9
LG5
M29
NPM
Q1R
R2-
RIG
ROL
RPZ
SAE
SSZ
UHS
WUQ
XPP
7X8
AAPII
ARTOV
H13
ID FETCH-LOGICAL-c278t-131a1bad49806b6188e4a69d5198e84c075ff2fb763f4a6fa8e913b13aa7dabb3
ISSN 0895-3996
1095-9114
IngestDate Sun Sep 28 10:54:13 EDT 2025
Thu Apr 03 06:57:32 EDT 2025
Tue Jul 01 05:25:42 EDT 2025
Tue Jun 17 22:30:26 EDT 2025
IsPeerReviewed true
IsScholarly true
Issue 4
Keywords cone beam CT
MOSFET
RPLGD
Absorbed dose
radiation dose
Language English
LinkModel OpenURL
MergedId FETCHMERGED-LOGICAL-c278t-131a1bad49806b6188e4a69d5198e84c075ff2fb763f4a6fa8e913b13aa7dabb3
Notes ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
PMID 37248944
PQID 2820968079
PQPubID 23479
PageCount 16
ParticipantIDs proquest_miscellaneous_2820968079
pubmed_primary_37248944
crossref_primary_10_3233_XST_230026
sage_journals_10_3233_XST_230026
ProviderPackageCode CITATION
AAYXX
PublicationCentury 2000
PublicationDate 2023-01-01
PublicationDateYYYYMMDD 2023-01-01
PublicationDate_xml – month: 01
  year: 2023
  text: 2023-01-01
  day: 01
PublicationDecade 2020
PublicationPlace London, England
PublicationPlace_xml – name: London, England
– name: Netherlands
PublicationTitle Journal of X-ray science and technology
PublicationTitleAlternate J Xray Sci Technol
PublicationYear 2023
Publisher SAGE Publications
Publisher_xml – name: SAGE Publications
References 2015; 35
2012; 81
2009; 82
2007; 188
2013; 42
2022; 45
2016; 32
2020; 15
2014; 46
2020; 58
2016; 121
2011; 38
2007; 34
2018; 49
2018; 48
1994; 21
2014; 43
2014; 64
2012; 150
2002; 28
2021; 15
2020; 130
2020; 175
1997; 37
2015; 44
2005; 32
2010; 194
2018; 73
2020; 21
2012; 114
2009; 38
2009; 39
Wood (10.3233/XST-230026_ref28) 2015; 35
Yoshizumi (10.3233/XST-230026_ref16) 2007; 188
Koivisto (10.3233/XST-230026_ref31) 2012; 114
Suomalainen (10.3233/XST-230026_ref22) 2009; 38
Attaia (10.3233/XST-230026_ref23) 2020; 130
Kim (10.3233/XST-230026_ref12) 2010; 194
Al-Senan (10.3233/XST-230026_ref5) 2011; 38
Lee (10.3233/XST-230026_ref32) 2020; 175
Lee (10.3233/XST-230026_ref26) 2020; 15
Goren (10.3233/XST-230026_ref24) 2013; 42
Fetin (10.3233/XST-230026_ref27) 2022; 45
Giglioli (10.3233/XST-230026_ref29) 2016; 32
Yadav (10.3233/XST-230026_ref9) 2020; 21
Kim (10.3233/XST-230026_ref18) 2018; 48
Koivisto (10.3233/XST-230026_ref20) 2014; 43
Ramani (10.3233/XST-230026_ref7) 1997; 37
Park (10.3233/XST-230026_ref33) 2018; 49
Lee (10.3233/XST-230026_ref21) 2021; 15
Roberts (10.3233/XST-230026_ref3) 2009; 82
Jursinic (10.3233/XST-230026_ref4) 2007; 34
Lee (10.3233/XST-230026_ref1) 2018; 73
Rottke (10.3233/XST-230026_ref2) 2013; 42
Chuang (10.3233/XST-230026_ref8) 2002; 28
Jones (10.3233/XST-230026_ref17) 2005; 32
Moon (10.3233/XST-230026_ref15) 2014; 46
Soubra (10.3233/XST-230026_ref13) 1994; 21
Chambers (10.3233/XST-230026_ref25) 2015; 44
Kim (10.3233/XST-230026_ref11) 2012; 150
Pauwels (10.3233/XST-230026_ref14) 2012; 81
Kim (10.3233/XST-230026_ref6) 2014; 64
Kim (10.3233/XST-230026_ref19) 2020; 58
Lee (10.3233/XST-230026_ref10) 2009; 39
Lee (10.3233/XST-230026_ref30) 2016; 121
References_xml – volume: 43
  start-page: 20140202
  year: 2014
  article-title: Effective dose assessment in the maxillofacial region using thermoluminescent (TLD) and metal oxide semiconductor field-effect transistor (MOSFET) dosemeters: a comparative study
  publication-title: Dentomaxillofac Radiol
– volume: 81
  start-page: 267
  year: 2012
  end-page: 271
  article-title: Effective dose range for dental cone beam computed tomography scanners
  publication-title: Eur J Radiol
– volume: 15
  start-page: 131
  year: 2021
  end-page: 139
  article-title: Analysis of national medical radiation exposure by dental radiography
  publication-title: J Radiat Indust
– volume: 49
  start-page: 139
  year: 2018
  end-page: 151
  article-title: Optimization of exposure parameters and relationship between subjective and technical image quality in cone-beam computed tomography
  publication-title: Imaging Sci Dent
– volume: 82
  start-page: 35
  year: 2009
  end-page: 40
  article-title: Effective dose from cone beam CT examinations in dentistry
  publication-title: Br J Radiol
– volume: 130
  start-page: 101
  year: 2020
  end-page: 109
  article-title: Dose reduction in head and neck organs through shielding and application of different scanning parameters in cone beam computed tomography: an effective dose study using an adult male anthropomorphic phantom
  publication-title: Oral Surg Oral Med Oral Pathol Oral Radiol
– volume: 32
  start-page: 3209
  year: 2005
  end-page: 3213
  article-title: MOSFET dosimeter depth-dose measurements in heterogeneous tissue-equivalent phantoms at diagnostic x-ray energies
  publication-title: Med Phys
– volume: 114
  start-page: 393
  year: 2012
  end-page: 400
  article-title: Assessment of radiation exposure in dental cone-beam computerized tomography with the use of metal-oxide semiconductor field-effect transistor (MOSFET) dosimeters and Monte Carlo simulations
  publication-title: Oral Surg Oral Med Oral Pathol Oral Radiol
– volume: 32
  start-page: 29
  year: 2016
  article-title: IGRT cone beam CT: A method to evaluate patient dose
  publication-title: European J Med Phys
– volume: 64
  start-page: 923
  year: 2014
  end-page: 928
  article-title: Adaptive statistical iterative reconstruction and bismuth shielding for evaluation of dose reduction to the eye and image quality during head CT
  publication-title: J Korean Phys Soc
– volume: 121
  start-page: 322
  year: 2016
  end-page: 329
  article-title: Comparison of dosimetry methods for panoramic radiography: thermoluminescent dosimeter measurement versus personal computer-based Monte Carlo method calculation
  publication-title: Oral Surg Oral Med Oral Pathol Oral Radiol
– volume: 150
  start-page: 50
  year: 2012
  end-page: 54
  article-title: Radiation dose from 3D rotational X-ray imaging: organ and effective dose with conversion factors
  publication-title: Radiat Prot Dosimetry
– volume: 44
  start-page: 20140287
  year: 2015
  article-title: The effective dose of different scanning protocols using the Sirona GALILEOS comfort CBCT scanner
  publication-title: Dentomaxillofacl Radiol
– volume: 15
  start-page: e02191063
  year: 2020
  article-title: Dose assessment in dental cone-beam computed tomography: Comparison of optically stimulated luminescence dosimetry with Monte Carlo method
  publication-title: PLos ONE
– volume: 21
  start-page: 567
  year: 1994
  end-page: 572
  article-title: Evaluation of a dual bias dual metal oxide-silicon semiconductor field effect transistor detector as radiation dosimeter
  publication-title: Med Phys
– volume: 42
  start-page: 20120417
  year: 2013
  article-title: Effective dose span of ten different cone beam CT devices
  publication-title: Dentomaxillofac Radiol
– volume: 188
  start-page: 1332
  year: 2007
  end-page: 1336
  article-title: Validation of metal oxide semiconductor field effect transistor technology for organ dose assessment during CT: Comparison with thermoluminescent dosimetry
  publication-title: Am J Roentgenol
– volume: 38
  start-page: 367
  year: 2009
  end-page: 378
  article-title: Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners
  publication-title: Dentomaxillofacial Radiol
– volume: 48
  start-page: 21
  year: 2018
  end-page: 30
  article-title: Estimation of the effective dose of dental cone-beam computed tomography using personal computer-based Monte Carlo software
  publication-title: Imaging Sci Dent
– volume: 58
  start-page: 388
  year: 2020
  end-page: 397
  article-title: Dose estimation of cone-beam computed tomography in children using personal computer-based Monte Carlo software
  publication-title: J Kor Dent Assoc
– volume: 34
  start-page: 4594
  year: 2007
  end-page: 4604
  article-title: Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements
  publication-title: Med Phys
– volume: 37
  start-page: 959
  year: 1997
  end-page: 964
  article-title: Clinical dosimetry using MOSFETs
  publication-title: Int J Radiat Oncol
– volume: 39
  start-page: 608
  year: 2009
  end-page: 615
  article-title: Effective dose estimation for pediatric voiding cystourethrography using an anthropomorphic phantom set and metal oxide semiconductor field-effect transistor (MOSFET) technology
  publication-title: Pediatr Radiol
– volume: 42
  start-page: 20120260
  year: 2013
  article-title: Effect of leaded glasses and thyroid shielding on cone beam CT radiation dose in an adult female phantom
  publication-title: Dentomaxillofacial Radiol
– volume: 46
  start-page: 255
  year: 2014
  end-page: 262
  article-title: Effective dose measurement for cone beam computed tomography using glass dosimeter
  publication-title: Nucl Eng Technol
– volume: 28
  start-page: 1109
  year: 2002
  end-page: 1115
  article-title: Investigation of the use of MOSFET for clinical IMRT dosimetric verification
  publication-title: Med Phys
– volume: 175
  start-page: 809
  year: 2020
  end-page: 826
  article-title: Assessment of dose reduction and image quality by barium composite shielding in head and chest CT
  publication-title: Radiat Eff Defects Solids
– volume: 45
  start-page: 205
  year: 2022
  end-page: 218
  article-title: PCXMC cone beam computed tomography dosimetry investigations
  publication-title: Phys Eng Sci
– volume: 35
  start-page: 153
  year: 2015
  end-page: 163
  article-title: Validation of a technique for estimating organ doses for kilovoltage cone-beam CT of the prostate using the PCXMC 2.0 patient dose calculator
  publication-title: J Radiol Prot
– volume: 38
  start-page: 4396
  year: 2011
  end-page: 4405
  article-title: Characteristics of an OSLD in the diagnostic energy range
  publication-title: Med Phys
– volume: 73
  start-page: 1377
  year: 2018
  end-page: 1384
  article-title: Comparative study of the effective dose from panoramic radiography in dentistry measured using a radiophotoluminescent glass dosimeter and an optically stimulated luminescence detector
  publication-title: J Korean Phys Soc
– volume: 21
  start-page: 127
  year: 2020
  end-page: 135
  article-title: MOSFET dosimeter characterization in MR-guided radiation therapy (MRgRT) Linac
  publication-title: J Appl Clin Med Phys
– volume: 194
  start-page: 186
  year: 2010
  end-page: 190
  article-title: Radiation dose from cone beam CT in a pediatric phantom: risk estimation of cancer incidence
  publication-title: Am J Roentgenol
– volume: 32
  start-page: 29
  year: 2016
  ident: 10.3233/XST-230026_ref29
  article-title: IGRT cone beam CT: A method to evaluate patient dose
  publication-title: European J Med Phys
– volume: 150
  start-page: 50
  year: 2012
  ident: 10.3233/XST-230026_ref11
  article-title: Radiation dose from 3D rotational X-ray imaging: organ and effective dose with conversion factors
  publication-title: Radiat Prot Dosimetry
  doi: 10.1093/rpd/ncr369
– volume: 46
  start-page: 255
  year: 2014
  ident: 10.3233/XST-230026_ref15
  article-title: Effective dose measurement for cone beam computed tomography using glass dosimeter
  publication-title: Nucl Eng Technol
  doi: 10.5516/NET.08.2012.080
– volume: 38
  start-page: 367
  year: 2009
  ident: 10.3233/XST-230026_ref22
  article-title: Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners
  publication-title: Dentomaxillofacial Radiol
  doi: 10.1259/dmfr/15779208
– volume: 42
  start-page: 20120260
  year: 2013
  ident: 10.3233/XST-230026_ref24
  article-title: Effect of leaded glasses and thyroid shielding on cone beam CT radiation dose in an adult female phantom
  publication-title: Dentomaxillofacial Radiol
  doi: 10.1259/dmfr.20120260
– volume: 21
  start-page: 567
  year: 1994
  ident: 10.3233/XST-230026_ref13
  article-title: Evaluation of a dual bias dual metal oxide-silicon semiconductor field effect transistor detector as radiation dosimeter
  publication-title: Med Phys
  doi: 10.1118/1.597314
– volume: 32
  start-page: 3209
  year: 2005
  ident: 10.3233/XST-230026_ref17
  article-title: MOSFET dosimeter depth-dose measurements in heterogeneous tissue-equivalent phantoms at diagnostic x-ray energies
  publication-title: Med Phys
  doi: 10.1118/1.2047827
– volume: 37
  start-page: 959
  year: 1997
  ident: 10.3233/XST-230026_ref7
  article-title: Clinical dosimetry using MOSFETs
  publication-title: Int J Radiat Oncol
  doi: 10.1016/S0360-3016(96)00600-1
– volume: 64
  start-page: 923
  year: 2014
  ident: 10.3233/XST-230026_ref6
  article-title: Adaptive statistical iterative reconstruction and bismuth shielding for evaluation of dose reduction to the eye and image quality during head CT
  publication-title: J Korean Phys Soc
  doi: 10.3938/jkps.64.923
– volume: 39
  start-page: 608
  year: 2009
  ident: 10.3233/XST-230026_ref10
  article-title: Effective dose estimation for pediatric voiding cystourethrography using an anthropomorphic phantom set and metal oxide semiconductor field-effect transistor (MOSFET) technology
  publication-title: Pediatr Radiol
  doi: 10.1007/s00247-009-1161-5
– volume: 175
  start-page: 809
  year: 2020
  ident: 10.3233/XST-230026_ref32
  article-title: Assessment of dose reduction and image quality by barium composite shielding in head and chest CT
  publication-title: Radiat Eff Defects Solids
  doi: 10.1080/10420150.2020.1768391
– volume: 42
  start-page: 20120417
  year: 2013
  ident: 10.3233/XST-230026_ref2
  article-title: Effective dose span of ten different cone beam CT devices
  publication-title: Dentomaxillofac Radiol
  doi: 10.1259/dmfr.20120417
– volume: 21
  start-page: 127
  year: 2020
  ident: 10.3233/XST-230026_ref9
  article-title: MOSFET dosimeter characterization in MR-guided radiation therapy (MRgRT) Linac
  publication-title: J Appl Clin Med Phys
  doi: 10.1002/acm2.12799
– volume: 49
  start-page: 139
  year: 2018
  ident: 10.3233/XST-230026_ref33
  article-title: Optimization of exposure parameters and relationship between subjective and technical image quality in cone-beam computed tomography
  publication-title: Imaging Sci Dent
  doi: 10.5624/isd.2019.49.2.139
– volume: 73
  start-page: 1377
  year: 2018
  ident: 10.3233/XST-230026_ref1
  article-title: Comparative study of the effective dose from panoramic radiography in dentistry measured using a radiophotoluminescent glass dosimeter and an optically stimulated luminescence detector
  publication-title: J Korean Phys Soc
  doi: 10.3938/jkps.73.1377
– volume: 15
  start-page: 131
  year: 2021
  ident: 10.3233/XST-230026_ref21
  article-title: Analysis of national medical radiation exposure by dental radiography
  publication-title: J Radiat Indust
– volume: 130
  start-page: 101
  year: 2020
  ident: 10.3233/XST-230026_ref23
  article-title: Dose reduction in head and neck organs through shielding and application of different scanning parameters in cone beam computed tomography: an effective dose study using an adult male anthropomorphic phantom
  publication-title: Oral Surg Oral Med Oral Pathol Oral Radiol
  doi: 10.1016/j.oooo.2019.11.012
– volume: 43
  start-page: 20140202
  year: 2014
  ident: 10.3233/XST-230026_ref20
  article-title: Effective dose assessment in the maxillofacial region using thermoluminescent (TLD) and metal oxide semiconductor field-effect transistor (MOSFET) dosemeters: a comparative study
  publication-title: Dentomaxillofac Radiol
  doi: 10.1259/dmfr.20140202
– volume: 114
  start-page: 393
  year: 2012
  ident: 10.3233/XST-230026_ref31
  article-title: Assessment of radiation exposure in dental cone-beam computerized tomography with the use of metal-oxide semiconductor field-effect transistor (MOSFET) dosimeters and Monte Carlo simulations
  publication-title: Oral Surg Oral Med Oral Pathol Oral Radiol
  doi: 10.1016/j.oooo.2012.06.003
– volume: 35
  start-page: 153
  year: 2015
  ident: 10.3233/XST-230026_ref28
  article-title: Validation of a technique for estimating organ doses for kilovoltage cone-beam CT of the prostate using the PCXMC 2.0 patient dose calculator
  publication-title: J Radiol Prot
  doi: 10.1088/0952-4746/35/1/153
– volume: 34
  start-page: 4594
  year: 2007
  ident: 10.3233/XST-230026_ref4
  article-title: Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements
  publication-title: Med Phys
  doi: 10.1118/1.2804555
– volume: 194
  start-page: 186
  year: 2010
  ident: 10.3233/XST-230026_ref12
  article-title: Radiation dose from cone beam CT in a pediatric phantom: risk estimation of cancer incidence
  publication-title: Am J Roentgenol
  doi: 10.2214/AJR.08.2168
– volume: 48
  start-page: 21
  year: 2018
  ident: 10.3233/XST-230026_ref18
  article-title: Estimation of the effective dose of dental cone-beam computed tomography using personal computer-based Monte Carlo software
  publication-title: Imaging Sci Dent
  doi: 10.5624/isd.2018.48.1.21
– volume: 82
  start-page: 35
  year: 2009
  ident: 10.3233/XST-230026_ref3
  article-title: Effective dose from cone beam CT examinations in dentistry
  publication-title: Br J Radiol
  doi: 10.1259/bjr/31419627
– volume: 15
  start-page: e02191063
  year: 2020
  ident: 10.3233/XST-230026_ref26
  article-title: Dose assessment in dental cone-beam computed tomography: Comparison of optically stimulated luminescence dosimetry with Monte Carlo method
  publication-title: PLos ONE
– volume: 121
  start-page: 322
  year: 2016
  ident: 10.3233/XST-230026_ref30
  article-title: Comparison of dosimetry methods for panoramic radiography: thermoluminescent dosimeter measurement versus personal computer-based Monte Carlo method calculation
  publication-title: Oral Surg Oral Med Oral Pathol Oral Radiol
  doi: 10.1016/j.oooo.2015.10.030
– volume: 188
  start-page: 1332
  year: 2007
  ident: 10.3233/XST-230026_ref16
  article-title: Validation of metal oxide semiconductor field effect transistor technology for organ dose assessment during CT: Comparison with thermoluminescent dosimetry
  publication-title: Am J Roentgenol
  doi: 10.2214/AJR.06.0742
– volume: 28
  start-page: 1109
  year: 2002
  ident: 10.3233/XST-230026_ref8
  article-title: Investigation of the use of MOSFET for clinical IMRT dosimetric verification
  publication-title: Med Phys
  doi: 10.1118/1.1481520
– volume: 38
  start-page: 4396
  year: 2011
  ident: 10.3233/XST-230026_ref5
  article-title: Characteristics of an OSLD in the diagnostic energy range
  publication-title: Med Phys
  doi: 10.1118/1.3602456
– volume: 81
  start-page: 267
  year: 2012
  ident: 10.3233/XST-230026_ref14
  article-title: Effective dose range for dental cone beam computed tomography scanners
  publication-title: Eur J Radiol
  doi: 10.1016/j.ejrad.2010.11.028
– volume: 45
  start-page: 205
  year: 2022
  ident: 10.3233/XST-230026_ref27
  article-title: PCXMC cone beam computed tomography dosimetry investigations
  publication-title: Phys Eng Sci
  doi: 10.1007/s13246-022-01103-9
– volume: 44
  start-page: 20140287
  year: 2015
  ident: 10.3233/XST-230026_ref25
  article-title: The effective dose of different scanning protocols using the Sirona GALILEOS® comfort CBCT scanner
  publication-title: Dentomaxillofacl Radiol
  doi: 10.1259/dmfr.20140287
– volume: 58
  start-page: 388
  year: 2020
  ident: 10.3233/XST-230026_ref19
  article-title: Dose estimation of cone-beam computed tomography in children using personal computer-based Monte Carlo software
  publication-title: J Kor Dent Assoc
  doi: 10.22974/jkda.2020.58.7.001
SSID ssj0011588
Score 2.27273
Snippet BACKGROUND: Due to the wide application of the cone beam computed tomography (CBCT) in clinical practice, it is important to assess radiation dose of CBCT more...
BACKGROUND: Due to the wide application of the cone beam computed tomography (CBCT) in clinical practice, it is important to assess radiation dose of CBCT more...
Due to the wide application of the cone beam computed tomography (CBCT) in clinical practice, it is important to assess radiation dose of CBCT more accurately...
SourceID proquest
pubmed
crossref
sage
SourceType Aggregation Database
Index Database
Publisher
StartPage 837
SubjectTerms Cone-Beam Computed Tomography - methods
Monte Carlo Method
Phantoms, Imaging
Radiation Dosage
Radiation Dosimeters
Radiation Exposure
Radiometry - methods
Title Radiation exposure in cone beam CT measured using a MOSFET and RPLGD dosimeter and Monte Carlo simulation in phantom
URI https://journals.sagepub.com/doi/full/10.3233/XST-230026
https://www.ncbi.nlm.nih.gov/pubmed/37248944
https://www.proquest.com/docview/2820968079
Volume 31
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
journalDatabaseRights – providerCode: PRVEBS
  databaseName: EBSCOhost Academic Search Ultimate
  customDbUrl: https://search.ebscohost.com/login.aspx?authtype=ip,shib&custid=s3936755&profile=ehost&defaultdb=asn
  eissn: 1095-9114
  dateEnd: 20250330
  omitProxy: true
  ssIdentifier: ssj0011588
  issn: 0895-3996
  databaseCode: ABDBF
  dateStart: 19970901
  isFulltext: true
  titleUrlDefault: https://search.ebscohost.com/direct.asp?db=asn
  providerName: EBSCOhost
– providerCode: PRVLSH
  databaseName: Elsevier Journals
  customDbUrl:
  mediaType: online
  eissn: 1095-9114
  dateEnd: 99991231
  omitProxy: true
  ssIdentifier: ssj0011588
  issn: 0895-3996
  databaseCode: AKRWK
  dateStart: 19890701
  isFulltext: true
  providerName: Library Specific Holdings
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3db9MwELdKJ6btAUH5Kl8ygrcoECdO4rwgTR1johTQ1om-RbbjaJWWplpTDfjrOdtJmn1JwEsS2bId-ffL5Xy-OyP0VoYsyEI_dgXn0qWSUniikRuK3NPJW0ROdLzz5Gt0eEI_z8JZr_eh47W0rsQ7-fvGuJL_QRXKAFcdJfsPyLadQgE8A75wBYTh-lcYH-nEAgZA9XNZalufY9zKQXEUihfOaOoU1gaYOWtjFODO5NvxwcepjUv8_uXTvpOVq3mhnWKsz4XOVqXdQM5KB8rrw72Mr_mpPm-4uEWbnbnn_JfTBAkZt8xrRvvxhbJ7_PschjhV5VnX5uAHHZuDsnLSM0c82vjPRpDW4nzetRLYX6rN63JVWge-tiYfzI6nLiyEPBs534FtWRjcgtinLLFZIq_kxm6q7qAtP44iv4-29sZHP8btPhIJGbNJafVg7zdD7aDtpvFljeTaMuOSi5_ROqb30b16gvGexf4B6qnFAO12kkgO0Pakdo8YoLvGn1euHqKqpQZuqIHnC6ypgTU18GiKG2pgQw3MsaUGBvCwoQZuqWHKDDWwoQbeUEN3W1PjETqB9qNDtz5gw5V-zCqXBIQTwTOaMC8SEWFMUR4lGWj1TDEqQZ3Mcz8X8A_KoSLnTCUkECTgPM64EMFj1F_Aiz9FmIIgp2Gig74I3KUgSnIlQukRGaiIDdGbZpLTpc2jksL6U6OSAiqpRWWIXjfzn4KY03tXfKHK9Sr1QVNNIubFyRA9scC0_TRADhHWSKX1F7q6YYBntzZ-jnY2TH-B-tX5Wr0EfbMSr2pS_QFHKn6z
linkProvider Library Specific Holdings
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Radiation+exposure+in+cone+beam+CT+measured+using+a+MOSFET+and+RPLGD+dosimeter+and+Monte+Carlo+simulation+in+phantom&rft.jtitle=Journal+of+X-ray+science+and+technology&rft.au=Kweon%2C+Dae+Cheol&rft.date=2023-01-01&rft.eissn=1095-9114&rft.volume=31&rft.issue=4&rft.spage=837&rft_id=info:doi/10.3233%2FXST-230026&rft_id=info%3Apmid%2F37248944&rft.externalDocID=37248944
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0895-3996&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0895-3996&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0895-3996&client=summon