Experimental measurement of radiation dose in a dedicated breast CT system

Radiation dose is an important performance indicator of a dedicated breast CT(DBCT).In this paper,the method of putting thermoluminescent dosimeters(TLD) into a breast shaped PMMA phantom to study the dose distribution in breasts was improved by using smaller TLDs and a new half-ellipsoid PMMA phant...

Full description

Saved in:
Bibliographic Details
Published inChinese physics C Vol. 38; no. 3; pp. 91 - 97
Main Author 申善威 王燕芳 舒航 唐晓 魏存峰 宋玉收 史戎坚 魏龙
Format Journal Article
LanguageEnglish
Published 01.03.2014
Subjects
Online AccessGet full text
ISSN1674-1137
0254-3052
DOI10.1088/1674-1137/38/3/038201

Cover

More Information
Summary:Radiation dose is an important performance indicator of a dedicated breast CT(DBCT).In this paper,the method of putting thermoluminescent dosimeters(TLD) into a breast shaped PMMA phantom to study the dose distribution in breasts was improved by using smaller TLDs and a new half-ellipsoid PMMA phantom.Then the weighted CT dose index(CTDI_w) was introduced to average glandular assessment in DBCT for the first time and two measurement modes were proposed for different sizes of breasts.The dose deviations caused by using cylindrical phantoms were simulated using the Monte Carlo method and a set of correction factors were calculated.The results of the confirmatory measurement with a cylindrical phantom(11 cm/8 cm) show that CTDI_w gives a relatively conservative overestimate of the average glandular dose comparing to the results of Monte Carlo simulation and TLDs measurement.But with better practicability and stability,the CTDI_w is suitable for dose evaluations in daily clinical practice.Both of the TLDs and CTDI_w measurements demonstrate that the radiation dose of our DBCT system is lower than conventional two-view mammography.
Bibliography:Radiation dose is an important performance indicator of a dedicated breast CT(DBCT).In this paper,the method of putting thermoluminescent dosimeters(TLD) into a breast shaped PMMA phantom to study the dose distribution in breasts was improved by using smaller TLDs and a new half-ellipsoid PMMA phantom.Then the weighted CT dose index(CTDI_w) was introduced to average glandular assessment in DBCT for the first time and two measurement modes were proposed for different sizes of breasts.The dose deviations caused by using cylindrical phantoms were simulated using the Monte Carlo method and a set of correction factors were calculated.The results of the confirmatory measurement with a cylindrical phantom(11 cm/8 cm) show that CTDI_w gives a relatively conservative overestimate of the average glandular dose comparing to the results of Monte Carlo simulation and TLDs measurement.But with better practicability and stability,the CTDI_w is suitable for dose evaluations in daily clinical practice.Both of the TLDs and CTDI_w measurements demonstrate that the radiation dose of our DBCT system is lower than conventional two-view mammography.
11-5641/O4
dedicated breast CT radiation dose experimental measurement
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1674-1137
0254-3052
DOI:10.1088/1674-1137/38/3/038201