Experimental measurement of radiation dose in a dedicated breast CT system
Radiation dose is an important performance indicator of a dedicated breast CT(DBCT).In this paper,the method of putting thermoluminescent dosimeters(TLD) into a breast shaped PMMA phantom to study the dose distribution in breasts was improved by using smaller TLDs and a new half-ellipsoid PMMA phant...
Saved in:
Published in | Chinese physics C Vol. 38; no. 3; pp. 91 - 97 |
---|---|
Main Author | |
Format | Journal Article |
Language | English |
Published |
01.03.2014
|
Subjects | |
Online Access | Get full text |
ISSN | 1674-1137 0254-3052 |
DOI | 10.1088/1674-1137/38/3/038201 |
Cover
Abstract | Radiation dose is an important performance indicator of a dedicated breast CT(DBCT).In this paper,the method of putting thermoluminescent dosimeters(TLD) into a breast shaped PMMA phantom to study the dose distribution in breasts was improved by using smaller TLDs and a new half-ellipsoid PMMA phantom.Then the weighted CT dose index(CTDI_w) was introduced to average glandular assessment in DBCT for the first time and two measurement modes were proposed for different sizes of breasts.The dose deviations caused by using cylindrical phantoms were simulated using the Monte Carlo method and a set of correction factors were calculated.The results of the confirmatory measurement with a cylindrical phantom(11 cm/8 cm) show that CTDI_w gives a relatively conservative overestimate of the average glandular dose comparing to the results of Monte Carlo simulation and TLDs measurement.But with better practicability and stability,the CTDI_w is suitable for dose evaluations in daily clinical practice.Both of the TLDs and CTDI_w measurements demonstrate that the radiation dose of our DBCT system is lower than conventional two-view mammography. |
---|---|
AbstractList | Radiation dose is an important performance indicator of a dedicated breast CT(DBCT).In this paper,the method of putting thermoluminescent dosimeters(TLD) into a breast shaped PMMA phantom to study the dose distribution in breasts was improved by using smaller TLDs and a new half-ellipsoid PMMA phantom.Then the weighted CT dose index(CTDI_w) was introduced to average glandular assessment in DBCT for the first time and two measurement modes were proposed for different sizes of breasts.The dose deviations caused by using cylindrical phantoms were simulated using the Monte Carlo method and a set of correction factors were calculated.The results of the confirmatory measurement with a cylindrical phantom(11 cm/8 cm) show that CTDI_w gives a relatively conservative overestimate of the average glandular dose comparing to the results of Monte Carlo simulation and TLDs measurement.But with better practicability and stability,the CTDI_w is suitable for dose evaluations in daily clinical practice.Both of the TLDs and CTDI_w measurements demonstrate that the radiation dose of our DBCT system is lower than conventional two-view mammography. Radiation dose is an important performance indicator of a dedicated breast CT (DBCT). In this paper, the method of putting thermoluminescent dosimeters (TLD) into a breast shaped PMMA phantom to study the dose distribution in breasts was improved by using smaller TLDs and a new half-ellipsoid PMMA phantom. Then the weighted CT dose index (CTDI sub(w)) was introduced to average glandular assessment in DBCT for the first time and two measurement modes were proposed for different sizes of breasts. The dose deviations caused by using cylindrical phantoms were simulated using the Monte Carlo method and a set of correction factors were calculated. The results of the confirmatory measurement with a cylindrical phantom (11 cm/8 cm) show that CTDI sub(w) gives a relatively conservative overestimate of the average glandular dose comparing to the results of Monte Carlo simulation and TLDs measurement. But with better practicability and stability, the CTDI sub(w) is suitable for dose evaluations in daily clinical practice. Both of the TLDs and CTDIw measurements demonstrate that the radiation dose of our DBCT system is lower than conventional two-view mammography. |
Author | 申善威 王燕芳 舒航 唐晓 魏存峰 宋玉收 史戎坚 魏龙 |
AuthorAffiliation | College of Nuclear Science and Technology, Harbin Engineering University, Harbin 150000, China Key Laboratory of Nuclear Analytical Techniques, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China Beijing Engineering Research Center of Radiographic Techniques and Equipment, Beijing 100049, China |
Author_xml | – sequence: 1 fullname: 申善威 王燕芳 舒航 唐晓 魏存峰 宋玉收 史戎坚 魏龙 |
BookMark | eNo9kE1LAzEQhoNUsK3-BCHevKyb72aPUuoXBS_1HLKbSV3ZTdpkC_bfu6VS5jAM87wD88zQJMQACN1T8kSJ1iVVC1FQyhcl1yUvCdeM0Cs0JUyKghPJJmh6YW7QLOcfQpQYo1P0sfrdQWp7CIPtcA82HxKcJhw9Tta1dmhjwC5mwG3AFjtwbWMHcLhOIz3g5QbnYx6gv0XX3nYZ7v77HH29rDbLt2L9-fq-fF4XDVNkKKCRzNZS1N4qah3XpGZAlax87ZQVjddSiaqWVFPtnVeeV25cu4YRv9Dc8Tl6PN_dpbg_QB5M3-YGus4GiIdsqCZjKcHkiMoz2qSYcwJvduOvNh0NJebkzpy8mJMXw7Xh5uxuzD38575j2O7bsL0Eha6EZLTif1nDbxk |
Cites_doi | 10.1088/1674-1137/36/7/019 10.1118/1.594995 10.1093/oxfordjournals.rpd.a082564 10.1118/1.3521469 10.1109/TNS.2009.2036848 10.1259/bjr/80446730 10.1118/1.2128126 10.1259/bjr/18118287 10.1088/0031-9155/49/24/003 10.1118/1.1636571 |
ContentType | Journal Article |
DBID | 2RA 92L CQIGP ~WA AAYXX CITATION 7U5 8FD H8D L7M |
DOI | 10.1088/1674-1137/38/3/038201 |
DatabaseName | 维普_期刊 中文科技期刊数据库-CALIS站点 维普中文期刊数据库 中文科技期刊数据库- 镜像站点 CrossRef Solid State and Superconductivity Abstracts Technology Research Database Aerospace Database Advanced Technologies Database with Aerospace |
DatabaseTitle | CrossRef Aerospace Database Solid State and Superconductivity Abstracts Technology Research Database Advanced Technologies Database with Aerospace |
DatabaseTitleList | Aerospace Database |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Physics |
DocumentTitleAlternate | Experimental measurement of radiation dose in a dedicated breast CT system |
EISSN | 0254-3052 |
EndPage | 97 |
ExternalDocumentID | 10_1088_1674_1137_38_3_038201 48945219 |
GroupedDBID | 02O 1JI 1WK 29B 2B. 2C. 2RA 4.4 5B3 5GY 5VR 5VS 7.M 92E 92I 92L 92Q 93N AAGCD AAGID AAJIO AAJKP AALHV AATNI ABCXL ABHWH ABJNI ACAFW ACGFS ACHIP AENEX AFUIB AFYNE AIBLX AKPSB ALMA_UNASSIGNED_HOLDINGS ATQHT BBWZM CCEZO CCVFK CEBXE CHBEP CJUJL CQIGP CRLBU CW9 DU5 EBS EDWGO EJD EPQRW EQZZN ER. FA0 FEDTE HVGLF IJHAN IOP IZVLO JCGBZ KNG KOT M45 N5L NT- NT. OK1 PJBAE Q02 RIN RNS ROL RPA RW3 SY9 TCJ TGP W28 ~WA -SA -SC -S~ AAYXX ADEQX AERVB AOAED CAJEA CAJEC CITATION Q-- U1G U5K U5M 7U5 8FD AEINN H8D L7M |
ID | FETCH-LOGICAL-c260t-ec52ab54bfa61ad380b2e1659fbd6a4cf85649b51818fdf6f39d165dc20f783d3 |
ISSN | 1674-1137 |
IngestDate | Fri Sep 05 00:05:51 EDT 2025 Tue Jul 01 03:03:10 EDT 2025 Wed Feb 14 10:37:50 EST 2024 |
IsDoiOpenAccess | false |
IsOpenAccess | true |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 3 |
Language | English |
License | http://iopscience.iop.org/info/page/text-and-data-mining http://iopscience.iop.org/page/copyright |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-c260t-ec52ab54bfa61ad380b2e1659fbd6a4cf85649b51818fdf6f39d165dc20f783d3 |
Notes | Radiation dose is an important performance indicator of a dedicated breast CT(DBCT).In this paper,the method of putting thermoluminescent dosimeters(TLD) into a breast shaped PMMA phantom to study the dose distribution in breasts was improved by using smaller TLDs and a new half-ellipsoid PMMA phantom.Then the weighted CT dose index(CTDI_w) was introduced to average glandular assessment in DBCT for the first time and two measurement modes were proposed for different sizes of breasts.The dose deviations caused by using cylindrical phantoms were simulated using the Monte Carlo method and a set of correction factors were calculated.The results of the confirmatory measurement with a cylindrical phantom(11 cm/8 cm) show that CTDI_w gives a relatively conservative overestimate of the average glandular dose comparing to the results of Monte Carlo simulation and TLDs measurement.But with better practicability and stability,the CTDI_w is suitable for dose evaluations in daily clinical practice.Both of the TLDs and CTDI_w measurements demonstrate that the radiation dose of our DBCT system is lower than conventional two-view mammography. 11-5641/O4 dedicated breast CT radiation dose experimental measurement ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
PQID | 1808086425 |
PQPubID | 23500 |
PageCount | 7 |
ParticipantIDs | proquest_miscellaneous_1808086425 crossref_primary_10_1088_1674_1137_38_3_038201 chongqing_primary_48945219 |
ProviderPackageCode | CITATION AAYXX |
PublicationCentury | 2000 |
PublicationDate | 2014-03-01 |
PublicationDateYYYYMMDD | 2014-03-01 |
PublicationDate_xml | – month: 03 year: 2014 text: 2014-03-01 day: 01 |
PublicationDecade | 2010 |
PublicationTitle | Chinese physics C |
PublicationTitleAlternate | Chinese Physica C |
PublicationYear | 2014 |
References | 11 12 13 Leitz W (7) 1995; 57 2 3 4 5 6 8 9 Thacker S C (1) 2004; 49 Tang X (10) 2012; 36 |
References_xml | – ident: 5 – volume: 36 start-page: 675 year: 2012 ident: 10 publication-title: Chinese Physics C (HEP & NP) doi: 10.1088/1674-1137/36/7/019 – ident: 12 – ident: 6 doi: 10.1118/1.594995 – volume: 57 start-page: 377 issn: 0144-8420 year: 1995 ident: 7 publication-title: Radiation Protection Dosimetry doi: 10.1093/oxfordjournals.rpd.a082564 – ident: 11 – ident: 2 doi: 10.1118/1.3521469 – ident: 3 doi: 10.1109/TNS.2009.2036848 – ident: 9 doi: 10.1259/bjr/80446730 – ident: 13 doi: 10.1118/1.2128126 – ident: 8 doi: 10.1259/bjr/18118287 – volume: 49 start-page: 5433 issn: 0031-9155 year: 2004 ident: 1 publication-title: Physics in Medicine and Biology doi: 10.1088/0031-9155/49/24/003 – ident: 4 doi: 10.1118/1.1636571 |
SSID | ssj0064088 |
Score | 1.9288418 |
Snippet | Radiation dose is an important performance indicator of a dedicated breast CT(DBCT).In this paper,the method of putting thermoluminescent dosimeters(TLD) into... Radiation dose is an important performance indicator of a dedicated breast CT (DBCT). In this paper, the method of putting thermoluminescent dosimeters (TLD)... |
SourceID | proquest crossref chongqing |
SourceType | Aggregation Database Index Database Publisher |
StartPage | 91 |
SubjectTerms | Assessments Breast Computer simulation CT系统 Deviation Mathematical analysis Monte Carlo methods Performance indices PMMA Polymethyl methacrylates TLD 乳腺 剂量评估 实验测量 蒙特卡罗方法 辐射剂量 |
Title | Experimental measurement of radiation dose in a dedicated breast CT system |
URI | http://lib.cqvip.com/qk/92043A/201403/48945219.html https://www.proquest.com/docview/1808086425 |
Volume | 38 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV1Lb9QwELZKERIXxFNsechI-LTKbhI7zviYbLOqKvE4bKXeojgP6IEstLsXfjM_ghnnsalACLhEI3s8tjwT-3MynmHsraRUNIWtPdyLfU_Fte-BAuNpWVUNbhFh4_KQvXuvzy7U-WV0eXT0Y-K1tN_ZRfn9t_dK_kerWIZ6pVuy_6DZUSgWII36xSdqGJ9_peNsGp7_y-FrHyHAawo64JRbkUv6VTsv5pX7K0MY05Iv-m6-2vShnKcYVWRKpCCSU5FFwqQiRSIWYERiiDC-AL0S2VqkK2ESkYEAZApEpoXRAuKeG0YnWdcKZUonEFkUEQm2C-aOOSMJRCBf5AQmyD13FAgT9kSSzJ0ERWOg3owwxGVorLB2Qk-FASKws9R37MkoPaNhUTsci3Z12CgNqQilYzUJRwHJVKYRaSa67ErD15FAHdzD-gVdx8oLgi6wzKLuFlY8FHu4zoUT-5aTRbxLH9bDgc57-JeNBhdn-uYxSEfaxZmnKBi-JFR12F9Hr0d8BxSiJXOH3Q1jRHp03fTDxwE3aOW7LKmjzOG-GcByLFtKWMpl1wNFA_m8bT99Q4xzG1XdBhUOKW0esgf9EYcnnb0-Ykd1-5jdc67G5c0Tdj61Wj6xWr5t-Gi1nKyWX7W84KPV8s5q-WrDO6t9yi7W2WZ15vUZPbwSz807ry6jsLCRsk2hg6KS4NuwDnRkGlvpQpUNRFoZGyHshKZqdCNNhdVVGfpNDLKSz9hxu23r54wXgY0VZSuKEKAWEEOFJ_8yChqrLcLqYsZOxtnJv3aRW_JBAzO2GKZrrHPeGAA5zXVOc51LyGXezfWMvRkmNcf1l36qFW293d_kAQVmBTzFRyd_6vEFu38wz5fseHe9r18hnN3Z184KfgIYXHZi |
linkProvider | IOP Publishing |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Experimental+measurement+of+radiation+dose+in+a+dedicated+breast+CT+system&rft.jtitle=%E4%B8%AD%E5%9B%BD%E7%89%A9%E7%90%86C%EF%BC%9A%E8%8B%B1%E6%96%87%E7%89%88&rft.au=%E7%94%B3%E5%96%84%E5%A8%81+%E7%8E%8B%E7%87%95%E8%8A%B3+%E8%88%92%E8%88%AA+%E5%94%90%E6%99%93+%E9%AD%8F%E5%AD%98%E5%B3%B0+%E5%AE%8B%E7%8E%89%E6%94%B6+%E5%8F%B2%E6%88%8E%E5%9D%9A+%E9%AD%8F%E9%BE%99&rft.date=2014-03-01&rft.issn=1674-1137&rft.eissn=0254-3052&rft.issue=3&rft.spage=91&rft.epage=97&rft_id=info:doi/10.1088%2F1674-1137%2F38%2F3%2F038201&rft.externalDocID=48945219 |
thumbnail_s | http://utb.summon.serialssolutions.com/2.0.0/image/custom?url=http%3A%2F%2Fimage.cqvip.com%2Fvip1000%2Fqk%2F92043A%2F92043A.jpg |