Comparing Powered Bone Marrow Biopsy Procedures to Manual Bone Marrow Biopsy Procedures Using Healthy Volunteers

Abstract 1523▪ A battery-powered bone marrow biopsy system was cleared by the U.S. Food and Drug Administration in 2007 for bone marrow sampling. A study was designed to determine if the powered device has advantages over traditional manual devices in terms of decreased pain, insertion time, and imp...

Full description

Saved in:
Bibliographic Details
Published inBlood Vol. 116; no. 21; p. 1523
Main Authors Miller, Larry, Philbeck, Thomas, Montez, Diana, Puga, Tatiana, Brenner, Andrew, Cohen, Stephen C., Spadaccini, Cathy
Format Journal Article
LanguageEnglish
Published Elsevier Inc 19.11.2010
Online AccessGet full text
ISSN0006-4971
1528-0020
DOI10.1182/blood.V116.21.1523.1523

Cover

Abstract Abstract 1523▪ A battery-powered bone marrow biopsy system was cleared by the U.S. Food and Drug Administration in 2007 for bone marrow sampling. A study was designed to determine if the powered device has advantages over traditional manual devices in terms of decreased pain, insertion time, and improved sample yield. The study protocol was approved by IntegReview Institutional Review Board. The study was conducted in a community cancer center. Pain-free adult volunteers, not on analgesics or anxiolytics, were enrolled in the study. Patients were not informed which needle was proprietary or being used with each procedure. Noise cancelling headphones and sham drill sound during manual insertions were employed. They were also uninformed as to which device was proprietary. After history, physical (including complete blood count) and consenting procedures, volunteer subjects were randomized to first pass device type and first-pass procedure side (left or right). Following local anesthesia administration using buffered Lidocaine, device operators (board-certified in hematology/oncology) performed the procedures using a Jamshidi® 11 gauge/4” Manual bone marrow biopsy needle (Havel, Inc. Richmond, VA) and an OnControl™ 11 gauge/102mm Powered bone marrow biopsy needle (Vidacare Corporation, Shavano Park, TX). A 100mm visual analog scale (VAS) was used to assess pain at initial biopsy needle insertion through the cortex, needle advancement for core sample capture, and needle removal—followed by a VAS measurement of the overall procedure pain and a final check 30 minutes after the procedure. After the second procedure, subjects were given a chance to change their overall pain score for the first procedure (“second look overall pain score”). Follow-up assessment for VAS and complications were made 1 day, 2 days, and 7 days after the biopsy procedures. Procedure time was measured from needle-to-skin contact to core specimen ejection from the needle. Core specimens were measured for length and submitted to pathology reference laboratory for grading. Pathology assessment included size measurement, post-fixation quality, medullary component quality, and overall quality. There were 24 evaluable subjects the study. The mean age was 33.8 ± 11.3 years, and 54.2% were male. Mean height and weight were 169.7 ± 10.8cm and 77.0 ± 17.7kg, respectively. Of the 24 evaluable subjects, there were 24 manual insertions and 24 powered insertions. Cores were obtained in 67% of the Manual insertions and 100% of Powered insertions—a significant difference (p=0.002). The mean time to core acquisition was 86 seconds for Manual device and 47 seconds for the Powered device, a significance difference (p<0.001). Using the second look overall pain score (for those that opted to change their score), the mean score for the Manual device was 33.3 and 20.9 for the Powered—a significant difference (p<0.05). There was no statistical or clinical difference in other pain scores between the two devices. The mean biopsy core length was 11.1 ± 4.5mm for the Manual device and 17.0 ± 6.8mm for the Powered device (Figure 1); a statistical difference (p<0.005). Pathology assessment for the Manual device showed mean length of 6.08 ± 5.59mm, width of 0.98 ± 0.74mm, and mean volume of 11.00 ± 10.76mm3. Measurements for the Powered device were mean length of 15.29 ± 6.12mm, width of 1.98 ± 0.31mm, and mean volume of 49.12 ± 21.45mm3. All differences were statistically significant with p-values <0.001. For overall quality, 33.3% of Manual samples and 79.2% of Powered samples were graded adequate—a statistical difference (p=0.002). There were no complications. Results suggest the superior size and overall quality of core specimens delivered by the Powered device may provide more material for pathological evaluation of hematopoietic and oncological disorders. The powered device was significantly faster in obtaining a biopsy than the manual device and its capture rate in obtaining a satisfactory sample was much higher. Use of the Powered bone marrow biopsy device significantly decreases overall procedure pain. [Display omitted] Miller:Vidacare Corporation: Employment, Equity Ownership, Patents & Royalties. Philbeck:Vidacare Corporation: Employment. Montez:Vidacare Corporation: Employment. Puga:Vidacare Corporation: Employment. Brenner:Vidacare Corporation: Consultancy, Research Funding. Cohen:Vidacare Corporation: Consultancy, Research Funding.
AbstractList Abstract 1523
Abstract 1523▪ A battery-powered bone marrow biopsy system was cleared by the U.S. Food and Drug Administration in 2007 for bone marrow sampling. A study was designed to determine if the powered device has advantages over traditional manual devices in terms of decreased pain, insertion time, and improved sample yield. The study protocol was approved by IntegReview Institutional Review Board. The study was conducted in a community cancer center. Pain-free adult volunteers, not on analgesics or anxiolytics, were enrolled in the study. Patients were not informed which needle was proprietary or being used with each procedure. Noise cancelling headphones and sham drill sound during manual insertions were employed. They were also uninformed as to which device was proprietary. After history, physical (including complete blood count) and consenting procedures, volunteer subjects were randomized to first pass device type and first-pass procedure side (left or right). Following local anesthesia administration using buffered Lidocaine, device operators (board-certified in hematology/oncology) performed the procedures using a Jamshidi® 11 gauge/4” Manual bone marrow biopsy needle (Havel, Inc. Richmond, VA) and an OnControl™ 11 gauge/102mm Powered bone marrow biopsy needle (Vidacare Corporation, Shavano Park, TX). A 100mm visual analog scale (VAS) was used to assess pain at initial biopsy needle insertion through the cortex, needle advancement for core sample capture, and needle removal—followed by a VAS measurement of the overall procedure pain and a final check 30 minutes after the procedure. After the second procedure, subjects were given a chance to change their overall pain score for the first procedure (“second look overall pain score”). Follow-up assessment for VAS and complications were made 1 day, 2 days, and 7 days after the biopsy procedures. Procedure time was measured from needle-to-skin contact to core specimen ejection from the needle. Core specimens were measured for length and submitted to pathology reference laboratory for grading. Pathology assessment included size measurement, post-fixation quality, medullary component quality, and overall quality. There were 24 evaluable subjects the study. The mean age was 33.8 ± 11.3 years, and 54.2% were male. Mean height and weight were 169.7 ± 10.8cm and 77.0 ± 17.7kg, respectively. Of the 24 evaluable subjects, there were 24 manual insertions and 24 powered insertions. Cores were obtained in 67% of the Manual insertions and 100% of Powered insertions—a significant difference (p=0.002). The mean time to core acquisition was 86 seconds for Manual device and 47 seconds for the Powered device, a significance difference (p<0.001). Using the second look overall pain score (for those that opted to change their score), the mean score for the Manual device was 33.3 and 20.9 for the Powered—a significant difference (p<0.05). There was no statistical or clinical difference in other pain scores between the two devices. The mean biopsy core length was 11.1 ± 4.5mm for the Manual device and 17.0 ± 6.8mm for the Powered device (Figure 1); a statistical difference (p<0.005). Pathology assessment for the Manual device showed mean length of 6.08 ± 5.59mm, width of 0.98 ± 0.74mm, and mean volume of 11.00 ± 10.76mm3. Measurements for the Powered device were mean length of 15.29 ± 6.12mm, width of 1.98 ± 0.31mm, and mean volume of 49.12 ± 21.45mm3. All differences were statistically significant with p-values <0.001. For overall quality, 33.3% of Manual samples and 79.2% of Powered samples were graded adequate—a statistical difference (p=0.002). There were no complications. Results suggest the superior size and overall quality of core specimens delivered by the Powered device may provide more material for pathological evaluation of hematopoietic and oncological disorders. The powered device was significantly faster in obtaining a biopsy than the manual device and its capture rate in obtaining a satisfactory sample was much higher. Use of the Powered bone marrow biopsy device significantly decreases overall procedure pain. [Display omitted] Miller:Vidacare Corporation: Employment, Equity Ownership, Patents & Royalties. Philbeck:Vidacare Corporation: Employment. Montez:Vidacare Corporation: Employment. Puga:Vidacare Corporation: Employment. Brenner:Vidacare Corporation: Consultancy, Research Funding. Cohen:Vidacare Corporation: Consultancy, Research Funding.
Author Cohen, Stephen C.
Spadaccini, Cathy
Brenner, Andrew
Puga, Tatiana
Montez, Diana
Miller, Larry
Philbeck, Thomas
Author_xml – sequence: 1
  givenname: Larry
  surname: Miller
  fullname: Miller, Larry
  organization: Vidacare Corporation, Shavano Park, TX, USA
– sequence: 2
  givenname: Thomas
  surname: Philbeck
  fullname: Philbeck, Thomas
  organization: Vidacare Corporation, Shavano Park, TX
– sequence: 3
  givenname: Diana
  surname: Montez
  fullname: Montez, Diana
  organization: Vidacare Corporation, Shavano Park, TX
– sequence: 4
  givenname: Tatiana
  surname: Puga
  fullname: Puga, Tatiana
  organization: Vidacare Corporation, Shavano Park, TX, USA
– sequence: 5
  givenname: Andrew
  surname: Brenner
  fullname: Brenner, Andrew
  organization: University of Texas Health Science Center at San Antonio, San Antonio, TX, USA
– sequence: 6
  givenname: Stephen C.
  surname: Cohen
  fullname: Cohen, Stephen C.
  organization: San Antonio Tumor & Blood, San Antonio, TX, USA
– sequence: 7
  givenname: Cathy
  surname: Spadaccini
  fullname: Spadaccini, Cathy
  organization: Ameripath, South Texas, San Antonio, TX, USA
BookMark eNqFkM1OwzAQhC1UJNrCM5AXSNh14vwc2wooUhE90F4tJ9mAURpHdkrVtydJufeyc9id0ew3Y5PGNMTYI0KAmPKnvDamDPaIccAxQMHDcdywaS-pD8BhwqYAEPtRluAdmzn3A4BRyMWUtStzaJXVzZe3NSeyVHrLPt97V9aak7fUpnVnb2tNQeXRkvM60--ao6qv3O3ckLkmVXffZ29v6mPTEVl3z24rVTt6-Nc52708f67W_ubj9W212PgFQhz6Wd-3UGWIVCoOhAhJoVIRl3me5ZSqRIVphFRBmkdJJvKKUg5VxeNMhJkQIpyz5JJbWOOcpUq2Vh-UPUsEOYCTIzg5gJMc5cBsHL1zcXFSX-9Xk5Wu0NT0j2lLRSdLo69m_AFRKXwP
ContentType Journal Article
Copyright 2010 American Society of Hematology
Copyright_xml – notice: 2010 American Society of Hematology
DBID 6I.
AAFTH
AAYXX
CITATION
DOI 10.1182/blood.V116.21.1523.1523
DatabaseName ScienceDirect Open Access Titles
Elsevier:ScienceDirect:Open Access
CrossRef
DatabaseTitle CrossRef
DatabaseTitleList CrossRef

DeliveryMethod fulltext_linktorsrc
Discipline Medicine
Chemistry
Biology
Anatomy & Physiology
EISSN 1528-0020
EndPage 1523
ExternalDocumentID 10_1182_blood_V116_21_1523_1523
S0006497119435921
GroupedDBID ---
-~X
.55
1CY
23N
2WC
34G
39C
4.4
53G
5GY
5RE
5VS
6I.
6J9
9M8
AAEDW
AAFTH
AAXUO
ABOCM
ABVKL
ACGFO
ADBBV
AENEX
AFFNX
AFOSN
AHPSJ
ALMA_UNASSIGNED_HOLDINGS
BAWUL
BTFSW
CS3
DIK
DU5
E3Z
EBS
EJD
EX3
F5P
FDB
FRP
GS5
GX1
IH2
K-O
KQ8
L7B
LSO
MJL
N4W
N9A
OK1
P2P
R.V
RHF
RHI
ROL
SJN
THE
TR2
TWZ
W2D
W8F
WH7
WOQ
WOW
X7M
YHG
YKV
ZA5
0R~
AALRI
AAYXX
ACVFH
ADCNI
ADVLN
AEUPX
AFPUW
AGCQF
AIGII
AITUG
AKBMS
AKRWK
AKYEP
AMRAJ
CITATION
H13
ID FETCH-LOGICAL-c1063-9006cad31eda20e1107ca856dbb9be8a7a3841ef08b4795bfe820ff2695395553
ISSN 0006-4971
IngestDate Tue Jul 01 01:45:40 EDT 2025
Fri Feb 23 02:43:42 EST 2024
IsDoiOpenAccess true
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Issue 21
Language English
License This article is made available under the Elsevier license.
LinkModel OpenURL
MergedId FETCHMERGED-LOGICAL-c1063-9006cad31eda20e1107ca856dbb9be8a7a3841ef08b4795bfe820ff2695395553
OpenAccessLink https://dx.doi.org/10.1182/blood.V116.21.1523.1523
PageCount 1
ParticipantIDs crossref_primary_10_1182_blood_V116_21_1523_1523
elsevier_sciencedirect_doi_10_1182_blood_V116_21_1523_1523
ProviderPackageCode CITATION
AAYXX
PublicationCentury 2000
PublicationDate 2010-11-19
PublicationDateYYYYMMDD 2010-11-19
PublicationDate_xml – month: 11
  year: 2010
  text: 2010-11-19
  day: 19
PublicationDecade 2010
PublicationTitle Blood
PublicationYear 2010
Publisher Elsevier Inc
Publisher_xml – name: Elsevier Inc
SSID ssj0014325
Score 1.9766155
Snippet Abstract 1523▪ A battery-powered bone marrow biopsy system was cleared by the U.S. Food and Drug Administration in 2007 for bone marrow sampling. A study was...
Abstract 1523
SourceID crossref
elsevier
SourceType Index Database
Publisher
StartPage 1523
Title Comparing Powered Bone Marrow Biopsy Procedures to Manual Bone Marrow Biopsy Procedures Using Healthy Volunteers
URI https://dx.doi.org/10.1182/blood.V116.21.1523.1523
Volume 116
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://utb.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwnV3NT9swFLc60D4u0yhDg43Jh2mXKZ3j2HWyG3QgNK2ISQVxi-zEvcDaiqWH7p_Zv8p7tpMUxijsYrmW-5r4_fr8_Pw-CPlQYhIuZkXETJZFQhsTZcqyCGuJKNghjHSlE4bH_aNT8e1cnnc6f5a8luaV6RW_74wr-R-uwhjwFaNkH8HZhigMQB_4Cy1wGNoH8XjgiwjCYf8Ei52B7rg_nWAxIUysiGUmZ78WPhSgnF_5XA5D7fLp3z_P-xH4CKXFpzN4B1j-4Crf3ABfhjLzNyMKvwPNxkqP1hrg3sUtTyTn4wsUne36K-Cz2RpO5t7OO9JVMxxMEujeEUdB8NVito-l6zxobJCsmAqbcXZD9Po4y4AxHyodJCnMT5Z25frj3xI_xQyyzsu_dwb0ejzGmk5JryWwnGP71t7XeCS6s1DKc0coR0I5j3Ok4ZonZJ0rUM7w1v9He00lEu5LZIT3DQ6EQOjzP57obvVnSaUZvSIvw1mE7nlgbZCOnXTJ5t5EV9OfC_qROu9gd-3SJU_3697zQV0jsEueDYNrxiaZNWCkAYwUQUY9yKgHGW1BRqsp9WBcMc-BkQYw0haMr8np4cFocBSFch5REYMiHGWwSoUuk9iWmjOLhodCp7JfGpMZm2qlk1TEdsxSI1QmzdiCdjoe834mk0xKmWyRtQk80BtCVWEU6J2i4GkpDFNa4MlcGhNLBb1ym7B6lfOZz9qSr-DwNvlScyMPyqdXKnNA2qov7zz-996SF-1f5x1Zq67mdhf028q8dxi7BknWoXM
linkProvider Colorado Alliance of Research Libraries
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Comparing+Powered+Bone+Marrow+Biopsy+Procedures+to+Manual+Bone+Marrow+Biopsy+Procedures+Using+Healthy+Volunteers&rft.jtitle=Blood&rft.au=Miller%2C+Larry&rft.au=Philbeck%2C+Thomas&rft.au=Montez%2C+Diana&rft.au=Puga%2C+Tatiana&rft.date=2010-11-19&rft.issn=0006-4971&rft.eissn=1528-0020&rft.volume=116&rft.issue=21&rft.spage=1523&rft.epage=1523&rft_id=info:doi/10.1182%2Fblood.V116.21.1523.1523&rft.externalDBID=n%2Fa&rft.externalDocID=10_1182_blood_V116_21_1523_1523
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0006-4971&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0006-4971&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0006-4971&client=summon