RESPIPulse: Machine learning assisted sensory device for pulsed mode delivery of oxygen bolus using surface electromyography (sEMG) signals

Continuous mode delivery of medical oxygen from oxygen concentrators and oxygen cylinders leads to wastage of precious medical oxygen during exhalation and rest phases of the respiratory cycle. Pulse mode oxygen delivery can address the stated problem, however, it is required to determine the number...

Full description

Saved in:
Bibliographic Details
Published inSensors and actuators. A. Physical. Vol. 369; p. 115121
Main Authors Mondal, Aruna, Dutta, Debeshi, Chanda, Nripen, Mandal, Nilrudra, Mandal, Soumen
Format Journal Article
LanguageEnglish
Published Elsevier B.V 16.04.2024
Subjects
Online AccessGet full text
ISSN0924-4247
1873-3069
DOI10.1016/j.sna.2024.115121

Cover

More Information
Summary:Continuous mode delivery of medical oxygen from oxygen concentrators and oxygen cylinders leads to wastage of precious medical oxygen during exhalation and rest phases of the respiratory cycle. Pulse mode oxygen delivery can address the stated problem, however, it is required to determine the number of oxygen release pulses and the exact instant of inhalation or exhalation. Herein we report the design and development of an intelligent pulsed mode respiratory device- “RESPIPulse,” which is capable of delivering oxygen bolus by automatically sensing the inhalation and exhalation instances from body mount surface electromyography (sEMG) electrodes without manual intervention or settings. The device comprises a set of miniature single-channel sEMG electrodes, an embedded machine-learning algorithm, a normally open solenoid valve, an airflow sensor, and necessary driving electronics. The solenoid valve opens or closes depending on the muscular inhalation or exhalation effort determined from the sEMG signals, thus preventing the wastage of respiratory oxygen. The sEMG signals are subjected to envelop extraction followed by feature extraction. Performances of k-nearest neighbor (kNN), support vector regression (SVR), and random forests (RF) regressors are initially tested in Python IDE to identify the best learning algorithm that is deployed in the microcontroller for determination of the instances of inhalation and exhalation. Trials are conducted on 20 healthy subjects and 10 dyspnea-affected patients. Based on the computed performance measures and evaluation time, the kNN algorithm estimates the respiratory instances more accurately than the other two algorithms. A significant amount of oxygen savings, ranging between 35.48–82.35%, is obtained using the RESPIPulse device which is much higher than the pulse mode delivery devices employing manual settings exhibiting maximum conservation of 48.2%. [Display omitted] •Automated pulsed mode oxygen delivery device based on sEMG and airflow is developed.•The device can save up to 82% of medical oxygen.•The device acts as oxygen conservation system.
ISSN:0924-4247
1873-3069
DOI:10.1016/j.sna.2024.115121